GYROZEN 624R User manual

2
Compact Bench-Top, Refrigerated Centrifuge 624R

3
Compact Bench-Top, Refrigerated Centrifuge 624R

4
Compact Bench-Top, Refrigerated Centrifuge 624R
CONTENTS
Page #
1. Meanings of Symbols & Safety Precautions................................................................................4
1-1. Meanings of Symbols.................................................................................................................5
1-2. Safety Precautions......................................................................................................................5
2. Product Description & Technical Specifications .....................................................................8
2-1. Product Description....................................................................................................................8
2-2. Technical Specifications............................................................................................................8
3. Unpacking..............................................................................................................................................9
4. Installation........................................................................................................................................10
4-1. Power On/Off and Lid Release...............................................................................................10
4-2. Rotor Coupling and Disassembling......................................................................................11
5. Operation..........................................................................................................................................13
5-1. Key Functions of Control Panel.............................................................................................13
5-2. Setting the RPM/RCF Value.....................................................................................................14
5-3. Setting the Time Value .............................................................................................................14
5-4. Setting Temperature and Fast Cool......................................................................................15
5-5. Acceleration / Deceleration.....................................................................................................16
5-6. Program Saving & Recalling...................................................................................................17
5-7. Start/Stop.....................................................................................................................................18
5-8. Repeat Account and Sound Pitch of End Alarm ...............................................................18
5-9. Pulse .............................................................................................................................................19
5-10. Emergency Lid-Lock Release...............................................................................................19
6. Maintenance........................................................................................................................................20
6-1. Outer part of Instrument..........................................................................................................20
6-2. Chamber.......................................................................................................................................20
6-3. Shaft..............................................................................................................................................20
6-4. Rotor .............................................................................................................................................20
6-5. Transportation of Instrument .................................................................................................20
7. Trouble Shooting...............................................................................................................................21
7-1. Check List....................................................................................................................................21
7-2. Error Code...................................................................................................................................21
8. CE
...........................................................................................................................................................24

5
Compact Bench-Top, Refrigerated Centrifuge 624R
1. Meanings of Symbols & Safety Precautions
1-1. Meanings of Symbols
1-1-1. Symbols on the device
Symbol
Meaning
Symbol
Meaning
Attention and warning.
Attention and warning for
electric shock
Attention and warning for correct way of sample
balancing in the rotor.
Attention and warning for rotor coupling.
Attention and warning for lid opening and closing
1-1-2. Symbols in this document
Symbol
Meaning
Symbol
Meaning
This symbol refers to safety
relevant warnings and indicates
possible dangerous outcomes.
☞
Note.
This symbol refers to the important
reminder.
1-2. Safety Precautions
Before using the instrument, please read this operation manual to ensure correct usage. Incorrect handling of the
instrument could possibly result in personal injury or physical damage on the instrument or its accessories.
1. ALWAYS locate the instrument on a flat, rigid and stable table capable of withstanding the weight of the
instrument and its spinning operation.
2. ALWAYS make a safety zone of 30 cm around the centrifuge to indicate that neither hazardous
materials nor persons should be permitted within the area during operation.
ALWAYS position the instrument with enough space on each side of instrument to ensure
proper air circulation.
3. ALWAYS install the instrument within a temperature and humidity controlled environment. (Permissible
ambient temperature: +5°C ~ +35 °C, Relative humidity: ≤ 85%)

6
Compact Bench-Top, Refrigerated Centrifuge 624R
4. Before connecting the power, check the rated voltage.
5. Should not use unapproved rotors and accessories.
Only use rotors from GYROZEN Co., Ltd. with appropriate centrifugal tubes and suitable
adaptors to embrace sample containers tightly enough inside rotors.
6. Before operating the instrument, check if the rotor and the rotor lid are securely fastened.
Should operate the instrument with a rotor properly installed and secured to the motor shaft.
7. Mount the rotor on the motor shaft properly, check it with spinning manually.
8. Do not stop the rotor by touching with hand during the instrument is running.
9. Emergency Lid-Lock Release should be performed only when spinning is completely stopped.
10. Should not exceed the rated speed or specific gravity. Samples whose density is greater than 1.2g/ml
must have reduced maximum rotational speed to avoid rotor failure.
11. The sample content should not exceed 80% of total capacity of a tube. Otherwise, it would cause
spillage of sample fluid and even the tube breakage.
12. ALWAYS load the tubes symmetrically with evenly weighted samples to avoid rotor imbalance. If
necessary, use the water blank to counterbalance the unpaired sample.
13. The operation speed should not exceed the highest value of the individual guaranteed g-forces of each
the centrifuge, rotor, bucket or adaptors and sample container, especially the guaranteed g-force of
sample container should not be neglected.
14. The rotors should be cleaned and kept dry after every use for longer life and safety.
15. ALWAYS disconnect the power supply prior to maintenance care and service to avoid electrical shock.
16. ALWAYS use proven disinfection procedures after centrifuging biohazardous materials.
17. Should not centrifuge flammable, toxic, radioactive, explosive, or corrosive materials.
18. When it is necessary to use toxic or radioactive materials or pathogenic micro-organisms which belong
to the Risk Group II of WHO: “Laboratory Bio- safety Manual,” should follow national regulations.
19. After using the centrifuge, turn the power switch off.
20. Unplug power cord before cleaning or whenever the centrifuge is to remain unused for a long period of
time.

7
Compact Bench-Top, Refrigerated Centrifuge 624R
Do not place dangerous materials within 30 cm distance around the instrument, and that is
also recommended by IEC 61010-2-020.
Use the Emergency Lid-Lock Release function only when the lid button on the control panel
is dumb under the condition of complete stop of rotor running.
Never try to open or move the instrument if it is not completely stopped.
If the power input is more than +/- 10% of the recommended voltage or fluctuates
frequently, it may cause malfunction of the instrument and often result serious damage.
Install the instrument at the place without any kind of corrosive gases.

8
Compact Bench-Top, Refrigerated Centrifuge 624R
2. Product Description & Technical Specifications
2-1. Product Description
1
Lid
2
Power Socket
3
Power Switch
4
Manual Lid Opening Hole
5
Display & Control Panel
6
Tool Box
2-2. Technical Specifications
Max. RPM/RCF
Fixed angle
6,000 rpm / 3,824 xg
Swing out
4,800 rpm / 3,888 xg
Max. capacity
Fixed angle
6 x 50 ml, 24 x 15 ml
Swing out
4 x 100 ml, 32 vacutainer tube, 4 MTPs
Temp. range (℃)
-10 ~ +40
FAST cool button
Yes
Time control
Pulse, timed < 10 hr or continuous
Time counting modes
Selectable, at set speed or from starting
RPM/RCF conversion
Yes
Noise level (dB)
~ 60
ACC/DEC
9/10 steps

9
Compact Bench-Top, Refrigerated Centrifuge 624R
Program memory
100
Parameters on display window
RPM (RCF), Oper Status
Lid Open/Close, Hr:Min, Temp, ACC, DEC
Display
White LCD
Automatic rotor Identification
Yes
Imbalance cut-off
Yes
Safety lid lock
Yes
Lid drop protection
Yes
Motorized lid open & close
Yes
Power supply(V/Hz)
230/50~60 (110V optional)
Power requirement
900 VA
Dimension(W x D x H, mm)
584 x 535 x 317
Weight without rotor (kg)
65
CE mark
YES
Cat. No.
GZ-0624R-B(Black display/ GZ-0624R-W (White display)
☞This instrument has following functions for safety.
1. Automatic rotor identification function.
2. Automatic detection and alarms for imbalance, excess speed and heating.
3. Automatic lid lock function.
3. Unpacking
Motor Fixing Cushion in 624R
☞The cushion should be removed before the installation of the instrument.
Motor Protecting Device: a motor fixing cushion (Polyethylene (PE) foam) is placed in the chamber
of 624R centrifuge to prevent possible damages to the imbalance sensor caused by motor shaking
during shipping or relocation.

10
Compact Bench-Top, Refrigerated Centrifuge 624R
4. Installation
4-1. Power On/Off and Lid Release
Action
4-1-1. Power On/Off
1. After connecting the AC Power cord at the power socket on the right back of the instrument, put the
plug into the outlet.
Check the proper power.
2. Turn on a power switch on the left side of the instrument.
With beeping sound, right before setting value is displayed.
The default values are Max. rpm, 10 min, ACC 7, DEC 7 and
25°C.
4-1-2. Lid Release
1. For opening the lid, touch the [LID] button.
Should touch the [LID] button When the lid is closed
(Lid LED shows off)
Close the lid until hearing clank shut.
When the lid is opened, the lid LED turns on.
☞
The lid is not opened while the instrument is running.
If the lid is opened, the instrument could not be operated even with pressing the [START] button.
For operational safety, this instrument has the automatic rotor recognition function.
When you supply the power, “Rotor Scan…” will be appeared. If the rotor is absent, the “Error 9”
will be appeared. This message will be cleared after rotor coupling and running.
The lid is not automatically opened after finishing operation to keep the sample at proper
temperature.
Power Failure: If there is any power failure during operation, lid is not opened with „Lid‟ button. Lid
can be opened only when the operation is completely stopped and the power is on again. If you
want to open the lid at the power failure, please refer to „5-9. Emergency Lid-Lock Release‟.

11
Compact Bench-Top, Refrigerated Centrifuge 624R
4-2. Rotor Coupling and Disassembling
Action
1. Before coupling a rotor, clean the motor shaft and chamber with soft dry towel.
4-2-1. Swing-Out Rotor
1. Mount a proper rotor into the motor shaft.
2. Grasp the rotor with one hand, and place Rotor Locking Tool
at the center hole of the rotor.
To assemble the rotor: Rotate the Rotor Locking Tool
clockwise until tightly assembled.
To disassemble the rotor: Rotate the Rotor Locking
Tool counterclockwise.
3. Hang the appropriate buckets into the rotor.
Load the identical bucket at each wing for safety. (Do not leave a vacant wing without
bucket. All wings should hold identical bucket. )
Remove dirt and dust around hooks of rotor and hanging part of bucket.
4. Spin the rotor manually to check if bucket swinging is free enough and ever. If they do not swing freely,
apply the Lubricant (grease) to the linking area.
4-2-2. Fixed Angle Rotor
1. Mount a proper rotor into the motor shaft.
Grasp the rotor with one hand, and place the Rotor Locking Tool at the center
hole of the rotor.
To assemble the rotor: Rotate the Rotor Locking Tool clockwise until
tightly assembled.
To disassemble the rotor: Rotate Rotor Locking Tool counterclockwise.
2. To close of the rotor lid, rotate the rotor lid nut clockwise.
For opening lid: rotate the rotor lid nut counterclockwise.
When you run a fixed angle rotor, make sure that the rotor lid is tightly closed. If you don‟t close the rotor lid
completely, it will be crushed.
☞For operational safety, this instrument has the automatic rotor recognition function.

12
Compact Bench-Top, Refrigerated Centrifuge 624R
4-3. Positioning of Sample Tubes
Action
1. Before loading sample tubes, check the water drop or dirt in the rotor hole or inner adaptor.
If there is a water drop or dirt in the rotor hole or inner adaptor, remove it with soft dry cloth.
2. Tubes should be placed in the rotor with same amount of samples at symmetrical positions.
Only use appropriate centrifugal tubes and do not exceed the speed beyond the tube‟s max
g-force.
Correct Way of Sample Balancing and Tube Usage
If the number of samples is not in pair, please load the control tubes at each symmetrical
position. Otherwise, it results noise and vibration, which eventually damage the instrument
☞For safety, the „Imbalance Cut Off‟ function will be occurred, if there is imbalance of loading tubes (Error 8,
Imbalance error). Please refer to 7. Trouble Shooting.

13
Compact Bench-Top, Refrigerated Centrifuge 624R
5. Operation
5-1. Key Functions of Control Panel
□ RPM/RCF
For automatic conversion of RPM/RCF and to set the speed
□ TIME
Use to set time, available range up to 9 hour 59 min (00:00: continuous)
□ Temp
Use to set temperature (-10°C ~ 40°C)
Use to reach rapid refrigeration up to the setting temperature. (touch for more than 2
seconds.)
□ ACC/DEC
Use to set the acceleration & deceleration level from 1 to 9 steps. „0‟ in deceleration step
means natural deceleration. Larger number means faster acceleration or deceleration.
□ PROG
Use to save a set of setting values or recall the saved program number
□ KEY LOCK
Use for key lock mode
□ AT SET SPEED
Use to count the run time once the actual run speed reaches to the set speed value
□ Pulse
Use for quick runs
□ Enter
Use for completion of data setting
□ Start/Stop
Use to start and stop operation
□ Lid
Use to open instrument lid
□ Sound
Use to set the number of sound and volume

14
Compact Bench-Top, Refrigerated Centrifuge 624R
AT SET SPEED Mode
For exact time control, this instrument can be set with AT SET SPEED mode which counts the run
time once the actual run speed reaches to the set speed value and stops when the deceleration begins
.
*[AT SET SPEED] lamp turns on: From t1 to t2
*[AT SET SPEED] lamp turns off: From t0 to t2
5-2. Setting the RPM/RCF Value
Action
5-2-1. Setting the RPM Value
▶Speed setting unit: 1 rpm/ 1xg
1. Touch the [RPM/ RCF] button once.
RPM MODE is generated with touching a [RPM/
RCF] button once.
RCF MODE is generated with touching the [RPM/
RCF] button twice.
RPM/ RCF LED is flickering at the display window.
2. Touch the number buttons to change input value.
If you do not touch the number button for 15
seconds, the setting mode is cleared.
3. Touch the [ENTER] button to complete the setting.
Touch [ENTER] to save the setting value.
If wrong number is entered, touch [] button and change the value again.
5-3. Setting the Time Value
Action
5-3-1. Setting the AT SET SPEED mode
1. Touch the [AT SET SPEED] button once.

15
Compact Bench-Top, Refrigerated Centrifuge 624R
5-3-2. Setting the ‘HOUR / MIN’Value
▶Time setting unit: 1hr. / 1 min.
Action
Time is down-counted after starting centrifugation.
1. Touch the [TIME] button once.
„MIN‟value on LED is flickering.
2. Touch the number buttons to change the minute value.
If you do not touch the number button for 15
seconds, the setting mode is cleared.
If wrong number is entered, touch [] button
and change the value again.
3. Touch the [ENTER] button to pass the „HOUR‟ value
setting.
4. Touch the number buttons to change the hour value.
If you do not touch the number button for 15
seconds, the setting mode is cleared.
If wrong number is entered, touch [] button and
change the value again.
5. Touch the [ENTER] button to complete the setting.
5-4. Setting Temperature and Fast Cool
Action
5-4-1. Setting Temperature
▶Temperature can be set from -10℃to 40℃
▶Temp setting unit: 1 ℃
1. Touch the [TEMP] button. Default or latest temperature
value blinks on the display window.
2. Touch the number buttons to change temperature.
3. Touch the [ENTER] button to complete setting.
.

16
Compact Bench-Top, Refrigerated Centrifuge 624R
5-4-2. Fast Cool
1. Setting the temperature. (Please refer to 5-4-1. Setting
Temperature)
2. After installation of the rotor and closing the lid of
instrument, touch the [TEMP] button for more than 2 seconds.
For fast cooling, the instrument is refrigerated down to the set temperature in a short time.
During the fast cooling, the rotor runs at low speed (1,000 rpm).
The passed time is showed on the display window.
☞
If you‟d like to load your sample tubes before pressing the [Fast Cool] button, please check if the
sample is safe during spinning at 1,000 rpm.
Before starting Fast Cooling, please check the rotor coupling and symmetry of sample tubes.
5-5. Acceleration / Deceleration
Use the adjustment function of acceleration (level: 1~9) & deceleration levels (level: 0~9) to protect sensitive
samples.
Action
1. Touch [ACC/DEC] button.
2. Touch the number buttons to change input ACC value.
Input the desired level of ACC from 1 to 9.
(Level 9: The fastest acceleration)
If you do not touch the number button for 15
seconds, the setting mode is cleared.
If wrong number is entered, touch [] button
and change the value again.
3. Fix the ACC level by touching [ENTER] button.
4. Touch the number buttons to change input DEC value.
Input the desired level of DEC from 0 to 9.
(Level 0: Natural deceleration / Level 9: The
fastest deceleration)
If you do not touch the number button for 15
seconds, the setting mode is cleared.
If wrong number is entered, touch [] button
and change the value again.
5. Fix the DEC level by touching [ENTER] button.

17
Compact Bench-Top, Refrigerated Centrifuge 624R
5-6. Program Saving & Recalling
Action
5-6-1. Program Saving
1. Set parameters. (Refer to 5-2 ~ 5-5)
2. Touch the [PROG] button twice.
„SAVE‟is turned on the display window.
3. Touch the number buttons to change input Program number.
If you do not touch the number button for 15
seconds, the setting mode is cleared.
Save up to 100 programs. (Program numbers
from 00 to 99).
If wrong number is entered, touch [] button
and change the value again.
4. Touch the [ENTER] button to complete the saving.
5-6-2. Program Recalling
1. To recall the saved program, touch the [PROG] button once.
„CALL‟is turned on the display window.
2. Touch the number buttons to select program number you
want to recall and then touch the [ENTER] button.
If you do not touch the number button for 15
seconds, the setting mode is cleared.
If wrong number is entered, touch [] button
and change the value again.

18
Compact Bench-Top, Refrigerated Centrifuge 624R
5-7. Start/Stop
Action
5-7-1. Start
1. After setting RPM/RCF and Time, touch
[START/STOP] button.
During running, a „Start LED‟ is turned on.
The instrument is running only when the lid is
closed.
When you touch the [ENTER] button during
operation, display window shows the saved
setting parameters
5-7-2. Stop
1. In case of touching the [START/STOP] button, the operation is stopped.
When you touch the [START/STOP] button twice, the operation is stopped with DEC 9.
5-8. Repeat Account and Sound Pitch of End Alarm
Repeat account and sound pitch of the end alarm can be selected.
Action
5-8-1. Setting the Sound Pitch of End Alarm
1. Touch [SOUND] button.
„Sound LEVEL_03‟ appears on the display
window.
2. Touch the number buttons to change the value
for the pitch of sound.
3. Fix the value by touching [ENTER] button.
Sound LEVEL: 0~10 (0: silent)
5-8-2. Setting the Repeat Count of End Alarm
1. Touch [AT SET SPEED] button for more than 2 seconds.
„Sound rPt‟ appears on the display window.

19
Compact Bench-Top, Refrigerated Centrifuge 624R
2. Touch the number buttons to change the value for the repeat count.
3. Fix the value by touching [ENTER] button.
The number of finishing sound: 0~99 (0: silent, 99: 99 times)
5-9. Pulse
It is for quick and short spin down.
Action
1. If you touch [PULSE] button and release at the point you
want to stop, the centrifuge decelerates immediately.
When the operation is completely stopped, the lid is
not automatically opened.
5-10. Emergency Lid-Lock Release
For emergency lid-lock release, you can use the Emergency Lid-Lock Release Tool when the instrument is
completely stopped.
The lid can be unlocked manually with the Emergency Lid-Lock Release Tool through the emergency lid-lock
release hole.
Action
1. Find the emergency lid-lock release hole in the left side of the instrument and take out the white rubber
closure.
2. Insert the Emergency Lid-Lock Release Tool into the hole and revolve it clockwise until the lid is released.
Manual opening should be performed only when spinning is completely stopped. Otherwise, harmful
damage will be accompanied to not only operators but samples.
After opening the lid manually, it is recommended to wait until normal electricity comes back.

20
Compact Bench-Top, Refrigerated Centrifuge 624R
6. Maintenance
6-1. Outer part of Instrument
1. Clean the outside of the instrument with dry soft cloth. If necessary, dip the cloth in neutral detergent
and clean contaminated area. Keep completely dry after cleaning.
2. Do not use any volatile chemicals such as alcohol and benzene, etc.
3. Be careful not to make scratches on the surface of the instrument. The scratches can cause corrosion
on the surface of the instrument.
If any rust appears, clean it with neutral detergents and keep dry.
6-2. Chamber
1. Keep dry inside the chamber after every use.
2. If the chamber is contaminated, dip the cloth in neutral detergent and clean contaminated area.
6-3. Shaft
1. Always make special attention to clean the motor shaft to avoid any imbalance problem due to the
contaminants.
2. After using the instrument, take out the rotor from the shaft, and clean the shaft with dry soft cloth to
keep dry.
6-4. Rotor
1. If any parts are contaminated with samples, clean the rotor with soft wet cloth and keep the rotor dry.
2. Be careful not to make scratches inside or on the surface of rotors. Any small scratches can cause
corrosion of the rotor and big damage to the instrument.
3. If you do not use the instrument, keep the rotor separately from the motor shaft and stand it upside
down.
6-5. Transportation of Instrument
1. If you need to move or ship the instrument, be cautious to protect the motor shaft from any physical
impact or turbulence.
2. Do not mount a rotor in any cases of movement. Fill inside the chamber with proper materials to keep
the motor shaft on place and not to be influenced by physical pressure.
Table of contents
Other GYROZEN Laboratory Equipment manuals

GYROZEN
GYROZEN 1696R User manual

GYROZEN
GYROZEN 1524 User manual

GYROZEN
GYROZEN 416 User manual

GYROZEN
GYROZEN 1730R User manual

GYROZEN
GYROZEN 1236R User manual

GYROZEN
GYROZEN 1580R User manual

GYROZEN
GYROZEN 406 User manual

GYROZEN
GYROZEN 1248R User manual

GYROZEN
GYROZEN MD406 User manual

GYROZEN
GYROZEN MicroCentrifuge mini User manual

GYROZEN
GYROZEN 416 User manual

GYROZEN
GYROZEN 1848R User manual

GYROZEN
GYROZEN 1730R User manual

GYROZEN
GYROZEN mini User manual

GYROZEN
GYROZEN 1096R User manual

GYROZEN
GYROZEN 1736R User manual

GYROZEN
GYROZEN 1248R User manual

GYROZEN
GYROZEN 2236R User manual

GYROZEN
GYROZEN 1580 User manual

GYROZEN
GYROZEN 2236R User manual