Siare SIARETRON 1000 User manual

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 1 / (72 ) ( rev.3) 30.06.2004
SIARETRON 1000
SIARETRON 1100
SIARETRON 3000
SERVICE MANUAL

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 2 / (72 ) ( rev.3) 30.06.2004
INDEX
1 USE OF SERVICE MANUAL ........................................................................................................................ 3
1.1 MANUAL USERS ....................................................................................................................................................... 3
1.2 CONSULTING NOTES................................................................................................................................................. 3
1.3 SAFETY NOTES................................................................................................................................................... 3
2 GENERAL MAINTENANCE PROCEDURES................................................................................................ 5
2.1 GENERALITY............................................................................................................................................................ 5
3TECHNICAL DESCRIPTION......................................................................................................................... 6
3.1 FRONT VIEW AND PARTS DESCRIPTION ....................................................................................................................... 6
3.2 FRONT PANEL DESCRIPTION ...................................................................................................................................... 7
3.3 REAR VIEW AND PARTS DESCRIPTION ......................................................................................................................... 8
3.4 OPERATING PRINCIPLE ............................................................................................................................................. 9
3.4.1 Hardware features....................................................................................................................................... 9
3.5 GAS ADMINISTRATION AND PROPORTIONING PNEUMATIC SYSTEM ................................................................................ 12
3.5.1 Operation description ................................................................................................................................ 13
4CALIBRATION AND TESTING.................................................................................................................... 15
4.1 LV200-MAIN/1 CPU BOARD ................................................................................................................................. 15
4.2 LV200 I/O BOARD .............................................................................................................................................. 21
4.3 PANEL BOARD .................................................................................................................................................. 25
4.4 POWER BOARD................................................................................................................................................. 28
4.5 PNEUMATIC CIRCUIT ....................................................................................................................................... 30
4.6 FINAL TESTING ................................................................................................................................................. 33
5 PNEUMATIC................................................................................................................................................ 41
5.1 PNEUMATIC DRAWING............................................................................................................................................. 41
5.2 PNEUMATIC COMPONENTS LIST................................................................................................................................42
5.3 VALVES GROUP ..................................................................................................................................................... 43
5.3.1 Components list......................................................................................................................................... 44
6ELECTRONIC.............................................................................................................................................. 45
6.1 ELECTRIC DRAWINGS FOR TEST ATS.098 TOOLS.................................................................................................... 71
6.2 SIARETRON 1100 - PNEUMATIC DRAWING................................................................................... 72
Warning
SIARE is used throughout the description inside this manual as an abbreviation for
SIARE HOSPITAL SUPPLIES S.r.l.
Via Giulio Pastore, 18 - 40056 Crespellano (BO) - ITALIA
Tel.: 051/969802 / Fax: 051/969101
Manufacturer of the equipment described in the manual hereto

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 3 / (72 ) ( rev.3) 30.06.2004
1 USE OF SERVICE MANUAL
1.1 Manual users
This manual is addressed to SIARE qualified personnel or to technical qualified personnel formerly authorized
by SIARE.
The SIARE authorized technician has the instruments and suitable spare parts and he is instructed to operate
respecting product safety.
SIARE declines any responsibility arising from technical interventions effected on the equipment without its
formal authorization.
The correct and safe use of the equipment both for patient an operator requires the knowledge of
recommendations and instructions contained in the present manual and in the User’s Manual.
1.2 Consulting notes
The mark !reported next to and instruction individuates an important remark for the safety both of patient
and operator.
The manual describes the equipment and the operating principle with the aid of electric and pneumatic
drawings.
The User’s Manual is part of this manual, the technician must have a copy of User’s Manual and must know its
content before effecting the operations described in the present manual.
1.3 SAFETY NOTES
WARNINGS !!
The equipment has been manufactured under the warranty and quality conditions of the product and its
components in order to assure the maximum reliability of the equipment and safety for the patient and operator.
Therefore, the safety is guaranteed only in case the technical interventions are effected in conformity to this
manual indications and those contained in the user’s manual which is unseparable par of the product
documentation currently used by the service.
Always in order to grant safety it is necessary to respect the maintenance schedule foreseen in the user’s
manual.
The maintenance and replacement of any part must be effected by personnel formerly authorized by SIARE
using the original SIARE spare parts or checked by the same.
SIARE is not civically or criminally liable in the following cases:
1) Use in conditions and for reasons not stated or prescribed in service and user’s manuals.
2) Lack or omission of scheduled maintenance as described in service and user’s manuals.
3) Maintenance performed by personnel not formerly authorized by SIARE.
4) Use of non original SIARE spare parts or not checked by the same.
5) Connection with equipment that is not in compliance with Safety Norms in effect for intended use.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 4 / (72 ) ( rev.3) 30.06.2004
!
Not connect the equipment to the patient by antistatic conductive hoses.
Since the use of antistatic hoses can cause burns when high frequency surgical equipment are used, their use
is not allowed in any application with this equipment.
For general safety of electro medical equipment, we underline the important aspect of interaction existing
between the equipment and the patient, the operator and environment.
For a correct and safe employ of the equipment it is absolutely necessary to follow the indications of the present
manual and user’s manual, observing the supplied indications.
Any operation on equipment presupposes the precise and correct knowledge of the present instructions.
The equipment must be submitted to an inspection and maintenance every 6 (six) months by SIARE qualified
technical personnel. All interventions carried out by qualified personnel are registered on a maintenance report.
Any repairing intervention must be carried out by SIARE qualified personnel or by qualified technical personnel
authorized by SIARE.
SIARE decline any responsibility for direct or indirect damages or prejudices to persons arising from not
authorized technical interventions or from an improper use of the same not in conformity to user’s and service
manual indications.
For repairing the equipment affected by malfunctioning, defect or failures, the dedicated personnel for use of the
equipment must contact SIARE or the local Service authorized to carry out the repairing or eliminate the
malfunctioning. It is necessary that at the moment of intervention request the model and serial number of the
involved equipment must be specified.
Use exclusively the suggested accessories.
It is allowed to use other accessories only in case of formal SIARE authorization, in conformity to safety norms
in force.
The equipment operation is only allowed in rooms equipped in accordance to the safety norms in force.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 5 / (72 ) ( rev.3) 30.06.2004
2 GENERAL MAINTENANCE PROCEDURES
2.1 Generality
!
In order to grant both patient and operator safety the equipment must be inspected and checked as soon as
1000 hours operation are reached or in case of not intensive use of the equipment every 6 months at least.
Such inspections and checking’s require a perfect knowledge of involved equipment, so they must be carried
out exclusively by highly qualified personnel and specifically instructed and formerly authorized by SIARE.
It will be care of the Clinician or Anesthetist, to carry out the ordinary maintenance on the equipment as per
indications of the user’s manual.
The cleaning, disinfections, and parts replacement operations will have to be effected respecting the instructions
of the user’s manual in order to avoid damages to the components which could also compromise the patient and
operator safety.
The employed materials have been carefully selected during projecting phase after specific verifications, and
comparative tests.
Furthermore, the same materials are constantly tested during manufacturing cycle in order to obtain the
maximum reliability and safety for patient and operator.
Therefore, any part or circuit must be replaced exclusively with original spare parts supplied or checked by
SIARE.
SIARE assumes all law responsibility foreseen in case the used equipment is submitted to periodic maintenance
in accordance to the modalities indicated in the user’s manual and in this manual. As a proof that maintenance
has been carried out a Maintenance Report made and signed by SIARE authorized technician is necessary.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 6 / (72 ) ( rev.3) 30.06.2004
3 TECHNICAL DESCRIPTION
3.1 Front view and parts description
1- Front panel
2- Ambient air safety valve
3- Inspiratory valve
4- Expiratory valve
1- Front panel
2- Ambient air safety valve
3- Inspiratory valve
4- Expiratory valve
1234

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 7 / (72 ) ( rev.3) 30.06.2004
3.2 Front panel description
For a better individuation of regulations and messages the control panel is divided in 7 sections listed here
below (from left to right and from top to bottom):
A) AIRWAY PRESSURE: airways pressure monitoring and control;
B) ALARMS: or indicators of alarm conditions and of acoustic alarm suspension;
C) MINUTE VOLUME: section of regulation of minute volume delivered to patient and CPAP flow;
D) MODE or section of operative modalities;
E) VENTILATION PARAMETERS REGULATIONS: for the regulation and displaying of ventilation
parameters such as: O2 Concentration, Breathing Rate, Inspiration percentage, SIMV Rate,
Sensitivity of spontaneous patient activity detection.
F) POWER SUPPLY: for the turn-on and off of equipment, as well as displaying of equipment
power supply modalities.
G) PEEP / CPAP: for end expiration positive pressure regulation or CPAP.
A
B
D
C
E F
G

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 8 / (72 ) ( rev.3) 30.06.2004
3.3 Rear view and parts description
HOSPITAL SUPPLIES
Crespellano (BOLOGNA) ITALY
TEL. 051 - 969802 FAX 051 - 969366
CL A SSE 1 TI P O B CE I 6 2 -5 D E L 1 99 1
MADE IN ITALY
Mod.
Volt
A
.
SERIE
Fuse
Fuse
7 - Main switch
8 - Protection fuses
9 - 220V power supply outlet
10 - RS 232 interface
11 - Cooling fan wheel
13 - Identification label
14 - Air inlet connection
15 - Oxygen inlet connection
18 - 12Vdc power supply inlet connector
If the fan wheel stops for a failure or an obstruction, an acoustic alarm sounds.
The same occurs in case of power failure.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 9 / (72 ) ( rev.3) 30.06.2004
3.4 Operating principle
The lung ventilators of SIARETRON series are devices for assisted ventilation equipped with a microprocessor
control system composed of a 5 printed circuits kit; such a system controls the pneumatic circuit which delivers
the medical gas to the patient and permits to the operator the setting of operating parameters by a button panel
and a display.
The printed circuits composing the control system are:
• Power supply (POWER)
• cpu (LV200-MAIN/1)
• panel and panel control (PANEL)
• input/output (LV200I/O)
3.4.1 Hardware features
POWER (POWER SUPPLY)
Device power supply board. It controls three different power sources:
1. 12 Vdc (external battery: it needs a “presence contact”, or a staple closed between two clamps)
2. 12 Vac (secondary of a main powered transformer at 220V/50Hz)
3. 12 Vdc (rechargeable internal battery)
Here below listed in decreasing priority order; on the board two 2,5A fuses are present on the inlets of the two
batteries, while the main inlet is protected by fuses on 220 V power line at the level of the main switch block.
In case of main power supply the board provide to recharging internal battery in a maximum of 10 hours; such a
recharging can be effected also when the machine is off, provided that main switch (on the rear side) is on.
The start-up of the machine (which is made by pressing the ON/OFF button on the front panel) and the choice
of power supply source are made by dedicate relay present on the board.
The board generates the following tensions:
• + 5 V / 1 A for the logic part of the system,
• + 8 V / 1,5 A for displays and panel leds powering;
• 20 V not stabilized for the polarization of analog electronic presents on LV200I/O board;
• 12 ÷15 V not stabilized (depends on which source is used) for powering of pneumatic circuit
electro valves, fan wheel, buzzer and relays.
The turn-on/off mechanism is effected by the circuit indicated in figure 1; with the ON/OFF switch of the panel
the two clamps get in contact the two CN2 connector clamps, starting up the Q3 transistor whom excites the K3
relay. Hence, the Vcc tension is generated (through the 8301L stabilizer) whom supply the micro controller on
the CPU board, this generates by its side an high logic level (Vcc) on the line denominated AUTO_ON whom
maintains the K3 relay excited (and so all the device is powered).
The turn-off by ON/OFF switch is possible thanks to the Q5 transistor, through whom the micro controller knows
switch status, the software carries out the turn-off when it reads the pressed switch for more than one second,
generating a low logic level (0 V) on the AUTO_ON line and turning-off all peripheral devices (display and
electro valves). The machine remains, anyway, physically powered since the ON/OFF switch is released.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 10 / (72 ) ( rev.3) 30.06.2004
LV200-MAIN/1 (CPU)
The Main board containing the control microprocessor (Hitachi H8/532 at 16 bit), an EPROM of program (max.
dimension 64 KB), a data RAM of 32 KB, a GAL 16V8 for addresses decodification, some I/O gate mapped on
the bus (among them an UART for panel communication) and a serial E2PROM of 128 byte for the storage of
operating parameters of the machine. The clock rate of microprocessor is generated by a quartz at 19,6608
MHz.
The board is equipped with a WATCH DOG circuit. This is not sensible to electric power supply and it is realized
with an NE555; such a circuit is able to generate a resetting of microprocessor if it not periodically “refreshed” by
the software, retaking the machine in a know status in case of not correct performance of the working program.
PANEL
Front panel containing a 7 segments display, indication leds and buttons. The keyboard is composed of 20
buttons and 4 dip switches organized in a 5 x 5 mother; on the board there also are the writing latch (374) and
the reading buffer (244).
The displays and leds are controlled by three M5451 integrated circuits, each of them drives two multiplexate
led groups.
These integrated circuits are serie/parallel converter which shares the clock signal keeping the data lines
separated; on the board there are three trimmers for the light intensity regulation of leds piloted by each
integrated circuit and adjusting them it is possible to conform the three groups.
COOLING FAN WHEEL

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 11 / (72 ) ( rev.3) 30.06.2004
PANEL (PANEL MANAGER)
The panel control board, contains an Hitachi H8/325 micro controller which effects the keyboard reading and
displays management.
It communicates with the main micro controller (H8/532 on the LV200-MAIN/1 board) through a serial port; it
forwards the acquired buttons and receives the data to be displayed using an internal protocol.
The board is equipped with WATCH DOG circuit not sensible to electric power supply realized with an NE555;
such a circuit is able to generate a microprocessor reset if it is not periodically “refreshed” by the software,
retaking the machine in a know status in case of not correct performance of working program.
The program, which is located in an EPROM inside the micro controller, performs a complete scanning of
keyboard approx. each 260 µs and manages the displays with a multiplexer frequency of approx. 47 Hz.
LV200I/O (INPUT/OUTPUT)
Input/output board managing the pneumatic circuit.
The printed circuit is divided in two sections, whom are kept separate as most as possible, an analog and a
digital one.
In the analog part there are stabilizers for the necessary electric power supplies (+10 V, -10 V) and all circuits
for input signals conditioning, namely the amplifiers, filters, protection devices.
The analog inlets are:
• pressure signal (from transducer MPX10 mounted on board edge);
• O2 concentration signal (from an external oxymetry sensor);
• MINUTE VOLUME setting signal (from potentiometer on the panel);
• measurement signals of O2and AIR flows (from two external flow meters).
The mass of the analog section is physically kept separate from the mass of digital section; the only contact
point of these two masses is on the LV200-MAIN/1 board on the microprocessor pins.
The digital section of the board includes:
• two digital inlets (contact cleaned by the two pressure switches);
• an outlet for signaling buzzer
• an ON/OFF electro valve outlet for medical gas commutation flow;
• two outlets for flow by-pass;
• two control outlets for proportional electro valves which regulate the flow of the two medical gas.
The proportional electro valves control is performed in PWM by two integrated circuits TL494, arranged on a
commutation frequency of approx. 30 KHz. The feedback of each of these integrated circuits is made on the
electric power of the relevant proportional electro valve; the set point (as electric power) is supplied by main
micro controller (H8/532 of LV200-MAIN/1 board) through an integrated low frequency signal PWM.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 12 / (72 ) ( rev.3) 30.06.2004
3.5 Gas administration and proportioning pneumatic system
The gas administration and proportioning pneumatic system is represented in the following figure and
mechanical study.
The system has the function to administrate in a correct and safe way the breathing gas to the patient, under the
total control of electronic system.
FR 3
MAX PAW
FR3.1
FR 3 PEEP/CPAP
EV 3
V.EXP.
R1 R2
R1.1 R2.1
LPV2
EXP. VALVE
PATIENT
PR 1
CSV 1
GAS SELECTOR
VALVE
PR 2
PR 3
AIRWAY
PRESSURE
TRANSDUCER
OXIGEN
SENSOR AIR SAFETY
VALVE
EV PROP. 1
O2 EV PROP. 2 INSP.
A
IR
FS 1
O2 FLOW
SENSOR
FS 2
A
IR FLOW
SENSOR
EV PART. 1
O2 EV PART. 2
A
IR
PSW 2
PSW 1
INLET O2 3.5 bar +/- 0.75
INLET AIR 3.5 bar +/- 0.75
EV4

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 13 / (72 ) ( rev.3) 30.06.2004
3.5.1 Operation description
O2 INLET – AIR INLET Medical gas inlets (Medical OXYGEN and AIR, coming from the central piping system
or from cylinders).
PR1 - PR2. Pressure reducers. They have the function to reduce the inlet gas pressure to the stable value of
approx. 2.6 bar.
PSW1 - PSW2. Pressure switches. They have the function of signaling the eventual gas failure or low gas
pressure.
CSV1. Circuit selecting valve. It has the function to guarantee the operation of pneumatic devices supplied by
PR3 reducer also with only one gas in the inlet.
PR3 Pressure reducer. It has the function to reduce the pressure of the gas coming from CSV1 to the stable
value of approx.1.5 bar.
Only for SIARETRON 1000 and SIARETRON 3000
EV PART 1 O2 - EV PART 2 AIR. 12 V Pilot electrovalve. These valves are supplied from PR3 and have the
function to control the opening and closure of pneumatic by-passes for the measurement of flows over 15 l/min.
(see description of gas proportioning assembly). They are controlled by microprocessor.
EV PROP 1 INSP O2 - EV PROP 2 INSP AIR. 6V 0-120 l/min proportional electro valves. These valves regulate
the gas delivered to the patient allowing to control both the O2-AIR concentration and the FLOW. They are
controlled by microprocessor through a feedback system on the flow guaranteed by PS1 and PS2 flow sensors
Only for SIARETRON 1000 and SIARETRON 3000
R1.1 - R1 - R2.1 - R2. Pneumatic resistances. They are made of holes calibrated in accordance to the function
of distributing the flow coming from the proportional valves.
When the flow is lower or equal to 15 l/min the EV PART 1 and 2 valves are excited and the communication
with R1 and R2 is interrupted. In this way the flow will completely pass through the PS1 and PS2 sensors for
measurement.
When the flow is higher than 15 l/min the EV PART 1 and 2 valves are disexcited and the communication with
R1 and R2 is open. The special connection between the holes’ diameters and the form of channels, permit to a
know portion of the flow to bypass PS1 and PS2 and to the remaining part to pass through them for the
measurement. In this way it is possible to measure the flow since from 0.1 to 99 l/min with sensors from 0.1 to
15 l/min.
AIR SAFETY VALVE. Unidirectional mechanic valve. It has the function to allow to the patient to breath ambient
air in emergency case. It opens with a depression of approx. 2 mBar.
OXIGEN SENSOR. It has the function of measure the real oxygen concentration delivered to the patient.
AIRWAY PRESSURE TRANSDUCER. Pressure transducer with semiconductor. It has the function to
continuously measure the pressure inside the breathing circuit.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 14 / (72 ) ( rev.3) 30.06.2004
FR3 PEEP/CPAP. Flow regulator. It has the function to keep inflated the LPV2 expiratory valve, during the
expiration phase in CPAP modality, so that the patient CFR (residual functional capacity) will rise.
FR3 MAX PAW. Flow regulator. It has the function to keep inflated the LPV2 expiratory valve during the
inspiration phase, so that the circuit discharge will be closed and the lungs inflating will be possible. This
regulator is tared in combination with FR3.1 regulator, so that the expiratory valve will automatically open if the
airways pressure is higher than 80 mBar.
EV3 V. EXP. 12 V pilot electrovalve. It has the function to activate the LPV2 expiratory valve in accordance with
the inspiratory and expiratory cycle. It is microprocessor controlled.
EV4. DISCHARGE. 12V pilot electrovalve. It has the function to discharge in the environment the exceeding
inspiratory portion during PS and CPAP. It is microprocessor controlled.
FR3.1. Flow regulator. It has the function to regulate the discharge of LPV2 expiratory valve inflating system; it
works in combination with FR3 and FR3 MAX PAW.
FS1 O2FLOW SENSOR, FS1 AIR FLOW SENSOR. Flow capacity transducers. They have the function to
measure the air and O2 flows delivered to the patient. Thanks to R1.1, R1, R2.1 and R2 it is possible the flow
measurement from 0.1 to 99 l/min with sensors from 0.1 to 15 l/min.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 15 / (72 ) ( rev.3) 30.06.2004
4 CALIBRATION AND TESTING
4.1 LV200-MAIN/1 CPU board
Connect the POWER, PANEL and LV200I/O boards (already tested) and the board to be tested by the
appropriate cables kit; referring to the following connecting drawings:
Connect the secondary of transformer to CN7 POWER connector; supply the primary of transformer.
To enter the testing program, press the ON/OFF switch keeping pressed the LOW AIRWAY PRESS. Íand HI
AIRWAY PRESS Íkeys until hearing a short sound, hence release the above indicated switches and press
the ALARM RESET key within 5 seconds. Another short acoustic signal will occur, on the MINUTE VOLUME
display message TESTING will scroll once, after that on the same display will appear alternatively the
messages: I-O | ofG .
220Vac
LV200PW
LV200_MAIN1
LV100-P/1
LV100
-
ESP
LV200I/O
CVL
-
056
CVL
-
053
CVL-072
CVL-052
CVL-071
CVL-058
CVL-054
CVL-055
Transformer
(
220Vac - 12 Vac
)
c
n-7
c
n-
8
cn-1
cn-2
P-8
P
-
3
P-10
P-2
P-5 P-4
JP1
P-3
JP1
cn-8
cn-3
cn-1
POT
EV ON/OFF
cn-5
1

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 16 / (72 ) ( rev.3) 30.06.2004
Testing Performance Procedure
Acting on the PANEL board, while the alternated scripts: I-O | ofG (on DS13/14/15) appear, perform the
following checks (the referrals under brakes are relevant to the positions on the PANEL board):
• By the AUT. TIME CYCLED (SW7) button the ON/OFF electrovalve must light ON and OFF
following the status of the LED (LD1) next to the (LED ON, LED OFF) button.
• By the AUT. PRESS CYCLED (SW2) button the BUZZER must light ON and OFF following the
status of LED (LD2) next to the (LED ON, LED OFF) button.
• With the main power supply the green ON LINE LED (LD12) must be light on.
• With the external battery the yellow ON BATT LED (LD11) must be light on, simulate the
external battery with power supply regulated at 12 Vdc connected as follows:
Power
supply
• Verify the connection of LOW BATT (LD10) alarm LED lowering the external power supply
whom simulates the battery. At an electric power between 10.5 and 11.2Vdc the alarm must
intervene.
• By pressing the ON/OFF switch (SW1) the yellow SPONT. BREATH DETECTOR LED (LD5)
must light on.
• Short-circuiting the CN6 connector feet of the LV200I/O board (Air inlet pressure switch) the red
LOW AIRWAY PRESS (APNEA).LED (LD7) must light off.
• Short-circuiting the CN7 connector feet of LV200I/O board (O2 inlet pressure switch) the red
GAS SUPPLY.LED (LD9) must light off.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 17 / (72 ) ( rev.3) 30.06.2004
A/D channels testing.
The testing phase of A/D channels starts by pressing the ALARM RESET (SW20) button. On O2CONC
(DS9/10) and SIMV RATE (DS11/12) displays respectively appear messages: An and In , on RATE (DS3/4)
display appears message: Ch , on % INSP (DS5/6) display appears the channel on which the monitoring is
effected and on MINUTE VOLUME (DS13/14/15) display appears the analog/digital conversion value. With
the % INSP Í(SW14) and Î(SW17) switches the A/D channel can be adjusted from 0 to 5; perform the
following checks:
a) Ch 0 (pressure): vary the inlet pressure of MPX10 sensor on the LV200I/O board from 0 to 80 cm of
H2O and verify that the A/D value on the MINUTE VOLUME (DS13/14/15) display changes from 20
± 1 to 100 ± 2.
b) Ch 1 (oxymeter): apply to the CN12 connector an electric power variable from 0 to 75 mV between
the pin 4 (+) and the pin 1 (GND); verify that the corresponding A/D value changes from 0 to 225 ± 2.
c) Ch 2 (MINUTE VOLUME potentiometer): rotate the potentiometer (POT) of MINUTE VOLUME
regulation and verify that the corresponding A/D value changes from 0 to 1023 (023 on display).
Ch 3 : not used.
d) Ch 4 (AIR flowmeter): apply to the CN13 connector an electric power variable from 0 to 5 V between
the 4 (+) pin and the 2 (GND) pin; verify that the corresponding A/D value changes from 0 to
1000 ± 2.
e) Ch 5 (O2 flowmeter): apply to the CN14 connector an electric power variable from 0 to 5 V between
the pin 4 (+) and the pin 2 (GND); verify that the corresponding A/D value changes from 0 to
1000 ± 2.
After the test verify the machine switch-off by pressing ON/OFF (SW1) button.
5) If the testing result is positive, put on the board a green label indicating the identification number, the date
and tester technician signature.

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 18 / (72 ) ( rev.3) 30.06.2004
Component lay out of LV200-MAIN/1 board

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 19 / (72 ) ( rev.3) 30.06.2004
Component lay out of LV200 I/O board
H
I
G
B
C
TP2
TP3
TP1 TP5

SIARETRON 1000 – 1100 - 3000
Service Manual
SIARETRON_SE3_en 20 / (72 ) ( rev.3) 30.06.2004
Component lay out of PANEL BOARD (component side):
This manual suits for next models
2
Table of contents
Other Siare Medical Equipment manuals

Siare
Siare Falco 101 User manual

Siare
Siare Morpheus Series User manual

Siare
Siare Falco 202 User manual

Siare
Siare AM5000 User manual

Siare
Siare Morpheus E User manual

Siare
Siare Falco 202 Evo User manual

Siare
Siare Siaretron 4000 User manual

Siare
Siare Morpheus ND User manual

Siare
Siare Aria 104 User manual
Popular Medical Equipment manuals by other brands

Reison
Reison 10-390 Series User instruction

PCE Instruments
PCE Instruments PCE-889A user manual

Bauerfeind
Bauerfeind EpiPoint manual

Bard Peripheral Vascular
Bard Peripheral Vascular LifeStent 5F Instructions for use

GlobalMed
GlobalMed CVL ClearSteth user manual

otometrics
otometrics HORTMANN Aquamatic II user guide