CHAPTER 9, ERROR MESSAGES
During pH calibration the following error messages can appear on the display:
"ERR1”: The zero (pH-7) calibration of the pH-electrode is out of range (± 1,00pH).
"ERR2”: The span (pH-4) calibration of the pH-electrode is out of range (90%-105%).
Press the pH-button to stop the error message. During 10 seconds one of the following messages
will appear on the display:
"AH”: The zero (pH-7) calibration of the electrode is out of range (± 1.00pH).
"SH”: The span (pH-4) calibration of the electrode is out of range (> 105%).
"SL”: The span (pH-4) calibration of the electrode is out of range (< 90%).
During EC calibration the following error messages can appear on the display:
"ERR3”: The calibration of the electrode is out of range (70%-100%).
Press the EC-button to stop the errormessage. During 10 secondsoneof the followingmessages
will appear on the display:
"SH”: The span (EC-4) calibration of the electrode is out of range (> 130%).
"SL”: The span (EC-4) calibration of the electrode is out of range (< 70%).
During temperature calibration the following error messages can appear on the display:
"ERR4” followed by “AH”: The calibration of the electrode is out of range (±2,0°C).
Press the Temp-button to stop the error message. The instrument will continue temperature
measurement.
When one of these error messages occurs, check if the buffer solutions have the right values and
that the electrode is properly placed into the liquid. Check also if the electrode is properly
connected to the instrument. Start the calibration procedure again. When the error message still
appears, contact your supplier.
CHAPTER 10, THE EPH-119RS
The EC/pH/°C-meter EPH119RS is provided with a serial port RS232 and can be connected
with any equipment with a RS232 serial port.
Data are sent every 0.3 seconds, i.e. every time the display of the EPH119 is updated.
Data format:
±xx.xx°C ±xx.xxpH ±xx.xxmS