
15fMRI User Guide
There are two ways in which one can synchronize activities: manual or automatic. Manual synchronization is
not accurate and is therefore not recommended. However, if it is absolutely necessary to manually synchronize
your stimulus presentation with the EPI sequence, one first needs to calculate how many dummy scans the EPI
sequence will be running (see Section 3.2.8). Based on this knowledge, one can start the sequence first, then
count the measurements (dummy scans) by hearing the gradient sounds (each measurement will typically be a
continuous sound, followed by a small break, and then followed by the next measurement), and start the stimulus
presentation when the required numbers of measurements = dummy scans are completed. If the stimulus needs
to be manually switched between “Baseline” and “Active” states (Figure 5, trace (a)), one can use the display at the
bottom left on the screen, which displays either “B” (Baseline) or “A” (Active) to indicate what part of the paradigm
is being acquired by the EPI sequence. This paradigm displays what is specified on the BOLD card in the protocol
menu of the sequence. For example, if the total numbers of measurements is 60 in the EPI measurement, and
the paradigm size is set to 20, with measurements 1 to 10 being “Baseline” and measurements 11 to 20 being
“Active”, then you will see the display “B” at the bottom left of the screen for the first 10 measurements. On the
11th measurement this display will switch to “A” until the 21st measurement, when it will switch back to “B”, and
so on. When “A” and “B” toggle, that is the cue for switching the stimulus between “Baseline” and “Active” states.
Automatic synchronization is usually done by means of a TTL signal that is fed to all the various equipments used,
and serves as a reference signal for timing of stimulus, EPI data, physiological measurements, etc. In general, a
separate computer, other than the scanner, controls the stimulus presentation. The physiological recordings can
either be controlled by another separate computer (preferred) or the stimulus computer. Since different events are
controlled by different machines, all events are synchronized by means of a timing pulse that is generated by the
EPI sequence on the scanner, and is fed as input to all other computers that control either stimulus presentation
or physiological monitoring. A TTL pulse can be read into a computer in several ways. Some data acquisition
peripherals come with an analog input option. If free slots are available on the motherboard, an analog I/O card
can be purchased. Or if one has access to an electronics shop, an adapter can be made for the parallel port that
would drive a pin high or low when a pulse is sent. For details on the synchronization signal, please refer to
Section 3.2.1.
4.7 Installing peripheral devices
The major challenges to installing peripheral devices within the scanner room are the introduction of RF noise
in the images, and the effect of a strong magnetic field on the device itself. The former may lead to a loss of
SNR in the images, and may cause artifacts, while the latter may damage the peripheral device. RF noise causes
degradation of image quality and may go undetected for a long time if it doesn’t cause readily visible artifacts.
Moreover, once all peripheral devices are installed, diagnosing the source of RF noise can be very laborious.
Therefore, it is critical that utmost care be taken when installing the devices.
The higher the magnetic field, the higher the likelihood that any components in the device may fail. In general,
the magnetic field drops rapidly as the distance from the scanner increases. Conversely, as the device is brought
closer to the magnet, the magnetic field that the device experiences rises rapidly. Typically, electronic devices
with inductors, transformers with iron cores, CRTs, metallic objects in motion, etc. will be affected the most by the
magnetic fields.