manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Rigol
  6. •
  7. Measuring Instrument
  8. •
  9. Rigol DS Demo User manual

Rigol DS Demo User manual

Rigol DS Demo Guide:
Date: 02.06.2014
Basic Demonstration
Items needed:
1 DS1Z, DS2, DS/MSO4, or DS6000 Series Scope
2 Probes
Qty 1 DS6000 Demo Board
USB memory stick files
Optional items:
DG1000, DG4000 or DG5000 Arb/Function Generator
1. Initial Setup and Demo
Start by connecting the two probes to the 2 channels and connect each to the square wave
generator on the front of the scope.
View the square waves by pushing “AUTO”. After a few seconds you should see this:
Single Wave (F1)
Multiple Wave (F2)
Rising Edge (F3)
Falling Edge (F4)
When the “AUTO” Menu is up just after auto has completed you will see options for looking at
multiple waves, single wave, rising edge, or falling edge. That is how the AUTO function works.
Whenever you get into trouble or don’t know what you are looking at you can hit AUTO and it
will usually find the meaningful signal.
If you still have trouble after that you can go to “SAVE” -> from the “MENU” soft key press it
and select FACTORY (F7) . Once FACTORY is pressed this will reload all factory settings and
you will recover from whatever settings were altered.
2. Sampling & Memory Depth Demonstration
After AUTO, we can easily demonstrate the Sampling and Memory Depth features.
Push “CH 2” twice so the light turns off and only channel 1 is shown.
Turn the Horizontal (time) knob until Time is set to 200 ms / div. This is a slow setting
and it looks like hash because the wave is going much faster. Then push the horizontal
knob. This enters zoom mode. We can now zoom in on the signal using the Horizontal
knob. When you zoom down to a 5 us / div window you should see this:
2 things to note in this image:
One: we are looking at just a sliver of our captured wave (the little black slice between the blue
bars on top).
Two: you can see the straight line in the zoom because there are no data points between that
time.
*Here is the math for those so inclined: DS6000 has 140Mpts per wave in normal mode so in
the full window you see 200 ms/div and 14 div = 2.8 sec of data. Divide that by 70 M points and
you get a point every 40 nsec or so. Just about what we are seeing above.
Turn the horizontal scale knob to 200ns/div you should see the signal change to this:
Push ACQ‐> and Mem Depth , then change from Auto to 14kPoint and switch between
140MKPoints and 14kPoint. That curve with 140MPoint is much more appropriate. So if
there was an event happening at even µsec timing you can miss it without deep memory.
This is especially true when you need to capture or trigger on data over a longer period of
time.
NOTE: You can also use the UltraVision Multifunction knob to scroll slowly or more quickly
through the waveform. The outer ring will fast-forward and rewind. The inner knob allows you
to scroll more slowly.
3. Measurement Demo
Start from an AUTO setup on the square wave signal.
Push MEAS then press the Menu button on left upper corner. Now the Quick
Measurement Menu appears.
You can change the waveform source by pressing the channel key on the front panel
(CH1, CH2, ..) By changing the Source from CH2 to CH1 you will see that the color will
change from blue to yellow. On the left side you can now select up to 5 items that will be
displayed on the bottom line.
Switch from frequency to voltage measurement, push the measurement Menu bottom
again.
Push All Measure Source and selecting CH1 and CH2 and set Display All to ON, you
will see this:
You can also show measurements between cursors… press Measurement > Down Arrow >
Meas. Range and select Cursor region. This can be helpful in Zoom mode, as you can measure
on the zoomed portion of the waveform of interest and get more detail.
4. Cursors
You can also set up a single measurement like FREQUENCY and then set the cursor to AUTO.
The AUTO cursor takes the last measurement and places the cursor on the screen to show how it
is making that measurement. This is an easy and great way to align cursors and highlight
characteristics of a signal for review or reporting.
Press MEAS, press Clear and select All Item to Delete, then add FREQUENCY as the
only measurement. That should display on the screen. Then push CURS and set mode to
AUTO. It should now look like this:
The cursor is locked to the zero crossing showing the time between consecutive waves.
You can also show the TRACK and MANUAL modes of the CURSOR.
Use the menu KNOB to move the cursor position.
In MANUAL mode note that you can use the soft keys to highlight Cursor A,Cursor B, or
Coursor AB. This determines which cursors will move with the knob. If you haven’t
shown the RECORDING & PLAYBACK yet, this is a great time to show the SINC or
SINE wave signals and you can use the cursors on the recorded data. That is a great way
to do post‐analysis of signals.
5. Reference Waveform
Sometimes, it is useful to have reference signals available on the screen. This can help designers
see how changes they make can effect the output of a circuit or design. It can also be helpful for
EMC and Compliance Engineers to capture different characteristics before and after changes
have taken place. The Reference Waveform function can help by capturing a trace and displaying
it along with newly captured data.
Press AUTO and select single
Push REF (below or next to the CH buttons on most scopes). Select SAVE. Note that you
can save and reload References from internal or external memory. This is a great way to
compare traces between scopes or to use as a guide for verification testing. Once saved,
the reference will appear as a white trace that you can then visually compare to live
signals on any channel. It looks like this:
6. FFT
One great feature available on all the scopes is FFT (Fast Fourier Transform). This provides a
basic Frequency domain measurement of a signal.
Push MATH and select FFT as the function
Select CH 1 as the source
Use the VERT and HORIZ knobs to set the scaling appropriately. If you do an FFT on the
square wave you should see a primary frequency at 1 KHz and harmonics at 3,5,7,…
KHz respectively like this:
7. PASS/FAIL Measurements
You can use Pass/Fail measurements for a number of applications. With Rigol scopes you can set
it up to stop or continue when there is a failure or a pass, you can set the output to be the speaker
or
through the optically isolated output on the rear of several models or both, you can create masks
with varying error budgets around a target signal, and you can even use the record mode to
record all the failures as frames. In this way, you could set up a test over night, let it run, and
come back and have all the failure states recorded for you.
PASS/FAIL push UTIL and push PASS/FAIL.
Set Enable to ON, automatically a mask will be created.
To change the mask parameter use X mask and Y mask and then press CREATE MASK.
You can turn on the Msg Display and then press OPERATE.
You will now see the Total Wfs counter counting up. If you remove the signal or adjust
the probe attenuation switch you can cause failures. If you set it to stop on the failure you
will see something like this:
8. Record Mode
This mode can be used to capture a large number of single waveforms and allows you to play
back and/or analyze them. It is basically stacking bitmap images of each triggered event in
memory. They can then be reviewed frame-by-frame or contiuously, like a movie.
To activate this mode press UTIL and Record, then set the Mode to Record
Press the Operate soft key the recording will start
By selecting the play back mode you can run captured signal again
You can play the record from and selected Start Frame to an End frame with a tunable
interval or scroll per Knob over the selected frames
Another feature is the Analyze function, available by choosing ModeAnalyze. By
pushing Setup Template and pressing Start you recorded waveform will be analyzed
You can also use Previous and Next buttons to select only the “failed” frames within the
recording
In addition, you can use the UltraVision Multifunction knob to scroll slowly or more
quickly through the waveform. The “Stop”, “Play/Pause”, and “Record” keys located
below the UltraVision Multifunction Knob are also active in these modes. They can be
used to quickly capture and view new frames.
In Playback mode, you can set the speed of playback between frames quickly using the
UltraVision Knob. The outer ring makes the interval increase quickly. 500ms/frame is a
good start.
You can enable timestamps in Playback mode as well. The “TimerTag” needs to be
enabled on page 2/2 of the Record menu.
9. Roll Mode for slower moving signals
You need a scope and a generator. Select an Arb function like Cardiac and set the signal to 5Vpp
and 1 Hz and make sure the output is ON.
Roll mode is a nice option for slower moving signals that helps you follow it visually.
To select ROLL MODE, push MENU next to the horizontal knob and change time base to
ROLL. Once your timing is set you should see something like this.
10. XY mode
XY mode is another way to look at 2 signals and how they interact.
To activate, push MENU next to the Horizontal knob and change time base to X‐Y.
Select CH1 and push the vertical position knob to center the signal, then adjust the scale
using the larger vertical scale knob until it fills one dimension of the screen without
clipping.
Then, select CH2 and repeat.
If you have a Dual ChannelArb Generator, like the DG1022, you can enable both outputs
and connect them to both scope channels. Set both outputs to SINE waves at the same
frequency and amplitude. Align the Phase of both channels and next change start phase of
the CH2 from the generator.
The ellipse below is generated in XY mode where both signals are sine waves and channel 1 is
on the Y axis and channel 2 is on the X‐Axis. This is one way to measure the phase between 2
sine waves. To complete the calculation follow the instructions from the manual shown below.
There are a number of uses for XY mode. This is just one example.
Consequently, if you change the PHASE DEVIATION between the two Generator Outputs on
the DG1022 you can alter the orientation of the ellipse. If you hook it up to 2 different generator
outputs that are on different units and not synced the ellipse will change shape over time.
NOTE: In X-Y mode, the data points are displayed as dots. There is no interpolation between
them. If the waveform is not as crisp, increase the number of samples by lowering the horizontal
scale. This will slow the waveform update rate.
11. Measurements Between Channels
Perhaps an easier way to measure the phase difference of the two signals is to use the Delay 1-
>2f MEASUREMENT function and PERIOD measurements. See the 2 sine waves in normal
view below. The phase difference is the (Delay 1->2f/ Periode)*360 in degrees. You can select
those items from the MEASURE menu as we did earlier. The Delay 1->2f MEASURE options
measure the time between zero crossings for channel 1 and channel 2. When they are the same
frequency we can use this to understand the phase difference. It should look like this:
The 2 Sine waves are 180 degrees out of phase.
You can calculate this as:
(Delay 1->2f/ Periode)*360 ‐> (500e‐6/1e‐3)*360 = 180
12. Persistance
Persistence allows you to view how the signal changes over time on a scope. This is most useful
when viewing signals that are dynamic such as a modulated waveform. To turn on Persistance,
push DISP and turn PERSIST Time to Infinite. Use a Arb Generator and leave it in SINE wave
mode as above, then PUSH the MOD button. Select the TYPE toAM or FM accordingly.
These images below showAM waveform modulations from a Generator while using
PERSISTANCE on the scope. PERSISTANCE allows you to see how signals are changing over
time more clearly. On the image you can see shows the depth of the AM modulation quite clearly
which can’t be seen from the single live waveform, how much the frequency is really changing.
13. Trigger and Decode the CAN bus signal (Optional)
NOTE: This requires that the scope has the CAN Serial Decoding option activated. You can
check the active decoding buses by pressing the DECODE1 or DECODE2 buttons located on the
front panel near the CH2 vertical position knob on the DS1Z, DS2, and DS/MSO4 or the MATH
> BUS1 or BUS2 on the DS6.
Starting from Factory Defaults (Button labeled “default” located in the upper-right-hand
side of the front panel for the DS1Z, DS2/A, and DS/MSO4. DS6 defaults are under
Storage > Factory), set the timbase to 5us
Connect the DS6000 Demo board USB connector to the USB connector on the scope
(DS/MSO4s and DS6s have a additional USB port on the rear panel. DS1Z and
DS2000/A have only a front USB port for powering the board)
Connect the signal output pin CAN_H and GND to CH1 of the oscilloscope properly
using the probe clip
Set the vertical scale of CH1 to 1V/div
In the trigger menu (press Menu in the Trigger area on the front panel) and set the trigger
type to “CAN” , the signal source to “CH1” , the signal type to “CAN_H” , the trigger
condition to “SOF”, the baud rate to “1Mb/s” and the trigger mode to “Auto”, and set the
trigger level to ~ 300mV
When the oscilloscope is in T’D state, press the DECODE1 button to open the decode
menu
Set the decoding type to “CAN”, Source CH1, Signal type CAN_H, Baud Rate 1Mb/s,
and the format to ASCII
On page 2/2 of the decode menu (accessed by pressing the light blue down arrow on the
lower right-hand side of the display), select threshold. Use the intensity/multifunction
knob to move the decode threshold to 2.7V
Set the BUS status to “ON” and the demonstration result is as shown in the figure below
You can see that the ASCII data reads “RIGOL”
Deep Memory with Decode:
You can also use the deep memory of the scope during the decode. This enables you to capture a
long string of communications with the ability to analyze and decode sections of the string.
Same configuration as above, then set the timebase to 2ms. Now, you can see that you
have collected quite a few packets of communication
Press Run/Stop and then press the Horizontal Scale knob in to enable Zoom
Rotate the Horizontal Scale Knob to set the Zoom scale to 20us/div
Note that the packet is decoded
You can use the Inner Navigation Knob to step through the data, or you can use the outer
knob to fastforward and rewind through the zoomed data.
You also enable the event table to view and export a table of the decoded data. You can
access the event table in the Decode menu
14. Trigger and Decode the SPI bus signal (Optional)
NOTE: SPI is not available on 2 channel DS4 scopes.
Starting from Factory Defaults (Button labeled “default” located in the upper-right-hand
side of the front panel for the DS1Z, DS2/A, and DS/MSO4. DS6 defaults are under
Storage > Factory), set the timbase to 5us
Connect the DS6000 Demo board USB connector to the USB connector on the scope.
The DS/MSO4s and DS6s have a additional USB port on the rear panel. The DS1Z and
DS2000/A have only a front USB port for powering the board
Connect CH1 probe to the SPI_SCLK loop on the Demo Board. Connect the probe
ground lead to the GND loop on the Demo Board. Set the CH1 vertical scale to 2V
Connect CH2 probe to the SPI_MOSI loop on the Demo Board. Connect the probe
ground lead to the GND loop on the Demo Board. Press CH2 to activate the channel and
set the vertical scale to 2V
Open the Trigger Menu by pressing the MENU button in the trigger area and set the
trigger type to “SPI”
Set the SCL to “CH1”
oSet the SCL trigger level to 1.5V by pressing SCL and adjusting the trigger level
using the trigger level knob located in the trigger section on the front panel
Set the SDA to “CH2”
oSet the SDA trigger level to 1.5V by selecting SDA and adjusting the trigger level
using the trigger level knob
Set the trigger condition to “Timeout”
Select Data and open the Trigger Data menu. Set the bits to 8, the data to “LLLH LLLH”
and the clock edge to “Leading”. The oscilloscope should be triggering stably
TIP: You can use the intensity/multifunction knob to adjust the CurrentBit selection by
rotating it. You can change the data (H, L, X) by pressing the knob.
Set the trigger mode to Normal by pressing the Mode button in the trigger menu area on
the front panel. The scope should trigger stably. This is shown as a green “T'D” indication
in the upper-right-hand portion of the display.

Other Rigol Measuring Instrument manuals

Rigol DSA800 Series User manual

Rigol

Rigol DSA800 Series User manual

Rigol DS6000 Series Installation and operating instructions

Rigol

Rigol DS6000 Series Installation and operating instructions

Rigol PCA1500 User manual

Rigol

Rigol PCA1500 User manual

Rigol DSA800E Series User manual

Rigol

Rigol DSA800E Series User manual

Rigol RPL1116 User manual

Rigol

Rigol RPL1116 User manual

Rigol RSA5000-VNA User manual

Rigol

Rigol RSA5000-VNA User manual

Rigol DSA800E Series Operating instructions

Rigol

Rigol DSA800E Series Operating instructions

Rigol RSA Series Operating instructions

Rigol

Rigol RSA Series Operating instructions

Rigol PLA2216 User manual

Rigol

Rigol PLA2216 User manual

Rigol DS6104 Administrator Guide

Rigol

Rigol DS6104 Administrator Guide

Rigol DSA800 Series User manual

Rigol

Rigol DSA800 Series User manual

Rigol DSG3030-IQ User manual

Rigol

Rigol DSG3030-IQ User manual

Rigol RP1000D Series User manual

Rigol

Rigol RP1000D Series User manual

Rigol RP1000D Series User manual

Rigol

Rigol RP1000D Series User manual

Rigol DSA815 User manual

Rigol

Rigol DSA815 User manual

Rigol DSA700 series User manual

Rigol

Rigol DSA700 series User manual

Rigol RP1003C User manual

Rigol

Rigol RP1003C User manual

Rigol DS1000CA series User manual

Rigol

Rigol DS1000CA series User manual

Rigol DS1202Z-E User manual

Rigol

Rigol DS1202Z-E User manual

Rigol MSO1000Z series User manual

Rigol

Rigol MSO1000Z series User manual

Rigol DSA800 Series User manual

Rigol

Rigol DSA800 Series User manual

Rigol RSA5000 Series User manual

Rigol

Rigol RSA5000 Series User manual

Rigol DM3058 User manual

Rigol

Rigol DM3058 User manual

Rigol DSA1000 Series User manual

Rigol

Rigol DSA1000 Series User manual

Popular Measuring Instrument manuals by other brands

VOLTCRAFT LWT-110 operating manual

VOLTCRAFT

VOLTCRAFT LWT-110 operating manual

Altronics PH-01 quick start guide

Altronics

Altronics PH-01 quick start guide

Banner Q5X quick start guide

Banner

Banner Q5X quick start guide

Vega WEIGHTRAC 31 operating instructions

Vega

Vega WEIGHTRAC 31 operating instructions

Voltech PM6000 user manual

Voltech

Voltech PM6000 user manual

Veto TU-2016 instruction manual

Veto

Veto TU-2016 instruction manual

Pitney Bowes SendPro C Lite user guide

Pitney Bowes

Pitney Bowes SendPro C Lite user guide

CALIMA gli 46065 operating instructions

CALIMA

CALIMA gli 46065 operating instructions

Trace2O HydroTest HTTURB instruction manual

Trace2O

Trace2O HydroTest HTTURB instruction manual

Racelogic VBOX 20SL manual

Racelogic

Racelogic VBOX 20SL manual

Buhler TC-Standard (+) X2 Installation and operation instructions

Buhler

Buhler TC-Standard (+) X2 Installation and operation instructions

Blue-White D-300 quick start guide

Blue-White

Blue-White D-300 quick start guide

Apogee Instruments MQ-100X owner's manual

Apogee Instruments

Apogee Instruments MQ-100X owner's manual

Rice Lake Measurement Systems International MSI9750A Operator's manual

Rice Lake

Rice Lake Measurement Systems International MSI9750A Operator's manual

ScienTECH CS750 operating manual

ScienTECH

ScienTECH CS750 operating manual

BRONKHORST MASS-STREAM D-6400 user manual

BRONKHORST

BRONKHORST MASS-STREAM D-6400 user manual

Wilcoxon MachineryMate 800-IS operating guide

Wilcoxon

Wilcoxon MachineryMate 800-IS operating guide

LOVATO ELECTRIC DME D100 T1 MID instruction manual

LOVATO ELECTRIC

LOVATO ELECTRIC DME D100 T1 MID instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.