
1–10 Operation
SRS QMS Gas Analyzer
The microcontroller is inactive for about 5 seconds after main power is turned on. The user cannot turn
on any of the system components during this period. If the user does turn on one of the front panel
switches during this period, it will be rejected and the adjacent light will blink. To recover from this
state, turn off all four switches and begin again. This restriction is related to power failures, which are
discussed later in this chapter.
The microcontroller was programmed to support two modes of sampling: continuous and batch.
Continuous mode is suited for samples of unlimited volume that change with time. Batch mode is for
small volume samples of fixed composition. The QMS system is strongly oriented towards the
continuous sampling mode, which makes good use of ability of the instrument to perform high speed
continuous measurements. For situations where this flowrate is unacceptable, the batch mode may be
useful. The next two sections describe how to setup the instrument to perform these two types of
sampling.
Continuous Sample Mode
Continuous sampling requires that the system be set to state 4 in the state diagram (pumps on and all
valves open). As the diagram shows, there are two main paths to get from Off to Sample Flow. The
preferred path follows the diagram through state 2A. The user should in sequence turn on all four
switches in the order Mechanical Pump, Capillary Flow Valve, Turbo Pump, and Sample Inlet Valve.
The user does not have to wait for each action to be completed; just toggle the switches in the order
you wish the actions to occur. The controller will perform the sequence, following the state diagram
through state 2A. When these actions are complete, the system will be sampling and ready to record
data. This is the main mode of operation. Other actions are described in the following sections.
Startup Just the Pumps
To prepare the system, but not draw any sample gas, the two pumps can be started. In this state the
RGA can be operated, although the mass spectrum will only show the chamber background. This state
is useful to prepare the software, test custom programs, and to idle the system (discussed further
below). Press just the Mechanical Pump and Turbo Pump switches, which will bring the system to state
2B in Figure 6. The microcontroller will start the diaphragm pump first. After a few seconds, the light
will become bright and the turbo pump will start. The turbo pump accelerates to full speed in under 2
minutes. During the acceleration of the pump, the current will reach full scale. Once it reaches full
speed, the light will become bright.
Startup the Sample Flow
To start the sample flow from state 2B, the two valves need to be opened. The user can toggle the
Capillary Flow Valve and then the Sample Inlet Valve switches to continue to state 4. At state 2B,
there is a volume of trapped gas that is at atmospheric pressure between the exit of the capillary and the
entrance of the capillary flow valve. This volume must be removed slowly. The solenoid valves are
either fully open or fully closed, so the microcontroller follows a pulse program to open the valve. The
valve will open momentarily and then close. This will be repeated as many times as necessary to start
the capillary flow without stopping the turbo pump. Capillaries with large volumes will require several
steps; this will be accompanied by much clicking, which is normal. Once the capillary flow is