
BKI16ATEX0014X
dbf5512a0600p_05
15 / 20
5.1.3 Parameters – descriptions and programming
WARNING! During installation the zero point of the transmitter may be shifted. It is suggested to make a correction in P7
P0: - - - a Assignment of 4 mA to the measured (lowest) value
P1: - - - a Assignment of 20 mA to the measured (highest) value
The lowest and highest (limit) value of the (pressure or level) range is to be assigned to the 4 and 20 mA output current. This can be performed
by two methods
a: manually i.e. by entering the relevant values in P0 and P1 (Make sure that values to be programmed does not exceed the range of the unit in
question, the modification of the parameter value will not be accepted i.e. after pressing
E
the parameter value will remain blinking.)
The procedure for assignment of 4 mA to 0 bar in P0: press
E
+ to enter programming mode, go to P0 with and press
E
to go to
parameter value, enter 0 as lowest pressure value and press
E
to go back to the parameter address, finally press
E
+ to exit
programming/return to measurement mode).
b: automatically i.e. using the GET VALUE function at being in P0 and P1. This method for which the sensor should be put under the required
(zero and high limit value) pressure or the relevant level in the tank should be available, provides higher accuracy.
FACTORY DEFAULT: limit values of the (pressure) range of the unit in question
P5: - - - a Density [kg/m3 or lb/ft3]
Selectable between 600 and 2000 kg/m3
FACTORY DEFAULT: 1000 kg/m3
P6: - - - a Gravitational acceleration [m/s2 or ft/s2]
For calculation of the level the gravitational acceleration should be entered if it is substantially different as the default value.
FACTORY DEFAULT: 9.806 m/s2
P7: - - - a Zero point correction and offset
As mentioned above the zero point correction is advised to perform once the unit is installed already. This can be best done with using the GET
VALUE function in P7 having zero pressure on the sensor. (press
E
+ to enter programming mode, go to P7 with and press
E
to go
to parameter value, press + GET VALUE to make the unit recognise the actual 0 bar, press
E
to go back to the parameter address,
finally press
E
+ to exit programming/return to measurement mode provided no other programming is needed)
With this procedure shift of the zero point will be displayed (with its – or + sign) and corrected.