The shell of this instrument should be properly grounded through the power ground wire. Any
damage to the internal or external grounding path of the instrument may be dangerous.
Note: If leakage is found, please cut off the power immediately and stop using the instrument.
Note: Please cut off the power supply before moving the instrument.
Use the power cord provided with this instrument. If the original power cord is broken, worn, or
disconnected, the same power cord should be replaced as soon as possible.
In order to avoid the danger of electric shock, the power cord of this instrument must be reliably
grounded. The power cord of this instrument is a standard plug power cord. Please insert the
power plug into the appropriate three-wire grounding socket (the external network power supply
connected by the socket must be reliably grounded) to ensure the safe use of the instrument.
The electrical supply in which the instrument is located must be connected with ground wire.
Please carefully check whether the power connection is firm. Do not hard pull the power cord.
When the plug is plugged into the power supply, make sure the plug is fully plugged into the
socket.
Please keep the power plug and the power cable away from high temperature objects such as
heater. When using this instrument, please do not place any items on the power cord, do not put
the power cord in the place where people often walk.
The fuse type of this instrument is T10AH250VP, placed in the backup box at the power outlet
at the back of the instrument. The use of improper fuses may damage the instrument's line system
and cause fire. Before turning on the instrument power switch, check and make sure the fuse is
properly installed.
In order to prevent fire hazards, use this instrument only with the indicated fuse. When replacing
the fuse, please cut off the power supply, pull out the power cord, pry open the fuse box with a
word screwdriver, and replace the old fuse with the same new fuse.
1.5 Environmental Safety and Precautions
This instrument is only suitable for indoor use, indoor ventilation should be good, and free of
corrosive gas.
EDM may cause explosion hazard and is prohibiting the use of this instrument in the presence or
possible presence of flammable and explosive gases.
The working environment temperature of this instrument should be between 10~30and the
reate umdt soud 5%.
The working environment pressure of this instrument should be normal atmospheric pressure
(altitude less than 2000 meters).