Suntex PC-3110-RS User manual

1
PC-3110-RS
Intelligent
pH/ORP
Transmitter
Operation
Manual

Precautions for installation
Brief Instruction 1
1. Specifications 4
2. Assembly and installation
2.1 Transmitter installation 5
2.2 Illustration of panel mounting 5
2.3 Illustration of wall mounting and pipe mounting 6
2.4 Assembly of electrode and housing 6
2.4.1 Cable set-up 7
2.4.2 Assembly of immersive electrode holder and junction box 7
2.5 Illustration and description of junction box 9
3. Overview of Intelligent pH/ORP transmitter PC-3110-RS
3.1 Illustration of rear panel 10
3.2 Illustration of terminal function 10
3.3 Description of terminal function 11
3.4 Installation of accessorial transmitter PH-300T(Optional) 12
3.5 Connection of transmitter and accessional transmitter PH-300T 13
3.6 Typical wirings 14
3.7 Illustration of electrical connection 14
4. Configuration
4.1 Illustration of front panel 15
4.2 Keypad 15
4.3 LED indicators 15
5. Operation
5.1 Measurement mode 16
5.2 Set-up menu 16
5.3 Calibration menu 16
5.4 Shortcuts 16
5.5 Default value
5.5.1 Setting default value 16
5.5.2 Calibration default value 16
6. Measurement display mode
6.1 Text mode 17
6.2 Real-time chart mode 18
6.3 Trace mode 19
6.4 Warning symbols and text 20
CONTENTS

7. Settings
Block diagram of settings 21
7.1 Entry of set-up menu 23
7.2 Security code of settings(Code) 24
7.3 Language 25
7.4 Mode 26
7.5 Multi-point calibration(Multi-Cal) 28
7.6 Product adjustment 29
7.7 Temperature 30
7.8 Relay 1 31
7.9 Relay 2 32
7.10 Clean 33
7.11 Analog output (pH/ORP) 34
7.12 Date/Time (Clock) 35
7.13 RS-485 communication 36
7.14 Sample average of measurements (Digital filter) 37
7.15 Backlight settings 38
7.16 Contrast settings 39
7.17 Logbook 40
7.18 Automatically back to measurement mode(Return) 41
8. Calibration
Block diagram of Calibration 42
8.1 Entry of calibration menu 43
8.2 Security password of calibration 44
8.3 pH calibration 45
8.3.1 TECH mode 45
8.3.2 NIST mode 45
8.3.3 Any mode 45
8.3.4 Define mode 46
8.3.5 Definition of calibration parameter 46
8.3.6 TECH, NIST buffer Calibration 47
8.3.7 Any Calibration 48
8.3.8 Define Calibration 49
8.4 ORP calibration 51
8.5 Automatically back to measurement mode(Return) 52
9. Modbus protocol and instructions 53
10. Error messages (Error code) 58
11. Maintenance 59
Appendix 60

Thank you for purchasing Suntex products. In order to continually improve and enhance the transmitter’s
function, Suntex reserves the right to modify the content and icon display of the product. The actual situation
is subject to the instrument without notice. The operation manual is only provided for function and
installation description, Suntex Instruments Co., Ltd. is not liable for any person or entity for any direct or
indirect loss or damage due to improper usage of this product. If you have any questions or find omission,
negligence or mistakes of the operation manual, please contact with our staff, thank you.
Precautionsforinstallation
Wrong wiring will lead to breakdown or electrical shock of the instrument, please read this
operation manual clearly before installation.
Make sure to remove AC power from the transmitter before wiring input, output connections,
and remove it before opening the transmitter’s housing.
The installation site of the transmitter should be good in ventilation and avoid direct
sunshine.
The material of signal cable should be special coaxial cable. Strongly recommend using our
coaxial cable. Do not use normal wires instead.
Avoid electrical surge when using power. Especially when using three-phase power, use
ground wire correctly. If the power surges interference occurs, separate the power supply of
transmitter from the control device, such as: dosing machines, mixers, etc. to make individual
power supply for the transmitter; or set surge absorber to reduce the power surges at all
electromagnetic switches and power control device coils.
The internal relay contact of the instruments is for alarm or control function. Due to safety,
please must connect to external relays which can stand enough ampere to make sure the
safety operation of the instrument.(Please refer to chapter 3.7“Illustration of electrical
connection”)

Brief Instruction
Description of set-up settings (see chapter 7 for details)
Press and simultaneously to see the overview of the set-up settings now. Then press I
if you would like to modify set-up settings. Press keypad according to index of keypad on the screen.
Index of keypad
keypad Accordingly item Description
Back to upper layer
Choose leftward of change to left page
Increase digit
Choose rightward of change to right page
Decrease digit
Confirm settings after modifications and then go through next
step
Selection of set-up items
keypad Accordingly item Description
Mode Measurement mode, to choose pH or ORP measurement
Multi-Cal. Multi-point calibration, to choose 1~5 points calibration
Product Adj.
Sample readings adjust settings--Users may make correction
without removing the electrode out of installation site. Adjust
the on-site measurements until the reading is as same as the
reading from the lab measurement to eliminate the doubt of
measurement error.
Temperature
Temperature measurement and compensation, including MTC,
PTC, NTC (3 types total). MTC---Manual temperature
compensation, PTC/NTC--- auto temperature compensation
Relay 1 First relay setting, to choose action off or Hi/Lo alarm
1

Relay 2 Second relay setting, to choose action off or Hi/Lo alarm
Clean Automatic wash time setting, to choose electrode clean
equipment’s ON and OFF duration
Analog 1 Current output according to pH or ORP setting range
Clock
Clock setting (When out of power and reboot it, the
instrument’s time setting can maintain to the real time. If
not, please replace the inner 3V CR2025 battery.)
RS-485 RS485 serial interface (Modbus protocol)
Digital Filter Take every serial 1~60 measurements, average them
continuously, and make it as the readings
Black-light Backlight setting, to set Auto/ON/OFF backlight, brightness,
and sensitivity
Contrast Contrast of screen setting
Logbook Event recorder logbook (50 data)
Return Setting of returning to the measurement mode
Code
Security code of set-up mode. The set-up code is precedential to
calibration code, thus it can pass a different security code of
calibration.
Language Available for English, Traditional Chinese, Simplified Chinese
2

Description of calibration settings (see chapter 8 for details)
Press and simultaneously to see the last calibration information. Then press if you
would like to make a new calibration or modify setting of calibration. Press keypad according to
index of keypad on the screen.
Index of keypad:
keypad Accordingly item Description
Back to upper layer
Choose leftward of change to left page
Increase digit
Choose rightward of change to right page
Decrease digit
Confirm settings after modifications and then go through next
step
Selection of calibration items (up to five-point calibration)
keypad Accordingly item Description
Code Security code of calibration mode.
Return Time interval setting of returning to the measurement mode
TECH Use tech buffer as standard solution for calibration
NIST Use NIST standard buffers(DIN 19266) as standard solution for
calibration
Any Use any buffer solution by users’definition for calibration
Define
There are five default standard buffer sets, The user can change
the default, create and save the appropriate standard buffer pH /
temperature relation curve to do calibration,.(memory up to
five)
Note
Due to the need for continuous improvement of the transmitter function, we reserve the right to modify the
content and the icon of the function. The actual icons and contents are subject to the instrument without notice.
3

1.Specifications
Model PC-3110-RS
Measuring modes pH / ORP / Temp.
Ranges
pH -2.00~16.00 pH
ORP -1999~1999 mV
Temp. -30.0~130.0 C
Resolutions
pH 0.01 pH
ORP 1 mV
Temp. 0.1 C
Accuracy
pH ±0.01 pH ±1 Digit
ORP ±0.1% ±1 Digit
Temp. ±0.2C (±1 Digit),
equipped with temperature error modification function
Temperature
Compensation
NTC30K/ PT 1000 auto temperature compensation
Manual adjustment temperature compensation
Calibration mode TECH, NIST, Any Buffers, up to five point calibration
Ambient Temp. 0~50C
Storage Temp. -20~70C
Input Impedance > 1012 Ω
Display
Large LCM with sensitization sensor for auto/manual illumination
function and contract function
Text mode: Numerical display
Chart mode: 3 mins real-time dynamic graph
Trace mode: Set up from 3 mins to four weeks duration of the
measured value trend graph
Analog output Isolated DC 0/4~20mA corresponding to main measurement,
max. load 500
Serial interface RS-485 (MODBUS RTU or ASCII)
Logbook 50 event records
Settings
Contact RELAY contact,240VAC 0.5A Max.(recommended)
Activate Hi/Lo. Hi/Hi. Lo/Lo selectable two limited programmable,
ON/OFF
Wash RELAY contact: ON 0~99min. 59sec. / OFF 0~999hr 59min.
Voltage Output DC
±
12V,1W max.
Power Supply 100V~240VAC±10%,9W max.,50/60Hz
Installation Wall or Pipe or Panel Mounting
Dimensions 96m × 96mm × 132mm (H
WD)
Cut off Dimensions 93 mm
93 mm (H
W)
Weight 0.5Kg
Protection IP65(NEMA 4X)
Note: The specifications are subject to change without notice.
4

2.Assemblyandinstallation
2.1 Transmitter installation: This Transmitter can be installed through panel mounting, wall
mounting and 2” pipe mounting.
Installation of panel mounting: First,prepare a square hole of 93 x 93mm on the panel box,
and then insert the controller directly into the panel box. Insert the accessorial mounting bracket
from the rear, and make it be fixed into pickup groove.
2.2 Illustration of panel mounting
93 mm
93 mm
Hole size
87 mm
+
87 mm
Hole distances on the
p
anel box
Mounting bracket
Illustration of panel mounting,
fixed with mounting bracket
+
++
112 mm
5

2.3 Illustration of wall mounting and pipe mounting
Penetrate the two prepared holes in the rear
cover and fix the U-shaped pipe clip, and
then install two waterproof squeezed caps
(the transmitter’s standard accessory kit) into
the holes from inner rear cover to prevent
from the water vapor.
Installation of pipe mounting
Fixed with U-shaped pipe clamp.
(Optional, Order Number: 5333027)
Installation of wall mounting
Fixed with 4 x M4 screws
Sun Shield (Pipe mounting, Optional)
( Order No.: 8-35 + 8-35-3 + 8-35-1) Sun Shield (Wall mounting, Optional)
(Order No.: 8-35 + 8-35-3 + 8-35-2)
Insert the single hole rubber plug into
the unused cable gland, and tighten up
the cable gland to prevent from the
penetration of water vapor.
6

2.4 Assembly of electrode and housing
2.4.1 Cable set-up
a. Make sure to remove the conductive rubber or aluminum-foil layer between the electrode signal
wire and the coaxial shield.
b. Extend the cable to the transmitter without any joint except specific junction box. Connect the
transparent coaxial inner directly to the Glass terminal on the back of transmitter, and metal
connect coaxial shield to Ref.terminal.
2.4.2 Assembly of immersive electrode holder and junction box 8-09-5+ PP-100A(Optional)
1. Insert the electrode(H) through PP Electrode Protective Housing(G)
2. Rinse the electrode(H) properly, so that it can be easily pass through Rubber electrode
holder(I),leave about 5cm bellow.
3. Install the prepared Rubber electrode holder (I) into PP Electrode Protective Holder(G) and
fix with PP pipe protective cover(J) tightly.
A------ Upper cover of round junction box
B------ O-ring
C------ Cable fixing gland MG16A
D------ Lower cover of round junction box
E------ Cable fixing point MG16A
F------ O-ring
G------ PP Electrode Protective Housing
H------ Electrode (Sensor)
I------- Rubber electrode holder
J------- PP pipe protective cover
Set-up diagram of coaxial cable:
See the correct set-up method on the left:
Note: The black conductive rubber
covering on the coaxial inner should be
removed.
Metal coaxial shield
Transparent coaxial inner
(Remove the conductive rubber)
7

4. Insert the Electrode cable(H) through Lower cover of round junction box(D) and Cable fixing
gland(C), and use Lower cover of round junction box(D) to fix PP Electrode Protective
Housing(G) tightly.
5. Prepare 15cm cable in the PP pipe, and then fix Cable fixing gland MG16A(C) tightly. Leave
Electrode cable(H) for about 12-14cm, and split it carefully.
6. Fix the terminal of electrode coaxial inner on terminal block 1 of round holder; Fix the
terminal of electrode coaxial shield on terminal block 3. (See the instruction of junction box)
7. Extend the cable to pass through Cable fixing gland(E) on Lower cover of round junction
box(D), and fix Cable fixing gland MG16A(E) tightly, leaving 12-14cm in the box for split.
8. Extend the lead coaxial inner and electrode coaxial inner to connect them; extend the lead
coaxial shield to fix on the terminal block 3. Tight up Upper cover of round junction box(A)
to finish the installation.
Installation of holder support base
The L-shaped electrode holder support base is installed by finding a appropriate position in
the edge of a pool according to the field’s needs with nails or expansion screws.
1. Fix the holder fixed plate(C) into the support base (D)
2. Fix the U-shaped clamp (A) into immersive holder (B)
3. Combine the item 1. and item 2., tighten it up by the plastic screws (E)
A------Holder U-shaped clamp
B------Immersive holder(PP-100A)
C------Holder fixed plate
D------Support base
E------Plastic screws
8

2.5 Illustration and description of junction box
(Two-wire distributing system
and Three-wire distributing system)
(1) Two-wire distributing system
INPUT terminals Terminal
No. OUTPUT terminals Terminals on
Transmitter
Coaxial inner 1 Coaxial inner’s extension wire for
electrode GLASS
Shield (forbidden) 2 Shield (forbidden) ------
Coaxial shield 3 Coaxial shield’s extension wire for
electrode REF
Temperature probe
red wire 4 Red wire’s extension wire for
electrode T/P
Temperature probe
green wire 5 Green wire’s extension wire for
electrode SG
Alternative 6, 7 Alternative ------
Note: 1. Extension cable, Order number: 7202-F94009-BK or 7202-RG-58
1.) If temperature probe is not used, the Order No. is 7202-RG-58.
2.) If temperature probe is used, the Order No. is 7202-F94009-BK.
2. If temperatures probe 8-26-3(NTC30K) or 8-26-8(PT1000) is used for two-wire distribution, the
black wire terminal should be forbidden.
(2) Three-wire distributing system
IN terminals Terminal
No. OUT terminals Terminals on
Transmitter
Coaxial inner 1 Coaxial inner’s extension wire for
electrode GLASS
Solution ground wire
(Shield) 2 Solution ground wire SG
Coaxial Shield 3 Coaxial Shield’s extension wire for
electrode REF
Temperature probe
red wire 4 Red wire’s extension wire for
electrode T/P
Temperature probe
green wire 5 Green wire’s extension wire for
electrode SG
Alternative 6, 7 Alternative ----
Note: 1. The black wire on the temperature probes of 8-26-3(NTC30K) or 8-26-8(PT-1000) is used as special
wire as solution ground rod which is to be connected at terminal 2.
2. The extension cable, Order Number: 7202-F94009-BK, is for system that apply a temperature probe or
solution ground rod.
9

3. Overview of pH transmitter PC-3110-RS
3.1 Illustration of rear panel
3.2 Illustration of terminal function
POWER
WASH/Cln
±12V
+
-
REL2REL1
pH/
ORP
GLASS
Ref.
AC
INPUT
POWER
4/20
m
A
+
-
Temp
SG
SG
T/P
RS-485
D+(B)
G
D-(A)
10

3.3 Description of terminal function
01 REL1: First alarm control, the contact for an
externalrelay
02
03 REL2: Second alarm control, the contact for an
external relay
04
05 WASH: Wash relay contact for an external relay
06
07 NC: None contact
08 Power(N): 100~240VAC Power supply terminal
09 NC: None contact
10 Power(L): 100~240VAC Power supply terminal
11 GLASS: Coaxial inner of pH/ORP electrode
signal wire
12
13 NC: None contact
14 Ref. : Coaxial shield of pH/ORP electrode
si
g
nal wire
15 SG: Solution ground wire. In two-wire distributing system, there should
be a jumper between this terminal and REF (a short circuit slice
is attached when going out the factory)
16 T/P: Connect with one of cable end of temperature probe
17 SG: The other cable end of temperature probe, or used as ±12V ground
potential.
18 DC±12V: Output terminal of direct current voltage ±12V (PH-300T only)
19
20 4~20mA +terminal: Master measure current output terminal +, for external
recorder or PLC control
21 4~20mA – terminal: Master measure current output terminal -, for external
recorder or PLC control
22 4~20mA + terminal/ D+(B) : RS-485 output D+(B)
23 4~20mA – terminal/ G: RS-485 output GND
24 NC / D- (A) : RS-485 output D-(A)
11

3.4 Installation of accessorial transmitter PH-300T (Optional)
Accessorial pH/ORP transmitter, PH-300T, is mainly installed on the electrode protective pipe,
but also can apply wall mounting and pipe mounting. For long distance transmission (100m), if
PC-3110 is more than 30m far away from the electrode, PH-300T accessorial transmitter is
recommended to avoid the attenuation of electrode signal, and for the convenience of onsite
observation, measurement, and calibration.
Illustration of wall mounting type
1. Combine the sun shield and PH-300T transmitter by round stainless steel screws 4x15
2. Fix item 1 combination on wall by self-tapping screws
Illustration of pipe mounting type
1. Fix immersive holder(B) into U-shaped clamp(A)
2. Combine the item 1. combination with PH-300T(C) by Round stainless steel screws 4×15
A------Sun shield of PH-300T
B------PH-300T transmitter
C------Round stainless steel screws 4x15
D------Self-tapping screws(by customers)
A------U-shaped clamp
B------Immersive holder
C------PH-300T transmitter
D------ Round stainless steel screws 4×15
12

3.5 Connection of transmitter PC-3110 and accessorial transmitter PH-300T
A. Connect the GLASS point of transmitter PH-300T’s terminal to the electrode coaxial inner.
(Note: Remove the black conductive rubber); connect the REF point of transmitter
PH-300T’s terminal to the electrode coaxial shield.
B. See the two-wire distributing system and three-wire distributing system in the following
page.
C. Sign “PT-1000” on transmitter PH-300T’s terminal is the connector for automatic
temperature compensation probe, PT-1000, or applies a fixed temperature compensation
resistance.
D. The V+ and V- of transmitter PH-300T’s terminal respectively connect to DC12V+ and –
of the controller.
E. The S+ and S- on transmitter PH-300T’s terminal respectively connect to GLASS and REF
of the controller.
F. The I+ and I- on transmitter PH-300T’s terminal are output (4-20mA), which can connect
to devices that receive current signals. (Note: The current output signal of this
transmitter is not insulating, and thus do not directly connect with a PLC!)
Note: Refer to the following table for proper fixed temperature compensation resistance
Temperature 0℃ 5℃ 10℃ 15℃ 20℃
R value 1000Ω1019.25Ω1038.5Ω1057.75Ω1077Ω
Temperature 25℃ 30℃ 35℃ 40℃ 45℃
R value 1096.25Ω1115.5Ω1134.75Ω1154Ω1173.25Ω
Temperature 50℃ 55℃ 60℃ 65℃ 70℃
R value 1192.5Ω1211.75Ω1231Ω1250.25Ω1269.5Ω
Temperature 75℃ 80℃ 85℃ 90℃ 100℃
R value 1288.75Ω1308Ω1327.25Ω1346.5Ω1385Ω
13

3.6 Typical wirings
3.7 Illustration of electrical connection
Three-wire distribution
GND REF
.
GLAS
pH/
ORP
Coaxial
inner
Coaxial
shield
Blac
k
Two-wire distribution
GND REF. GLAS
pH/
ORP
Coaxial
inner
Coaxial shiel
d
Short circuit slice
Surge absorber
Surge absorber
Surge absorber
Surge absorber
Transmitter
Cleaning device
Dose feeder
Dose feeder
100 ~ 240VAC
Surge absorber
Surge absorber
External relay
External relay
External relay
Note: The transmitter built-in miniature relays is necessary to be repaired and replaced by professional technicians.
It is recommended to use an external relay (Power Relay) to activate the external equipments.
14

4. Configuration:
4.1 Illustration of front panel
4.2 Keypad:
In order to prevent inappropriate operation by others, before the parameter setting and
calibration, the operation applies multi-keys, and coding protection if necessary. Description of
the key functions is in the following:
:In the parameter set-up mode, pressing this key allows you exit parameter set-up mode
and back to Measurement mode.
:In the Calibration mode, pressing this key allows you exit Calibration mode and back to
Measurement mode.
:1. In the parameter set-up mode and Calibration mode, pressing this key to select
leftward or change to another page.
2. When adjusting value, press this key to increase the value.
:1. In the parameter set-up mode and Calibration mode, pressing this key to select
rightward or change to another page.
2. When adjusting value, press this key to decrease the value.
:Key for confirmation; pressing this key is essential when modifying data value or
selecting the parameter setting items in the window.
4.3 LED indicators:
ACT: Washing device operation indicator and controlling operation indicator (Relay 1, Relay 2)
B.L. : Light sensor; in the automatic display backlit mode, the lamp will light or go out as the
change of environmental brightness.
15

5.Operation
5.1 Measurement mode
After all electrical connections are finished and tested, connect the instrument to the power
supply and turn it on. The transmitter will automatically entering measurement mode with the
factory default settings or the last settings from user.
5.2 Set-up menu
Please refer to the set-up instructions in Chapter 7. Press and simultaneously to
enter into set-up menu, and press to go press to back to measurement mode.
5.3 Calibration menu
Please refer to the calibration instructions inChapter8.Press and
simultaneously to enter into calibration menu, and press to go back to measurement mode.
5.4 Shortcuts:
1. In the measurement mode, if selecting MTC for temperature compensation mode, you may press
and to adjust MTC temperature value.
2. Under measurement mode, press continuously for 2 seconds to see the Logbook function
directly. Press key to back to measurement mode.
3. Under measurement mode, press continuously for 2 seconds to switch the display mode
from text mode, trace mode, and real-time chart display mode.
5.5 Default value:
5.4.1 Setting default value:
Measurement mode: pH
Multi-Cal: 2 points pre-setting
Temperature compensation: MTC 25°C
Relay 1: High point alarm: AUTO, SP1= 10.00 pH, Hys= 0.10 pH
Relay 2: Low point alarm: AUTO, SP2 =04.00 pH, Hys= 0.10 pH
Wash time: OFF
Analog 1 current output (pH/ORP): 4~20 mA, 0.00~14.00pH
RS-485: RTU, Even, 19200, ID: 01
Digital filter: 5
Backlight setting: Off
Code set-up: OFF
Date & Time: 2012/1/1 00:00:00
Contrast: 0
Logbook: None
Auto back: Auto, 3 minutes
5.4.2 Calibration default value:
Asy: 0 mV
Slope: -59.15 mV/pH @ 25.0˚C
Calibration type: TECH-No Cal
Calibration value: None data
Auto back: Auto, 3 minutes
Note: The factory default of calibration presetting is “No Cal”, and the calibration value is “None”. It means that the
user has not calibrated the sensor with the transmitter yet.After finishing every calibration, the display shows the
calibration type and the calibration value. If the equipments have not been calibrated yet, the measurement takes
pre-set Asy and Slope into calculation. The factory default values are subject to change without notice.
16
Table of contents
Other Suntex Transmitter manuals

Suntex
Suntex PC-3310 Installation manual

Suntex
Suntex PC-3310-RS User manual

Suntex
Suntex EC-4110-I User manual

Suntex
Suntex EC-4110 User manual

Suntex
Suntex CT-6110-POL User manual

Suntex
Suntex IT-8100 User manual

Suntex
Suntex TC-7310-RS-M User manual

Suntex
Suntex PC-3110-P User manual

Suntex
Suntex TC-7100 User manual

Suntex
Suntex TC-7310-RS User manual
Popular Transmitter manuals by other brands

Eaton
Eaton MTL83*C Modbus Software manual

Panametrics
Panametrics DigitalFlow XGS868i Service manual

KLAY-INSTRUMENTS
KLAY-INSTRUMENTS 2000 Series instruction manual

RKI Instruments
RKI Instruments 65-2643RK-05-04 Operator's manual

M-system
M-system M5MS instruction manual

Anywave
Anywave ATSC VHF 100W quick start guide