manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Ono Sokki
  6. •
  7. Measuring Instrument
  8. •
  9. Ono Sokki HR-6800 User manual

Ono Sokki HR-6800 User manual

Digital Tachometer (For high speed)
HR-6800
Electromagnetic Detector
MP-5350
(The sensor is not included in a price of HR-6800
)
Instruction Manual (Basic Operations)
Thank you for your selection of the HR-6800 Digital Ta-
chometer.
To ensure the performance of the HR-6800, please read
this manual thoroughly.
Warnings and Cautions
In this document precautions are classified into two cat-
egories: WARNING and CAUTION.This depends on the
degree of danger or damage possible if the precaution
is ignored and the product is used incorrectly.
1Power switch
Turns the power ON or OFF.
2Display
Displays the measurement value and various settings.
3Input connector
Connected with the Detector with the supplied connection
cable.
4RECALL & ↑switch
Used for memory call during measurement and numeri-
cal input in the parameter setup mode.
5MENU switch
Used to switch between the measurement mode and the
parameter setup mode.
The lock mode can be set and canceled by pressing this
switch for 3 seconds or longer.
6MEMORY & →switch
Used for memory writing during measurement and nu-
merical digit shift in the parameter setup mode.
7MODE & NEXT switch
Used for mode change during measurement and item
selection in the parameter setup mode.
8Indicator (input signal indicator)
When the detector circuit detects a rotational signal, the
LED indicator lights up.
9Type A/B selector switch
Used to select Type A or B.
0Trigger level adjustment volume
Volume for adjusting the trigger level.
qTripod mounting hole
Used to mount a tripod.
wBattery cover
eConnector cover
Cover for the power inlet and output connectors.
rDC power input
Input terminal for connecting the dedicated AC adapter.
(when the dedicated AC adapter and dry batteries are
used together, the AC adapter is given priority.)
tAnalog output
Terminal for connection with a recorder, etc. through
the optional AX-501 cord.
yPulse output
Terminal for connection with an FFT analyzer, etc.
through the optional AX-501 cord.
uCONDITION display
Displays the measurement mode, LOW battery, and
errors.
iMAIN display
Displays measurement values, selections, settings, etc.
oSUB display
Displays memory addresses, settings, etc.
pUNIT display
Multiplies the measured value in the MAIN display by
ten for readout in "r/min."
aDetecting element
Bring the detecting element at the end of the Detector
close to the body of revolution under measurement.
sOutput connector
Connected with the HR-6800 unit using the supplied
connection cable.
1.Overview
The HR-6800 Digital Tachometer is a handy tachometer with
built-in batteries integrating the measurement unit and dis-
play unit. It is designed for measurement of bodies of revolu-
tion for dental instruments, textile machines, and general-
purpose high-speed machine tools, etc.
With the use a dedicated Detector, the HR-6800 makes it
possible to measure the rotational speed of the body of revo-
lution under measurement.
Up to 20 measurement data items can be stored in the HR-
6800 unit using the memory function, which is useful for sub-
sequent confirmation.
The display unit displays the rotational speed (x10 r/min) with
a 5-digit number.
The equipment is provided with an analog output (inter-
changeable with a monitor output) and a pulse output. The
analog output is used for recording with a data recorder, etc.,
and the pulse output for tracking analysis with an FFT ana-
lyzer.
2.Features
・Measurement from low to high rotational speeds (100 to
999,990 r/min)
・High sensitivity
・MAX and MIN modes for displaying the maximum and
minimum values
・Convenient memory function (up to 20 items can be
memorized) for confirmation of measurement results
・Over mark function for indicating measurement values
exceeding a specified value
・Analog output, sensor signal monitor output, and pulse
output
・Type AAA batteries and AC adapter commonly used
・Back light function which is convenient for use in dark
places
3.Measurement Principle
The Detector incorporates a permanent magnet and a de-
tection coil to detect the rotational speed of an object under
measurement based on the principle of electromagnetic in-
duction.
4.Unpacking
When you unpack the unit, make sure that you have all the
following:
【HR-6800】
①HR-6800 main unit .................................................. x1
②Type AAA alkaline dry battery ................................ x4
③Instruction manual ................ x2 (2 different manuals)
④Carrying case ......................................................... x1
⑤MI-0301 Tripod attachment adapter ........................ x1
【MP-5350】
(The sensor is not included in a price of HR-6800
)
①Detector (MP-5350) ................................................ x1
②Connection cable (1m) ........................................... x1
Overview
o
i
u
p
6
7
4
2
1
8
3
e
5
ANALOG PULSE
6
V
0.5
A
r t y
q
w
180.5
66
14
47.5
TRIG LEVEL
Type
A B
9
0
189.5
10713
a
s
1HR-6800 main unit ②Type AAA
alkaline dry batteries
3Instruction manuals 4Carrying case
High-speedTachometer
HR-6800
InstructionManual(FunctionReference)
Copyright© ONO SOKKI CO.,LTD.
2005All rights reserved.
ThankyouforyourselectionoftheHR-6800High-
speedTachometer.
ToensuretheperformanceoftheHR-6800,pleaseread
thismanualthoroughly.
High-speedTachometer
HR-6800
ElectromagneticDetector
MP-5350
InstructionManual(BasicOperations)
Copyright© ONO SOKKI CO.,LTD.
2005All rights reserved.
ThankyouforyourselectionoftheHR-6800High-
speedTachometer.
ToensuretheperformanceoftheHR-6800,pleaseread
thismanualthoroughly.
⑤Tripod attachment adapter (MI-0301)
Name and Function of Each SectionObserve the Following Points before Use
CAUTION
●Do not drop the equipment or apply excessive shock to
it.
Since this equipment incorporates high-precision elec-
tronic parts, be careful not to drop it or apply strong shock.
●Wipe dirt off using a dry cloth or a cloth dampened with
neutral detergent and squeezed firmly. Do not use vola-
tile oils (thinner or benzine) or alcohols.
●Be sure to use the supplied connection cables.
Using cables other than the supplied one may make it
impossible to obtain specified performance or, in the worst
case, cause damage to the equipment.
●Measurement in strong magnetic field or near an AC line
is not possible.
Since this equipment has a high input sensitivity to en-
able measurement with a minute signal, measurement in
strong magnetic field or near an AC line is not possible.
WARNING
●Perform measurement using enough care with the target
under measurement rotating at high speeds.
Be careful not to allow connection cables to be caught by
rotating sections.
●When using an AC adapter, be sure to use a dedicated
AC adapter (option).
Using adapters other than the optional one may cause
failure.
●Avoid using the equipment in places subject to rapid tem-
perature change.
Do not move the equipment rapidly from a hot place to a
cold one or vice versa. Condensation can form inside the
unit which may cause failure.
●Be careful not to get water, oil, dust, or foreign materials
inside the unit.
Do not use the equipment in places where you may get
water or oil or places which are humid or dusty.
1MP-5350 Detector ②Connection cable
Copyright©ONOSOKKICO.,LTD.
2007Allrightsreserved.
■Omission of Issuance of Certificate
This product has been tested under strict inspections for
correct operation before shipment. Please note that the
issuance of certificate is omitted.
■Warranty
1. This product is covered by a warranty for a period of one
year from the date of delivery.
2. This warranty covers free-of-charge repair during the
warranty period for defects occurred while the product is
used under correct operating conditions according to de-
scriptions in this manual and notices on the unit label.
3. For free-of-charge repair during the warranty period, con-
tact your dealer or your nearest Ono Sokki sales office
nearby.
4. Even during the warranty period, the following failures
will be handled on a fee basis.
(a) Failures or damages occurring through misuse,
misoperation, repairing without ONO SOKKI’S ap-
proval.
(b) Failures or damages occurring through mishandling
(dropping) during transportation after purchase.
(c) Failures or damages occurring by an Act of God (fires,
earthquakes, flooding, and lightening), environmen-
tal disruption, or abnormal voltage.
(d) Replenishment of expendable supplies, spare parts,
and accessories.
WARNING
CAUTION
Thissymbolisusedtoindicateprecautions
wherethereisariskofdeathorserious
personalinjurytotheoperatoriftheproductis

handledincorrectly.
Thissymbolisusedtoindicateprecautions
wherethereisariskofsomepersonalinjury
totheoperatororonlymaterialdamagetothe

productiftheproductishandledincorrectly.
This guarantee covers only the performance of the product itself only.
All inconvenience by the trouble of this product is not included.
*Outer appearance and specifications are subject to change without prior notice.
HOME PAGE: http://www.onosokki.co.jp/English/english.htm
WORLDWIDE
ONO SOKKI CO., LTD.
1-16-1 Hakusan, Midori-ku,
Yokohama 226-8507, Japan
Phone : 045-935-3976
Fax : 045-930-1906
E-mail : [email protected]
B00002033 / IM07062203(2) 081(MS)020
1.Power Supply
The HR-6800 operates on four Type AAA batteries or the
dedicated AC adapter (option: PB-7080).
If the batteries are consumed and the LOW mark " "
appears, replace them with new ones. Be sure to replace
all the four batteries at the same time.
Battery replacement procedure
①While pushing lightly the two (anti-slip) slots of the
battery cover with your finger, slide it to remove.
②Put the batteries properly in the battery compartment
with the correct polarity (+/-).
③Put the battery cover.
2.MI-0301 Tripod Attachment Adapter
With the use of the supplied tripod attachment adapter,
the Detector can be installed using a commercially avail-
able tripod.
Attach the Detector to and detach it from the tripod attach-
ment adapter as shown below.
1.Measurement Section
Objects under measurement: Bodies of revolution for dental in-
struments
Bodies of revolution for textile ma-
chines
Bodies of revolution for high-speed
machine tools, etc.
* The object under measurement needs
to be polarized.
Operation method : Periodic operation method
Measuring time : 50 ms + Input signal 10-period time
or less
Input frequency range : 1.6 Hz to 16.666 kHz
Measuring unit : 10 r/min (rotational speed)
Measuring range : 100 to 999990 r/min (selectable)
Measurement accuracy : Display value x (±0.02 %) ±1 count
Over range function : If the measurement value exceeds the
display range, over range " "
appears.
Upper limit rotation warning function: If the rotational speed ex-
ceeds a preset upper limit, upper limit
warning " " appears.
Trigger adjustment function: Can be adjusted using the rotary
volume on the right-hand side face of
the HR-6800 unit.
2.Detecting Element
(The sensor is not included in a price of HR-6800
)
Detection method : Electromagnetic induction
DC resistance : 25 to 40 Ω[20 ˚C]
Operating temperature range: 0 ˚C to +40 ˚C
Storage temperature range: -10 ˚C to +50 ˚C
Resistance to vibration : 19.6 m/s
2
Resistance to shock : 490 m/s
2
Mass : About 50 g (detecting element only)
Dimensions : 107 x ø14 mm
3.Display Section
Number of display digits : 5 digits
Character height : 10.2 mm
Indicator : 7-segment LCD with back light
Refresh time : About 1 s/0.5 s selectable
4.Measurement Mode
MAX (peak hold) : Displays the maximum value during mea-
surement.
MIN (peak hold) : Displays the minimum value during mea-
surement.
Others : Displays the current measurement value.
Memory function : Up to 20 measurement values can be
memorized each time the memory switch is
pressed. Since these values are stored in
non-volatile memory, they are retained even
after the power is turned OFF.
Specifications
5.Analog Output Section
【REVO】
Output contents : Output to the display value of rotational
speed.
Voltage range : 0 to F.S./0 to 1 V
(The full scale can be set with a setup
parameter.)
Conversion method : 10-bit D/A conversion method
Linearity : ±1 % of F.S.
Output refresh time : 50 ms + Input signal 10-period time or
less
Temperature stability : ±0.05 % of F.S./˚C (ZERO & SPAN)
Setup error : ±0.5 % of F.S. (adjustment setup error
at the time of shipment, ZERO & SPAN)
Load resistance : 100 kΩor more
Output connector : Super mini jack (ø2.5)
【SIG】
Output contents : Analog output for monitoring after wave-
form shaping of the sensor signal (be-
fore pulse waveform conversion)
Load resistance : 100 kΩor more
Output connector : Super mini jack (ø2.5/Commonly used
with REVO output)
6.Pulse Output Section
Output timing : Outputs one pulse for each signal de-
tection.
(For bodies of revolution with one out-
put per rotation, outputs one pulse per
rotation.)
Output voltage : Hi level=4.5 V or higher
Lo level=0.5 V or lower
Output logic : Positive logic pulse
Load resistance : 100 kΩor more
Output connector : Super mini jack (ø2.5)
7.General Specifications
Power supply : Type AAA dry battery (x4) or dedi-
cated AC adapter (PB-7080)
Continuous operating time: About 13 hours (back light OFF)
About 8 hours (back light ON)
(When alkaline batteries are used at
20 ˚C)
Battery LOW display : Lights up at about 4.5 V.
Operating temperature range: 0 ˚C to +40 ˚C
Storage temperature range: -10 ˚C to +50 ˚C
Operating humidity range: +35 to +85 %RH (without conden-
sation)
Storage humidity range : +35 to +85 %RH (without conden-
sation)
Mass : About 230 g (HR-6800 unit only, bat-
teries not included)
Dimensions : 189.5 x 66.0 x 47.5 mm (HR-6800
unit)
Before Use
Battery polarity display
4.Measurement
①Connect the HR-6800 unit with the Detector using the
supplied cable.
②Select Type A or B using the Type selector switch on the
side face.
③Slide the power switch to turn the power ON.
④Set the number of pulses per rotation (P/R) and other
parameters according to the object under measurement.
⑤Bring the Detector close to the object under measurement.
Bring the end or its side face of the Detector to the body
of revolution as shown below.
【Detection at the end】
【Detection at side face of the end】
⑥During measurement, turn the trigger level adjustment
volume so that the indicator blinks stably and the rotational
speed be displayed. Since the center value of the trigger
level adjustment volume is 5, gradually increase or
decrease the trigger level from 5 to adjust the trigger level.
⑦Perform measurement.
TRIG LEVEL
5
Options Storage
The storage temperature range of the HR-6800 is -10˚C
to +50˚C. When you store it, avoid locations where the
temperature is extremely high or low or the humidity is
high. Store it in a place which is well-ventilated and not
exposed to direct sunlight. If you do not use it for a pro-
longed period of time, be sure to remove the batteries to
prevent accident caused by battery leakage, etc.
①AX-501: Output cord
②
PB-7080: AC adapter
(IN: 100 to 120 VAC, OUT: 6 VDC)
Body of revolution
under measurement
5〜20 mm
3.Measurable Bodies of Revolution
The HR-6800 enables measurement of the following bodies
of revolution:
・A body of revolution with a magnet embedded
Most stabilized measurement is possible.
This system is recommended for a drill for rotation test.
・A body of revolution partially polarized
A magnet is rubbed on a part of the body of revolution for
polarization before measurement. If the magnetic force
weakens with time, polarization is performed again.
・A body of revolution with a bumpy shape
Since a magnet is embedded in the Detector of the HR-
6800, whether a body of revolution can be measured
depends on its shape. However, depending on the material
and shape of the body of revolution, a detected signal may
become minute making detection of rotation difficult.
Therefore, if possible, magnet-based detection is
recommended.
Note: Since the Detector has frequency characteristics,
it needs strong magnetic force particularly when
measuring low rotational speed.
Refer to the following rough standard of magnetic
force:
100 r/min: 10 mT or more
10000 r/min: 5 mT or more
100000 r/min: 1 mT or more
Note: Measurement is not possible in places where
strong magnetic field is present around the mea-
suring point.
RANGE
Hi
Lo
−
Type selection
A
B
10,000 to 999,990 r/min
100 to 50,000 r/min
200,000 to 999,990 r/min
Rotational speed range
* When the number of pulses is set to "1"
* Generally, use Type A.
Attachment Detachment
10
5〜20 mm
Applicable Standards
①CE Marking
EN61326
EN61010-1
: This mark declares the compliance with
EC instructions.

This manual suits for next models

1

Other Ono Sokki Measuring Instrument manuals

Ono Sokki CF-4700 User manual

Ono Sokki

Ono Sokki CF-4700 User manual

Ono Sokki DS-2000 Series User manual

Ono Sokki

Ono Sokki DS-2000 Series User manual

Ono Sokki FT-2500 User manual

Ono Sokki

Ono Sokki FT-2500 User manual

Ono Sokki SE-1200 User manual

Ono Sokki

Ono Sokki SE-1200 User manual

Ono Sokki HT-3200 User manual

Ono Sokki

Ono Sokki HT-3200 User manual

Ono Sokki SE-2400 User manual

Ono Sokki

Ono Sokki SE-2400 User manual

Ono Sokki TS-3200A User manual

Ono Sokki

Ono Sokki TS-3200A User manual

Ono Sokki SE-1520 User manual

Ono Sokki

Ono Sokki SE-1520 User manual

Ono Sokki DS-2000 Series User manual

Ono Sokki

Ono Sokki DS-2000 Series User manual

Ono Sokki LV-1800 User manual

Ono Sokki

Ono Sokki LV-1800 User manual

Ono Sokki DS-2000 Series User manual

Ono Sokki

Ono Sokki DS-2000 Series User manual

Ono Sokki GE-1400 User manual

Ono Sokki

Ono Sokki GE-1400 User manual

Ono Sokki DS 0297 User manual

Ono Sokki

Ono Sokki DS 0297 User manual

Ono Sokki DS-2000 Series User manual

Ono Sokki

Ono Sokki DS-2000 Series User manual

Ono Sokki HT-5500 User manual

Ono Sokki

Ono Sokki HT-5500 User manual

Ono Sokki CF-7200 Reference guide

Ono Sokki

Ono Sokki CF-7200 Reference guide

Ono Sokki FT-7100 User manual

Ono Sokki

Ono Sokki FT-7100 User manual

Ono Sokki HT-6200 User manual

Ono Sokki

Ono Sokki HT-6200 User manual

Ono Sokki HT-5200 User manual

Ono Sokki

Ono Sokki HT-5200 User manual

Ono Sokki FT-7200 User manual

Ono Sokki

Ono Sokki FT-7200 User manual

Ono Sokki HT-6200 User manual

Ono Sokki

Ono Sokki HT-6200 User manual

Popular Measuring Instrument manuals by other brands

az-instrument 8921 Operation manuals

az-instrument

az-instrument 8921 Operation manuals

ELBRO LX-1148SD Operation manual

ELBRO

ELBRO LX-1148SD Operation manual

Hauber Elektronik HE100 Series operating instructions

Hauber Elektronik

Hauber Elektronik HE100 Series operating instructions

Victron energy ET340 manual

Victron energy

Victron energy ET340 manual

Dostmann Electronic 31.5008.02 Operating	 instruction

Dostmann Electronic

Dostmann Electronic 31.5008.02 Operating instruction

Rohde & Schwarz ZNL6-COM Getting started

Rohde & Schwarz

Rohde & Schwarz ZNL6-COM Getting started

Sharper Image 208667 user guide

Sharper Image

Sharper Image 208667 user guide

SOR 701 Series manual

SOR

SOR 701 Series manual

Elgama EPQS user manual

Elgama

Elgama EPQS user manual

Eaton EMC3P-P2C1 Product profile

Eaton

Eaton EMC3P-P2C1 Product profile

Trotec BY10 operating manual

Trotec

Trotec BY10 operating manual

Continental Refrigerator ContiPressureCheck Brief instructions

Continental Refrigerator

Continental Refrigerator ContiPressureCheck Brief instructions

Emerson Rosemount 3051 quick start guide

Emerson

Emerson Rosemount 3051 quick start guide

Simex CCM-P-01 operating manual

Simex

Simex CCM-P-01 operating manual

Eastron SDM630-EV user manual

Eastron

Eastron SDM630-EV user manual

Optoma MirrorBoard Q300 user guide

Optoma

Optoma MirrorBoard Q300 user guide

Kobold PSR series operating instructions

Kobold

Kobold PSR series operating instructions

Steren HER-400 manual

Steren

Steren HER-400 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.