manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Physio Logic
  6. •
  7. Thermometer
  8. •
  9. Physio Logic DiGiPro User manual

Physio Logic DiGiPro User manual

Other Physio Logic Thermometer manuals

Physio Logic Accuflex PRO Plus User manual

Physio Logic

Physio Logic Accuflex PRO Plus User manual

Physio Logic Accuflex 5 VU User manual

Physio Logic

Physio Logic Accuflex 5 VU User manual

Physio Logic QuickScan 016-670 User manual

Physio Logic

Physio Logic QuickScan 016-670 User manual

Physio Logic Accuflex Pro User manual

Physio Logic

Physio Logic Accuflex Pro User manual

Physio Logic Insta-Therm User manual

Physio Logic

Physio Logic Insta-Therm User manual

Physio Logic DiGiPro User manual

Physio Logic

Physio Logic DiGiPro User manual

Physio Logic Insta-Therm User manual

Physio Logic

Physio Logic Insta-Therm User manual

Physio Logic Accuflex Pro User manual

Physio Logic

Physio Logic Accuflex Pro User manual

Physio Logic Accuflex User manual

Physio Logic

Physio Logic Accuflex User manual

Physio Logic ProScan 016-660 User manual

Physio Logic

Physio Logic ProScan 016-660 User manual

Physio Logic Accuflex 5 User manual

Physio Logic

Physio Logic Accuflex 5 User manual

Physio Logic Accuflexpro User manual

Physio Logic

Physio Logic Accuflexpro User manual

Popular Thermometer manuals by other brands

Benetech GM900 instruction manual

Benetech

Benetech GM900 instruction manual

Parton PTD-400 manual

Parton

Parton PTD-400 manual

GAME 15900 owner's manual

GAME

GAME 15900 owner's manual

scigiene FP-180 instruction manual

scigiene

scigiene FP-180 instruction manual

Ametek Land UNO U1 600/1600C-L user guide

Ametek Land

Ametek Land UNO U1 600/1600C-L user guide

Delton 981A instruction manual

Delton

Delton 981A instruction manual

Taylor 5286498 instruction manual

Taylor

Taylor 5286498 instruction manual

Baker MM2P instruction manual

Baker

Baker MM2P instruction manual

Chino IR-FA Series instructions

Chino

Chino IR-FA Series instructions

Vive DMD1010 user manual

Vive

Vive DMD1010 user manual

Tarmo 320281 manual

Tarmo

Tarmo 320281 manual

Citizen CTD 504 instruction manual

Citizen

Citizen CTD 504 instruction manual

TFA 14.1500 instruction manual

TFA

TFA 14.1500 instruction manual

Gotemp THB4002 user guide

Gotemp

Gotemp THB4002 user guide

GREISINGER G 1202 Quick reference guide

GREISINGER

GREISINGER G 1202 Quick reference guide

Amprobe TMD-53W user manual

Amprobe

Amprobe TMD-53W user manual

Trotec BP10 operating manual

Trotec

Trotec BP10 operating manual

Radiance instruments RT616 Series product manual

Radiance instruments

Radiance instruments RT616 Series product manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

016-625
Features:
• For use orally, rectally or underarm
• Accurate within 0.1°C / 0.2°F
• Displays in Celsius
• Memory recall / display of last reading
• Soft beep signals peak temperature
• Waterproof / washable for easy cleaning
• Auto shut-off for extended battery life
• Includes storage case
Note:
Note: The patient’s mouth must remain closed during the measurement for to
ensure the most accurate reading. Opening the mouth could result in an extended
measurement time and may affect the reading.
Taking a rectal temperature
Approximate measurement time: 30 seconds.
Note: Rectal temperatures are an appropriate and reliable method for infants and
small children. Otherwise, this method should only be used when it is impossible
or impractical to take an oral or axillary temperature.
. Apply a water-soluble lubricant to the tip and insert the probe to a MAXIMUM
of 1/2” (1.25 cm) into the rectum.
Note: The t
Note: Rectal temperatures are generally 0.6°C (1.0°F) higher than oral readings.
Taking an axillary temperature (under the arm)
Approximate measurement time: 101 seconds.
Wipe the underarm with a dry towel.
2.
4. Position the arm across the chest so the probe tip is well covered by the arm.
This also insures the probe is not affected by the room's air.
Note: Axillary temperatures are generally 0.6°C (1.0°F) lower than oral readings.
The temperature reading may increase - even after the “°C” has stopped
flashing - if the thermometer is moved into a warmer place. If this happens,
allow a few seconds for the thermometer to re-stabilize and show the
final reading.
1. Press the ON/OFF button to turn the unit on.
The unit will show the displays indicated on
the left and a °C symbol will start flashing,
which means it is now taking readings.
2
3. Once the “°C” symbol on the display has stopped flashing (usually within 60
seconds), the peak stable temperature will be indicated.
emperature reading may increase - even after the °C has stopped
flashing - if the thermometer is moved into a warmer place. If this happens,
allow a few seconds for the thermometer to re-stabilize and show the
final reading.
1.
Press the ON/OFF button to turn the unit on.
The unit will show the displays indicated on
the left and a °C symbol will start flashing,
which means it is now taking readings.
3. Place the probe tip under the arm so the tip is touching the skin, with the
thermometer perpendicular to the body.
5. Once the °C symbol on the display has stopped flashing (usually within 60
seconds), the peak stable temperature will be indicated.
Note: The temperature reading may increase even after the “°C” has stopped
flashing if the thermometer is moved into a warmer place. If this happens,
allow a few seconds for the thermometer to re-stabilize and show the
final reading.
To turn the unit off
Press the ON/OFF button. However, the unit will automatically shut off after
approximately 10 minutes if it is not turned off manually.
To recall memory
To recall memory of the last reading, hold the ON/OFF button for approx. 2
seconds when turning the thermometer on.
General care
1. Clean the thermometer before and after each use.
2. Store the unit in the protective case when it is not in use.
3. Do not bite the probe.
4. Do not store the unit where it will be exposed to direct sunlight, dust or
humidity. Avoid extreme temperatures.
5. Dropping or subjecting the thermometer to strong shocks should be avoided.
6. Do not attempt to disassemble the unit.
7. The thermometer should only be used under the supervision of an adult.
8. The patient should not walk, run or talk during temperature taking.
Specifications
Display: Liquid Crystal Display
Display resolution: 4 Digits (1/10th of a degree increments)
Measurement range: 32°C to 44°C
Accuracy: 0.1°C to 35.0°C
and 39.0°C
± 0.2°C otherwise
Average measurement time: Oral: 60 seconds, rectal: 30 seconds,
axillary: 101 seconds
Tone: Thermometer beeps softly when activated and
deactivated. A series of four beeps signal
when peak temperature is reached
Memory: Displays last measured temperature
Battery: One, 1.55V, alkaline Manganese battery LR41 or
UCC 192
Battery life: Approximately 90 hours of continuous operation
or approx. 18 months of 10 minutes per day
Automatic shut-off: After approximately 10 minutes
Ambient operating temperature: 0°C to 50°C (humidity: 10 % ~ 83 %)
Dimensions: 12.6 cm x 1.8 cm x 1.0 cm
Weight : 7 grams with battery inside, approximately
Specifications are subject to change without notice.
For further questions or comments, visit www.amgphysiologic.com
Cleaning instructions
Clean the thermometer with one of the solutions listed below, then store it in
its plastic case. Any other cleaning method may damage the thermometer.
1. Press the ON/OFF button to turn the unit on.
The unit will show the displays indicated on
the left and a °C symbol will start flashing,
which means it is now taking readings.
3. Once the °C symbol on the display has stopped flashing (usually within 60
seconds), the peak stable temperature will be indicated.
How to take the patient’s temperature
1. Clean and disinfect the thermometer according to the instructions provided
above.
2. Decide whether the measurement will be oral, rectal or underarm.
3. If desired, place sterile probe cover on the tip of the thermometer (not
necessary)
Note: Measurement time may vary based on the selected site.
Taking an oral temperature
Approximate measurement time: 60 seconds.
Note: Drinking hot or cold fluids, exercising,
smoking or performing other activities may raise
or lower your temperature. It is important for the
patient to relax for approximately 5 minutes,
mouth closed, prior to taking a reading.
2. Place the probe tip under the tongue as near as possible to the heat pocket,
close the patient’s mouth and avoid moving the unit.
TM
®
Thank you for purchasing the Physio Logic DiGiPro
Thermometers. Please read the following instructions to achieve
the most accurate temperature readings and safe operation.
REV. 1 0506 P/N 804-039
User Guide
Digital Clinical Thermometer
HEAT
POCKET
HEAT
POCKET
TM
Solution Dilution Method
Isopropyl Alcohol (IPA) 70% - 90% Wipe with soaked cloth
Ethanol 70% - 80% Wipe with soaked cloth
(do not soak unit)
Sodium Hypochlorite 0.02% - 0.05%
Gluconate Chlorhexidine 0.1 % - 0.5%
Benzethonium Chloride 0.1% Soak for up to 30 minutes
Benzalkonium Chloride 0.1%
Glutaraldehyde 0.5% - 2%
Alkylpolyaminoethylglycine 0.05% - 0.2%
PROBE TIP
ON/OFF BUTTON
DIGITAL DISPLAY WINDOW
TM
Features of the DiGiPro digital clinical thermometer
016-625
Features:
• For use orally, rectally or underarm
• Accurate within 0.1°C / 0.2°F
• Displays in Celsius
• Memory recall / display of last reading
• Soft beep signals peak temperature
• Waterproof / washable for easy cleaning
• Auto shut-off for extended battery life
• Includes storage case
Note:
Note: The patient’s mouth must remain closed during the measurement for to
ensure the most accurate reading. Opening the mouth could result in an extended
measurement time and may affect the reading.
Taking a rectal temperature
Approximate measurement time: 30 seconds.
Note: Rectal temperatures are an appropriate and reliable method for infants and
small children. Otherwise, this method should only be used when it is impossible
or impractical to take an oral or axillary temperature.
. Apply a water-soluble lubricant to the tip and insert the probe to a MAXIMUM
of 1/2” (1.25 cm) into the rectum.
Note: The t
Note: Rectal temperatures are generally 0.6°C (1.0°F) higher than oral readings.
Taking an axillary temperature (under the arm)
Approximate measurement time: 101 seconds.
Wipe the underarm with a dry towel.
2.
4. Position the arm across the chest so the probe tip is well covered by the arm.
This also insures the probe is not affected by the room's air.
Note: Axillary temperatures are generally 0.6°C (1.0°F) lower than oral readings.
The temperature reading may increase - even after the “°C” has stopped
flashing - if the thermometer is moved into a warmer place. If this happens,
allow a few seconds for the thermometer to re-stabilize and show the
final reading.
1. Press the ON/OFF button to turn the unit on.
The unit will show the displays indicated on
the left and a °C symbol will start flashing,
which means it is now taking readings.
2
3. Once the “°C” symbol on the display has stopped flashing (usually within 60
seconds), the peak stable temperature will be indicated.
emperature reading may increase - even after the °C has stopped
flashing - if the thermometer is moved into a warmer place. If this happens,
allow a few seconds for the thermometer to re-stabilize and show the
final reading.
1.
Press the ON/OFF button to turn the unit on.
The unit will show the displays indicated on
the left and a °C symbol will start flashing,
which means it is now taking readings.
3. Place the probe tip under the arm so the tip is touching the skin, with the
thermometer perpendicular to the body.
5. Once the °C symbol on the display has stopped flashing (usually within 60
seconds), the peak stable temperature will be indicated.
Note: The temperature reading may increase even after the “°C” has stopped
flashing if the thermometer is moved into a warmer place. If this happens,
allow a few seconds for the thermometer to re-stabilize and show the
final reading.
To turn the unit off
Press the ON/OFF button. However, the unit will automatically shut off after
approximately 10 minutes if it is not turned off manually.
To recall memory
To recall memory of the last reading, hold the ON/OFF button for approx. 2
seconds when turning the thermometer on.
General care
1. Clean the thermometer before and after each use.
2. Store the unit in the protective case when it is not in use.
3. Do not bite the probe.
4. Do not store the unit where it will be exposed to direct sunlight, dust or
humidity. Avoid extreme temperatures.
5. Dropping or subjecting the thermometer to strong shocks should be avoided.
6. Do not attempt to disassemble the unit.
7. The thermometer should only be used under the supervision of an adult.
8. The patient should not walk, run or talk during temperature taking.
Specifications
Display: Liquid Crystal Display
Display resolution: 4 Digits (1/10th of a degree increments)
Measurement range: 32°C to 44°C
Accuracy: 0.1°C to 35.0°C
and 39.0°C
± 0.2°C otherwise
Average measurement time: Oral: 60 seconds, rectal: 30 seconds,
axillary: 101 seconds
Tone: Thermometer beeps softly when activated and
deactivated. A series of four beeps signal
when peak temperature is reached
Memory: Displays last measured temperature
Battery: One, 1.55V, alkaline Manganese battery LR41 or
UCC 192
Battery life: Approximately 90 hours of continuous operation
or approx. 18 months of 10 minutes per day
Automatic shut-off: After approximately 10 minutes
Ambient operating temperature: 0°C to 50°C (humidity: 10 % ~ 83 %)
Dimensions: 12.6 cm x 1.8 cm x 1.0 cm
Weight : 7 grams with battery inside, approximately
Specifications are subject to change without notice.
For further questions or comments, visit www.amgphysiologic.com
Cleaning instructions
Clean the thermometer with one of the solutions listed below, then store it in
its plastic case. Any other cleaning method may damage the thermometer.
1. Press the ON/OFF button to turn the unit on.
The unit will show the displays indicated on
the left and a °C symbol will start flashing,
which means it is now taking readings.
3. Once the °C symbol on the display has stopped flashing (usually within 60
seconds), the peak stable temperature will be indicated.
How to take the patient’s temperature
1. Clean and disinfect the thermometer according to the instructions provided
above.
2. Decide whether the measurement will be oral, rectal or underarm.
3. If desired, place sterile probe cover on the tip of the thermometer (not
necessary)
Note: Measurement time may vary based on the selected site.
Taking an oral temperature
Approximate measurement time: 60 seconds.
Note: Drinking hot or cold fluids, exercising,
smoking or performing other activities may raise
or lower your temperature. It is important for the
patient to relax for approximately 5 minutes,
mouth closed, prior to taking a reading.
2. Place the probe tip under the tongue as near as possible to the heat pocket,
close the patient’s mouth and avoid moving the unit.
TM
®
Thank you for purchasing the Physio Logic DiGiPro
Thermometers. Please read the following instructions to achieve
the most accurate temperature readings and safe operation.
REV. 1 0506 P/N 804-039
User Guide
Digital Clinical Thermometer
HEAT
POCKET
HEAT
POCKET
TM
Solution Dilution Method
Isopropyl Alcohol (IPA) 70% - 90% Wipe with soaked cloth
Ethanol 70% - 80% Wipe with soaked cloth
(do not soak unit)
Sodium Hypochlorite 0.02% - 0.05%
Gluconate Chlorhexidine 0.1 % - 0.5%
Benzethonium Chloride 0.1% Soak for up to 30 minutes
Benzalkonium Chloride 0.1%
Glutaraldehyde 0.5% - 2%
Alkylpolyaminoethylglycine 0.05% - 0.2%
PROBE TIP
ON/OFF BUTTON
DIGITAL DISPLAY WINDOW
TM
Features of the DiGiPro digital clinical thermometer