WARNING: This training guide should be used by clinicians only. Do not permit patients to have
access to this guide, as the information would allow access to all programming and operating
functions. Read the entire Operator’s Manual before operating the CADD-Legacy®ambulatory
infusion pump. Failure to properly follow warnings, cautions and instructions could result in
death or serious injury to the patient. If the pump is dropped or hit, inspect it for damage. Do
not use a pump that is damaged or is not functioning properly.
To Start The Pump: Press and hold STOP/
START until Starting and (– – – – – – – – – – – –)
disappear one by one from the display. Release
the STOP/START key. RUN appears on the display.
To Stop The Pump: Press and hold STOP/START
until Stopping and (– – – – – – – – – – – –) appear
one by one on the display. Release the STOP/
START key. STOPPED appears on the display.
To Prime: With the pump stopped and in LL0 or
LL1, press and hold PRIME until the word Prime
appears on the display, followed by (– – – – – – –
– – – – –) that appear one by one. Release Prime.
Press and hold PRIME until fluid path is free of air.
Press NEXT to exit priming.
Warning: Do not prime the fluid path with the tubing
connected to a patient as this could result in
over-delivery of medication or air embolism.
To Turn The Pump On: Press ON/OFF. The pump
beeps as it cycles through its power-up routine.
To Turn The Pump Off: Press and hold ON/OFF
until (•••• •••• ••••) appear one by one and
Turning Off appears on the display. The screen will
go blank as the pump goes into a lower power state.
OTHER PUMP FUNCTIONS CHANGING LOCK LEVELS BIOMED FUNCTIONS
The pump must be stopped.
1. Press LOCK to view the current lock level.
2. Press ´orÎuntil the desired lock level
appears (LL0, LL1, LL2).
3. Press LOCK or ENTER/CLEAR.
Code and 0 appear.
4. Press ´orÎuntil the lock level
code appears.
5. Press LOCK or ENTER/CLEAR
to set the new lock level. Press LOCK
again to verify the new lock level.
LOCK LEVEL 0 (LL0) is the programming level.
LOCK LEVEL 1 (LL1) allows the
patient to start and stop the pump, prime
the tubing, reset the reservoir volume, clear
the GIVEN screens and program the DOSE
STARTS In screen. The Continuous Rate (ml/
hr) can be decreased and increased up to the
value programmed in LL0.
LOCK LEVEL 2 (LL2) allows the patient
to stop and start the pump and reset the
reservoir volume.
The pump must be stopped and in LLO.
To Access:
1. Press LOCK.
2. Press LOCK or ENTER/CLEAR. Code and 0 appear.
3. Press
´orÎuntil the biomed code appears.
4. Press LOCK or ENTER/CLEAR.
5. Press NEXT to go through settings.
To Turn Air Detector ON-HIGH, ON-LOW or OFF:
6. Press NEXT until AIR DETECTOR appears.
7. Press ´orÎuntil desired setting appears.
8. Press ENTER/CLEAR.
WARNING: When the air detector is turned off , the pump will not detect
air in the fluid path. Periodically inspect the fluid path and remove any air
to prevent air embolism.
To Turn Upstream Sensor ON/OFF:
9. Press NEXT until UPSTREAM SENSOR appears.
10. Press ´orÎuntil desired setting appears.
11. Press ENTER/CLEAR.
WARNING: When the upstream occlusion sensor is turned off, the pump will
not detect occlusions upstream (between pump and reservoir). Periodically
inspect the reservoir for decreasing reservoir volume and inspect the fluid
path for kinks, a closed clamp, or other upstream occlusions. Upstream
occlusions could result in under- or non-delivery of medications.
To Change Delivery Mode:
12. Press NEXT until DELIVERY MODE appears.
13. Press ´orÎuntil desired setting appears
(Intermittent
or Continuous).
14. Press ENTER/CLEAR.
To Exit:
15. Press NEXT until NEXT FOR BIOMED, ENTER FOR MAIN appears.
16. Press ENTER/CLEAR to return to main screen.
5
Continuous Delivery Mode
In LLO, STOPPED
To change delivery mode, see BIOMED FUNCTIONS.
Install new batteries in battery compartment. Slide battery door closed.
1Reservoir Volume (ml)
Scroll to the volume contained in the
Medication Cassette Reservoir. Press
ENTER/CLEAR to enter value. Continuous Rate (ml/hr)
Select the infusion rate. Press
ENTER/CLEAR to enter value.
2
3
Air Detector (Status Only)Verify setting is
correct. Screen shows whether sensor is
turned On (High or Low) or turned Off. To
change setting, see BIOMED FUNCTIONS.
4
Upstream Sensor (Status Only)
Verify setting is correct. Screen shows
whether sensor is turned On or Off. To
change setting, see BIOMED FUNCTIONS.
Given (ml) Displays the total amount of
medication the pump delivered since last
cleared. Press ENTER/CLEAR to clear value.
THE DETAILS GIVEN IN THIS LEAFLET ARE CORRECT AT THE TIME OF GOING TO PRESS. THE COMPANY RESERVES THE RIGHT TO IMPROVE THE EQUIPMENT SHOWN.
CADD-Legacy and the Smiths Medical and CADD design marks are trademarks of Smiths Medical.
The symbol ® indicates the trademark is registered in the U.S. Patent and Trademark Office and certain other countries.
©2010 Smiths Medical. All rights reserved. 11/10 IN19894
Product(s) described may not be licensed or available for sale in Canada or other countries outside of the United States.
www.smiths-medical.com
Smiths Medical is part of the global technology business Smiths Group plc.
CADD-Legacy®PLUS Pump
Model 6500
Continuous Mode Training Guide
WARNING: This pump delivers medication at a continuous rate or intermittent rate in milliliters per hour (ml/hr).
Programming the pump at a delivery rate other than what is prescribed will cause over- or under-delivery of medication.
SCROLLING
Use the up and
down arrow keys
to increase or
decrease the value
on the screen.
Rate is in
ml/hr
Smiths Medical ASD, Inc.
St. Paul, MN 55112, USA
Phone: 1-214-618-0218
Toll-Free USA: 1-800-258-5361
Smiths Medical Canada Ltd.
Markham, Ontario, Canada, L3R 4Y8
Phone: 905-477-2000
Toll-Free: 1-800-387-4346
EC Authorized Representative
Smiths Medical International Ltd.
TN25 4BF, UK
Phone: +44 (0) 1233 722100
IN19894.indd 3 28-10-11 14:21