2.3.3 General Performance
NOTES: ※1:CEmarkcompliantmodelsapplyonlytoDCpowered,non-explosionproofspecifications.
※2:Fortransmissionlengthofoutputsignal,seesec.5.7"OutputSignalandPowerTerminalldentification".
※3:Ifsignalcablelengthexceeds200meters,consultthefactory.
※4:Thisconfigurationmeetstherequirementsforexplosionproofcertifiedsystemapprovedforthesensorunitand
transmitter combined.
※5:Thisconfigurationmeetstherequirementsforexplosionproofcertifiedequipmentapprovedonthetransmitter
alone (valid only for use in combination with high temperature service sensor except for CN003).
※6:Thisexplosionproofconfigurationappliesonlytoremotelylocatedmodels.
※7:Bell202signalissuperimposedoveranaogoutput1(whencombinedwithSmartCommunicationUnit"LinkTop").
※8:Flowsignalsbelow0.3%ofthemaximumallowableflowrateiscutoff.
※9:Ifvoltagepulseisselected,analogoutputisnotavailable.
※10:Oferroroutputs,"autozeroinprogress"statusoutputcanbesetup.
※:Insulationresistanceanddielectricstrengthtestsareunacceptable.
Item Description
Power source 85to264VAC,50/60Hzor20to30VDC(※1)
Power consumption 25VA max. or 10W max.
Ambient temperature -20to+50℃(freeofdewcondensation)
Transmissionlength(remotely-located)(※2) 200metersmax.(usingdedicatedinterconnectcable)(※3)
Applicable EU directives EMC directive : 89/336/EEC, 92/31/EEC, 93/68/EEC
ATEX Directive : 94/9/EC
Applicable EN directives EMC : EN55011 : 1998/A1 : 1999 Group 1 Class B EN61000-6-2 : 1999
ATEX : EN50021 : 1999
Explosionproof housing TIIS
Combined explosionproof Exd [ib] IIB T3 or Exd [ib] IIB T4
(CN080...Exd[ib]IIAT3orExd[ib]IIAT4)(※4)
Exd(ib)IIBT2(※5)
ATEX EexnC[nL]IIBT3(※6)
Dusttight and waterproof IP66
Installation Integrally mounted or remotely mounted (remote mount only for CN003 and one
combined with high temp. sensor unit)
Finish Munsell 7.5G 7/2.5
Display LCD display (7-segment 8-digit) provided
Weight Integrally mounted type 4.6kg app. Remotely mounted type 5.8kg app.
Communication interface Bell202(underHARTprotocol)(※7)
Pulseoutput(※8)
Open collector output (10V min. to 30V max., 50mA DC)
orvoltagepulse(※9)"0":1.5Vmax."1":15Vmin.
Outputimpedance2.2kΩ
FS : 0.1 to10000Hz
Analogoutput(※8)
4to20mADCMax.load600Ω
Select two outputs from instant flowrate (mass or volume), temperature, or density.
Damping (default) : Flowrate 1 s, density 4 s, temperature 2.5 s.
Status output Open collector output Normal : ON Error : OFF (30V, 50mA max.)
Selectoneoutputfromerror(default)(※10),flowdirection,orhi/lowalarm.
Status input
Contact-closureinput(Form"a"contact)Close:200Ωmax.Open:100Ωmin.
Select one input from function OFF (default), remote zero, total count reset, or 0%
signal lock.
Item Description
Model CN003 CN006 CN010 CN015 CN025 CN050 CN080
Flowrate
Max. service range, kg/min 0 to 1.2 0 to 6 0 to 20 0 to 60 0 to 180 0 to 650 0 to 2000
Max. permissible range, kg/min 0 to 2.4 0 to 12 0 to 40 0 to 120 0 to 360 0 to 3000 0 to 4000
Min. range, kg/min 0 to 0.06 0 to 0.3 0 to 1 0 to 3 0 to 9 0 to 32.5 0 to 100
Factory
calibration
accuracy
Liquid 〔±0.1% ± zero stability error〕of reading
Gas 〔±0.5% ± zero stability error〕of reading
Reproducibility Liquid 〔±0.05% ±1/2 zero stability error〕of reading
Gas 〔±0.25% ±1/2 zero stability error〕of reading
Zero stability, kg/min 0.00012 0.0006 0.002 0.006 0.018 0.065 0.2
Density
(Liquid) Factory calib. accuracy(option) 0.3 to 2 g/mL
±0.001 g/mL
Analog accuracy Respective accuracies ±0.1% of full scale
2.3.2 Transmitter (CT9401 : Sophisticated model)