Olympus CK40-TBI Operating instructions

f
TILTING BINOCULAR TUBE
CK40-TBI
REPAIR MANUAL
OLYMPUS
All rights reserved. Reproduction in whole or in part without written permission is prohibited. (8H)
f-‘

CK40-TBI
CONTENTS
A. OUTLINE OF PRODUCT
1. Outline ............................................................................................................................. A-l
2. Features........................................................................................................................... A-l
3. Using conditions .............................................................................................................. A-l
4. Specifications .................................................................................................................. A-l
5. Dimensions ...................................................................................................................... A-2
B. INSPECTION STANDARD
1. Inspection items and methods .......................................................................~................. B-l
C. DISASSEMBLY AND ASSEMBLY PROCEDURES
1. TBIC unit ........................................................................................................................ C-l
2. BI unit .............................................................................................................................. c-3
D. REPAlR PROCEDURE
1. Adjustment procedure ..................................................................................................... D-l
2. TBIC unit repair procedure ............................................................................................. D-2
3. Preparation for adjustment .............................................................................................. D-6
4. Adjusting revolving axis ................................................................................................. D-7
5. Adjusting left/right optical axis....................................................................................... D-12
6. Adjusting trapezoidal prism ............................................................................................ D-15
7. Adjusting absolute optical axis (center of standard objective) ....................................... D-19
8. Checking exit pupil center .............................................................................................. D-22
9. Adjusting parfocality ....................................................................................................... D-23
10. Adjusting tube ass’y ....................................................................................................... D-27
E. JIGS AND TOOLS
1. List of jigs and tools ........................................................................................................ E-l
F. LUBRICANTS AND CHEMICALS
1. List of lubricants ............................................................................................................. F-l
2. List of chemicals ............................................................................................................. F-l

CK40-TBI A. OUTLINE OF PRODUCT
1. Outline
This product is a tilting binocular tube to be usedin combination with a CK 40 culture microscope
for observation in the standing position.
2. Features
(1) Since the eyepoint height can be changed from 390 mm to 465 mm (observation tube tilt
angle 30” - 60”) the best eyepoint height can be setfor individual users.
(2) At an observation tube tilt angle of 60” (eye point height: 465 mm), a standing user can
observe a specimen through the microscope in a comfortable posture.
(3) At an observation tube tilt angle of 45” (eye point height: 430 mm), the sameeye point as
IX canbe obtained.
f‘
3. Using conditions
(1) A CK40-TBI can be combined a CK40 microscope only. It cannot be combined with any
other model.
(2) In somecases,an eyepieceof field number 20 causesinsufficient peripheral light intensity
during 10X observation.
(3) A CK40-TBI cannot be combined with CK40-EPA.
(4) An infinity correction tube cannot be combined with any other UIS observation tube.
(5) In somecases,a specimenof high reflectance or high contrast produces aghost.
4. Specifications
tube can be stopped at any angle within
9 Left/right diopter The left sleeve has a helicoid for diopter.
difference Adjustment range: r5 diopter
10 Mounting on microscope Observation tube: Circular mounting dovetail fitting $34; height 6 mm
Infinity correction tube: Drop fitting $30z,0j00 [mm]
11 Weight Observation tube: 1.1 kg
Infinity correction tube: 70 g
A-l

CK40-TBI A. OUTLINE OF PRODUCT
5. Dimensions
CK40-TBI
t
C
C.
I
v
179 (at interpupillary distance of 62 mm)
-4
Infinity correction tube
Observation tube Unit: mm
A-2

CK40-TB I B. INSPECTION STANDARD
c-
1. Inspection items and methods
(f- ?
P
Part Item Standard Method
H unit Interpupillary 50 mm to 76 mm In the observation state, insert a thin sheet of paper
Binocular distance with graduations at the eyepoint position and measure
ube adjustment the interpupillary distance.
ection) range
Interpupillary 1000 - 2000 g Put a string on the sleeve and measure the working
distance (9800 mN - 19600 mN) force to actuate the interpupillary distance using a
working force tension gauge.
Revolving axis 0.2 mm or less Using the standard eyepiece (KNO048 with Adapter-l)
center and a 4X or 10X objective, observe a specimen whose
center can be identified (ex. concentric circles) on the
stage.
Taking the right sleeve asthe reference, match the
specimen center and the visual field center, open and
close the interpupillary distance to read the
displacement between the specimen center and the
visual field center in the right sleeveusing the reticle
scale (1 graduation=O.l mm) of the KNO048.
Left/right
optical axis On image surface: Using the standard eyepiece (KNO048 with Adapter-l)
0.2 mm or less in vertical and a 4X or 10X objective, observe a specimen whose
direction center can be identified (ex. concentric circles) on the
0.2 mm or less in outward stage.
direction Taking the right sleeve asthe reference, match the
0.4 mm or less in inward specimen center and the visual field center and read
direction the displacement between the specimen center and the
visual field center in the left sleeveusing the reticle
scale (1 graduation=O.l mm) of the KNO048.
Absolute optical 0.1 mm or less on the Combining the standard eyepiece (RN0048 with
axis (right image surface Adapter-l), a microscope stand (product), the standard
sleeveonly) objective (KNOO15), read the displacement between the
specimen center and the visual field center in the right
sleeveusing the reticle scale (1 graduation=O.l mm) of
the RN0048 (with adapter-l).
Exit pupil
center Within 20% of objective’s Combining the centering telescope (KNOO29), a 4X or
exit pupil diameter 10X objective, and a microscope stand (product), read
the displacement between the exit pupil center of the
objective and the cross hairs center of centering
telescope.
Left/right
parfocality
difference
0.1 mm or less Combine the standard eyepiece (KNOO48; with
Adapter-l), the focusing telescope @T-36), a
microscope stand (product), and the standard objective
(KNO016).
Operating the helicoid of the RN0048 (with Adapter-l)
separately in the left and right optical paths, focus on
the specimen in the RN0016 and read the parfocality
difference using the helicoid scale (1 graduation=O.l
mm) of the RN0048 (with Adapter-l).
B-i

CK40-TBI B. INSPECTION STANDARD
Part Item Standard Method
BI unit Difference 1.5 mm or less Set the diopter at “0.” Put a rigid Hat plate acrossthe
CBinocular between left and left and right tube surface, insert a 1 mm thickness
tllbe
right tube gauge between the flat plate and the tube surface. If
section) length the clearance is greater than 1 mm, the unit is out of
the specification.
Absolute 0.2 mm or less on the
parfocality image surface Combine the standard eyepiece KNO048 (with
Adapter-l), the focusing telescope @T-36), a
microscope stand (product), and the standard objective
(KNO016). Set the interpupillary distance at 62 mm,
read the parfocality difference between the reticle scale
of KNO048 (with Adapter-l) and the specimen of
KNO016 in right sleeve using the helicoid scale (1
graduation=O.l mm) of the KNO048.
TBIC unit Optical axis 0.5 mm or less on the In the observation state, observe a specimen whose
(Tilting displacement by image surface when the center can be identified (ex. concentric circles) on the
section) tilting interpupillary distance is stageusing the KNO048 with Adapter-l. Match the
62 mm specimen center and the visual field center. While
tilting, read the displacement between’the specimen
center and the visual field center using the reticle scale
(1 graduation=O.l mm) of the KNO048 with Adapter-l.
Working force up-Down400-9oog Using a tension gauge, measure the working force after
(3920 mN - 8820 mN) mounting the TBIC unit on the microscope stand
Down - Up 200 - 700 g (CK40).
(1960 mN - 6860 mN)
B-2

CK40-TBI C. DISASSEMBLY AND ASSEMBLY PROCEDURES
1. TBIC unit
i$-%!J$
J
tff?a
y-y&f @
0
1 @*
A
“;
ED
2 0
/ ’
@?!f2?/-
Lx f
&(g .“i”
P if 8
0
+3il @3 Q
0 c&?-i’@\
7
*6 1
*6

CK40-TBI C. DISASSEMBLY AND ASSEMBLY PROCEDURES
‘2. TBIC unit repair
ly adhesiveto the screw head
c-2

CK40-TBI C. DISASSEMBLY AND ASSEMBLY PROCEDURES
BI unit
al-I @
0
@
707
\ \-
c-3

CK40-TBI C. DISASSEMBLY AND ASSEMBLY PROCEDURES
vo.
Parts name Screw Grease Adhesive Remarks
0 CAP (2 PCS.)
DIOPTER RING ACU3X3SA (*l) 3 PCS. Just enough to loosen.
Refer to “9. Adjusting parfocality”
on page D-23.
ADJ. RING
INSIDE EYE
TUBE
0 SLEEVER
@ COLUMNED
PRISM
ACU3X3SA (*2)
CUK3XGSB (*3) 2 PCS.
Refer to “9. Adjusting parfocality”
on page D-23.
Just enough to loosen.
Refer to “4. Adjusting revolving
axis” on page D-7.
0 DIAPHRAGM
@ HELICOID ASS’Y CUK3X6SB (*4) 2 pcs.
@ COVERL 3PUTS2X3SB (*5) 4 PCS.
@ COVERR 3PUTS2X3SB (*6) 4 PCS.
@ PRISML ABSK3X8SA (*7) 2 PCS.
Refer to “5. Adjusting left/right
optical axis” on page D-12.
Refer to “8. Checking exit pupil
center” on page D-22.
@ DIAPHRAGM
COVER
Push it in the arrow directions (C-3).
@ DIAPHRAGM
SHEET
@ PRIMSR ABSK3X8SA (*8) 2 PCS. Refer to “8. Checking exit pupil
center” on page D-22.
Push it in the arrow directions (C-3).
@ ADJ. RING ACU3X3SA (*9) Just enough to loosen.
@ SPRING WASHER Set it in correct direction.
(2 PCS.)
@ WASHER
@ PRISM MOUNT L 3PUTS2X3S.B (*lo) OT1595
@ WASHER OT1595
PRISM MOUNT R 3PUTS2X3SB (*ll) OT1595
@ WASHER OT1595
@ SHAIT OT1595 Refer to “7. Adjusting absolute
optical axis” on page D-19.
c-4

CK40-TBI D. REPAIR PROCEDURE
1. Adjustment procedure
2. TBIC unit repair
3. Preparation for adjustment
(Refer to D-6.)
4. Adjusting revolving axis
(Refer to D-7.)
5. Adjusting left/right optical axis
(Refer to D-l 2.)
6. Adjusting trapezoidal prism
(Refer to D-l 5.)
7. Adjusting absolute optical axis
(Refer to D-l 9.)
8. Checking exit pupil center
(Refer to D-22.)
9. Adjusting patfocality
(Refer to D-23.)
10. Adjusting tube ass’y
(Refer to D-27.)
D-l

CK40-TBI D. REPAIR PROCEDURE
2. TBIC unit repair procedure
2-1 Gear and balance-spring repair procedure (Disassembly procedure)
upperGEAR 0 andGEARFIXING PLATE 0
X Don’t dismountthelower GEAR 0 andGEAR
FIXING PLATE @in thisstepbecausethe
BALANCE-SPRING 0 mayfly out.
x Usingthe SPRINGFIXING JIG 0 asalever,
removetheBALANCE-SPRING 0 from the
hole. Thejig settingpositionis shownin the
A. Fulcrum: Contactbetweenthecurvepart of
thejig andtheprism2 ass’y
B. Powerpoint: Handleof thejig
C. Pointof application: Tip of thejig that is
hookedaroundtheprotruding part of the
BALANCE-SPRING 0 and
Formaclearance.
BALANCE-SPRING 0 andPRISM2 ASS’Y
0 andhookthetip of thejig aroundthe
protruding part of BALANCE-SPRING 0.
D-2

CK40-TBI D. REPAIR PROCEDURE
f
Step5) 5) LeavingtheSPRINGFIXING JIG 0 hooked
Igr Turnthejig
slightlyin the aroundtheBALANCE-SPRING 0, turn it
counterclock- slightly in thecounterclockwisedirectionso
wisedirection. that thespringforcemaynotbeappliedto the
PRISM 2 ASS’Y 0.
Step6)
Lift thejig 6)
%!?
Keepingthestateof 5), forcefully lift the
forcefullyto SPRINGFIXING JIG 0 topull out the
pull outthe protrudingpart of BALANCE-SPRING 0
spring. from thehole.
Step7) 7)
(r Turn90” Keepingthestateof 6), turn theSPRING
.
34 clockwiseand FIXING JIG 0 90” clockwiseandfurther lift
furtherlift the it forcefully to removetheBALANCE-
jig forcefully SPRING0 from thePRISM 2ASS’Y 0.
toremovethe
spring. 8) Removethetwo screws(*l) anddismountthe
lower GEAR FIXING PLATE 0 andGEAR
0, andtheBALANCE-SPRING 0 canbe
completelydismounted.
D-3

CK40-TBI D. REPAIR PROCEDURE
2-2 Gear and balance-spring repair procedure (Assembly procedure)
0
1) SettheBALANCE-SPRING 0 suchthat its
1 protrudingpart isinserted
to
thespringholeof
thePRISM 2 ASS’Y 0. Mount theGEAR 0
andGEAR FIXING PLATE @andfix them
nearlyat thecenterof thescrewplay.
x Tightenthescrews(*l) quitefirmly.
*1
Hook the protruding
part of the spring
around the tip of
the jig.
2)
SettingtheSPRINGFIXING JIG 0 in the
positionshownin theillustration,hold it with the
left hand.
HoldingtheBALANCE-SPRING 0 with the
right hand,carry it towardthearrow directionin
theillustration to hooktheprotrudingpart of the
balance-springonthetip of thespringfixing jig.
D-4

CK40-TBI D. REPAIR PROCEDURE
@
0
@
0
-II
c7
@
0
@
(A)
Upper gear
- Lower gear
Q
3
&
% 0 0 *
Upper gear - @
G
@
@
0 Lower gear ?I
0
00
@
@
3) Holdingthetop of thebalance-springto prevent
it from beingreleasedfrom thespringfixing jig,
turn thejig in thearrow directionto insertthe
protrudingpart of thebalance-springinto the
springhole.
4) Removethespringfixing jig andfurther push
thebalance-springfrom thetop to insertthe
protrudingpart securelyinto thehole.
5) SettheTBIC unit to themaximumtilt angle
with thestoppercontactingit securely,and
mounttheupperGERA 0 andlower GEAR 0
sothat no engagementplay existsbetween
them.
X It isthebestto mountthetwo gearsin the
positionsshownin FigureA without any
engagementplaybetweenthem.
Avoid themountingpositionsshownin Figures
(B) and(C) becausethetilting rangemaybe
insufficientor aproblemsuchasgearslippingor
unevengearoperationmayariseevenif no
engagementplay existsbetweenthetwo gears.
x Engagethegearsatthepositionof themaximum
tilt angle,or theimagemaybeinclinedby tilting
theTBIC unit.
x Tightenthescrews(*l) quitefirmly.
x Checkthefollowing itemsafter assembling:
1) Makesurethatthegearsareengagedwith each
otherwithout anyplay andtheworking force
meetsthestandardin theentiretilting range.
2)
3)
Standard: Up - Down:400 - 900g
(3920mN - 8820mN)
Down- Up: 200 - 700g
(1960mN - 6860mN)
Makesurethat thetilting rangeissufficientand
theunit stopssecurelyatthebothends.
Theimageshouldnotbeinclinedby tilting the
TBIC unit.
(Cl
D-5

CK40-TBI D. REPAIR PROCEDURE
3. Preparation for adjustment
Test plate
(KN0003)
KN0048
(with Adapter-l)
, ,
7 L z I I 5
- I i ! I
Standard obiective for ootical axis
(KN0015) ’ ’
Standard objective for tube length
(KN0016)
EDAch4X or 1OX objective
Gauge for checking stage tilt alignment
(SKN0002)
Adjustment item
Revolvingaxis
Left/rightopticalaxis
Absoluteoptical
axis
Exitpupilcenter
Parfocality
Stagetilt alignment
Eyepiece (jig)
Standardeyepiece
(KNOO48;withAdapter-l)
Standardeyepiece
(KNOO48;withAdapter-l)
Standardeyepiece
(KNOO48;withAdapter-l)
Centeringtelescope
(KNOO29)
Standardeyepiece
(KNOO48;withAdapter-l)
FT-36
Objective (jig)
4X or 10Xobjective
4X or 10Xobjective
Standardobjectivefor
opticalaxis
(KNOO15)
4X objective
Standardobjectivefor tube
length
(KNOO16)
Gaugefor checkingstage
tilt alignment
(sIsNooo2)
Jig and tool
Testplate
(KNOOO3)
Testplate
(KNOOO3)
Thicknessgauge
(0.3mm)
x Besureto checktheoptical axisof theCK40 mainbodyusingthestandardtube(BKN0003) before
adjustingtheCK30-TBI.
x If CK40 body is not available,usetheB2KC0403for adjustingwork. (UsetheKNOO41asobjective.)
D-6

CK40-TBI D. REPAIR PROCEDURE
r
4. Adjusting revolving axis
4-1 Set the following jigs to adjust the revolving axis:
l
Eyepiece.......... KNO048: Standard eyepiece(with Adapter-l)
l
Objective ........ EDAch4X or 10X
l
Specimen ........ KNOO03:Test plate (5/100 mm concentric circles)
l
Insert the standard eyepiece(KNO048 with Adapter-l) to the right sleeve.
l
Keep the TBIC unit fixed (stopper position) during the adjustment.
4-2 Set the interpupillary distance by opening the right side fully.
4-3 Turning the control knob of the stage, align the specimencenter with the cross hairs center of
the standard eyepiece.
Before adjustment After adjustment
L
Cross hairs center of L Specimen center
standard eyepiece
D-7

CK40-TBI D. REPAIR PROCEDURE
4-4 Check the optical axis displacementdue to revolving axis.
,
Work I
I Image seenthrough the standard eyepiece
I
I
I
I
I
I
I
I
I
I
I 0
I
I
I
I
I
Closetheinterpupillary distance I
slowly while observingthroughthe 1
standardeyepiece.
I
Thecentersarealigned Thecentersbecomedisplacedlike
I
atfirst. theillustration, for example.
I
4-5 Find the revolving axis center from the distance and direction of the optical axis displacement.
Taking the displacement distancein “4-4” asanexample, the position of the revolving axis
center is asshown below.
Whentheinterpupillary distanceis reduced,theconcentric
circlescentermovesfrom B to A.
Revolvingaxis center“0” existsontheleft sideof the
movingdirection.
(Whentheinterpupillary distanceis increased,revolving
axiscenter“0” existsontheright sideof themoving
direction.)
Thetriangleconsistingof A-B-O shownontheleft always
hasafixed shape.(Revolving axistriangle)
Therevolving axis centercanbe foundsimplyby adopting
this triangle.
r-------------------------------------------------- -------------I
I
Revolving axis triangle I
I
I
O Rewlting tis center
Remembertheshapeof this triangle anduseit for finding I
I
I
n
therevolving axis centerafter step4-4. I
I
I I
I
Reference
I
I I
I I
I I
I I
I I
I I
I
B A
AB:AO+1:1.3 I
I
I
Cross hairs center of Specimen center
I I
standard eyepiece I
I
I
I
I
-_-______-__________---------------------------- ---_--_---_-----
D-8

CK40-TBI D. REPAIR PROCEDURE
f -“\
4-6 Align the specimen center with the revolving axis center found by “step 4-5.”
I
Work
I
I
Image seenthrough the standard eyepiece
I
Turning the stagecontrol knob, move the
I
I
Image at first
specimencenterfrom A to 0.
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
Imageatendof adjustment
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
D-9

CK40-TBI D. REPAIR PROCEDURE
c
4-7 Align the cross hairs center of standard eyepiecewith the specimen center.
Work
1
I
I Image seenthrough the standard eyepiece
(1) Slightly loosenthe two screwsthatfix the / Imageat first
right sleeve. I
(Don’t usealarge diameterscrewdriver I
becauseit easilydamagestheoutersurface
of thesleeve.) 1
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
- I
(2) Whileobservingthrough thestandard
eyepiece,align the crosshairscenterof Align thetwo centers.
standardeyepiecewith therevolvingaxis I
centerlocatedin thelast step. I
(Adjust by manually changingthemounting
positionof theright sleeve.) i
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
(3) Firmly tightenthe two screwsthatfix the ; Imageatendof adjustment
right sleeve.
I
(Don’t usealarge diameterscrewdriver I
becauseit easilydamagestheoutersurface I
of thesleeve.) I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
I
D-l 0
Table of contents
Other Olympus Microscope manuals

Olympus
Olympus BX3-SSU User manual

Olympus
Olympus VM Series User manual

Olympus
Olympus VS120 User manual

Olympus
Olympus CHA-P User manual

Olympus
Olympus ORBEYE Manual

Olympus
Olympus IX2-DSU User manual

Olympus
Olympus DP26 User manual

Olympus
Olympus FLUOVIEW FVMPE-RS User manual

Olympus
Olympus SZ-III User manual

Olympus
Olympus CHBS User manual

Olympus
Olympus BX43 User manual

Olympus
Olympus CX21 User manual

Olympus
Olympus CX41 User manual

Olympus
Olympus U-CBS User manual

Olympus
Olympus SZ61 User manual

Olympus
Olympus BX51WI User manual

Olympus
Olympus Fluoview FV1000 Installation guide

Olympus
Olympus KHC User manual

Olympus
Olympus Fluoview-1000 User manual

Olympus
Olympus Fluoview FV1000 User manual