6. SUGGESTED INSTRUCTIONS FOR USE:
6.1 Initial Settings:
6.1.1 If this is the first use of the acapella®choice
system, ensure that the frequency adjustment
dial is turned counterclockwise to its lowest
frequency-resistance setting, indicated at “1”.
(See Figure 2)
6.1.2 Instruct the patient to relax while performing
diaphragmatic breathing and inspiring a
volume of air larger than normal tidal volume
(but not to total lung capacity).
6.1.3 Direct the patient to exhale to Functional
Residual Capacity (FRC) actively, but not
forcefully, through the device.
6.1.4 Adjust the dial to increase the resistance.
6.1.5 The patient should be able to exhale for 3-4
seconds while the device vibrates. Clockwise
adjustment increases the resistance of the
vibrating orifice; which will allow the patient to
exhale at a lower flowrate. Counter-clockwise
adjustment decreases the resistance.
6.1.6 Once the proper range has been identified, the
patient may be instructed to exhale harder or
softer, or dial adjustments may be made to
optimize the response the user “feels” from the
vibratory pressure.
6.2 Procedure for the User:
6.2.1 Ensure the adjustment dial is set to the correct
range as identified by your clinician.
6.2.2 Sit with your elbows resting comfortably on
the table.
6.2.3 Place the mouthpiece lightly in your mouth.
• Be sure to maintain a tight seal on the mouthpiece
during exhalation.
• Your clinician may recommend the use of a nose
clip, if necessary.
• If using a mask, apply the mask tightly but
comfortably over your nose and mouth.
6.2.4 Breathe from the diaphragm, as directed by your
clinician, taking in a larger than normal breath,
but not filling your lungs to capacity.
6.2.5 Hold your breath for 2-3 seconds.
6.2.6 Exhale actively, but not forcefully, through the
device. Exhalation should last approximately 3 to 4
times longer than inhalation.
6.2.7 Perform 10-20 PEP breaths as recommended by
your clinician.
6.2.8 Remove the mouthpiece (mask) and perform
2-3 “huff” coughs to raise secretions as
needed. Your clinician may direct you on proper
cough technique.
6.2.9 Repeat steps 6.2.2 to 6.2.7 as prescribed.
Note: See nebulizer set-up section.
6.3 Set-up Nebulizer
6.3.1 Review the diagrams contained with the devices.
6.3.2 A possible nebulizer and acapella®choice system
set up is reflected below. (See Figure 3)
6.3.3 Follow set-up instructions for each device.
6.3.4 Follow cleaning instructions contained with
each device.
6.3.5 Inspect device(s) on a routine basis to ensure
proper use and function.
6.3.6 If damaged, do not use.
6.3.7 Verify all connections are secure.
7. CLEANING AND DISINFECTING
INSTRUCTIONS:
Precaution: Bleach is not recommended for use on the
acapella®choice system. It may deteriorate the nickel
plated mechanism located in the interior of the device.
7.1 There are four parts to clean:
A. The mouthpiece
B. The base unit (with frequency adjustment dial)
C. The cover
D. The rocker assembly – DO NOT disassemble the
rocker arm from its platform base
7.2 Disassembly
7.2.1 Detach the mouthpiece or mask. (See Figure 4)
7.2.2 Detach the cover by placing your forefinger and
thumb on the (3) ribs located on the unit toward
the mouthpiece.
7.2.3 Gently press on the ribbed surface and lift
the cover up.
Note: The cover should only be lifted to a 90°
angle. (See Figure 5)
Warning: The rocker assembly is not designed to
be disassembled - do not attempt to separate the
rocker arm from its platform base. Separating
these components may cause the acapella®
choice system to function improperly, which may
deprive the patient of therapy and cause injury to
the patient.
7.2.4 Detach the rocker assembly by placing your
forefinger and thumb.
7.3 Cleaning/Disinfecting Instructions
As per the Cystic Fibrosis Foundation’s cleaning
and disinfecting guidelines entitled, “Respiratory,
Stopping the Spread of Germs” 2003, below are
the guidelines for the acapella®choice system.
Note: The acapella®choice system is for single
patient use
7.3.1 Cleaning: This should be done prior to
Disinfecting (7.3.2)
Cleaning with Liquid Dish Detergent:
As needed, detach the mouthpiece (mask) then
soak the device and mouthpiece in warm, soapy
water as required to remove visible contaminants.
Use a liquid dish detergent (Dawn®or equivalent),
mixing two (2) tablespoons of detergent per one (1)
gallon water. Rinse thoroughly with sterile water,
and allow parts to air dry.
Drain the device by placing it in a normal resting
position. (See Figure 2)
7.3.2 Disinfecting:
• Boiling – Boil the device in water (100°C/212°F)
up to twice daily for five (5) minutes. In addition,
Smiths Medical suggests the use of distilled
or sterile water to lessen the potential of local
community, mineral-rich tap waters from
calcifying the acapella®choice systems metallic
components.
• Autoclaving – The acapella®choice system
will functionally withstand autoclaving at
temperatures not to exceed 136°C/277°F
for a maximum of 30 cycles.
• Automatic Dishwasher – The acapella®choice
system is dishwasher safe. It is recommended to
place the parts on the top shelf.
• Alcohol – Soak five (5) minutes, twice daily.
The acapella®choice system is compatible with
70% isopropyl alcohol. Rinse with sterile water.
You can make water sterile by boiling for five
(5) minutes.
• Glutaraldehydes (Cidex®or equivalent) –
The acapella®choice system will maintain its
integrity using cold sterilizing solutions such
as glutaraldehydes.
8. DISPOSAL:
Dispose of the acapella®choice system in a safe
manner according to Federal/State/Local
regulations and guidelines for disposal of
contaminated medical waste.
9. ESTIMATED DEVICE LIFETIME:
Assuming the manufacturer recommended
cleaning protocol is followed, the acapella®choice
system devise should have a useful life of six (6)
months under normal and customary usage.
The six (6) month usage duration is measured
from the date of initial use.
Ref
acapella®choice
system Catalog
Number
Quantity Per
Carton
Mouthpiece 27-7000 10
Caution •Latex Free •Do not use if package is
damaged •Non-sterile •Contains or Presence of
Phthalate: bis(2-ethylhexyl) phthalate (DEHP)
[www.smiths-medical.com/phthalates] •Catalogue
Number •Batch Code •Date of Manufacture
Caution: Federal (U.S.A.) law restricts this device to sale
by or on the order of a physician. SINGLE PATIENT USE
f ~ : ; ‚
< = J 2 6
Smiths Medical ASD, Inc.
6000 Nathan Lane North
Minneapolis, MN 55442, USA
Tel: 1-614-210-7300
Toll-Free USA: 1-800-258-5361
www.smiths-medical.com