manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Olympus
  6. •
  7. Measuring Instrument
  8. •
  9. Olympus OMCL-AC160TN-C3 User manual

Olympus OMCL-AC160TN-C3 User manual

user’s manual 書
Cantilever
Case label
Products name Cantilever chip
(substrate)
Thank you for purchasing OLYMPUS Micro cantilever.
Please read this manual carefully before use.
- 1e -
Please obey the following to the OLYMPUS micro cantilevers
Warning
Use protective eye glasses when handling to avoid
damage to the eyes from breakage of the cantilever
chips.
Caution
Please handle our cantilevers carefully because they are fragile.
Caution
Do not drop or shake the cantilever case. Even when the cantilever
chips are contained in the cantilever case, the cantilevers may break
if the case is handled roughly or jarred.
Caution
It is recommended that precautions be taken to prevent damage to the
cantilever tips from electrostatic discharge.
Caution
When discarding, please obey the laws and regulations in your
country and/or your company. These cantilever are made from
silicon.
- 2e -
Case (Tac carrier) Adhesive sheet Cantilever chips
Upside
Down side
Tape
Case (closed) Case (open)
Cantilevers are tip-side-up
as viewed in the case.
One cantilever is elongated from the
side of round shoulders of the chip.
Case Label
Manual and Spec sheet
Plastic pack (blue colored)
Plastic case
(Cantilevers are in the case)
Cantilever chips in a plastic case are included in the envelope with manual and spec sheet.
<Explanation of the each part of the products>
Cantilever chips are contained in the plastic case (Tac carrier)
OMCL-AC160TN-C3
300
26
Magnified illust.
Lot Number
user’s manual
Label Inside
Envelope for transportation
Inspection stamp
Special feature of OLYMPUS Micro cantilever
(OMCL-
A
C160TN-C3)
- 3e -
1 Preparation
1) Please prepare the followings before using OLYMPUS cantilevers.
2) To gain a better understanding of how cantilevers and chips are connected, cantilevers
1. AC mode cantilever – This type of micro cantilever is for AC (dynamic) mode AFM operation.
should be inspected under the microscope.
High mechanical Q factor of the cantilevers allows you high sensitivity measurement to
1) Work environment
- 4e -
probe your sample surface gently.
3. Tip View – The tip is located on the very end of the cantilever.
This feature allows you to set the tip over a point of interest on the sample, easily and
precisely, if you use an AFM combined with an optical microscope.
4. Point terminated tip – The tip radius is typically 7nm
A sharpened tetrahedral tip of single crystal silicon is employed for high-resolution
measurements. The tetrahedral shape is ideal for achieving a point terminated tip. In
addition to the geometrical dimensions of the tip, the tip is further sharpened with our
exclusive sharpening process.
5. Symmetric tip – The tetrahedral tip is thin with the macroscopic tip angle of 35 degrees
viewing from both its front and side. Tip shows good symmetry in viewing along the
cantilever axis (see Chapter 5).
Symmetric shape tip results AFM images less distortion which reduce misinterpret of
the data.
6. No metal reflex coating
2. Pre-separated chip – Each cantilever chips is isolated in the case.
The chips can be attached to the AFM instrument as soon as the case is open.
7. Compatible chip – The thickness of the chip (substrate) is 0.3mm.
This type of the chip can be attached to most of AFM instruments.
2) For hazard avoidance
3) For cantilever treatment
4) For inspection
2 Open the case
Table of contents
1 Preparation 4e
2 Open the case
3 Picking up the cantilever chip
4 Tip shape of Tetrahedral tip
5 Information
7
6
8 Specification 1s
page
Please handle our cantilevers carefully because they are fragile.
Caution
It is recommended that precautions be taken to prevent damage to the
:Clean bench
(Use of an electrical charge neutralizer of ionizer is recommended.)
:Protective eye glasses
:Tweezers
(Use of anti-electrostatic discharge mat and a wrist band is
recommended.)
:Stereo microscope
cantilever tips from electrostatic discharge.
Caution
1) It is recommended that the cantilever case be opened in a clean environment like a clean bench
in order to avoid the cantilever being contaminated. Handling under an ionizer is recommended.
2) Avoid wearing clothes like woolen sweaters, fleece etc that give off the static electricity when
See the specification sheet of OLYMPUS Micro cantilevers at the last page of this manual. handling the cantilever cases and chips. Use of an anti-electrostatic mat and wrist band is
preferable.
3) In opening the case, put the plastic case label-side down on a desk.
The cantilevers are tip-side-up as viewed in the case.
4) Open the case
3 Picking up the cantilever chip from the case
4e
4e
5e
Avoid any contact with the cantilevers when you pull up the
cantilever tip from the case.
Caution
6e
1) Pick up the chip by the long side with the tweezers and mount it in the AFM.
Ver.1.1 July. 3, 2009
OLYMPUS CORPORATION
The apex of the tetrahedral tip becomes sharper due to an oxide sharpening process. The
tip angle around a few hundreds nano meter down from the apex, is about 15 to 25
degrees (see below).
- 5e -
- 6e -
4 Tip shape of Tetrahedral tip 5 Information
Please contact following if you have any question on this user’s manual.
OLYMPUS CORPORATION
Microtechnology R&D Division
2-3 Kuboyama-cho Hachioji-shi Tokyo 192-8512 Japan
email : [email protected]
Please access to the web page of OLYMPUS micro cantilevers.
http://www.olympus.co.jp/probe
As can be seen in the left illustration, a
tetrahedral tip is located at the exact end of
the cantilever.
The finite tip shape will determine the scan
line profile as in the illustrations below.
The tip profile is symmetric with a half tip
angle of 18 degrees macroscopically (see left
below).
The side tip profile is asymmetric with a tip
angle of 35 degrees. Then the cantilever chip
is attached to a chip holder in your AFM with
an angle, about 10 degrees, the asymmetry is
improved (see right below).
Tip shape and scan line profile

Other Olympus Measuring Instrument manuals

Olympus OMCL-AC240TN-R3 User manual

Olympus

Olympus OMCL-AC240TN-R3 User manual

Olympus OMCL-AC240TS-C3 User manual

Olympus

Olympus OMCL-AC240TS-C3 User manual

Olympus 27MG User manual

Olympus

Olympus 27MG User manual

Olympus USPM-RU-W Series User manual

Olympus

Olympus USPM-RU-W Series User manual

Olympus ORBEYE OME-V200 User manual

Olympus

Olympus ORBEYE OME-V200 User manual

Olympus OMCL-AC240TS-R3 User manual

Olympus

Olympus OMCL-AC240TS-R3 User manual

Olympus SOLTIVE User manual

Olympus

Olympus SOLTIVE User manual

Olympus BondMaster 600 User manual

Olympus

Olympus BondMaster 600 User manual

Olympus USPM-RU-W Series User manual

Olympus

Olympus USPM-RU-W Series User manual

Olympus TERRA II User manual

Olympus

Olympus TERRA II User manual

Olympus USPM-RUIII Series User manual

Olympus

Olympus USPM-RUIII Series User manual

Olympus DS-3000 User manual

Olympus

Olympus DS-3000 User manual

Olympus 45MG User manual

Olympus

Olympus 45MG User manual

Olympus Vanta L Series User manual

Olympus

Olympus Vanta L Series User manual

Olympus BIX III User manual

Olympus

Olympus BIX III User manual

Olympus DELTA 50 User manual

Olympus

Olympus DELTA 50 User manual

Olympus CIX100 User manual

Olympus

Olympus CIX100 User manual

Olympus STM7-AF User manual

Olympus

Olympus STM7-AF User manual

Olympus 45MG User manual

Olympus

Olympus 45MG User manual

Olympus WS 100 - 64 MB Digital Voice Recorder User manual

Olympus

Olympus WS 100 - 64 MB Digital Voice Recorder User manual

Olympus BTX III User manual

Olympus

Olympus BTX III User manual

Olympus VN-850PC User manual

Olympus

Olympus VN-850PC User manual

Olympus 27MG User manual

Olympus

Olympus 27MG User manual

Olympus 38DL PLUS User manual

Olympus

Olympus 38DL PLUS User manual

Popular Measuring Instrument manuals by other brands

Murphy SELECTRONIC 1400 Series Installation and operation manual

Murphy

Murphy SELECTRONIC 1400 Series Installation and operation manual

ELNet LT Installation & operation manual

ELNet

ELNet LT Installation & operation manual

Konica Minolta CM-23d instruction manual

Konica Minolta

Konica Minolta CM-23d instruction manual

Satec BFM II EM Operation and installation manual

Satec

Satec BFM II EM Operation and installation manual

AII GPR-1500 DA owner's manual

AII

AII GPR-1500 DA owner's manual

Minebea CSD-892B-73 instruction manual

Minebea

Minebea CSD-892B-73 instruction manual

Dinel ULM-53 series instruction manual

Dinel

Dinel ULM-53 series instruction manual

3M PL300 Quick reference guide

3M

3M PL300 Quick reference guide

Parkside PKLL 7 A1 operating instructions

Parkside

Parkside PKLL 7 A1 operating instructions

Elenco Electronics XK-700 instruction manual

Elenco Electronics

Elenco Electronics XK-700 instruction manual

Fieldpiece HS26 Operator's manual

Fieldpiece

Fieldpiece HS26 Operator's manual

PCB Piezotronics 317A61 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 317A61 Installation and operating manual

Siemens ULTRAMAT 23 operating instructions

Siemens

Siemens ULTRAMAT 23 operating instructions

Dickey-John GAC500XT Operator's manual

Dickey-John

Dickey-John GAC500XT Operator's manual

Seneca S203TA-D user manual

Seneca

Seneca S203TA-D user manual

TESTING 1.0235 operating manual

TESTING

TESTING 1.0235 operating manual

YOKOGAWA 2041 user manual

YOKOGAWA

YOKOGAWA 2041 user manual

Pogo Pro Start here

Pogo

Pogo Pro Start here

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.