Features
• Large easy-to-read LCD with 3-color backlight
• Replaceable single-junction sensor saves you money
• Stability and battery status icons
• Auto power-off function conserves battery life
• IP67 waterproof rating
Operating Manual
Thank you for purchasing the Oakton® ORPTestr® 5 Pocket Tester.
Please carefully read this operating manual before using the product
to obtain an accurate and reliable test result and avoid unnecessary
damage to the tester or probe.
ORP Calibration
1. Clean electrode in distilled water and dry it.
2. Pour a small amount of 222 mV ORP stan-
dard solution (SKU 80055-90, 500-mL bottle)
into the calibration bottle.
3. Short press /MEAS key to turn the meter
on. Long press CAL/ key to enter calibra-
tion mode. If you need to exit calibration
mode, short press /MEAS key.
4. Dip electrode in the ORP standard solution
(Zobell’s solution), stir gently, and allow it
to stand still in the calibration solution until
a stable reading is reached. When measure-
ment stability icon () appears and remains
on the screen (see Fig. 1), press MODE/
key to adjust the calibration value according
to the table below (also on the buffer bottle’s label).
ORP Standard Buffer Solution Calibration Reference Table
Temperature (ºC) mV Temperature (ºC) mV
10 242 30 215
15 235 35 209
20 227 38 205
25 222 40 201
For example, if the temperature is around 25ºC, then adjust the cali-
bration value to 222 mV. If the temperature is around 20ºC, then
adjust the calibration value to 227 mV. Short press CAL/ key to
complete the calibration.
ORP Measurement
Press /MEAS key to power on, clean the probe in distilled water,
and gently dry it. Place the probe in the tested solution and stir
gently and let it stand still. Read the measurement after the stability
icon () appears and remains on.
Notes
Usually there’s no need to calibrate for ORP testing. When not used
for a long time, users can test the unit in 222 mV ORP standard buf-
fer. If the error is large, recalibrate according to the steps above.
Cleaning and Activating the ORP Probes
After long-time use, the surface of the platinum sensor can be con-
taminated and cause inaccuracy and slow response time. If that hap-
pens, the followed methods can be adopted to clean and activate:
A. For contamination caused by inorganic substances, dip the
probe into 0.1 mol/L dilute hydrochloric acid for 30 minutes.
Clean it with distilled water, soak in KCL storage solution
(SKU 00653-04) for 6 hours, and then it will be ready for use.
B. For contamination caused by organic substances and oiliness,
clean the platinum surface with dishwashing liquid and wash
with distilled water. Soak for 6 hours in storage solution
(SKU 00653-04) and then it will be ready for use.
C. If the platinum surface is so severely contaminated that an
oxide lm forms, use toothpaste to clean the platinum surface,
and then wash it with distilled water. The probe will be ready
for use again after 6 hours of soaking in the storage solution
(SKU 00653-04).
Setting the Parameters
When tester is off, long press /MEAS key to enter parameter setting
mode. Short press MODE/ key to switch from P1 to P2…P6. Short
press CAL/ key and parameter will ash, then short press MODE/
key to choose desired parameter. Short press CAL/ key to conrm
selection. Long press /MEAS key to exit parameter setting mode.
Symbol Menu setting Selection Factory default
P1 Set low measurement alarm –1000 to +1000 mV –1000
P2 Set high measurement alarm –1000 to +1000 mV 1000
P3 Select Automatic Lock (HOLD) Off – On Off
P4 Select backlight Off – 1 – On 1 (1-min auto-off)
P5 Select temperature unit °C – °F °C
P6 Restore to factory default No – Yes No
A. High and Low Measurement Heads-Up Examples (P1 and P2)
Heads-up when measured value ≤120 mV: Preset low value = 120
mV, highest value = 1000 mV, when measured value is ≤120 mV (sta-
bility icon () displays on LCD) the red backlight appears on display.
Heads-up when measured value ≥150 mV: Preset high value = 150
mV, lowest value = –1000 mV, when measured value is ≥150 mV (sta-
bility icon () displays on LCD) the red backlight appears on display.
Heads-up when measured value in range of 120 mV and 150 mV:
Preset low value = 120 mV, highest value = 150 mV, when the
measured value is <120 mV or >150 mV (stability icon () displays
on LCD) the red backlight appears on display.
B. Automatic Lock (P3)
Select “On” to activate auto lock function. When reading is stable for
more than 10 seconds, the tester will lock the value automatically,
and “HOLD” icon will appear on the bottom left of the screen. Press
CAL/ key to cancel HOLD on reading.
C. Backlight (P4)
Select “Off” to turn off backlight function, “On” to turn on backlight
function, or “1” to have backlight last for 1 minute.
D. Temperature Unit (P5)
Select °C or °F; the factory default is °C.
E. Factory Default Setting (P6)
Select “Yes” to recover tester calibration to theoretical value (pH
value in zero potential is 7.00, slope is 100%), parameter setting
return to initial value. This function can be used when instrument
does not work well in calibration or measurement. Calibrate and
measure again after resetting the unit to factory default status.
220
25.8
mV
ORP
☺
☺
°c
Figure 1
Keypad Functions
Short press = <2 seconds Long press = >2 seconds
MEAS
1. Short press to turn on the tester and long press to
turn off the tester.
2. When turned off, long press to enter parameter setting mode.
3. In measurement mode, short press to turn on backlight.
MODE

1. In calibration mode, short press to change calibration
value (unidirectional).
2. In mode setting, short press to change parameter
(unidirectional).
CAL
<
1. Long press to enter calibration mode.
2. In calibration mode, short press to confirm calibration.
3. When measured value is locked (HOLD icon), short press
to unlock.
Preparation Before Use
1. Pour certain amount of storage solution (KCl) into the probe
cap, soak the probe for 3 to 5 minutes to activate the sensor
(if frequently used, there is no need to soak).
2. The soaking solution is 3M KCl solution; preferably Oakton
Electrode Storage Solution (SKU 00653-04). Users can also make
it themselves by dissolving 25 g KCl in 100 ml of distilled water.
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
1065O101_MAN_ORP5_35634-54 March 2018
Sensor Replacement
Screw off the sensor ring, unplug the sensor, plug in the new
replacement sensor (pay attention to the probe’s position), and
rescrew on the sensor ring
Battery Installation
The tester uses four AAA batteries. Please install batteries according to
the following steps. Note the correct direction of battery installation: the
positive side (+) of every single battery must face up. Incorrect installa-
tion of batteries will cause damage to the tester and create a potential
hazard.
1. Pull the battery cap up.
2. Slide the battery cap along the direction of arrow.
3. Open the battery cap.
4. Insert the batteries (ALL POSITIVE SIDES FACING UP).
5. Close the battery cap.
6. Slide and lock the battery cap along the direction of arrow.
7. Fit the tester’s cap while making sure to push all the way
down. The tester’s waterproof design may be compromised if
the cap is not tted correctly.
4 x AAA Batteries
+
+
+
+
__
__
1
2
3
4
5
6
7
Warranty
We warrant this instrument to be free from defects in material and
workmanship and agrees to repair or replace free of charge, at
option of Oakton Instruments, any malfunctioned or damaged
product attributable to responsibility of Oakton Instruments, for a
period of two years from the delivery (a six-month limited warranty
applies to sensors). This warranty does not apply to defects resulting
from actions such as misuse (violation of the instructions in this
manual or operations in the manner not specied in this manual),
improper maintenance, and unauthorized repairs. Warranty period is
the time limit to provide free service for the products purchased by
customers, not the service life of the tester or probe.
Oakton Instruments reserves the right to update the information in
this manual without giving notice in advance.
Self-Diagnostic Messages
Symbol Self-diagnosis
information How to fix
Er2 CAL/is pressed
before measurement
is stable (appears)
Wait for the measurement stability
icon () to appear and stay, then
press CAL/.
Specifications
ORP (mV)
Range ±1000 mV
Resolution 1 mV
Accuracy ±0.2% full-scale
Temperature
Range 32 to 122°F (0 to 50°C)
Resolution 0.1°F/°C
Accuracy ±0.9°F (0.5°C)
Display: LCD with three-color backlight. Blue = measurement;
Green = calibration; Red = alarm
Reading lock: HOLD icon
Power: four AAA batteries (included); >400 hours of continuous
operation
Low-voltage warning: battery status icon ashes
Auto power-off: tester automatically turns off after 8 minutes
of nonuse
IP rating: IP67 (waterproof), oats on water when sensor cap is on
Dimensions (L x W x H): 7" x 1.5" x 1. 5" (17.8 x 4 x 4 cm)
Weight: 4.7 oz (133 g)
Ordering Information
Model Product description Catalog number
ORP5 ORPTestr 5 pocket tester 35634-54
— Replacement sensor for ORPTestr 5 35634-58
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

This manual suits for next models

1

Other Oakton Measuring Instrument manuals

Oakton PH550 User manual

Oakton

Oakton PH550 User manual

Oakton PC 700 User manual

Oakton

Oakton PC 700 User manual

Oakton CON 10 User manual

Oakton

Oakton CON 10 User manual

Oakton CON550 User manual

Oakton

Oakton CON550 User manual

Oakton pHTestr 50 User manual

Oakton

Oakton pHTestr 50 User manual

Oakton PD 300 User manual

Oakton

Oakton PD 300 User manual

Oakton 450 Series User manual

Oakton

Oakton 450 Series User manual

Oakton PCD 650 User manual

Oakton

Oakton PCD 650 User manual

Oakton EcoTestr pH1 User manual

Oakton

Oakton EcoTestr pH1 User manual

Oakton WD-35427-10 User manual

Oakton

Oakton WD-35427-10 User manual

Oakton T-100IR User manual

Oakton

Oakton T-100IR User manual

Oakton CON 700 User manual

Oakton

Oakton CON 700 User manual

Oakton pH 310 User manual

Oakton

Oakton pH 310 User manual

Oakton pH 610 User manual

Oakton

Oakton pH 610 User manual

Oakton 300 Series User manual

Oakton

Oakton 300 Series User manual

Oakton 35700-00 User manual

Oakton

Oakton 35700-00 User manual

Oakton 35612-series User manual

Oakton

Oakton 35612-series User manual

Oakton 150 Series User manual

Oakton

Oakton 150 Series User manual

Oakton CON 400 User manual

Oakton

Oakton CON 400 User manual

Oakton PH200 User manual

Oakton

Oakton PH200 User manual

Oakton Salt 6 User manual

Oakton

Oakton Salt 6 User manual

Oakton 35616-00 User manual

Oakton

Oakton 35616-00 User manual

Oakton ION 6 User manual

Oakton

Oakton ION 6 User manual

Oakton 35631-00 User manual

Oakton

Oakton 35631-00 User manual

Popular Measuring Instrument manuals by other brands

Doris Schmithals Messwerkzeuge WP 201 066ECO user manual

Doris Schmithals Messwerkzeuge

Doris Schmithals Messwerkzeuge WP 201 066ECO user manual

CEMB N200 Use and maintenance instruction manual

CEMB

CEMB N200 Use and maintenance instruction manual

finder 7M.38.8.400.0112 user manual

finder

finder 7M.38.8.400.0112 user manual

BASETech BTTH-1014 operating instructions

BASETech

BASETech BTTH-1014 operating instructions

Michell Instruments Easidew Online user manual

Michell Instruments

Michell Instruments Easidew Online user manual

Brouwland pH110 manual

Brouwland

Brouwland pH110 manual

Tractel dynafor LLZS Series Operating and maintenance instructions

Tractel

Tractel dynafor LLZS Series Operating and maintenance instructions

Diablo 5000B Getting started manual

Diablo

Diablo 5000B Getting started manual

Nissei TX-1202 user manual

Nissei

Nissei TX-1202 user manual

DOLD IP 5883 Installation / monitoring technique

DOLD

DOLD IP 5883 Installation / monitoring technique

twintex TI512 Operation manual

twintex

twintex TI512 Operation manual

DataCan 109684 user manual

DataCan

DataCan 109684 user manual

UEi C75 instruction manual

UEi

UEi C75 instruction manual

Ecotech Serinus 30 instruction manual

Ecotech

Ecotech Serinus 30 instruction manual

PROPOINT 8319683 user manual

PROPOINT

PROPOINT 8319683 user manual

ProAir FGP Series manual

ProAir

ProAir FGP Series manual

Delta-T Devices WET150 Kit quick start guide

Delta-T Devices

Delta-T Devices WET150 Kit quick start guide

red lion PAXLPT manual

red lion

red lion PAXLPT manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.