RS PRO ICM A6N User manual

Instruction Manual
ICM A6N
Clamp Meter
EN DE ES IT FR
1000Vdc
CAT
750Vac
CAT
600V

EN - 2
Safety Information
Read and understand this Instruction
Manual completely before using this
instrument. Failure to observe the warnings
and cautions in this Instruction Manual may
result in injury or death, or damage to the
instrument and other equipment or property.
If this instrument is used in a manner not
specified in these instructions, the protec-
tion provided by the instrument may be
impaired.
The following symbols may appear on
the instrument and in this Instruction
Manual:
Risk of electric shock
Refer to Instruction Manual
Alternating Current (ac)
<Battery
Earth
Equipment protected throughout by
double or reinforced insulation
Conforms to EU directives
Dispose of this equipment in
accordance with local regulations.
EApplication around and removal
from hazardous Live conductors is
permitted.
WARNING
-Examine the instrument and probes and
leads before use. Do not use the instru-
ment if it is wet or damaged, or if you
suspect it is not operating correctly.
-When using the instrument, test leads or
probes, keep your fingers behind the
finger guards.
-Remove the test lead from the instrument
before opening the battery cover or instru-
ment case.
-Always use the correct terminals, switch
position and range for measurements.

EN - 3
-Verify the instrument is operating correctly
by measuring a known voltage before
use. If in doubt, have the instrument
serviced.
-Do not apply more than the rated voltage
as marked on the instrument between
terminals, or between any terminal and
earth ground.
-Use caution when measuring voltages
above 30 Vac rms or 60 Vdc. These
voltages pose a shock hazard.
-To avoid incorrect readings that can lead
to electric shock, replace the battery as
soon as low battery indicator <appears
in the display.
-Disconnect the circuit power and
discharge all high-voltage capacitors
before making resistance, continuity, or
diode measurements.
-Do not use the instrument in a Hazardous
Area or around explosive gasses or
vapours.
-Wear suitable Personal Protective
Equipment when working around or near
hazardous Live conductors which could
be accessible.
Caution
-Disconnect the test leads from the test
points before changing the position of the
function rotary switch.
-Never connect the instrument to a source
of voltage with the function rotary switch in
Ω/ ;/Aposition.
- If possible, do not work alone, so
assistance can be given if required.
-Do not expose the instrument to extremes
of temperature or high humidity.

EN - 4
Maintenance
Do not attempt to repair this Instrument. It
contains no user-serviceable parts. Repair
or servicing should only be performed by
qualified personnel. This instrument
should be calibrated yearly, or more
frequently if used in harsh conditions or if
it is suspected of being inaccurate.
For calibration and repair contact RS
Components - the address is given at the
end of these instructions.
Cleaning
Periodically wipe the case with a damp
cloth and detergent. Do not use abrasives
or solvents.

EN - 5
Power On / Off
• When first turned on, the instrument
displays the remaining capacity of the
battery. Replace the battery when 0% is
shown or the low battery symbol “<”
remains visible in the display.
After a successful battery self-test, the
instrument will automatically select resis-
tance (Ω)/continuity mode.
Auto Hold Enable / Disable
Auto Power Off Enable / Disable

EN - 6
Self-Testing Aid
Fail of usage
Ready to use
CAT.Ⅲ.600V
with respect to
earth for the jaw.
Barrier for
Hand Guard
Do not hold
the meter above
the Barrier
• Do not attempt to measure while power-
ing up, as it will cause the self-test to fail.
Caution
If “FAIL” is shown in the display, do not use
the instrument as an internal fault has
occurred.
AC Current measurement

EN - 7
AC V / DC V
Caution
When connecting the test leads to the
circuit or device, connect the black lead
before the red lead. When removing the
test leads, remove the red lead before
the black lead.
•Input Impedance
Note : The input impedance is≧4K for an
input voltage up to 30V and increases
automatically with input voltage to approxi-
mately 375KΩat 750V.

EN - 8
Continuity
Resistance / Diode

EN - 9
Priority of function
The LCD display shows volt-
age mode, when the voltage at
the input terminals satisfies the
following conditions:
ACV:
Between 1.3V and 999.9V
DCV:
Between 2.1V and 999.9V,
or between -0.7V ~ -999.9V
The LCD display shows
resistance (Ω) mode when the
resistance at the input termi-
nals satisfies the following
conditions:
Resistance:
Between 0 and ∞Ω
ACV: Between 0 and 0.9V
DCV: Between -0.4 and -0.2V
DCV: Between 1.0 and 2.0V
The LCD display shows diode
mode when the voltage at the
input terminals satisfies the
following conditions:
DCV: Between 0.4 and 0.8V
The LCD display shows AC
current (A) mode when the
current sensed by the jaws
satisfies the following
conditions :
ACA:
Between 0.9A and 600.0A
Data Hold
Data-hold mode is not available when
“OL KΩ“ is displayed with no input
Signal present.

EN - 10
Data Hold Beep Guard
The internal sounder will beep continuously
and the LCD display will flash if either of
the following two situations are met in Data
-hold mode :
1. The instrument measures a
characteristic different from the LCD
reading; e.g. a voltage reading is held
on the display and an attempt is made to
measure an AC current .
2. The measured signal is the same
characteristic as the LCD reading, but
is 50 counts or more greater than the
held LCD reading.
Auto Hold ( ACA only )

EN - 11
Auto Power-off
Battery Replacement
When the low battery indicator "<"
appears on the LCD, replace the batteries
with the type given in the specifications.
WARNING
Disconnect the test leads from the circuit
and the instrument before removing the
battery cover.
Refer to the following figure to replace the
batteries:

EN - 12
Specifications
1-1 General Specifications
LCD display :
4 full large scale LCD readout.
Measuring rate : 5 times / sec.
Overrange display : “OL” is displayed for
“Ω” functions, shows the real value is
shown for the “A” and “V” functions.
Automatic power off time :
Approximately 20 minutes after power on
without operations.
Low battery indicator : <is displayed.
Replace the battery when the indicator <
appears in the display.
Power requirement : 9V battery
Battery life : Approximately 250 hours
approximately.
1-2 Environmental Conditions
Indoor use only.
Calibration : One year calibration cycle.
Operating temperature :
0°C to 30°C (32°F to 86°F) @ ≦80% RH
30°C to 40°C (86°F to104°F) @≦75% RH)
40°C to 50°C (104°F to 122°F) ≦45%RH
Storage temperature : -20 to + 60°C
(-4°F to 140°F) @ 0 to 80% RH with
batteries removed from the instrument.
Measurement Category (Installation
Category) :
per IEC 61010-1:2001: CAT.Ⅲ600V.
Measurement Category l is for measure-
ments performed on circuits not directly
connected to mains. Examples include :
Measurements on battery powered equip-
ment and specially protected (internal)
mains-derived circuits.
Measurement Category ll is for measure-
ments on circuits directly connected to the
low voltage installation. Examples include:
Household appliances, portable tools and
similar equipment.

EN - 13
Measurement Category lll is for measure-
ments performed in the building installation.
Examples include measurements on
distribution boards, junction boxes, socket-
outlets and wiring and cables in the fixed
installation.
Measurement Category lV is for measure-
ments performed at the source of the
low-voltage installation. Examples include
measurements on primary overcurrent
protection devices and electricity meters.
Conductor Size :
27mm diameter maximum.
Pollution degree : 2
Operating altitude : 2000m (6562 ft)
EMC : EN 61326-1
Shock vibration: Sinusoidal vibration per
MIL-T-28800E (5 to 55 Hz, 3g maximum).
Drop Protection :
4 feet drop onto hardwood or concrete floor.

EN - 14
1-3 Electrical Specifications
Accuracy is ±(% reading + number of
digits) at 23°C ±5°C < 80%RH.
Temperature coefficient :
Add 0.2 x (Specified accuracy) / °C ,< 18°C
> 28°C .
Voltage
Function Range Accuracy
V 1.3V ~ 750.0V ±(1.5%+ 3 dgt)
50Hz ~ 500Hz
V
2.1V ~ 1000V ±(1%+ 2 dgt)
-0.7V ~ 1000V ±(1%+ 4 dgt)
Overload protection :
1000 VDC / 750 VAC
Max Operation time: DT = 30s for ≧30V
Input impedance : ≧4K for input voltage
up to 30V. Impedance increases with input
voltage to approximate 375KΩat 750V.
Resistance & Continuity & Diode
Function Range Accuracy
Ω ;0.0Ω~ 9999Ω±(2% + 2 dgt)
0.4V ~ 0.8V ±(1% + 3 dgt)
Overload protection :
1000 VDC / 750 VAC
Max. open circuit voltage : 2.0V
Continuity check :
Internal sounder activates if the resistance of
the circuit under test is less than 25Ω.
It will then turn off if the resistance is
increased beyond 400Ω. Specified from
0°C ~ 40°C.
AC Current
Function Range Accuracy
A0.9A ~ 600.0A ±(1.9% + 5 dgt)*
50Hz ~ 60Hz
Overload protection : 600.0A
0 ~ 500A : Continuous Operation.
* 501A ~ 600A: 10 minutes maximum
Followed by 10 minutes cooling period.

EN - 15
Limited Warranty
This instrument is warranted to the original
purchaser against defects in material and
workmanship for 3 year from the date of
purchase. During this warranty period, RS
Components will, at its option, replace or
repair the defective unit, subject to verifica-
tion of the defect or malfunction. This
warranty does not cover disposable
batteries, or damage from abuse, neglect,
accident, unauthorized repair, alteration,
contamination, or abnormal conditions of
operation or handling.
Any implied warranties arising out of the
sale of this product, including but not lim-
ited to implied warranties of merchantability
and fitness for a particular purpose, are
limited to the above. RS Components shall
not be liable for loss of use of the instru-
ment or other incidental or consequential
damages, expenses, or economic loss, or
for any claim or claims for such damage,
expense or economic loss. Some states or
countries laws vary, so the above limita-
tions or exclusions may not apply to you.
For full terms and conditions, refer to the
current RS Catalogue.

EN - 16

Africa
RS Components SA
P.O. Box 12182,
Vorna Valley, 1686
20 Indianapolis Street,
Kyalami Business Park,
Kyalami, Midrand
South Africa
www.rs-components.com
Asia
RS Components Pte Ltd.
31 Tech Park Crescent
Singapore 638040
www.rs-components.com
China
RS Components Ltd.
Suite 23 A-C
East Sea Business Centre
Phase 2
No. 618 Yan'an Eastern Road
Shanghai, 200001
China
www.rs-components.com
Europe
RS Components Ltd.
PO Box 99, Corby,
Northants. NN17 9RS
United Kingdom
www.rs-components.com
Japan
RS Components Ltd.
West Tower (12th Floor),
Yokohama Business Park,
134 Godocho, Hodogaya,
Yokohama, Kanagawa 240-0005
Japan
www.rs-components.com
U.S.A
Allied Electronics
7151 Jack Newell Blvd. S.
Fort Worth, Texas 76118
U.S.A.
www.alliedelec.com
South America
RS Componentes Limitada
Av. Pdte. Eduardo Frei M. 6001-71
Centro Empresas El Cortijo
Conchali, Santiago, Chile
www.rs-components.com
Table of contents
Other RS PRO Measuring Instrument manuals

RS PRO
RS PRO ICM 20 User manual

RS PRO
RS PRO ILM 1335 User manual

RS PRO
RS PRO RS-1151 User manual

RS PRO
RS PRO RS 155B User manual

RS PRO
RS PRO IM-195 User manual

RS PRO
RS PRO RS-95 User manual

RS PRO
RS PRO RS-128M User manual

RS PRO
RS PRO RS-9883M User manual

RS PRO
RS PRO ICM 134 User manual

RS PRO
RS PRO RS-1361C User manual

RS PRO
RS PRO RS-3893 User manual

RS PRO
RS PRO RS-171V User manual

RS PRO
RS PRO RS-3809 User manual

RS PRO
RS PRO IM-191 User manual

RS PRO
RS PRO RSSA3000X User manual

RS PRO
RS PRO RS-1260 User manual

RS PRO
RS PRO 123-8776 User manual

RS PRO
RS PRO 236-9298 User manual

RS PRO
RS PRO RS-96 User manual

RS PRO
RS PRO 232.50.100 User manual