1. SSCOR sucon units are not designed or intended for use in extended procedures that require prolonged
high vacuum/low airflow applicaons, as is the case in wound drainage or endoscopic use or in any other
procedure that produces high vacuum levels within an occluded system for an extended period of me.
Turn the sucon unit offwhen it is not in use.
2. Federal law restricts this device to sale, distribuon, and use by, or on the order of a physician, emergency
medical technician, or other medical praconer. For use by medical personnel trained in suconing
techniques and in the use of medical sucon equipment.
3. Operator should be thoroughly familiar with these operang instrucons before this device is used.
4. Do not use in the presence of flammable agents or anesthecs.
5. External equipment intended for connecon to signal input, signal output or other connectors, shall
comply with relevant IEC standard (e.g. IEC 60950 for IT equipment and the IEC 60601 series for medical
electrical equipment). In addion, all such combinaons –systems – shall comply with the standard IEC
60601‐1‐1, safety requirements for medical electrical systems. Equipment not complying with IEC 60601
shall be kept outside the paent environment, as defined in the standard. Any person who connects
external equipment to signal input, signal output or other connectors has formed a system and is there
fore responsible for the system to comply with the requirements of IEC 60601‐1‐1. If in doubt, contact a
qualified technician or your local representave. For further informaon, reference the SSCOR Technical
Manual.
6.This device is not intended for suconing neonates.
7. To prevent fire of injury when baeries are not in the device or charger, always place the baeries in a
protecve pack to cover the terminals.
8. The hydrophobic filter in the non‐sterile, single use, canister will close the system when fluids contact the
filter. ONCE THE FILTER COMES INTO CONTACT WITH FLUIDS, THE UNIT WILL NOT SUCTION UNTIL A
NEW CANISTER IS INSTALLED. CHANGE CANISTER IMMEDIATELY AFTER MOISTURE SHUTS DOWN THE
UNIT.
9. Do not point the catheter directly upward when fluids are present in the canister. Poinng the catheter
upward will allow fluids to reach and occlude the shut‐offfilter in the canister.
10. Do not remove the baery from the sucon unit while the sucon unit is connected to the charging
source. This may result in a false reading from the baery condion indicator.
Cauon‐Noce Model 2402
11. Grounding reliability can only be achieved when connected to an equivalent receptacle marked “Hospital
Only” or “Hospital Grade”.
SSCOR Quickdraw®
US Patents 7,063,688 B2 ‐D564,654 S ‐7,938,794 B2
©2015 Quickdraw, HI‐D and Big Sck are registered trademarks of SSCOR, Inc
Quickdraw - Model 2400 Series
2
Clearing The Airway Is Our #1 Priority
Caution-Notice Revision M (02/17)
Rev. M (02/17)