Resodyn LabRAM I User manual

Resodyn Acoustic Mixers 130 N. Main Street Butte, MT, 59701 USA +1-406-497-5333 ResodynMixers.com
LabRAM I
Installation
and User
Manual

LabRAM I Installation and User Manual
Resodyn Acoustic Mixers 2 Rev. 072420
Table of Contents
Revision History......................................................................................................................... 4
1Safety............................................................................................................................. 5
1.1 General Safety Considerations .................................................................................... 5
1.2 Warnings and Cautions................................................................................................ 5
1.3 Important Safety Notes................................................................................................ 6
1.4 Location Advisory ........................................................................................................ 6
1.5 Component Replacement ............................................................................................ 6
1.6 Customer Responsibility .............................................................................................. 6
1.7 Guards and Safety Devices ......................................................................................... 7
1.8 Handling/Lifting............................................................................................................ 7
2ResonantAcoustic®Mixing (RAM) Basics....................................................................... 8
3Definitions .....................................................................................................................10
4Installation.....................................................................................................................11
4.1 Unpacking, Lifting, and Placing LabRAM I..................................................................11
4.2 Using the Retractable Handles....................................................................................12
4.3 Mounting.....................................................................................................................12
4.4 Lid Operation..............................................................................................................13
4.5 Remove the Shipping Lock Bar...................................................................................14
4.6 Install Vessel Holder Assembly...................................................................................15
4.7 Electrical Connections ................................................................................................17
5Technical Support .........................................................................................................17
6System Specifications ...................................................................................................18
6.1 Voltage Level..............................................................................................................18
6.2 Power Consumption....................................................................................................18
7System Overview ..........................................................................................................19
Acoustic Enclosure................................................................................................................20
7.1 Touch Screen Human Machine Interface (HMI)..........................................................20
7.2 Standard Vessel Holder Fixture and Vessels..............................................................21
7.3 Processing Accessories..............................................................................................22
8Recipe Demonstration Quick Start.................................................................................23
8.1 Setup..........................................................................................................................23
8.1.1 Open the Lid.....................................................................................................23
8.1.2 Loosen the Vessel Holder ................................................................................23
8.1.3 Secure Mixing Vessel in the Vessel Holder ......................................................23
8.1.4 Close the Lid....................................................................................................23
8.2 Power-Up....................................................................................................................24
8.3 Login...........................................................................................................................24
8.4 Select the Recipe........................................................................................................25

LabRAM I Installation and User Manual
Resodyn Acoustic Mixers 3 Rev. 072420
8.5 Run the Recipe...........................................................................................................26
8.6 Completion .................................................................................................................27
9HMI (Human Machine Interface) –Overview.................................................................28
9.1 Main Screen................................................................................................................28
9.2 Login...........................................................................................................................29
9.3 Operational Modes (Recipe and Auto)........................................................................29
9.4 Alarming .....................................................................................................................29
9.5 Setup and Configuration Screens ...............................................................................29
9.6 Viewer.........................................................................................................................30
10 Detailed HMI Operation –Security and Login................................................................30
10.1 Login Levels and Privileges ........................................................................................30
10.2 Screen Operations - Logging In/Out............................................................................31
10.3 Log Files.....................................................................................................................32
11 HMI Operation –Modes of Operation............................................................................33
11.1 Working in Recipe Mode.............................................................................................33
11.2 Screen Operations - Auto Mode..................................................................................36
12 HMI Operation –Alarms................................................................................................41
12.1 Screen Operations - Alarms........................................................................................41
13 HMI Operation –Viewer................................................................................................42
13.1 Screen Operations - Viewer........................................................................................42
14 HMI Operation –Setup and Configuration.....................................................................44
14.1 Status Tab..................................................................................................................45
14.2 Calibration Tab ...........................................................................................................45
14.3 Recipe Tab .................................................................................................................45
14.3.1 Screen Operations –Edit an Existing Recipe...................................................45
14.3.2 Screen Operations –Create New, Rename, and Delete...................................49
14.4 Setup1 Tab.................................................................................................................54
14.4.1 Auto Timer Group Box......................................................................................54
14.4.2 Machine Options Group Box ............................................................................54
14.4.3 Operational Options Group Box........................................................................55
14.5 Setup2 Tab.................................................................................................................55
14.5.1 Screen Operations –User Management ..........................................................56
14.6 Setting Date and Time................................................................................................57
14.7 Config Tab..................................................................................................................57
14.7.1 Screen Operations –Edit a Configuration ........................................................58
14.7.2 Screen Operations –Create a Configuration....................................................59
14.7.3 Configuration Usage.........................................................................................60
14.8 Data Tab.....................................................................................................................61
15 Care and Maintenance..................................................................................................62
15.1 Preventive Maintenance .............................................................................................62

LabRAM I Installation and User Manual
Resodyn Acoustic Mixers 4 Rev. 072420
15.2 Cleaning Procedures ..................................................................................................62
15.3 Accelerometer Calibration...........................................................................................62
15.4 Maintenance...............................................................................................................62
15.4.1 Major Repairs...................................................................................................62
15.4.2 Recommended Maintenance Schedule............................................................63
15.5 Replacement of Serviceable Parts..............................................................................63
15.5.1 Fuse Replacement...........................................................................................63
15.5.2 Bellows Replacement.......................................................................................64
15.5.3 Accelerometer Replacement............................................................................64
16 Troubleshooting and Service.........................................................................................65
Revision History
Periodically, as required, this manual will be updated to correct errors and document added
features. Table 1 documents the revision number, description, and release date.
Revision
Description
Date
Rev 1.0.080719
Initial Release
7 August 2019
Table 1 Manual Revision History

LabRAM I Installation and User Manual
Safety
Resodyn Acoustic Mixers 5 Rev. 072420
1 Safety
Every effort has been made to assure that the LabRAM I is easy to use, reliable, and safe. This
section outlines the general safety considerations and defines caution and warning symbols
used throughout this manual.
1.1 General Safety Considerations
For safe operation, the LabRAM I must be operated only within the limits outlined in the system
specifications. The following classifications define acceptable use for the LabRAM I.
•Indoor use only.
•Main supply voltage fluctuations are not to exceed ±10% of the nominal supply
voltage.
•This equipment is suitable for continuous operation.
•For use in non-hazardous environments as classified by the United States National
Electrical Code (NEC).
1.2 Warnings and Cautions
Throughout the manual, the following symbols are used to identify warnings and cautions:
The caution symbol indicates a potentially hazardous situation which could result
in minor injury or damage to the product.
•The high voltage symbol indicates the possibility of electrical shock.
•This symbol means to turn off the electrical supply before removing any cover
with the symbol attached. Failure to do so may cause exposure to electrical
shock hazard.
•Access to electrical components should only be carried out by a licensed
electrician or qualified electrical engineer.
•NEVER attempt tomodify anyelectrical components or circuits asthismay affect
the safe operation of the machine.
Electrostatic Discharge (ESD) sensitive components.

LabRAM I Installation and User Manual
Safety
Resodyn Acoustic Mixers 6 Rev. 072420
1.3 Important Safety Notes
This equipment should only be operated by trained and qualified personnel.
Mixing operations can generate heat and internal pressure depending on the
material and mixer accelerations. Do not mix materials longer than what has been
characterized as safe.
Do not attempt to run the machine without a Vessel Holder secured to the resonator
plate, and a properly installed vessel. Equipment damage may result.
Do not disable or tamper with any safety items, such as safety interlock switches,
located on the LabRAM I system.
NOTE: Changes in noise level or frequency are common during mixing operations.
If rattling, metal-on-metal, or obnoxious noises occur, turn off the LabRAM I and call
or email Resodyn Acoustic Mixers Technical Support at +1-406-497-5333 or
1.4 Location Advisory
1.5 Component Replacement
The LabRAM I system is designed with components that are specifically rated for use with the
LabRAM I machine. Replace worn or damaged components only with factory-replacement parts
or parts approved by Resodyn Acoustic Mixers. Incorrect component replacement can impair
the safety of the equipment and risk injury to personnel.
1.6 Customer Responsibility
Ensure that:
•All operators have received the correct training to operate the mixer.
•All operators have read and understood this Installation and User Manual.
•Only a qualified electrician performs electrical maintenance.
•Repairs to LabRAM I only by Resodyn Acoustic Mixers Service Engineers or via
return to Resodyn after receiving a Return Authorization number.
The LabRAM I system is designed to be operated in non-hazardous environments.
The LabRAM I system is protected by an IP20 enclosure which is not rated for
operation in dusty environments.
The LabRAM I system is rated for operating in ambient environments between 0°C –
40°C (32°F –104°F).
Non-factory replacement parts may not meet the LabRAM I operating requirements,
which may compromise the safety of the operating personnel and may cause damage
to the equipment.

LabRAM I Installation and User Manual
Safety
Resodyn Acoustic Mixers 7 Rev. 072420
1.7 Guards and Safety Devices
Ensure all safety devices (limit switches, etc.) are functioning correctly, and regularly check their
operation.
1.8 Handling/Lifting
The LabRAM I mixer weighs 180 lb. (82 kg) with no vessel holder, payload, or additional
options. The mixer should be lifted by at least four (4) people, one on each side. The two
primary lifting positions are on the left and right-hand sides of the LabRAM I machine where
retractable, integral lifting handles are located (Figures 1 and 2 below). Two remaining
personnel can assist in lifting the LabRAM I mixer at the front and back base plate next to the
machine feet. Exercise caution when lifting and moving the LabRAM I to avoid striking other
objects.
There is ~1 ½” (~38 mm) clearance under the LabRAM I. This space allows clearance for
fingers, and for the retractable handles to be securely positioned under the machine when not in
use. The images below show the recommended locations for lifting the mixer using a
combination of base plate handholds and retractable handles.
Do not lift using any location other than the retractable handles and the mixer’s baseplate
handholds as indicated above. Lifting it any other way may cause damage to the LabRAM I
mixer and may result in personal injury.
The LabRAM I mixer weighs 180 lb. (82 kg) with no payload or
additional options. For safe lifting and transporting of the mixer, it
should be lifted by at least four (4) people via the handles and lift
points on each side, as shown below.
Figure 2 - Underside View, lifting positions and handles
Figure 1 –Overhead View, lifting positions and handles

LabRAM I Installation and User Manual
ResonantAcoustic® Mixing (RAM) Basics
Resodyn Acoustic Mixers 8 Rev. 072420
2 ResonantAcoustic® Mixing (RAM) Basics
ResonantAcoustic®Mixing (RAM) is an advanced mixing and processing technology that utilizes
low-frequency acoustic energy. The LabRAM I is an innovative approach to multiple mixing
applications from an initial pilot stage to production scale. These applications include powder-
powder systems, powder-liquid systems, liquid-gas systems, as well as combinations thereof.
RAM technology has also been adopted for use in many processing functions, such as coating,
degassing, grinding and sieving.
RAM Technology Delivers:
•Exceptionally uniform mixing and/or processing of virtually any materials.
•Significantly faster mixing and processing than traditional methods.
•Completely sealed, dust-free process resulting in cleanliness and enhanced operator
safety.
•Highly repeatable results, ensuring batch-to-batch product consistency and quality.
•The ability to process materials not possible by other methods.
How RAM Works
The primary mechanical driver of RAM mixing technology is Acceleration (rate of change of
velocity). The unit of measure of acceleration in the International System of Units (SI) is meters
per second squared or m/s2. A way to describe this is the acceleration of gravity on earth or g.
One gis the acceleration of gravity at the Earth's surface and uses the standard gravity symbol:
g, defined as ~9.81 meters per second squared.
Acceleration is generated by the vertical motion of the mixing/processing vessel configured in a
RAM machine. This vertical, sinusoidal motion operates at a frequency of between 58 and 62
Hz and at a peak displacement of up to 0.55” when operated at 100 g. The combined low-
frequency vibration and resulting displacement creates a physical mixing and processing motion
that is uniform, repeatable, and intense, yet generates low shear.
All RAM systems operate at this frequency range in a resonant condition, which enables a
process that requires low levels of energy for operation. This condition of the system is
established over a range of frequencies that enables a balance between the kinetic energy of
the springs and potential energy of the oscillating plates that comprise the Resonator platform.
The materials being mixed are the impedance for the resonant system.
When the system is operating at resonance, virtually all the energy from the system motors is
applied directly to the materials being mixed and/or processed, resulting in a highly energy-
efficient system. This resonant condition is frequency-dependent and requires the frequency to
allow a near-perfect balance between the kinetic energy forces, the potential energy forces and
the impedance of the process. Both the kinetic and potential energy forces are set by the
system design, but the impedance depends upon the nature of the process that is contained in
the mixing/processing vessel that is agitated by the plate/spring Resonator. Moreover, as the

LabRAM I Installation and User Manual
ResonantAcoustic® Mixing (RAM) Basics
Resodyn Acoustic Mixers 9 Rev. 072420
materials’ properties in the processing vessel change, the impedance to the system changes as
well, requiring careful control and adjustment of the system frequency in order to keep the RAM
system at resonance. This operating frequency is automatically sensed and controlled by a
proprietary system/methodology that is an integral feature of the RAM technology. The
system/methodology continuously and automatically adjusts to changes in the
mixing/processing conditions within the mixing vessel or processing container. The continuous
maintenance of the resonant operation provides both power and economy to all RAM devices.
RAM Processing Parameters
The RAM mixing platform’s intense vertical motion also generates acoustic effects within the
mixing vessel. This “sound mixing” has different effects on different materials that are being
processed.
In addition to the vessel acceleration, g, other salient factors that affect the mixing and/or
processing performance include:
•pressure inside of the mixing/processing vessel (which can range from vacuum to
pressurized)
•temperature of the vessel contents
•vessel design
•processed material fill height of the vessel
•order of material added to the vessel for some processes.
For more information on the basics and practice of RAM technology, visit the Resodyn Acoustic
Mixers website at https://resodynmixers.com, or contact Technical Support (see Section 5
Technical Support).
How RAM Processing Scales
The processing settings used for the same material perform similar across all RAM technology
platforms, no matter the scale. This means that a finding of optimal settings on the bench scale
LabRAM I or LabRAM II mixer will be nominally equivalent to settings on the OmniRAM, RAM 5,
or RAM 55 mixers to process the same materials. As such, transitioning from discoveries made
in the laboratory to the pilot or production scale is seamless for most processes.

LabRAM I Installation and User Manual
Definitions
Resodyn Acoustic Mixers 10 Rev. 072420
3 Definitions
Acceleration: A change in velocity with respect to time that is expressed in “g.” Different
materials mix at different accelerations, based on the material properties and
other factors. (See “g” below.)
Damping: A term to quantify the instantaneous amount of energy absorbed by the mix
media while mixing.
g: A measure of acceleration expressed as a multiple of the gravitational force
applied to the vessel and materials during mixing. For example, 10 gis 10
times the force of gravity. It may also be shown as g. Another term that may
be used to describe acceleration is “gravities.” (See “Acceleration” above.)
HMI: Human Machine Interface. The control and input system screen used to
operate RAM Mixers. Also called an operating system.
Headspace: The volume of compressible air, gas, or void space above the mixing media
in the vessel that is not occupied.
Impedance: See “Damping” above.
Mix Media: The material inside the mixing vessel.
Mixing Power: The amount of power being delivered by the RAM system to perform mixing.
Mixing Regime: This term identifies certain modes of mixing. The mixing regime defines the
bulk flow pattern of the ingredients during mixing. A regime changes when
the bulk patterns change. For example, a stable fluid mixing regime for thick
pastes would be one that flows from the bottom of the vessel up the center
then across the top and down the sides. Another mixing regime, for example,
is characterized by a thick paste that does not couple with the vessel sides.
RAM: ResonantAcoustic®Mixing or ResonantAcoustic®Mixer.
RAMWare: Control software and its versions that run on a Personal Computer (PC) and
provides control to, and feedback from, the mixing process.
Resonance: The frequency (nominally 60 Hz) at which the RAM mechanical system
operates in the most efficient manner.
Vacuum: Internal vessel pressure that is less than atmospheric pressure

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 11 Rev. 072420
4 Installation
4.1 Unpacking, Lifting, and Placing LabRAM I
Upon receipt of the LabRAM I, do not unpack or remove any packaging materials until the
carton has been positioned as close to the installation location as possible. When in position,
remove all wrapping and strapping, and file shipping paperwork for your records.
Retain all packaging material for at least one year in case mixer transportation or shipment for
factory service is necessary. Original packaging is required for service shipment during the
warranty period and is highly recommended for any shipment performed during the life of the
product.
•Open the top of the box and remove any packing materials, bubble wrap, or other items.
When all packaging material is removed, slide the packaging box/sleeve up and over the
LabRAM I. The LabRAM I should be the only remaining item, resting on cushioning bags
in a corrugated tray on the shipping pallet.
•Make sure to remove any loose packaging material from the area, and check for tripping
hazards before attempting to move or install the LabRAM I.
•With a minimum of four people lifting at all four lifting locations, and using appropriate
safe lifting practices, lift the LabRAM I straight up and carry to its installation location.
See Section 1.8 Handling/Lifting on the previous page for lifting locations.
•When placing the LabRAM I at its location, allow enough room around the machine to
complete the installation, route wires and tubing, and allow for working space, etc.
The LabRAM I mixer weighs 180 lb. (82 kg) with no vessel holder
or additional options. For safe lifting and transporting of the mixer, it
should be lifted by at least four (4) people via the handles and lift
points on each side as shown in Section 1.8 Handling/Lifting on the
previous page.

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 12 Rev. 072420
4.2 Using the Retractable Handles
Retractable handles are located on both sides of the LabRAM I, installed in the base plate and
conveniently out of the way when not in use.
When you are prepared to lift and place the machine:
1. Locate the locking pins on either end of the handle brackets.
2. Pull straight down to release the locking pins while pulling the handles horizontally toward
you until they click-lock into the open, lifting position. Make sure the locked position is
attained before attempting to lift.
3. Reverse the procedure when the LabRAM I has been placed in its intended position. The
locking pins will automatically click-lock back in place when the handles are fully stowed
back under the machine. Make sure the locked position is attained before using the mixer.
4. Check for any blockage below the machine if the handles do not smoothly slide to their
locked storage position.
4.3 Mounting
The LabRAM I should be installed on a level, stable, and rigid surface capable of supporting the
machine weight of 180lbs (82kg) plus the weight of the vessel holder, vessel, and mix payload.
If other equipment or materials are to be placed or used on the same surface, a higher weight
rating should be used as appropriate.
ResonantAcoustic®Mixing (RAM) technology imparts near-zero forces in addition to the
machine’s weight to the surface on which it is mounted. Careful placement and leveling of the
surface by a qualified technician are recommended, as is a periodic inspection of the position
and condition of the surface. Surrounding instruments and accessories should be considered in
an overall placement plan. Contact Resodyn customer service with any questions regarding
appropriate mounting surfaces and your placement plan at +1-406-497-5333 or email

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 13 Rev. 072420
4.4 Lid Operation
1. To access the mix platform, mixing chamber, and install the vessel holder, press down
firmly on the latch located on the front of the mixer, above the HMI. The lid will release,
and gently open automatically via the lift-assist cylinder.
2. To close the lid, pull the lid down with one hand while holding the latch open with the other.
3. Push downward on the lid until the lower edge of the lid bracket has cleared the upper
edge of the latch, and then release the latch. Check to confirm the latch is securely
engaged.
NOTE: The lid is equipped with a Safety Interlock Switch. If the lid is not fully secured, the
LabRAM I will not operate. If the lid is opened while mixing, the mixer will stop.
Do not attempt to override the Safety Interlock Switch as serious injury may
result. During operation the Payload Plate, Driver Plate, Fixture, and
Vessel move up and down rapidly, accelerating up to 100 gat nominally 60
times per second. Serious bodily injury is possible if one comes into
contact with these moving surfaces.

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 14 Rev. 072420
4.5 Remove the Shipping Lock Bar
The LabRAM I is shipped with a Lock Bar that secures the resonator during transportation. The
Lock Bar must be removed before proceeding with installation or attempting to operate the
LabRAM I.
1. Open the lid and remove the 3/16” Hex Key from its shipping bag attached to the Warning
Label. Use the Hex Key to remove the two ¼-20 Socket Head Cap Screws securing the
Lock Bar.
2. Lift the Lock Bar off the platform and retain it along with screws for future re-positioning or
transport of mixer. A black-padded Resodyn case is provided for your documentation,
manuals, and other materials, like the Lock Bar and hardware.
Once the Lock Bar has been removed, the Vessel Holder assembly can be installed using the
instructions that follow.
Use provided Hex Key
to Remove Cap Screws

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 15 Rev. 072420
4.6 Install Vessel Holder Assembly
1. Remove the Vessel Holder Assembly from the accessory box accompanying the
shipment. The Vessel Holder ships loosely assembled.
NOTE: The Vessel Holder is shipped with nuts fastened to the bottom of the ¼-20 socket head
cap screws. These nuts hold the Vessel Holder together during shipment. Prior to installation,
remove the nuts and store them with the other hardware. Do not remove the cap screws or
change the configuration of the components until familiar with the parts and the assembly.
2. Follow the positioning and sequence below for installation, or if reassembly or vessel
holder movement is needed in the future.
3. Make sure the long, ¼-20 socket head cap screws are positioned properly throughout the
overall Vessel Holder assembly. The cap screws must pass through the Top Cross Bar,
Upper Base, Standoffs, and Lower Base as shown.
4. Thread the two ¼-20 socket head cap screws into the mix platform loosely, alternating
between screws. When fully-seated, torque each cap screw to 8 ft-lb (10.8 Nm) using a
3/16” Hex Wrench tool.
On its highest setting, the LabRAM I mixer payload operates at 100 gof acceleration.
The vessel holder screws must be torqued to 8 ft-lb (10.8 Nm) to prevent the Vessel
Holder and Vessel from loosening during operation.
Figure 3 Vessel Holder & Vessel Holder with Mixing Vessel
Hold Down Knob
Jam Nut
Nylon Washer
Cross Bar
Upper Base
Bolts & Standoffs
Lower Base
Bottom Dot

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 16 Rev. 072420
All Vessel Holder assemblies will match a hole pattern in the Top Plate of the LabRAM I. The
mounting hole pattern for the standard 4” Vessel Holder is ¼-20 threads measuring 4.375”
(111.13 mm) center-to-center. See the detailed illustration below for the position of these
mounting holes.
Only use the Vessel Holder Mounting Holes, shown in Figure 4, to mount the Vessel
Holder to the LabRAM I. Never use the mounting holes for mounting any other
equipment to the LabRAM I.
Never use other screw locations on the Payload or Cancellation Plates, see Figure 5,
to mount any other equipment. Never allow hardware to span from the Payload Plate
to the Cancellation Plate. Failure to comply is likely to damage the LabRAM I.
Figure 4 Payload Plate Mounting Hole Locations
Standard Vessel Holder Fixture Mounting
Holes 4.375” Center-to-Center
Outer Bellows
Inner Bellows
Payload Plate
Cancellation Plate
Figure 4 Payload Plate Mounting Hole Locations
Figure 5 Resonator Platform Components

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 17 Rev. 072420
4.7 Electrical Connections
Electrical power is supplied to the LabRAM I through the provided cord which inserts into the
Power Cord Socket on the back panel of the LabRAM I. Plug the other end into a power
receptacle delivering 100-120 VAC or 200-240 VAC.
NOTE: The cord and fuse ratings for supported voltages are as follows: 10 Amps for 100-120
VAC and 8 Amps for 200-240 VAC.
5 Technical Support
Every effort is made to ensure your LabRAM I is intuitive to use, and that it provides a long life
of reliable operation. If you have questions about its operation, or if your LabRAM I appears to
be malfunctioning, please do not hesitate to contact Resodyn Acoustic Mixers. You may contact
your customer service representative directly or utilize the general contact information provided
below.
Technical Support Contact Information:
Phone: +1-406-497-5333
Fax: +1-406-497-5206
Email: service@resodynmixers.com
Web: https://resodynmixers.com/support
Resodyn Acoustic Mixers, Inc.
130 North Main, Suite 630
Butte, Montana 59701
USA
NOTE: Please do not ship any products back to Resodyn Acoustic Mixers without express
authority, and a Return Authorization (RA) number from your customer service representative.
Any shipments made to Resodyn without an RA number clearly written on the shipping
container will be refused.
Figure 5 Resonator Platform Components
Vacuum System
Communications Port
USB Mini Port
Fuses
Power Switch
Power Cord Socket
Figure 6 LabRAM I Back Panel

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 18 Rev. 072420
6 System Specifications
The LabRAM I is equipped with the following external electrical connections:
•AC Power Cord Inlet
•Mini USB
•Communication Connector for Automatic Vacuum System
All external connections are found in the back panel of the LabRAM I machine. See Figure 6 on
Page 17 for a detailed illustration.
Maximum Payload Capacity
1.1 lb (500 gram)
Sound Emitted
<70 dB at 3’ at 100 gwith an empty vessel
Vessel Temperature Monitoring
1 RTD monitored and recorded (optional)
Automatic/Programmable Vacuum
Yes (optional)
Data Logging
Yes
Computerized Operator Interface (HMI)
Yes
Weight
180 lb (82 kg) plus vessel holder fixture
Dimensions
18” wide x 26” deep x 25” high
Electrical @ 50/60 Hz, CE Certified
100-120 / 200-240 VAC
6.1 Voltage Level
The LabRAM I is designed to utilize single-phase 100-120 or 200-240 VAC 50Hz/60 Hz supply
for use worldwide.
6.2 Power Consumption
The LabRAM I is designed to be energy efficient by operating at a resonant condition when
mixing. Depending on the acceleration setpoint, vessel selection, and mix media, the LabRAM I
will draw between 80 and 1200 Watts
Table 6-1. LabRAM I System Specification.

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 19 Rev. 072420
7 System Overview
The LabRAM I is a ResonantAcoustic®Mixer
(RAM) specifically designed and manufactured
for use in processing up to 1.1 lb (500 g) of
materials per batch. The LabRAM I is a highly
versatile mixing system designed for laboratory
and development environments. The mixer is
equipped with proprietary mixing control circuitry
and an acoustically insulated, compact
enclosure, enabling powerful, yet quiet operation.
Features include an adjustable Vessel Holder,
Safety Interlocked Lift-Assisted Lid with viewing
window, and a Touch Screen Human Machine
Interface (HMI). Optional features include RTD
temperature monitoring, Manual & Automatic
Vacuum System integration, and custom Vessels
and Vessel Holders.
The LabRAM I delivers energy to the material being mixed by moving the mixing vessel up and
down, 60 times a second, at up to 100 gof acceleration.
Electrical &
Communication
Connections
Fan
Rear
View
Air Vents
Vacuum Ports
Lift-Assist
Lid
Vessel
Holder
Acoustic
Enclosure
HMI
Touch
Screen
Front
Vi`e
Figure 7 External Dimensions
Figure 8 LabRAM I External Features
18”
26”
25”

LabRAM I Installation and User Manual
Installation
Resodyn Acoustic Mixers 20 Rev. 072420
Acoustic Enclosure
The acoustic enclosure surrounds the resonator
system. The enclosure reduces the sound emitted
by the LabRAM I and covers all the moving parts
that could pose a risk to operators.
7.1 Touch Screen Human Machine Interface (HMI)
The LabRAM I is digitally controlled through the HMI. All data entry and operations are provided
through software keypads and buttons on the screen. The touch screen and operating system
are described in detail in Section 9 HMI (Human Machine Interface) –Overview.
During operation the Payload Plate,
Cancellation Plate, Vessel Holder, and
Vessel move up and down rapidly,
accelerating up to 100 gat nominally
sixty times per second. Serious bodily
injury is possible if one comes into
contact with these moving surfaces.
Lift-Assisted
Lid
Resonator
Platform
Lid Latch
Acoustic
Enclosure
Figure 9 LabRAM I Acoustic Enclosure
Figure 10 LabRAM I Touch Screen HMI
Table of contents
Other Resodyn Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Delphi
Delphi Multec 3.5 Applications manual

Scientific Industries
Scientific Industries Enviro-Genie instruction manual

Okolab
Okolab 158206046 installation guide

CASHIDO
CASHIDO 10 Second Machine Commercial OH6800 Series user manual

MITS
MITS Eleven Auto FPZ-31AT Command manual

dymax
dymax BlueWave VisiCure user guide