’08.11.05 S
Caution
• Never clean the handpiece (EX-5B, EX-6B, EX-6D) in boiling water.
• Always place the handpiece in the center or upper shelf of the
chamber, as the local temperature at the bottom of chamber could
rise beyond the set value.
Caution
• Be sure to hold the handpiece firmly to prevent the handpiece from
slipping out by the spray pressure when lubricating.
• Supply lubricant until it comes out of the handpiece nose (for
approx. 2 seconds).
• Keep the PANA-SPRAY Plus can upright.
Tips for safe use and for maximum service life
• Never use bent, off-centered or damaged burs. Excessive loads
generated by defective burs rotating at 40,000min-1(rpm) will quickly
damage ball bearings.
• This handpiece is designed for (dental) treatment and precision
technical use. Longtime excavation with high torque with thick burs
such as 4mm or greater in diameter may cause a short life to the
bur, hard-to-pull-off bur or bur drop-off, which results in injury.
• Always insert a bur all the way into chuck. This will prevent “BUR-
WALK-OUT”, and/or damage to the ball bearing.
• Never operate Straight Nosecone without a bur or bur blank
securely in place, or bur lock ring at different positions.
• Make sure if the Os of the bur lock ring line up after fixing the bur.
Rotating the bur in an improper position may cause heat generation,
which results in burns.
(1) Cleaning
Wipe clean the handpiece sheath with alcohol-dipped soft tissue or
cotton. Never clean the handpiece in chemicals, with wire brushes, or
in ultrasonic cleaners.
(2) Sterilization
Steam autoclave is recommended.
Sterilization is required first time you use and after each patient.
Autoclaving procedure
Clean the handpiece sheath as described above.
Lubricate the handpiece with PANA-SPRAY Plus. (Ref. 4. Lubrication)
Place handpiece into autoclaving pouch and seal it in accordance with
instructions on the pouch.
Autoclavable up to max. 135˚C (275˚F).
ex.) Autoclave for 20 min. at 121˚C, or 15 min. at 132˚C.
Keep the handpiece in the autoclave pouch to keep it clean until you
use it.
Sterilization at 121˚C for more than 15 minutes is recommended by
EN13060 or EN554.
Care3 Plus
NSK Care3 Plus automatic handpiece lubrication system.
After cleaning the handpiece simply connect the handpiece to the
correct adaptor and activate the Care3 System per the Care3 System
instructions.
5. Cleaning and Sterilization
Supply PANA-SPRAY Plus and/or Care3 Plus after each use and/or before
autoclaving.
PANA-SPRAY Plus
Push E-type spray nozzle attachment over the PANA-SPRAY Plus
nozzle until it firmly seats.
Shake the can 3-4 times to well mix lubricant and propellant.
Insert the E-type spray nozzle in the rear of the handpiece and spray for
approximately 2-3 seconds until the oil comes out of the handpiece
head or nose.
4. Lubrication
To mount the bur, insert it into the chuck and turn the Bur Lock Ring in the
direction of the mark [C] until it clicks and the Black Dots align (*See the
picture below). To remove the bur, turn the Bur Lock Ring in the direction
of mark [O] until it clicks.
3. Mounting Burs
Fig.2
Fig.3
PANA-SPRAY PlusPANA-SPRAY Plus
E-type Spray Nozzle Handpiece
The products listed below are washable with Thermo Disinfector.
EX-5BM, EX-6BM, EX-6DM
Align
Bur Lock Ring
Black Dot
Close
Open
1
2
3
1
2
3
4
5