Tissue-Tek® AutoTEC® a120
Quick Reference Guide
Tissue-Tek® AutoTEC® a120
page 2 | 2 QRG | AutoTEC | V-version
1. Press the LOAD MAGAZINE icon on the main
Monitor screen
2. When the message DOOR IS UNLOCKED is
displayed, the input door can be opened.
3. Up to 4 Magazines can be loaded, slide the
magazine to the bottom.
4. Securely close the input door, automatic
embedding will commence after the
programmable warm-up time has
completed.
Note: Cold magazines should not be loaded
into the instrument, lids leveled to the
posts of the input door.
Up to 4 magazines can be
loaded
The Monitor screen provides the applicable
information on Input Door, Output Doors and
Basemold area during the embedding cycle.
Cassettes placed on the recovery area due to open
lid or cassette assembly issues can be retrieved by
pressing the INSTRUMENT ACCESS icon on the left
bottom corner.
When the embedding of the blocks has been
completed the blocks are placed in one of the
Output door(s). The output door icon indicates the
door can be removed of the instrument.
Up to 20 blocks are placed in
the output door
Note the Operating Manual Tissue-Tek®AutoTEC®
a120, chapter 7.
Dropped cassettes: In case a cassette is dropped
from the gripper a sensor will be triggered notifying
the user and allow recovery of the cassette.
Error’s, Messages or attentions: If displayed on
the screen follow press RESUME or shutdown and
follow further instructions displayed on the screen.
Insufficient Paraffin: In case of insufficient
molten paraffin in the paraffin reservoir, the
instrument will stop pulling cassettes from the
magazine in the input door and display 0
cassettes in the dashboard.
Cassette assembly: In case the Tissue-Tek®
Paraform®cassette is not properly assembled or
the lid is not closed completely, the instrument will
palce the cassette in the recovery area and will
display its presence in the Monitor Screen.
Note the Operating Manual Tissue-Tek®AutoTEC®a120, chapter 5 or the Application Bulletin Tissue-Tek®
AutoTEC®a120 Cleaning instructions.