Warranty
All hardware purchased from Opentrons is covered
under a 1-year standard warranty. Opentrons warrants
to the end-user of the products that they will be free
of manufacturing defects due to part quality issues or
poor workmanship and also warrants that the products
will materially conform to Opentrons’ published
specications.
Support
Opentrons Support can help you with questions about
our products and services. If you discover a defect, or
believe your product is not functioning to published
specications, contact us at support@opentrons.com.
Please have the pipette’s serial number available when
contacting support. You can nd the serial number
on the back of the pipette or via the touchscreen (tap
Instruments and tap the pipette name).
© OPENTRONS 2023
Opentrons Flex™(Opentrons Labworks, Inc.)
Registered names, trademarks, etc. used in this document, even when not
specically marked as such, are not to be considered unprotected by law.
Installation and Maintenance
Information about pipette installation and care.
ATTACH
Instructions on the touchscreen, or in the Opentrons
App, will guide you through the attachment process.
On the touchscreen, tap Instruments and then tap
an empty mount. To attach your pipette:
1. Clear the deck area so it is free from
obstructions. The gantry will move during the
installation process.
2. Connect the pipette to the left or right mount
on the gantry. To attach, align the mounting
holes on the back of the pipette with the pins
on the gantry.
3. Secure the pipette by tightening the two 2.5
mm mounting screws. These are captive screws
recessed into holes located on the front of the
pipette.
CALIBRATE
The pipette requires calibration before use.
Instructions on the touchscreen, or in the Opentrons
App, will guide you through the calibration process.
To calibrate your pipette:
1. Clear the deck area so it is free from
obstructions. The gantry and pipette will move
during calibration.
2. Attach the calibration probe to the pipette and
start the automated calibration procedure.
3. Remove and store the probe when calibration is
complete. Your pipette is ready for use.
CLEAN
Clean the outside of the pipette with a soft cloth and
isopropyl alcohol or ethyl alcohol. Do not use acetone
or other aggressive solvents. For information about
cleaning and decontamination, scan the QR code or
search for “Flex pipette cleaning” on
support.opentrons.com.