RADWAG PUE C32 User manual


2
AUGUST 2020

3
PRECAUTIONS
Prior to installation, use or maintenance activities, carefully read this user
manual. Use the PUE C32 Indicator only as intended.
Prior to the first use, carefully read this user manual. Use the device only
as intended.
Protect the indicator against considerable temperature variation, solar and
UV radiation, substances causing chemical reactions.
The PUE C32 indicator must not be operated in hazardous areas
endangered with explosion of gases, and in dusty environments.
In case of damage, immediately unplug the device from the mains.
Scales to be decommissioned must be decommissioned in accordance
with valid legal regulations.
Do not let battery discharge in case of prolonged storage of the device
in low temperature.
A worn out battery can be replaced only by the manufacturer or by the
authorized service.
Accumulators do not belong to regular household waste. The European
legislation requires discharged accumulators to be collected and disposed
separately from other communal waste with the aim of being recycled.
Symbols on batteries identify harmful compounds: Pb = lead, Cd =
cadmium, Hg = mercury. Dear user, you are obliged to dispose of the worn
out batteries as regulated.

4
Contents
1. INTENDED USE .................................................................................................................................................5
2. WARRANTY CONDITIONS ...............................................................................................................................5
3. MAINTENANCE ACTIVITIES ............................................................................................................................5
3.1. Cleaning ABS Components.........................................................................................................................6
3.2. Cleaning Stainless Steel Components........................................................................................................6
4. MECHANICAL DESIGN.....................................................................................................................................6
4.1. Dimensions..................................................................................................................................................6
4.2. Connectors Arrangement ............................................................................................................................7
4.3. RS232 Connectors......................................................................................................................................7
4.4. Inputs / Outputs ...........................................................................................................................................7
4.4.1. Technical Specifications...................................................................................................................8
4.4.2. I/O Schematic Diagrams ..................................................................................................................8
4.5. Operation Panel...........................................................................................................................................9
4.6. Technical Specifications............................................................................................................................10
5. INDICATOR INSTALLATION ..........................................................................................................................10
5.1. Unpacking and Installation........................................................................................................................10
5.2. Start-Up .....................................................................................................................................................11
5.3. Battery Status............................................................................................................................................11
6. HOME SCREEN ...............................................................................................................................................12
6.1. Top Bar......................................................................................................................................................12
6.2. Weighing Result Window ..........................................................................................................................13
6.3. Workspace.................................................................................................................................................13
6.4. Pictograms.................................................................................................................................................13
7. OPERATING THE MENU.................................................................................................................................13
7.1. Entering the Menu .....................................................................................................................................13
7.2. Menu Keys.................................................................................................................................................14
7.3. Entering Numbers / Text ...........................................................................................................................15
7.3.1. Numerical Box................................................................................................................................15
7.3.2. Text Box..........................................................................................................................................16
7.3.3. Diacritical Sign Table......................................................................................................................18
7.3.4. Special Sign Table..........................................................................................................................18
7.4. Return to Weighing....................................................................................................................................19
8. INSTALLER INSTRUCTION............................................................................................................................ 19
8.1. Connecting 6-Wire Load Cell ....................................................................................................................19
8.2. Connecting 4-Wire Load Cell ....................................................................................................................20
8.3. Load Cell Shield Connection.....................................................................................................................21
9. FACTORY SETUP............................................................................................................................................ 21
9.1. Factory Setup Access ...............................................................................................................................21
9.2. Global Parameters.....................................................................................................................................22
9.3. Factory Parameters...................................................................................................................................23
9.4. Return to Weighing....................................................................................................................................25
9.5. Factory Adjustment....................................................................................................................................25
9.5.1. External Adjustment .......................................................................................................................25
9.5.2. Start Mass Determination...............................................................................................................25
9.6. Linearity Correction ...................................................................................................................................25
9.6.1. Linearity Determination ..................................................................................................................25
9.6.2. Corrections .....................................................................................................................................26
9.6.3. Deleting Linearity............................................................................................................................ 27
9.7. Gravitational Coefficient............................................................................................................................27
10. DIAGRAMS OF CONNECTION CABLES.....................................................................................................27

5
1.INTENDED USE
The PUE C32 weighing indicator is a part of construction of industrial scales
operating on the basis of load cells. The indicator features an ABS plastic
housing and a 5'' colour graphic display ensuring perfect readability. It is
operated using 22-key membrane keypad equipped with programmable
function keys.
Standard design offers two RS232 interfaces, USB type A, USB type B,
Ethernet, wireless communication, 2 proximity sensors, 4 I/O. The indicator can
be operated in areas where there is no access to the mains. It can optionally be
supplied by NiMH battery (nickel-metal-hydride) of 1800 - 2800 mAh capacity.
The device integrates with receipt and label printers, barcode scanners and PC
accessories (mouse, keyboard, USB flash drive).
2.WARRANTY CONDITIONS
A. RADWAG feels obliged to repair or exchange all elements that appear to
be faulty by production or by construction.
B. Defining defects of unclear origin and means of their elimination can only
be realized with the assistance of the manufacturer and the user
representatives.
C. RADWAG does not bear any responsibility for damage or losses
resulting from unauthorized or inadequate performing of production
or service processes.
D. The warranty does not cover:
mechanical damage caused by product exploitation other than
intended, damage of thermal and chemical origin, damage caused by
lightning, overvoltage in the power network or other random event,
inappropriate cleaning habits.
E. Loss of warranty takes place if:
a repair is carried out outside RADWAG authorized service point,
service claims intrusion into mechanical or electronic construction by
unauthorized people,
the scale does not bear security seal stickers.
F. Warranty conditions outline the warranty period for rechargeable
batteries attached to the device for 12 months.
G. For detailed warranty conditions read the warranty certificate.
H. Contact with the central authorized service: +48 (48) 386 63 30.
3.MAINTENANCE ACTIVITIES
In order to ensure safety in the course of cleaning, it is necessary to disconnect
the device from the mains. With this condition met, uninstall the weighing pan
and other detachable components.
Other manuals for PUE C32
3
Table of contents
Other RADWAG Accessories manuals

RADWAG
RADWAG HRP 16 User manual

RADWAG
RADWAG PUE HY10 User manual

RADWAG
RADWAG PUE C315 User manual

RADWAG
RADWAG MAS Y User manual

RADWAG
RADWAG PUE H315 User manual

RADWAG
RADWAG PUE 5.15 User manual

RADWAG
RADWAG PUE 7.1 User manual

RADWAG
RADWAG THB Series User manual

RADWAG
RADWAG PUE C/31 User manual

RADWAG
RADWAG PUE HX5.EX Mounting instructions