Olympus BX60 User manual

OfMMRUB
INSTRUCTIONS
BX60
SYSTEM
MICROSCOPE
This
instruction
manual
to
for
tho
Olympus
System
Microscope
Model
BX60.
To
obtain
optimum
performanca
and
to
familiariie
yourself
fully
whh
the
use
of
your
microscope,
we
recommend
that
you
read
this
manual
thoroughly
before
operating
the
microscope.

BX60
IMPORTANT
This
unit
employs
the
UIS
(universal
infinit
system)
optical
design
and
should
be
used
only
with
UIS
eyepieces,
objectives,
and
condensers.
Less
than
optimum
performance
may
result
if
inappropriate
accessories
are
used.
BX6O
is
a
versatile
microscope
which
can
be
used
both with
transmitted
light
for
use
in
the
biological
field
and
with
reflected
light
for
metallurgical
research.
Accordingly,
this
instruction
manual
contains:
•
Separate
outlines
of
the
observation
procedures
when
the
microscope
is
used
in
biological
or
metallurgical
research
purposes.
•
Explanation
of
the
use
of
controls
is
divided
into
sections.
•
The
most
important
points
of
the
various
observation
methods
are
explained.
For
reflected
light
fluorescence
microscopy
also
refer
to
the
manual
pertaining
to
the
reflected
light
fluorescence
attachment
BX-fLA.
Getting
Ready
1.
A
microscope
is
a
precision
instrument.
Handle
it
with
care
and
avoid
subjecting
it
to
sudden
or
severe
impact
2.
The
8X60
series
can be
used
with
any
kind
of
one
additional
inter
mediate
tube
excluding
U-DO
dual-viewing
attachment
and
U-MDO10
multi-viewing
attachment.
3.
Do
not
use
the
microscope
where
the
following
conditions
exist:
direct
sunlight
high
temperature
and
humidity,
dust
or
vibrations.
(Working
environment
ambient
temperature
and
hunrmfity
should
be
in
the
range
of
0-40°C.
30-90%.)
4.
When
moving
the
microscope,
carefully
cany
it
with
both
hands
by
grasping
the
arm
as
shown
in
the
figure
on
the
left-
*
You
can
damage
the
microscope
if
It
it
lifted
from
the
ttage.
coarse
adjustment
knob,
or
lamp
housing.
5.
Set
the
voltage
selector
switch
on
the
rear
of
the
base
to
the
100
-120V
or
220-240V
position
to
match
the
local
line
voltage,
using
a
flat-head
screwdriver.
(Before
shipment
from
the
factory,
the
voltage
selector
switch
is
set
to
220-240V
position).
(See
figure
on
the
left)
6.
To
avoid
potential
shock
hazard,
be
sure
to
properly
ground
the
power
cord.
7.
Always
turn
OFF
the
main
switch
and
disconnect
the
power
coed
before
replacing
the
halogen
bulb or
fuses.

I
Maintenance
and
Storage;
1
Clean
lenses
by
wiping
gently
with
gauze.
To
remove
fingerprints
or
oil
stains,
wipe
with
gauze
slightty
moistened
with
xylene
or
a
mixture
of
ether
(70%)
and
alcohol
130%).
*
SIneo
ethar
and
alcohol
are
highly
flammable,
be
careful
to
keep
these
chemicals
away
from
an
open
fire
and
potential
sources
of
electrical
sparks,
such
as
main
switches.
2.
Do
not
attempt
to
use
organic
solvents
to
clean
the
microscope
components
especially
plastic
pans
To
dean
use
a
neutral
detergent
3.
Do
not
disassemble
any
pan
of
the
microscope.
4.
When
not
using
the
microscope,
keep
it
covered
with
the
provided
dust
cover.
Symbols
on
the
Microscope
Frame
Caution
"...
If
the
microscope
is
operated
in
a
manner
not
specified
by
this
manual,
the
safety
of
the
equipment
may
be
impaired.
In
addition,
the
equipment
may
also
be
damaged.
Always
operate
the
equipment
as
outlined
in
this
instruction
manual.

,•
M
BX60
CONTENTS
2-1
Assembly
Diagram
2-2
Detailed
Assembly
Procedure.
Biological
Microscopy
13
Metallurgical
Microscopy
15
5-1
Base
17
5-2
Brightfield/Darkfield
Vertical
Illuminator
20
5-3
Stage
22
5-4
Observation
Tube
25
5-5
Condenser
27
5-6
Focusing
Adjustment
Knobs
29
5-7
Immersion
Objectives
(for
biological
purposes
only)
30
5-8
Photomicrography
30
!
6-1
Transmitted
Light
Brightfield
Observation
32
6-2
Reflected
Light
Brightfield/Darkfield
Observation
32
6-3
Reflected
Light
NcmarsH
Differential
Interference
Contrast
Observation
33
6-4
Reflected
Light
Simple
Polarized
Light
Observation
34
6-5
Reflected
Light
Fluorescence
Observation
35

NOMENCIA
TVRE
1
u
.Eyepiece
Brightfield/darkfleld
mirror
cuba
housing
or
universal
I
cube
housing
Revolving
INosepiece
CondansarI
Observation
tuba
I
Brightflold/darkfieki
collector
lens
unit
or
universal
collector
lens
unit
^\
Microsccpo
frame
|
Lamp
Housing
for
reflected
light
Objective
Lamp
housing
for
transmitted
light
Largo
stage
Inghi-r
1

BX60
ASSEMBLY
The
diagram
below
shows
how
to
assem&te
the
various
components.
The
numbers
indicate
the
order
of
assembly.
*
When
assembling
the
microscopo,
make
sura
that
all
parts
are
free
of dust
and
dirt
and
avoid
scratching
any
parts
or
touching
glass
surfaces.
Ump
housing
BriohtfWdMtrkfiald
Collector
tans
unit
IBrightfldd/dark-
ftakf
or
unhmsal)
Mtchtniol
nag*
with
WWund
(right-hind)
low
driv*
co»xl»l
controls
,
Largi
nMchtnlctl
sum
with
l«fl-h«nd
(right-hand)
low
(Irhw
coaxial
controls
*
The
DF
(dsrkfield)
insert
is
attached
when
the
microscopo
is
used
for
reflected
light
microscopy.
Be
sure
to
remove
the
Insert
when
the
microscope
is
used
for
transmitted
light
microscopy.

Attaching
the
Stage
(U-SIC4,
U-SV)
,Rg.,,
1.
FuUy
loosen
the
clamping
screw
©
at
the
rear of
the
stage
(at
the
front
n
the
case
of
U-SV).
2.
then
tighten
the
clamping
screw.
fig.
1
Rg.2
Mounting
the
Condenser
)FIa.
a
1.
Turn
the
coarse
adjustment
knob
©
to
raise
the
stage
to
its
upper
limit
lAssuming
there
are
no
objectives
mounted
on
the
nosepiece
yet.)
2.
Tumiheco^r^heiBht
holder
to
its
tower
Emit
3.
Loosen
the
condenser
clamping
screw.
4.
Position
the
condenser
with
the
scale
markings
in
front
and
insert
it
into
the
substage
fork
as
far
as
it
wB
go.
5.
Tighten
the
condenser
clamping
screw,
then
raise
the
condenser
to
its
upper
limit
in
tha
substage
fork.
™Ullin9
*•
U'SC
«•**•<"«
«chrom«le
condenser
or
the
U-UCO
universal
condenser,
swing
the
top
lens
out
of
the
way
before
inserting
the
condenser.

BX60
Fig.
3
fio-4
Mounting
tKS
(Universal
COBe
Housing)-
.
(Rg*.
3,*)
O
The
following
procedure
does
not
apply
to
the
briflhfieMWartfield
mirror
cube
housing.
*
Loosen
the
screw
that
secures
tha
cube
housing
to
the
frame
with
tha
provided
Allan
screwdriver.
It
is
located
In
the
hole
at
the
right
side of
the
arm.
1.
Invert
the
cube
housing
so
that
the
cube
dovetail
mounts
on
the
turret
@
point
upward.
(Dummy
plates
are
mounted
in
three
of
the
four
cube
positions.
When
you
wish
to
use
only
one
cube,
mount
it
in
tha
empty
position.
When
using
tw6
or
more
cubes,
loosen
the
clamping
screw
<D
and
remove
the
dummy
ptatels)
by
puffing
in
the
direction
infested
by
the
arrow,
and
then
mount
tha
actual
eubslsl
in
its
place.)
2.
Hold
tha
cube
to
be
mounted
with
its
index
side
facing
upward and
slide
it
all
the
way
onto
the
dovetail
mount
Next be
sure
to
tighten
the
cube
damping
screw
©
immediately.
(Tighten
all
(our
cube
damp
ing
screws.)
3.
Remove
the
cube's
magnetic
index
sticker
rj>*
end
affix
it
to
the
corre
sponding
turret
position.
(Fig.
4)
*U»
a
sharp
abject
such
as
the
tip
of
a
ballpoint
pan
or
macfiinical
pared
to
lift
tha
Cuba's
mfignatic
indax
sockar.
*
The
cube
indices
A, B.C.
D®
on
the
dovetail
mount
correspond
to
the
turret's
A,
B,
C,
D
Indices
(JX
Make
sure
to
match
tha
attached
cube
correctly
with
the
position
of
the
removed
magnetic
index
sticker
on
the
turret
(Fig.
5)
Mounting
the
Cub«
Housing.
(Ro,«
1.
Align
the
cube
housing
dovetail
©
with
the
dovetai)®
at
the
front
of
the
arm and
fully
slide
it
into
tha
arm.
2.
insert
the
Allen
screwdriver
through
the
hole
®
on
the
right
side of
the
arm.
then
securely
tighten
the
cube
housing
damping
screw.
\
Fig.
8

Mounting
the
Collector
Lens
Unit
(Fig.
6)
1.
Gently
insert
the
collector
lens
unit
as
far
as
it
will
go
into
the
opening
located
at
the
rear of
the
arm.
2.
Insert
the
provided
screws
in
the
four
screw
holes
ana
tighten
securery
with
the
Allen
screwdriver
Fig.
6
Attaching
the
Field
Iris
Diaphragm
and
Aperture
Iris
Diaphragm
Knobs
(Rg.7»
Rg.7
1.
Insert
the
provided
field
iris
diaphragm
and
aperture
iris
diaphragm
knobs
through
(he
holes
0
on
the
right
side
of
the
aim.
then
screw
in
the
knobs
until
tight
O
When
using
the
LMJflBL
universal
vertical
illuminator
lens
unit
the
shutter
knob
should
also
be
attached
in
the
same
manner.
2.
Insert
the
provided
field
iris
diaphragm
and
aperture
iris
diaphragm
centering
knobs
through
the
holes
(D
on
both
sides
of
the
arm.
then
screw
in
the
knobs
until
light.
3.
Insert
the
provided
pinhda
knob
through
the
hole
on
the
left
side
of
the
arm, then
screw
in
the
knob
until
tight
(U-RLBL
only).

-j
BX60
Fig.
B
fig.
9
Installing
therHalogen
Bulb
(Rga.
B,
(for
both
Transmitted/Reflected
Ughts)
9.10)
O
The
appropriate
bulb
is
a
12V.
100WHAL
halogen
bulb
(PhBips
77241.
1.
Futry
loosen
the
lamp
housing
damping
screw©
on
top
of
the
lamp
housing
cover
with
the
provided
AQen
screwdriver.
2.
Lift
the
lamp
housing cover
Q) upward
to
remove
it
(fig.
81
3.
Turn
the
lamp
socket
Q)
90*
in
the
direction
indicated
by
the
arrow.
4.
Holdtng
the
bulb®
with
gloves
or
a
piece
of
gauze,
depress
the
bulb
damping
levers
®
and
insert
the
bulb
pins
®
MV
into
the
pin
holes
ffl.
Gently
release
the
bub
damping
levers
®
to
their
original
positions
to
secure
the
bulb.
(figs.
9.
10)
*
Do
not
touch
the
bulb
with
bare
hands.
If
fingerprints
are
accidentally
left
on
the
bulb,
wipe
the
bulb
with
toft
doth.
S.
Slide
the
lamp
housing cover onto
the
housing
base from
above.
Tighten
the
damping
screw
®
whle
pressing
downward
on
the
cover.
(Fig.
3)
*
Whenever
you
replace
the
bulb,
first
tum
OFF
the
main
switch
and
wait
for
ths
bulb
and
lamp
socket
to
cool
down.
fig.
10

Attaching
the
Lamp
Housing
to
the
Microscope
|
Transmitted
Light
Lamp
Homing
|
(Figs.
11.
12.13)
fig,
11
1.
Using
the
Allen
screwdriver,
fully
loosen
the
lamp
housing
clamping
screw
©
on
the
microscope.
2.
Insert
the
lamp
housing
eoUactor
unit
mto
the
microscope
unU
it
touches
tha
lamp
housing
bracket
on
the
back
of
tha
base
(lower
part),
then
tighten
the
clamping
screw
©
(Fig.
11)
Rg.
12
|
Reflected
Light
Lamp
Homing
|
1.
Using
the
Allen
screwdriver,
fully
loosen
the
lamp
housing
damping
screw
©
on
the
reflected
Tight
port
on
the
microscope.
(Fig.
12)
2.
Insert
the
lamp
housing
collector
unit
into
the
lamp
port
unfl
it
touches
the
lamp
housing
bracket
on
the
back
of
the
base
(upper
pan),
than
tighten
the
clamping
screw
®.
(Fig.
13)
3.
Insert
the
cord plug
from
the transmitted
light
lamp
housing
securely
into
tha
power
outlet
©.
Insert
the
cord
plug
from
tha
reflected
light
lamp
housing
securely
into
the
power
outlet
©.
IFig.
13)
Fig.
13
Fig.
14
Mounting
the
Objectives
(Fig.
14)
C
For
reflected
tght
observation,
make
sure
to
mount
metallurgical
UIS
objectives.
Mount
the
objectives
on
the
revolving
nosepiece
in
such
a
manner
lhat
the
magnification
increases
from
tow
to
high
power
in
a
clockwise
direction.
(Fig.
14)

BX60
Mounting
tfie
Revolving
Noseplece*
*
(Rg.15)
*
Whan
used
for
metallurgical
obaarvatien:
Mount
tha
DF
insert
©.
The
insert
will
be
held
in
place
by
the
magnet on
tha
underside
of
the
arm.
1.
Turn
tha
coarse
adjustment
knob
to
lower
the
stage
all
the
way.
2.
Using
the
Allen
screwdriver,
loosen
the
nosaptece
clamping
screw
@
on
the
microscope.
3.
Carefully
slide
the
nosepisce
along
the
dovatail,
in
the
direction
of
the
arrow,
all
the
way
in
4.
Clamp
the
nosepiece
by
tightening
the
nosepiace
clamping
screw.
Mounting
the
Observation
Tuber
*
r
"
{Rg.
is>
1.
Using
ihe
ASen
screwdriver,
loosen
the
observation
tube
damping
screw
©■
2.
Insert
the
circular
dovetail
mount
at
the
bottom
of
the
observation
tube
into
the
opening
on
the
microscope
frame,
placing
the
observation
tube
to
point
the
binocular
eyepieces
towards
the front
Fasten the
observation
tube
by
tightening
the
clamping
screw.
Mounting
the
Eyepieces;-/
(Fig*
17,18)
go.
(Fig.
17)
Whan
Using
the
Trinocular
Observation
Tube
(U-TR30)
or
the
Supar-widaflold
Trlnocular
Observation
Tuba
IU-SWTR)
When
using
a
finder
eyepiece
or
an
eyepiece
with
irtci
xijustn
insert
this
eyepiece
into
the
right-hand
eyepiece
sleeve.
When
doing
so.
make
sure
that
the
eyepiece
positioning
pin
<j)
fits
into
the
notch
®
at
the
bottom
of
tha
eyepiece
sleeve.
(Fig.
18)
8

I
Fig.
19
Rg.
20
Connecting
the
Power
Cord
(R9.
1.
Make
sure
that
the
main
switch
(?)
a
on
OFF.
2.
Before
shipment
from
the
factory,
the
voltage
selector
switch
fj)
u
set
to the
200-240V
position.
In
C33e
your
local
line
vortaga
is
10O-120V,
move
the
switch
to
tha
100-120V
position
using
a
flat-head
screwdriver.
3.
Insen
the
power
cord
into
the
AC
receptacle
®
on
the
frame.
4.
Connect
tha
power
cord's
ground
wire
to
the
ground
terminal
on
the
wall
outlet
then
plug
the
power
cord
into
the
wall
outlet,
(in
case
of
a
three-prong
power
cord
and
wall
outlet separate
grounding
is
not
necessary.)
Fuse
Replacement
(Fig*.
20, 21)
O
Before
replacing
fuses,
set
tha
main
switch
to
OFF
and
unplug
the
power
cord,
(the
power
cord should
be
unptugged from
the
AC
recepocle
to
allow
removal
of
the
fuse
holder.)
1.
Remove
the
fuse
holder
(t)
by
squeezing
it
at
both
sides
and
puling
outward.
fFtg.
20)
2.
Replace
both
fuse3
©
with
new
ones.
{Fig.
21)
*
Use
only
specified
fuses.
Applicable
fuse:
260V,
SA
slow-blow
(Tims
Lag)
High-Breaking-Capacity,
2
fuses
(UTTLEFUSE
215005)
tnu
Dm
ad.
Cut
Spt
COr
■ni
Fie
Rg.21

BX60
CONTROLS
Antlyisr
insertion
slot
Iwhtouh
safacur
knob
Potariwr
tfldor
insertion
slot
Raid
trie
dtaphreofn
knob
Aperture
int
■uaphnom
knob
tnterpupitlery
distance
adjustment
tctlt
Diopter
adjustment
ring
Aperture
m»
duphraem
ccntsrino
knob
Fwld
ifis
csnttring
knob
Tnn«mltudA*fl«ciad
Saht
(elector
witeh
Mjcfoacop#
mountiiig
bolt
holn
filter
lever
(LBD)
Filter
lover
INWI
Teneion
ediuetment
knob
O
The
above
illustration
shows
the
microscope
with
the
U-RIBC
mirror
cube
housing
for
the
brightfiQkVdarfcfiold
vertical
illuminator.
In
the
case
ol
the
IHJRBC
cube
housing
lor
the
universal
vertical
illuminator,
the
cube
selector
knob
is
replaced
by
a
turret.

[I
Thaixfletononly
Be
used
for
both
modal*
U-00
and
U-MOOB.
11

I
BrightfleloVDarfcfield
Vertical
Illuminator
Attachment
(U-RLBC/U-RLBU
Collector
Lens
Unit
(U-RLBU
O
Refer
to
section
6-5.
Reflacted
Light
Fluorescence
Observation,
for
operation
details.
|
Universal
Vertical
Illuminator
Attachment
(U-RLBC/U-RLBL)
BX60
Hoto
for
imchflMfit
of
the
»p»rtur«
Ut»
dliprtrtflro
fcnob
Hoi*
for
ttuchment
of
the
tpanura
Hols
for
m#thimiu
of
\lh«
fitid
lrl»
(Utphrwm
Imeb
Hole
for
•ttKhrmnt
of
trw
fWd
Irii
^diaphnflm
etnmtng
knob
Collector
lent
unit
(U-URBU
8
12

Biological
Microscopy
o
The
Jcflcwng
section
outfines
the
operations)
procures
v^
me
microscope
o
used
fw rwmid
(transmitted
fight
brightfieW
observation).
^^
*
UlS
objectives
tor
biological
microscopy
should
be
used.
Using
the
universal
cube
housing,
rotate
the
cube
turret
to
en
empty
position.
Using
the mirror
cube
housing,
slide
tho
cube
selector
cube
to
the
DF
position.
1.
Press
the
transrruttedAellected
light
selector
switch
to
select
the
transmitted
light
mode.
Turn
ON
the
main
switch
and
adjust
the
tight
intensity
with
the
light
intensity
lever.
(At
this
point
leave the
tight
intensity
preset
button
OFF)
(Page
17)
2.
Disengage
aU
filters
from
the
light
path.
(Pages
17.
181
a.
Accesssory
fitter
cassette
b.
Fitters
built
into
the
base
3.
Turn
the
revolving
noseptece
to
engage
the
10X
objective.
Make
sure
the
revolving
rtosepiece
dicks
into
position.
4.
Place
a
specimen
on
the
stage.
(Page
22)
(Using
a
Trinocular
Observation
Tubel
6.
Push
the
observation
tube's
tight
path
selector
knob
to "binocular
eyepiece
100%'
(the
pushed-in
position).
(Page
26)
&
Looking through
the
nght
eyepiece
with
your
tight
eye.
turn
the coarse
adjustment
knob
to
bring
the
specimen
into
focus.
After
obtaining
approximate
focus,
use
the
fine
adjustment
knob
to
make
final
adjustments.
(Page
25)
13

j.
ait
BX60
7.
Looking
through
the
left
eyepiece with
your
left
eye.
turn
ttie
diopter
adjustment
ring
to
focus
on
the
specimen.
(Page
25)
8
Adjust
the
werpuftflary
distance
of
the
eyepieces.
IPage
25)
9.
Adjust
condenser
centering
and
focusing.
(Page
27)
I
10.
Engage
the
objective
to
be used
and
readjust
the
tight
intensity
to
the
desired
level
for
observation,
then
readjust
the
focus.
11.
Engage
your choice
of
fitters
into
the
fight
path.
(Pages
17, 18, 19)
a.
Accessory
filter
cassette
b.
Filters
built
into
the
base
12.
Adjust
the
field
iris
diaphragm.
(Page
27)
13.
Adiusi
the
aperture
ens
diaphragm, (Page
28)
14

Metallurgical
Microscopy
The
following
section
outlines
the
operational
procedures
when
the
microscope
o
used
to
normal
metallurgical
rrucroscopy
(reflected
light
brigritfteldAJarkfield
observation)
with
the
brightfield/darkfiald
vertical
illuminator.
In
the
case
of
the
universal
vertical
illuminator.
the
cube
selector
knob
is
replaced
by
a
turret.
l.
Select
the
cube
according
to
the
observation
purpose.
Set
the
cube
selector
knob
©
to
BF.
I
2
m
15
2.
Disengage
the
filter
slider
©
(left
side of
arm),
analyzer
i
slider
0.
and
NO
filter
from
the
light
path.
(Page
33)
3.
Press
the
transmitted/reflected
light
selector
switch
to
select
the
reflected
light
mode.
Turn
ON
the
main
switch
and
adjust
the
light
intensity
with
the
light
intensity
lever.
(At
this
point
leave
the
light
intensity
preset button
OFF.)
(Page
171
4.
Place a
specimen
on
the
stage.
(Page
22)
5.
Turn
the
revolving
nosepiece
to
engage
the
10X
objective.
Bring the
specie'
into
focus.
6.
Adiusi
the
mterpupillary
distance
of
the
eyepieces.
(Page
2SI

BX60
7.
Looking
through
the
right
eyepcece
with
your
light
eye.
turn
the
coarse adjustment
knob
to
bring
the
specimen
into
focus.
After
obtaining
approximate
focus,
use
the
fine
adjustment
knob
to
make
final
focus
adjustments.
(Page
25)
B.
Looking
through
the
left
eyepiece
with
your
left
eye,
turn
the
diopter
adjustment
ring
to
focus
on
the
specimen.
(Page
25)
9.
Confirm
that
the
illumination
is
adequate
for
the
observation
purpose.
IPage
32)
10.
Engage
the
required
filter.
11.
Engage
the
objective
to
be
used
for
observation,
then
readjust
the
focus
and
adjust
the
light
intensity
to
a
suitable
level
for
observation.
12.
In
case
of
reflected
light
brightfteU
observation,
select
the
best
field
and
aperture
iris
diaphragm
settings
in
accordance
with
the
objective
and
specimen.
(Pages
20.
21)
1
16
Other manuals for BX60
1
Table of contents
Other Olympus Microscope manuals

Olympus
Olympus CX22LED User manual

Olympus
Olympus U-LGPS User manual

Olympus
Olympus BX51M User manual

Olympus
Olympus SZX7 User manual

Olympus
Olympus SZX-DA User manual

Olympus
Olympus DSX1000 User manual

Olympus
Olympus Fluoview FV1000 User manual

Olympus
Olympus SZ2-ILA User manual

Olympus
Olympus U-DA User manual

Olympus
Olympus CKX41 User manual

Olympus
Olympus DP73 User manual

Olympus
Olympus U-MDOB3 User manual

Olympus
Olympus SZX16 Installation guide

Olympus
Olympus MVX10 User manual

Olympus
Olympus U-D7RES User manual

Olympus
Olympus SZ-III User manual

Olympus
Olympus BX63 User manual

Olympus
Olympus CKX53 User manual

Olympus
Olympus 1X71 User manual

Olympus
Olympus FLUOVIEW FVMPE-RS User manual