(Table 6)
Recommended separation distance between portable and mobile RF
communications equipment and the OMRON Digital Weight Scale
OMRON Digital Weight Scale is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The
customer or the user of this OMRON Digital Weight Scale can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the
OMRON Digital Weight Scale as recommended below, according to the
maximum output power of the communications equipment.
Output Power
of Transmitter
in Watt
Separation distance according to frequency of
transmitter in meter
150 kHz to 80 MHz
Not Applicable.
80 MHz to 800 MHz
d= 1.2 √P
800 MHz to 2.5GHz
d= 2.3 √P
0.01 Not Applicable. 0.12 0.23
0.1 Not Applicable. 0.38 0.73
1Not Applicable. 1.2 2.3
10 Not Applicable. 3.8 7.3
100 Not Applicable. 12 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reection from structures,
objects, and people.
Product description Digital Weight Scale
Model HN-290T
Display 10 to 100 kg with an increment of 0.1 kg
(22 to 220 lb with an increment of 0.2 lb)
100 to 250 kg with an increment of 0.2 kg
(220 to 550 lb with an increment of 0.5 lb)
Weight Accuracy 10 kg to 40 kg: ± 0.4 kg
(22 lb to 88 lb: ± 0.88 lb)
40 kg to 250 kg: ± 1%
(88 lb to 550 lb: ± 1%)
IP Classication IP21
Power Supply 4 "AA" alkaline batteries (LR6)
Battery Life Approximately 1 year (when used once per day)
Operating Temperature/
Humidity
+10℃to +40℃(+50℉to +104℉), 15% to 90% RH
Storage Temperature/
Humidity/Air Pressure
-20℃to +60℃(-4℉to +140℉), 10% to 95% RH,
860 hPa to 1060 hPa
Weight Approximately 3.5 kg (7.7 lb)
including batteries
External Dimensions Approximately 41 (H) × 415 (W) × 360 (D) mm
1 3/5" (H) × 16 1/3" (W) × 14 1/5" (D)
Contents Digital Weight Scale, 4 "AA" alkaline batteries,
instruction manual, and quick start guide
Applied Part = Type BF
8. Specications
Note: Subject to technical modication without prior notice.
This device fullls the provisions of the EMC Directive 2004/108/EC.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD
OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specic legal rights, and you may have other
rights which vary from state to state. Some states do not allow limitations on
how long an implied warranty lasts, so the above limitation may not apply to
you. Some states do not allow the exclusion or limitation of incidental or
consequential damages, so the above limitation or exclusion may not apply
to you.
OMRON Digital Weight Scale
Information for accompanying documents in the scope of IEC60601-1-2:2007
FCC CAUTION
Changes or modications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to part 15 of the FCC Rules. These limits
are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates, uses and can radiate
radio frequency energy and, if not installed and used in accordance with
the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and
on, the user is encouraged to try to correct the interference by one or more of
the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which
the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
This transmitter must not be co-located or operated in conjunction with any
other antenna or transmitter.
This equipment complies with FCC/IC radiation exposure limits set forth for an
uncontrolled environment and meets the FCC radio frequency (RF) Exposure
Guidelines and RSS-102 of the IC radio frequency (RF) Exposure rules. This
equipment has very low levels of RF energy that are deemed to comply without
testing of specic absorption ratio (SAR).
This device complies with Part 15 of the FCC Rules and Industry Canada
license-exempt RSS standard(s). Operation is subject to the following two
conditions: (1) this device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may
cause undesired operation.
Changes or modications made to this equipment not expressly approved by
NeuroSky Inc. may void the FCC authorization to operate this equipment.
Radio frequency radiation exposure Information:
The radiated output power of the device is far below the FCC radio frequency
exposure limits. Nevertheless, the device shall be used in such a manner that
the potential for human contact during normal operation is minimized.
This Product operates in the unlicensed ISM band at
2.4GHz. In case this Product is used around the other
wireless devices including microwave and wireless LAN,
which operate same frequency band of this Product,
there is a possibility that interference occurs between
this Product and such other devices. If such interference
occurs, please stop the operation of other devices or
relocate this Product before using this Product or do not
use this Product around the other wireless devices.
The Bluetooth®Smart word mark and logos are registered
trademarks owned by Bluetooth SIG, Inc. and any use of
such marks by OMRON HEALTHCARE Co., Ltd. is under
license.
9. FCC/IC Statement and Trademarks
11. Guidance and Manufacturer’s Declaration
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC’s and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation. Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic
Compatibility) with the aim to prevent unsafe product situations, the
IEC60601-1-2 standard has been implemented. This standard denes the
levels of immunity to electromagnetic interferences as well as maximum
levels of electromagnetic emissions for medical devices.
Medical devices manufactured by OMRON Healthcare conform to this
IEC60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• The use of accessories and cables other than those specied by OMRON,
with the exception of cables sold by OMRON as replacement parts for
internal components, may result in increased emission or decreased
immunity of the device.
• The medical devices should not be used adjacent to or stacked with other
equipment. In case adjacent or stacked use is necessary, the medical
device should be observed to verify normal operation in the conguration in
which it will be used.
• Refer to further guidance below regarding the EMC environment in which
the device should be used.
• The MEDICAL ELECTRICAL EQUIPMENT Digital Weight Scale needs
special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this documentations.
• The Essential Performance of the Digital Weight Scale is to measure weight.
The Digital Weight Scale may be interfered with by other equipment, even if
that other equipment complies with CISPR EMISSION requirements.
(Table 1)
Guidance and manufacturer’s declaration – electromagnetic emissions
OMRON Digital Weight Scale is intended for use in the electromagnetic
environment specied below. The customer or the user of this OMRON Digital
Weight Scale should assure that it is used in such environment.
Emissions test Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
The OMRON Digital Weight Scale uses
RF energy only for its internal function.
Therefore, its RF emissions are very low
and are not likely to cause any interference
in nearby electronic equipment.
RF emissions CISPR 11
Class B The OMRON Digital Weight Scale is
suitable for use in all establishments,
including domestic establishments and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not
Applicable.
Voltage uctuations/
icker emissions
IEC61000-3-3
Not
Applicable.
(Table 2)
Guidance and manufacturer’s declaration – electromagnetic immunity
OMRON Digital Weight Scale is intended for use in the electromagnetic
environment specied below. The customer or the user of this OMRON Digital
Weight Scale should assure that it is used in such environment.
Immunity test IEC 60601
Test level
Compliance
level
Electromagnetic environment
– guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±6 kV
contact
±8 kV air
±6 kV
contact
±8 kV air
Floor should be wood, concrete, or
ceramic tile. If oors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power
supply lines
±1 kV for
input/output
lines
Not
Applicable. Not Applicable.
Surge
IEC 61000-4-5
±1 kV
line(s) to
line(s)
±2 kV
line(s) to
earth
Not
Applicable. Not Applicable.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
inputlines
IEC 61000-4-11
<5 % UT
(>95 % dip in
UT)
for 0.5 cycle
Not
Applicable. Not Applicable.
40 % UT
(60 % dip in
UT)
for 5 cycles
70 % UT
(30 % dip in
UT)
for 25 cycles
<5 % UT
(>95 % dip in
UT)
for 5 sec.
Power
frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic elds
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
Note: UTis the A.C. mains voltage prior to application of the test level.
(Table 4)
Guidance and manufacturer’s declaration – electromagnetic immunity
OMRON Digital Weight Scale is intended for use in the electromagnetic
environment specied below. The customer or the user of this OMRON Digital
Weight Scale should assure that it is used in such environment.
Immunity test IEC 60601
Test level
Compliance
level
Electromagnetic environment
– guidance
Portable and mobile RF
communications equipment
should be used no closer to any
part of the OMRON Digital Weight
Scale including cables, than the
recommended separation distance
calculated from the equation
appropriate to the frequency of the
transmitter.
Conducted RF
IEC 61000-4-6
3 V rms
150 kHz to
80 MHz
Not
Applicable.
Recommend
separation distance
Not Applicable.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m d= 1.2 √P
80 MHz to 800 MHz
d= 2.3 √P
800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from xed RF
transmitters as determined by
an electromagnetic site survey,a
should be less than the
compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Contact your Telehealth service provider if you have any questions.
Manufactured for :
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho
Muko, Kyoto, 617-0002 JAPAN
Distributed by :
OMRON HEALTHCARE, INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2015 OMRON HEALTHCARE, INC.
Made in China
Note1:At 80 MHz and 800 MHz, the higher frequency range applies.
Note2:These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reection from structures,
objects, and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radio, AM and FM radio broadcast, and
TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to xed RF transmitters, an electromagnetic
site survey should be considered. If the measured eld strength in the location
in which the OMRON Digital Weight Scale is used exceeds the applicable
RF compliance level above, the OMRON Digital Weight Scale should be
observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the OMRON Digital Weight Scale.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less
than 3 V/m.
10. Limited Warranty
Your HN-290T Digital Weight Scale excluding batteries, is warranted to be free
from defects in materials and workmanship appearing within 2 years from the
date of purchase, when used in accordance with the instructions provided with
the scale. The above warranty extends only to the original retail purchaser.
We will, at our option, replace without charge any unit covered by the above
warranty. Replacement is our only responsibility and your only remedy under
the above warranty.