SSCOR S-SCORT III 74000 Instructions for use

Model 74000
Clearing The Airway Is Our #1 Priority
Operating Instructions
& Maintenance Manual
Tel 1+818-504-4054
Tel 1+800-434-5211
, INC.
11064 Randall Street
Sun Valley, CA 91352 USA
www.sscor.com
shop.sscor.com
techsupport@sscor.com

1. SSCOR sucon units are not designed or intended for use in extended procedures that require prolonged
high vacuum/low airflow applicaons, as is the case in wound drainage or endoscopic use or in any other
procedure that produces high vacuum levels within an occluded system for an extended period of me.
Turn the sucon unit offwhen it is not in use.
2. Federal law restricts this device to sale, distribuon, and use by, or on the order of a physician, emergency
medical technician, or other medical praconer. For use by medical personnel trained in suconing
techniques and in the use of medical sucon equipment.
3. This manual is restricted to the discussion of the use and maintenance of this device. It does not aempt
to discuss professional techniques in suconing procedures.
4. Operator should be thoroughly familiar with these operang instrucons before this device is used.
5. Do not use in the presence of flammable agents or anesthecs.
6. The S‐SCORT III produces a powerful vacuum. Do not use the S‐SCORT III to sucon neonates.
7. Before tesng for vacuum over ‐300mmHg look for an expiraon date on the canister (where applicable)
and change the canister if the canister has passed the expiraon date to minimize the possibility of
implosion, which can occur when a canister is aged or damaged.
8. The sucon pump must be reconnected to the charging source aer each use and remain on charge unl
needed. Disconnect only for portable use. Check the condion of the baery frequently , see the baery
test on page 6. SSCOR’s AC‐DC charger only charges the baery, it will not power the sucon device. The
sucon device will run on vehicle power when the unit is connected directly to the vehicle baery via the
DC power cord.
NOTE: DO NOT ATTEMPT TO CHANGE THE ELECTRICAL SYSTEM.
THIS UNIT OPERATES ON 12‐14V DC ONLY
S‐SCORT III Model 74000
Baery Operated Portable Sucon Pump
United States Patent No. 5134994
©2015 HI‐D and Big Sck are registered trademarks of SSCOR, Inc.
S-SCORT III - Model 74000
2
Clearing The Airway Is Our #1 Priority
Caution-Notice Revision K (02/17)
Rev. K (02/17)

S-SCORT III - Model 74000
3
Clearing The Airway Is Our #1 Priority
Table of Contents
General Description
Cauon‐Noce Page2
General Descripon Page3
ImportantFeatures Pages4‐5
Operang Instrucons Pages6‐7
Troubleshoong Page8
Warranty Page9
Maintenance Pages9‐10
General Specificaons Page11
Replacing the BaeryPack Page12
The S‐SCORT® III is a portable, 12V DC baery operated sucon pump to be used by professional personnel
trained in Emergency Care techniques for opening the airway by removing bodily fluids and parculate
maer.
The Model 74000 is powered by sealed lead acid baeries, capable of driving the unit for 30 ‐45 minutes.
The baeries are recharged by direct connecon to the vehicle, or by a AC/DC Dual Model Baery Charger
(included with the unit). A sealed lead acid baery is a very stable and reliable baery. Many factors can
affect the life of a baery:
• Leaving a unit switched on aer there is no longer enough power to run the pump can cause a baery to
deep discharge. This can reduce the life of, or destroy the baery.
Failing to charge a baery for an extended period of me will also cause the baery to go into deep
discharge.
• Low temperatures may reduce the available capacity.
High temperatures may cause deformaon of the baery case and damage the baery.
Sealed lead acid baeries can easily be maintained to permit proper operaon of the equipment. To protect
the baery, aer each procedure turn the unit off, put the unit on charge and always store the pump at room
temperature. The only way to assure the baery has funconal capacity, even if it is indicated the baery is
fully charged, is to perform the baery check suggested on page 6 of this operaons manual. Replacement
baery may be purchased at shop.sscor.com.
The S‐SCORT III is designed to provide instant, effecve suconing, independent of external sources of power
and can be pre‐set to be acvated immediately upon reaching the distressed paent, with no set up me
required during the first few crical minutes of the code. Sucon power is controllable for those instances
when full power would be considered harmful to the paent. All controls are clearly labeled and easily
accessible.
The S‐SCORT III is equipped with a disposable collecon canister which features a bacterial filter to screen
airborne parculates and a mechanical shut‐offvalve to prevent fluid overflow.
Rev. K (02/17)

S-SCORT III - Model 74000
4
Clearing The Airway Is Our #1 Priority
Important Features
Rev. K (02/17)

Key # Component Funcon
1 Vacuum Line Connects regulator to canister
2 Vacuum Port Connects canister to vacuum line
3 Paent Port Connects canister to paent tube
4 Paent Connecng Tubing 9/32” ID, SSCOR part #43200
5 Disposable Canister 1200cc/ml, SSCOR part #48041
6 Vented Sucon Tip HI‐D® “Big Sck”® Large Bore
Sucon Tip SSCOR part #44241
7 Charging Indicator Light Signifies electrical connecon from charger to
the unit
8 On/OffSwitch Turns unit on or off
9 Charging Receptacle To connect charger to baery
10 Carrying Handle
11 Regulator Control Adjusts negave pressure
12 Baery Connecon Quick disconnect
13 Pump Provides vacuum source
14 Baery Supplies electrical power to pump
15 Baery Charger Charges baery aer each use
16 12V DC Electrical Plug To connect charger to baery
17 DC Power Cord Connects power cord to vehicle
S-SCORT III - Model 74000
5
Clearing The Airway Is Our #1 Priority
Important Features
Rev. K (02/17)

For opmal performance the sucon device is to remain plugged in whenever the device is not in use. If for
any reason poor baery quality is suspected, a baery test should be conducted.
Baery Test
Run the following test whenever poor baery quality is suspected to ensure proper performance of the
device.
1. Confirm the power cord is supplying power to the device. Check the power indicator light on the control
panel.
2. Remove the power cord and run the unit from its internal DC baery.
3. Check for vacuum by occluding the paent tube and set the vacuum regulator to the maximum vacuum
seng. Un‐occlude the paent tubing.
4. Allow the unit to run for 15 minutes on DC power. If the unit stops or slows during the 15 minutes, it is
possible the baery capacity has been depleted. It is me to replace the baery.
5. If the unit is sll running at full power aer 15 minutes, adjust the regulator to the desired seng, turn
the device offand put it back on charge.
6. Reconnect the unit to the charging source as soon as possible aer each use
SSCOR recommends replacing the baery aer 3 years. Replacement baeries may be purchased at
shop.sscor.com.
BATTERY CHARGING SUGGESTIONS
It is important to keep the baery connected to the charging source at all mes when the unit is not in use.
In order for the S‐SCORT® III baery to operate at its maximum efficiency, it must be charged from a power
source that will bring the voltage level of the baery up to the 13V DC operang range. This S‐SCORT III
model is equipped to charge the baery two ways, with a AC/DC baery charger (15) or a DC power cord for
direct connecon to the vehicle (17).
An acve vehicle, running calls around the clock, will do well to charge the S‐SCORT III from the vehicle’s DC
charging system. The preferred method of charging the S‐SCORT III is to wire the DC power cord to the
vehicle DC power system on a properly fused line in front of the master switch. This charging method is
designed to keep the baery charging at all mes. If the sucon unit is operated while it is hooked up to the
vehicle it will ulize the vehicle power and save its own baery for emergency use. All SSCOR, Inc. sucon
units have a diode to prevent drawdown from the pump to the vehicle electrical system and a fuse to protect
the pump from vehicle electrical surges. If your sucon unit is wired to the vehicle baery via an automac
load switch power supply, be sure to use a filter in order to eliminate any voltage spikes.
A vehicle running infrequent calls does not charge itself oen enough to keep baeries in the 13V DC
operang range and should rely on the S‐SCORT III AC/DC baery charger when the vehicle is connected to
the shore line. In this instance the DC power cord should be used as a backup source of power to run the
S‐SCORT III from the vehicle during paent transport. To ulize the 115V AC baery charger, the sucon
pump must be removed from the vehicle and charged from an AC outlet or the charger must be plugged into
a “shore‐line” outlet in the vehicle. Whichever charging method is chosen, remember to keep the unit con‐
nected to the charging source whenever the unit is not in use.
For replacement parts and accessories, please visit shop.sscor.com
S-SCORT III - Model 74000
6
Clearing The Airway Is Our #1 Priority
Operating Instructions
Rev. K (02/17)

NOTICE: Component key numbers are indicated within ( )s. Reference pages 4 and 5.
Electrical operaon: Constant charging is required.
CHARGING THE S‐SCORT III FROM A VEHICLE
Connect the DC power cord to the cigaree lighter receptacle in the vehicle or hard wire the DC power cord
(17) to the hot DC electrical system of the vehicle on a properly fused line in front of the master switch.
Connect the electrical line cord to the S‐SCORT® III by securely aaching the charging plug (16) into the
receptacle (9). The S‐SCORT III baery is charged by the vehicle electrical system and the pump is powered
by the vehicle current and the S‐SCORT III baery running in parallel.
DC CHARGING FROM AC/DC DUAL MODE BATTERY CHARGER
If 115V AC charging is required, connect the S‐SCORT III to a SSCOR AC baery charger (15) SSCOR Part
#80533 (included with unit). AC charging should bring the baery to a dependable working charge in 4 to 6
hours. The green LED on the charger will light when the baery is fully charged.
The charging indicator light (7) on the front panel indicates the S‐SCORT III is receiving electrical input from
the AC charger or vehicle. It does not reflect the condion of the baery. Check the charging Indicator light
to be sure you have a good electrical connecon.
Operaon of Unit
Negave pressure on the Model 74000 is controlled by a two posion regulator (11). When fully depressed
(pushed toward the pump), the negave pressure exceeds ‐525mmHg. To reduce negave pressure, pull the
regulator straight out to the stop. In this posion, the negave pressure will be ‐120mmHg (±15%).
Aach the paent connecng tube (4) to the paent port (3) on the canister (5). Use 9/32" I.D. tubing. Turn
the On/Offswitch (8) to the On posion and occlude the end of the paent connecng tube. Keep it
occluded for 10 seconds. Release the occlusion and observe evidence of negave pressure. If the pump is
running and no negave pressure is observed, check to be sure the lid on the disposable canister is ght and
vacuum connecons are secure. Dispose of the canister aer use according to local / regional / naonal re‐
quirements for the disposal of hazardous waste materials.
Reconnect the S‐SCORT III to the vehicle or to the baery charger (15) as soon as possible following the code
by securely aaching charging plug (16) into the receptacle on the unit (9).
CAUTION ‐Always turn the pump switch “Off” as soon as possible aer the procedure.
Rechargeable baeries will short out if they connue to discharge aer they have reached the point where
they do not have sufficient power to drive the pump. Be sure the switch is in the “Off” posion unl needed.
S-SCORT III - Model 74000
7
Clearing The Airway Is Our #1 Priority
Operating Instructions
Rev. K (02/17)

PROBLEM POSSIBLE CAUSE CORRECTIVE ACTION
Unit does not funcon
when switch is in the
“ON” posion
Molex connecons disconnected
Baery discharged
Baery damaged
Charger damaged
Reconnect Molex connecons
Recharge baery
Replace baery
Check output of charger
Unite does not sucon
when pump is running
Vacuum line loose
Canister damaged
Canister lid is not ght
Thumb vent on sucon p is not
occluded
Check connecons
Replace canister
Re‐secure canister lid
Occlude thumb vent with thumb
Pump is sluggish Residual materials have collected
in the pump head
Baery is unable to retain a
charge
Loose connecons
Replace pump
Replace baery
Check connecons
System shuts down while
suconing heavy
parculate maer
Vacuum line clogged at canister
lid
Float valve has closed
Remove connector or canister lid and
loosen obstrucon
Loosen float valve, empty contents,
or replace canister
Regulator stem is difficult
to pull
Stem requires lubricaon Lubricate stem with pneumacs
lubricant
S-SCORT III - Model 74000
8
Clearing The Airway Is Our #1 Priority
Troubleshooting
Rev. K (02/17)

SSCOR warrants that each new product is free from defects in material and workmanship under normal use
and service for a period of one year from date of purchase. If returned to SSCOR, we will arrange for repairs
or replacement within the terms of the warranty. The product should be decontaminated and returned
properly packaged and postage prepaid. Loss or damage in transit to the factory shall be at the purchaser’s
risk. Please call 800‐434‐5211 or internaonal +1 818‐504‐4054 for return authorizaon or for the locaon of
an authorized repair center. Loss or damage in return shipment from SSCOR shall be at the purchaser’s risk.
The warranty shall not apply to any SSCOR product which has been repaired by anyone other than an
authorized SSCOR representave, or altered in any way so as, in SSCOR’s judgment, to affect its safety or
efficacy, nor which has been subject to misuse, negligence, or accident, nor which has had the serial number
altered, effaced or removed. Neither shall this warranty apply to any SSCOR product which has been
connected otherwise than in accordance with the instrucons furnished by SSCOR.
This warranty is in lieu of all other warranes expressed or implied and of all other obligaons or liabilies on
SSCOR’s part, and SSCOR neither assumes, nor authorizes any representave or other persons to assume for
it, any other liability in connecon with the sale of SSCOR products.
This warranty gives you specific legal rights and you may also have other rights that vary from jurisdicon to
jurisdicon. For countries where minimum warranty terms are determined by statute, the warranty term is
the longer of the statutory period or the term listed above.
Baeries, disposable items including collecon canisters, paent tubing and catheters are excluded from this
warranty.
Prevenve Care: Observe the following maintenance roune to ensure readiness at any me:
1. When the SSCOR aspirator is not in use, keep baeries on connuous charge.
2. Test the SSCOR aspirator at regular intervals; See page 6.
3. Make sure the SSCOR aspirator is always clean and ready for use.
4. If the procedure produced an excessive quanty of fluids, check the vacuum line (1) for evidence of
moisture. If the vacuum line between the pump and canister is moist, it is possible that fluids have
reached the vacuum pump. See Disinfecon Instrucons (page 10).
5. If the vacuum pump appears defecve, return the unit to the factory for repair. Do not aempt to repair
the vacuum pump.
6. For technical assistance, call (800) 434‐5211 or internaonal +1 (818) 504‐4054. For replacement parts
and accessories, please visit shop.sscor.com.
S-SCORT III - Model 74000
9
Clearing The Airway Is Our #1 Priority
Warranty
Maintenance
Rev. K (02/17)

As soon as possible aer use, the single use disposable canister, paent tubing and catheter should be
discarded according to local / regional / naonal requirements for the disposal of hazardous waste
materials. Clean the exterior of the SSCOR sucon unit using a mild detergent and clear water by
dampening a clean lint free cloth. Rinse using clear water and another damp clean lint free cloth to remove
any detergent residue.
NOTE: The hydrophobic filter in the canister helps to ensure that no moisture or parculate maer reach‐
es the inside of the device. When fluids fill the canister, the posive (mechanical float) shutoffvalve closes
immediately, shung the vacuum port offso as to prevent fluid from contacng the pump. The filter has
been tested by the manufacturer (Bemis) to screen out aerosolized microorganisms and parculate maer
at a bacterial efficiency rang of 99.99% DOP. The canister also has sidewall gradaon marks starng at
100 ml/cc and at every 50 ml/cc up to 1200 ml/cc indicang the fill level of the canister.
In the unlikely event that fluids may have reached the vacuum pump, read the disinfecon secon. Your
engineering department will have to open the unit to check the condion of the pump.
Do not reuse any single use disposable parts; do not submerge the device into any liquid, this will void the
warranty and cause the device to malfuncon.
Part Cleaning and Disinfecng
Collecon Canister Disposable item, re‐use not permied. Use new canister for each paent.
Paent Tubing
Disposable item, re‐use not permied. Use new paent tubing for each
paent.
HI‐D® Sck
Disposable item, re‐use not permied. Use new HI‐D Sck for each pa‐
ent.
Vacuum Pump
Wipe with damp cloth or disinfectant wipe. Sterilizaon not permied.
Vacuum pump should be replaced if contaminated
Chassis Wipe with damp cloth or disinfectant wipe. Sterilizaon not permied.
Cauon: Disconnect the unit from any power source prior to cleaning the unit.
Disinfect the unit using a mild surface disinfectant, such as a 10:1 mixture of water and bleach. The unit is
designed to sucon contaminated fluids, which should be removed from the system immediately aer use.
In the unlikely event that fluids may have reached the vacuum pump, your engineering department will
have to open the unit to check the condion of the pump. When cleaning the interior of the chassis,
disconnect the baery from the wiring harness. The only foreseeable way fluids may reach the vacuum
pump is the filter in the canister has been compromised or bypassed.
For technical assistance, call (800) 434‐5211 or +1 (818) 504‐4054. For replacement parts and accessories,
please visit shop.sscor.com.
S-SCORT III - Model 74000
10
Clearing The Airway Is Our #1 Priority
Maintenance
Disinfection
Use personal protecve equipment such as gloves, a smock, and face and eye protecon when handling
units that are suspected to be contaminated.
Rev. K (02/17)

CHARACTERISTICS SPECIFICATIONS
Size 11"L x 8"H x 5.25"W
Weight 7 pounds
Vacuum Pump 12V DC., 3.0 A
Clinical Airflow ≥30LPM
Exceeds 525mmHg
Regulator Controls negave pressure
Power Source:
Baery (DC Power)
DC Charging from Vehicle
Rechargeable Sealed Lead Acid. SSCOR part # 80638. Please read
pages 6 and 7 for baery care suggesons. At full capacity a fully
charged baery will power the sucon device for 30‐45 minutes.
Vehicle electrical system powers the pump and charges the baery
unl disconnected and then the unit switches automacally to its self
contained baery.
AC Charger U.L. Listed AC/DC Dual Mode Baery Charger
Switch On/OffRocker
Collecon Canister 1200cc/ml SSCOR part #48041
Paent tubing Vinyl tubing 9/32”ID x 72”L SSCOR part #43200
Sucon Tip HI‐D® “Big Sck®” Large Bore Sucon Tip with thumb control.
SSCOR part #44241C
S-SCORT III - Model 74000
11
Clearing The Airway Is Our #1 Priority
General Specifications
Rev. K (02/17)

Refer to parts diagram pages 4 and 5.
Instrucons to replace the baery
• Unzip the top flap and pull it back exposing the foam chassis.
• Remove the top poron of the foam chassis exposing the components.
• Lithe baery out of the chassis and disconnect the baery’s Molex fing from the wiring harness.
Dispose of the old baery according to local / regional / naonal requirements for the disposal of sealed
lead acid baeries.
Instrucons for disassembly to wash the fabric case
• Remove the four screws around the outside switch plate and remove the plate.
• Unzip the top flap and pull it back exposing the foam chassis.
• Remove the top poron of the foam chassis exposing the components.
• Remove the screw stop that aaches to the white regulator stem and remove the regulator stem.
• Use a flat head screwdriver to remove the “E” clip from the outside of the regulator.
• Pull the vacuum tubing back to the pump head, under the flap.
• All of the components should liout of the chassis and remove the foam chassis from the fabric case.
To reassemble the device
• Put the foam chassis back into the fabric case.
• Reaach the switch plate with the four screws.
• Put the pump and baery back into the foam chassis.
• Replace the white regulator stem into the regulator. Rotate the stem and place the set screw through the
regulator slot & into the set screw hole on the white regulator stem. Do not screw all of the way down
(approx. 1/16" from regulator stem). Check the regulator stem moves freely.
• Pull the stem all of the way out & place the “E” clip onto the end of the brass regulator (on the outside of
the case) using channel locks to snap the “E” clip into place.
• Reaach the vacuum tubing to the vacuum port of the canister. Close the zipper on the top of the case.
Turn the unit on and verify there is a vacuum
S-SCORT III - Model 74000
12
Clearing The Airway Is Our #1 Priority
Rev. K (02/17)
Replacing the Battery (Available for purchase on shop.sscor.com)
Table of contents
Other SSCOR Medical Equipment manuals

SSCOR
SSCOR DCell Suction Instructions for use

SSCOR
SSCOR 2109 Instructions for use

SSCOR
SSCOR VX-2 2310 Series Instructions for use

SSCOR
SSCOR Quickdraw 2400 Series Instructions for use

SSCOR
SSCOR S-SCORT newDuet 2314 Instructions for use

SSCOR
SSCOR 2315 Instructions for use

SSCOR
SSCOR S-SCORT new-Sentinel 2315 Instructions for use

SSCOR
SSCOR EVX 2309 Instructions for use

SSCOR
SSCOR S-SCORT VX2 2310 Instructions for use