10
Operation and Field Use
1. ToactivatetheinstrumentpressandholdthePOWER(A)buttonuntilthedisplayilluminates.
Thefollowingwarmupsequencewilloccuronthedisplay(a-g):
a. SensitLogowillappear
b. SystemCheck
i. Failureswillbelisted,ifanyoccur
1. PressingtheAbuttonwillturnofftheinstrument
2. PressingtheBbuttonwilldisplaythenexterrorifmorethanonehasoccurred
3. PressingtheCbuttonwillskippasttheerrortocontinuewarmup.Apassword
isrequiredtoskiptheerrormessage.Wrongpasswordwillcausetheerror
messagescreentobedisplayed.
ii. UsetheBand/orCbuttontoadjustthepassword.
iii. PresstheAbuttontoaccepttheentry.
c. Serialnumber,SoftwareRevision,DateandTime.
d. FlowCheck
i. IftheFlowCheckissuccessful,PASSEDwillbedisplayed.
ii. IftheFlowCheckisunsuccessful,ensurethatnothingisobstructingtheinletor
exhaust,andchecktheintegrityofthehydrophobiclter.
e. Warmupisindicated“PLEASEWAIT”witha600secondcountdowntimer.Thecountdown
timerwillbeginafteraninitialwarmupof5-10minutes.
f. CellCheck
i. Thisisa5stepprocessthatteststheintegrityoftheinternalcell.
ii. Theprocesstakesapproximately7.5minutestocomplete.
1. CleanAir
2. MethaneCellIn
3. TemperatureAdjustment
4. MethaneCellIn
5. CleanAir
iii. IftheCellCheckfails,restarttheinstrument.
Iftheprocessisunabletosuccessfullycomplete,serviceisrequired.
g. CalibrationduedateindicatingwhencalibrationwillneedtobedoneorPASTDUE
indicatingcalibrationdatehasbeenexceeded.Ifcalibrationisrequired:
i. PressingtheAbuttonwillturnofftheinstrument
ii. PressingtheBbuttonwillbeginthecalibrationprocess
iii. PressingtheCbuttonwillskippastthecalibrationrequiredmessagetocontinuewarmup.
h AutoBumptestfunctiontovalidatesensitivitytoethane(ifoptionisactivated)
i. PressingtheAbuttonwillturnofftheinstrument.
ii. PressingtheBbuttonwillbeginthebumptestprocess.
iii. PressingtheCbuttonwillskippastthebumptestrequirementtocontinuewarmup.
i. Workingmodedisplayindicatedby“READYFORMEASUREMENT”onthedisplay
andreadyLEDillunimated.