Observe the following maintenance roune to ensure readiness at any me:
1. When the SSCOR aspirator is not in use, keep baeries on connuous charge.
2. Test the SSCOR aspirator at regular intervals; See page 6.
3. Make sure the SSCOR aspirator is always clean and ready for use.
Note : No part requires lubricaon and lubricants should not be used.
Sanitaon
As soon as possible aer use, the single use disposable canister, paent tubing and catheter should be
discarded according to local / regional / naonal requirements for the disposal of hazardous waste
materials. Clean using a mild detergent and if necessary disinfect with a mild disinfectant such as 10 to 1
bleach and rinse using clear water to remove any residue.
NOTE: The hydrophobic filter in the canister helps to ensure that no moisture or parculate maer reaches
the inside of the device. In the unlikely event that fluids may have reached the vacuum pump, read the
disinfecon secon. Do not reuse any single use disposable parts; do not submerge the device into any
liquid, this will void the warranty and cause the device to malfuncon.
Disinfecon
Use personal protecve equipment such as gloves, a smock, and face and eye protecon when handling
units that are suspected to be contaminated.
Cauon: Disconnect the unit from any power source prior to cleaning the unit. When cleaning the interi‐
or of the chassis, disconnect the baery from the PC Board to prevent damaging the PC Board.
Part Cleaning and Disinfecng
Collecon Canister Disposable item, re‐use not permied. Use new canister for each paent.
Paent Tubing Disposable item, re‐use not permied. Use new paent tubing for each paent.
HI‐D® Sck Disposable item, re‐use not permied. Use new HI‐D Sck for each paent.
Vacuum Pump
Wipe with damp cloth or disinfectant wipe. Sterilizaon not permied. Vacuum
pump should be replaced if contaminated
PC Board Sterilizaon not permied. PC Board should be replaced if contaminated
Plasc Chassis Wipe with damp cloth or disinfectant wipe. Sterilizaon not permied.
Disinfect the unit using a mild surface disinfectant, such as a 10:1 mixture of water and bleach. The unit is
designed to sucon contaminated fluids, which should be removed from the system immediately aer use.
In the unlikely event that fluids may have reached the vacuum pump, your engineering department will have
to open the unit to check the condion of the pump. When cleaning the interior of the chassis, disconnect
the baery from the PC Board to prevent damaging the PC Board. The only foreseeable way fluids may reach
the vacuum pump is that the filter in the canister has been compromised or bypassed.
If the PC Board appears defecve, return the unit to the factory for repair. Do not aempt to repair the PC
Board. If the vacuum pump appears defecve, return the unit to the factory for repair. Do not aempt to
repair the vacuum pump. For technical assistance, call (800) 434‐5211 or +1 (818) 504‐4054.
WARNING:
To avoid any contaminants reaching the interior of the device, always remove the canister per the
instrucons in the sanitaon secon. Only use SSCOR, Inc. canisters affixed with hydrophobic filters.