4/4
npk411en2006h
INSTALLATION EXAMPLE USING OVERVOLTAGE PROTECTION UNITS
4.1 SPECIAL CONDITIONS OF SAFE USE
Hydrostatic level transmitters with Ex ia IIC and Ex ia IIB protection type can only
be used in intrinsically safe loops, powered by a certified sparkless power supply,
in accordance with the specified technical data.
The metal housing of the device must be connected to the EP network using the
connection cable marked 1. The maximum cable length of the transmitter is 5 m
(16.4 ft).
5. PUTTING INTO OPERATION, CONFIGURATION
A properly installed and suspended probe needs some time to stabilize, which
depends on the cable's length. With a 300 m (985 ft) cable, stabilization might take
up to several hours, depending on the circumstances. The device is ready for
operation as soon as it is switched on, but it can only measure accurately after the
stabilization period.
If correction of insertion length is needed, loosen the cable holder sliding sleeve,
adjust the probe to the desired level, and fasten the cable holder sliding sleeve.
5.1 HART®OUTPUT
HART®-capable transmitters can communicate with NIVELCO's MultiCONT
universal process controller or through a HART®-USB modem with a PC using
standard HART®commands; the devices can be programmed remotely with
EView2 configuration software. The MultiCONT can power the transmitters, it
provides a remote programming possibility and the measured values can be
transmitted via RS485 communication line if needed. (See the details in the User's
and Programming manual of MultiCONT). The current loop output of the units can
be also configured with the EView2 software in the pressure range from 2% to
130%.
The damping time of the units can be also configured with the EView2 software or
with any HART®standard programming interface. The damping time is a time
constant of a time period. Its minimal value: 0 sec., maximal value: 99 sec.
5.2 PARAMETERS AND PROGRAMMING
P0:- - - a (lowest) Pressure value assigned to 4 mA
P1:- - - a (highest) Pressure value assigned to 20 mA
P0 and P1 (lowest and highest) pressure values can be assigned to the
4 mA and 20 mA current loop output values.
When changing the factory set values make sure that the entered values
fall within the specified range of the pressure transmitter otherwise the
device will indicate error.
FACTORY DEFAULT:
P0 = [minimum measurable pressure value of the sensor] mmH2O
(usually 0000)
P1 = [ maximum measurable pressure value of the sensor] mmH2O
(usually the possible max. value of the measurement range)
P5: Medium density [g/cm3]
Minimum value: 0.5 g/cm3. Maximum value: 2 g/cm3.
(If thermal compensation is active, it is always determined by the density
curve of water, regardless of the stored density value.)
FACTORY DEFAULT: 1 g/cm3.
P9: Current generator test (mA)
With this parameter the user can test the current loop output by entering a
value between 3.9 mA and 20.5 mA and test it with an ammeter.
Warning: the test mode can be cancelled only by entering 0000 to P9.
P10:- - - a Measuring mode
a Measuring mode
0 mbar Pressure
1 psi
2 mm H2O
Level (water head)
3 ft H2O
4 cm H2O
5 m H2O
FACTORY DEFAULT: 2
P10:- - b - Thermal compensation for water
Digit "b" of the P10 parameter toggles between normal and thermally
compensated mode.
The thermally compensated mode applies to water medium and is only
active for level measurement. In compensated mode, the current output,
and PV is the compensated level. SV is the temperature. TV is always an
uncompensated value, and QV is always a compensated value.
bThermal compensation
0Level measurement without
thermal compensation
1Level measurement with
thermal compensation
FACTORY DEFAULT: 0
P12:- - - a Error indication by the current loop output
aError indication
0< 3.9 mA
1> 21 mA
FACTORY DEFAULT: 0
P13: HART short address (Polling address)
If multiple HART®-capable transmitters are used in a loop, the instruments
are distinguished by their polling addresses. If the polling address is 0
(default), the current output is 4…20 mA, and HART®communication works
on the 4…20 mA current signal. Conforming to the HART®standard max.
15 HART®devices can be connected to a HART®loop with polling
addresses between 1 and 15. Thus, the output current will be set to 4 mA
and only the digital HART®communication will work. Instruments connected
to the same loop should not have same polling addresses or 0 polling
address set.
FACTORY DEFAULT: 0
P24: Damping Time
There is a possibility to setting the output damping time. Damping time is a
constant value for a time period. The minimum value is 0 seconds, the
maximum is 99 seconds.
FACTORY DEFAULT: 0
6. MAINTENANCE AND REPAIR
Sometimes the orifices of the NPK’s protective cap and they need cleaning to
remove surface deposits. Do not touch the sensor membrane when cleaning!
The membrane cap of threaded probes must be taken off before cleaning, and the
filter behind it must also be removed. Do not touch the sensor membrane when
cleaning!
7. STORAGE CONDITIONS
Storage temperature: −10…+50 °C (+14…+122 °F)
Relative humidity: max. 85%
npk411en2006h
November 2020
NIVELCO reserves the right to change technical specifications without notice.