Suntex CT-6300RS User manual

1
CT-6300RS
Microprocessor
Residual Chlorine/
Ozone
Transmitter
Setup Cal. Mode Enter
Operation
Manual

1. Specifications 1
2. Precautions for installation 2
3. Assembly and installation
3.1 Controller installation 2
3.2 Illustration of panel mounting 2
3.3 Illustration of wall mounting and pipe mounting 3
4. Overview of Residual Chlorine/Ozone Transmitter CT-6300RS
4.1 Illustration of rear panel 4
4.2 Illustration of terminal function 4
4.3 Description of terminal function 5
4.4 Illustration of cable wiring 5
4.5 Illustration of electrical connection 6
5. Configuration
5.1 Illustration of front panel 7
5.2 Keypad 7
5.3 LED indicators 8
5.4 Display 8
6. Operation
6.1 Measurement mode 9
6.2 Set-up mode 9
6.3 Calibration mode 9
6.4 Reset 9
7. Settings
Block diagram of settings 10
7.1 Entry of set-up mode 12
7.2 Security code of settings 12
7.3 Measurement parameter set-up 13
7.4 Temperature set-up 14
7.5 Auto return set-up 14
7.6 Hi point Setting 15
7.7 Lo point Setting 16
7.8 Wash time 17
7.9 Analog output 1 (Residual Chlorine/Ozone) 18
7.10 Date/Time set-up 19
7.11 RS-485 set-up 20
7.12 Averaged signal value set-up 21
CONTENTS

7.13 Backlight LCD 22
8. Calibration
Block diagram of Calibration 23
8.1 Security code of calibration 24
8.1.1 Authority of Security Code 24
8.1.2 Security Code Settings 24
8.2 Calibration principle 25
8.3 Current type sensor calibration mode 25
8.4 Voltage type sensor calibration mode 26
8.5 Last calibration data 27
8.6 Attention of sensor usage 28
9. Modbus protocol and instructions for CT-6300RS 29
10. Error messages (Error code) 34
11. Maintenance 35

1.Specifications
Model CT-6300RS
Measuring modes Residual(Free) Chlorine / Ozone / Temperature
Cl20.00~100.00 ppm(mg/l), depends on sensor
O30.00~20.00 ppm(mg/l), depends on sensor
Ranges
TEMP -30.0~130.0°C
Cl20.01 ppm(mg/l)
O30.01 ppm(mg/l)
Resolutions
TEMP 0.1°C
Cl2±0.5%±1Digit
O3±0.5%±1Digit
Accuracy
TEMP ±0.2°C±1Digit
Temperature
Compensation Automatic with PT-1000 / NTC-30K or manual adjustment
Ambient Temp. 0~50°C
Storage Temp. -10~70°C
Display LCD display with sensitization sensorfor
auto/manual illumination function
Analog output Isolated DC 0/4~20mA corresponding to
residual chlorine/ ozone measurement, max. load 500Ω
Serial Interface RS-485(MODBUS RTU orASCII)
Contact 240VAC, 0.5A Max. (recommend)
Settings Activate Hi/Lo two limited programmable
Contact 240VAC, 0.5A Max. (recommend)
Wash Time ON: 0∼9999 sec. / OFF: 0.0∼999.9 hours
Voltage output DC±8V , 0.5W Max.
Power Supply 100V~240VAC±10%, 5W Max., 50/60Hz
Installation Wall or Pipe or Panel Mounting
Dimensions 144 mm ×144 mm ×115 mm (H×W×D)
Cut off Dimensions 138 mm ×138 mm (H×W)
Weight 0.82Kg
Protection IP 65 (NEMX 4X)
Note: The specifications and appearance of the instrument are subject to change without notice.
1

Wrong wiring will lead to breakdown or electrical shock of the instrument, please read this
operation manual clearly before installation.
zMake sure to remove AC power from the transmitter before wiring input, output connections,
and remove it before opening the transmitter housing.
zThe installation site of the transmitter should be good in ventilation and avoid direct
sunshine.
zThe material of signal cable should be special coaxial cable. Strongly recommend using our
coaxial cable. Do not use normal wires instead.
zAvoid electrical surge when using power. Especially when using three-phase power, use
ground wire correctly.
zThe internal relay contact of the instruments is for alarm or control function. Due to
safety, please must connect to external relay which can stand enough ampere to make
sure the safety operation of the instruments. (Please refer to chapter 4.5“Illustration of
electrical connection”)
3.Assemblyandinstallation
3.1 Transmitter installation: This transmitter can be installed through panel mounting,
wall mounting and pipe mounting.
Installation of panel mounting: First,prepare a square hole of 138 x 138mm on the panel box,
and then insert the transmitter directly into the panel box. Insert the accessorial mounting
bracket from the rear, and make it be fixed in to pickup groove.
3.2 Illustration of panel mounting:
2. Precautions for installation
138mm
138m
m
Hole size
42mm
+
+
42mm
Hole distances on the
p
anel box
Mounting bracket
Illustration of panel mounting,
fixed with Mounting bracket
+
+
2

3.3 Illustration of Wall mounting and pipe mounting
Installation of pipe mounting
Fixed with U-shaped pipe clip
Order No.: 8-34
Installation of wall mounting
Fixed with 4 x M5 screws
Sun Shield (Pipe mounting)
Order No.: 8-35 and 8-35-1 Sun Shield (Wall mounting)
Order No.: 8-35 and 8-35-2
3

4. Overview of residual chlorine/ozone transmitter CT-6300RS
4.1 Illustration of rear panel
4.2 Illustration of terminal function
+
-
4/20m
A
S+
S-
GND
TP
T
HI LO WASH
POWER
AC
INPUT POWER +
-
±8V
+
-
4/20m
A
GND
Chlorine/O³
4 wire electrode (voltage type)
Chlorine/O³
2 wire electrode (current type)
L1
L2
4

4.3 Description of terminal function
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
4.4 Illustration of cable wiring
Chlorine /O3
L2
2 wire electrode
(
current t
yp
e
)
PT1000
N
TC30
K
T/P SG L1
12
Chlorine /O3
S+
4 wire electrode
(
volta
g
e t
yp
e
)
PT1000
N
TC30
K
T/P SG S-
br
w
g
bl
+8 -8
AC100~240V: Power supply terminal
WASH: External wash relay terminal
REL2: External relay terminal Low Point control
REL1: External relay terminal High Point control
RS-485 output: GND
4~20mA:Master measurement current output terminal, for external
recorderorPLCcontrol
DC-8V:Connect to voltage type electrode (CLE2.2) terminal V-(brown wire)
RS-485 output: D-(A)
T/P: Connect to earth end of temperature probe
DC+8V/ L1:Connect to a voltage type electrode (CLE2.2)terminal V+(green wire),
or connect to a current type electrode(CLE3, or OZE3) terminal 1.
S-: Connect to black signal wire of the voltage type of electrode (CLE2.2)
L2: Connect to terminal 2 of the current type electrode (CLE3, or OZE3)
S+: Connect to white signal wire of the voltage type of electrode (CLE2.2)
g: green wire
br: brown wire
bl: black wire
w: white wire
Wiring of 4 wire electrode
(voltage type) Wiring of 2 wire electrode (current type),
leverage transmitter’s DC ±8V
14
15
RS-485 output: D+(B)
5

4.5 Illustration of electrical connection:
Surge absorber
Surge absorber
Surge absorber
Surge absorber
Transmitter
Cleaning device
Dose feeder
Dose feeder
100 ~ 240VAC
Surge absorber
Surge absorber
External relay
External relay
External relay
6

5. Configuration:
5.1 Illustration of front panel:
5.2 Keypad:
In order to prevent inappropriate operation by others, before the parameter setting and
calibration, the operation applies multi-keys, and coding protection if necessary.
Description of the key functions is in the following:
In the parameter set-up mode, pressing this key allows you exit parameter set-up
mode and back to Measurement mode.
In the Calibration mode, pressing this key allows you exit Calibration mode and
back to Measurement mode.
In the parameter set-up mode and Calibration mode, pressing this key to
increase the value or to scroll to other function.
In the parameter set-up mode and Calibration mode, pressing this key to
decrease the value or to scroll to other function.
:Key for confirmation; pressing this key is essential when modifying data value
or selecting the parameter setting items in the window.
: In the Measurement mode, pressing these two keys simultaneously
allows you enter Calibration mode.
: In the Measurement mode, pressing these two keys simultaneously
allows you enter parameter set-up mode.
+
+
Setup Cal. Mode Enter
7

In the Measurement mode, press the two keys simultaneously for
five seconds, and then press until you see a clock signal appearing on the display;
then loose all keys to restore factory default settings.
In the Measurement mode, press the two keys simultaneously for
five seconds, and then press until you see a clock signal appearing on the display;
then loose all keys to restore factory default calibrations.
5.3 LED indicators:
WASH: Washing device operation indicator lamp; when the washing device is started up,
the Alarm indicator will light.
HI : Controlling operation indicator lamp; when the high setting point is reached, the
REL1 indicator will light.
LO : Controlling operation indicator lamp; when the low setting point is started up, the
REL2 indicator will light.
B.L. :Light sensor; in the automatic display backlit mode, the lamp will light or go out
as the change of environmental brightness.
5.4 Display:
(Master Reset)Restore factory default parameter’s settings
+
+
(Calibration Reset)Restore factory default calibration’s settings
Calibration
mode
Controlling
function
locked
Set-up
mode
Measurement
mode
Relay 1
alarm
activated
Relay 2
alarm
ac
t
ivated
Clean device
activated
Output
current
Over ran
g
e
Measurement
uni
t
Recovery
indicator
mA
mA : Measurement is over-range than the setting which is corresponding to output 20mA
: Measurement is over-range than the setting which is corresponding to output 0/4mA
Auto temp.
co
m
p
ensa
t
ion
8

6.Operation
6.1 Measurement mode:
After all electrical connections are finished and tested, connect the instrument to the
power supply and turn it on. The transmitter will automatically entering measurement
mode with the factory default settings or the last settings from user.
6.2 Set-up mode:
Please refer to the set-up instructions in Chapter 7, and press to back to
measurement mode.
6.3 Calibration mode:
Please refer to the calibration instruction in chapter 8, and press to back to
measurement mode.
6.4 Reset:
Factory default value:
Measurement mode: CLE2, 20.00 ppm
Temperature compensation: MTC
Auto return: AUTO
Relay 1(High point alarm):AUTO,SP1= 10.00 ppm,db1= 0.10 ppm
Relay 2(Low point alarm):AUTO,SP2= 1.00 ppm,db2= 0.10 ppm
Wash time: OFF(ON.S=0S,OFF.H=O.OH,DB.S=10S)
Analog output4~20 mA,0.0~20.00 ppm
Date & Time: 2010/1/1, 00:00:00
RS-485: RTU,Even,19200,ID:01
Code: OFF
9

7.Settings
Block diagram of settings 1:
Parameter
Setting
Conf.
Setting)
Code
Setting
℃
Setting
(Temp.)
Relay 1
Setting Cln
(Clean)
Setting
Cl/O3
Setting
(Analog)
Default
(密碼設定)
OFF
Code ON
Code
Pass
code
change
Select
CL
(Chlorine)
OFF
(MTC
)
Input
Temp.
Temp.
Correct
(±5℃)
Imput
REL1
SP
Input
REL1
DB
REL1
Active or
Inactive
Relay1
Test
Cln
Off-time
Input
Cln
DB
Input
Cln
Test
Cln
On-time
Input
b.L.
Setting
(Backlight)
Pass code
Input
Active
Pass code
:Right forward key
:Up forward key
:ENTER key
Select
CLE2 Select
CLE3
PTC
REL1
ON REL1
OFF
Relay2
Setting
Input
REL2
SP
Input
REL2
DB
REL2
Active or
Inactive
Relay2
Test
REL2
ON REL2
OFF
Select
O3
(Ozone)
Select
O3E3
Std.b
(Auto
return)
OnAuto
Select
O3E2
ppm
or
mg/l
ppm
or
mg/l
ppm
or
mg/l
ppm
or
mg/l
Cln
Active or
Inactive
Cln
ON Cln
OFF
Range
2.00 Range
100.0
Range
20.00
Temo.
Correct
(±5℃)
NTC
Continued on next page Continued on next page
10

Cl/O3
Setting
(Analog)
SEC
Setting
(Digital
filter)
Relative
value to
20mA
0/4 and
20mA
Output Test
0/20mA or
4/20mA
Output Signal
average
:Right forward key
:Up forward key
:ENTER key
Relative
value to
0 or 4mA
b.L.
Setting
(Backlight)
Select
Aut o
mode
Select
ON
mode OFF
Bright
-ness
setting
Bright
-ness
setting
Sensi-
tivity
setting
RTC
Setting
(Time)
SERL
Set
(RS-485)
Input year Select
communica
-tion mode
Input date Select Odd
or Even
parity
Input time Select
Baud rate
Set
ID
Cln
Setting
(Clean)
Code
Setting
Connected with previous page Connected with previous page
Block diagram of settings 2:
11

7.1 Entry of set-up mode
In the measurement mode, pressing the two keys simultaneously allows
you enter the parameter set-up mode. You can back to the measurement mode at any
time by pressing the key . The original code is 1111.
7.2 Security code of settings
In the set-up mode, you can set up the code by pressing the key , and confirm by
pressing the key .
+
Enter parameter selection set-up
Press for confirmation
Press for confirmation
Press or
Press forconfirmation
When the first “0” of the digits 0000
sparkles, press upward key or rightward
key to input the code, and press ENTER
for confirmation. Then, continue
inputting the second digit, and by
analogy.
(If security code is
cancelled, you can enter
the set-up mode directly
without inputting the
password any more. You
can reset the security code
by pressing “Mode” key
by double pressing.)
Modify the code =>
If you input the wrong
code, the CODE will
sparkle 4 times, and
then the display will
back for another input.
12

7.3 Measurement parameter set-up
Please select” CL” for residual chlorine measurement. Then, please select “CLE2” for
4-wire (voltage type) electrode; select “CLE3” for 2-wire(current type) electrode. Then,
according to sensor model, select measuring limit “2.00ppm, “20.00ppm”, or “100.0ppm.”
Please select” O3” for ozone measurement. Then, please select “O3E2” for 4-wire
(voltage type) electrode; or select “O3E3” for 2-wire(current type) electrode.
to select O
3
o
r
CL
Enter “temperature” set-up
When select CLE3
When select O3E3, O3E2, CLE2
Press for confirmation
Press for confirmation
Press or
Press for confirmation Press for confirmation Press for confirmation
to select mg/l or ppm
Press or
to select 2.00, 20.00, or 100.0ppm
Press or
13

7.4 Temperature parameter
7.5 Auto return set-up
Enter setup of auto return mode (Return) to set the function that the instrument
automatically exit the setup menu after a period of time without pressing any key.
Enter “Auto return” set-up
Enter “High point alarm” set-up
If select auto, the display
automatically exit the setting or
calibration mode and back to
measurement mode after a period
of time without pressing any key.
Same as left procedure
Use standard thermometer to test
the actual temperature of the
solution, and press or
to input the correct temperature
value.
If necessary, compare with the
actual temperature value tested by
standard thermometer. Press
or to input the modified
value.
Press for confirmation
press
or
for selection
press
or
for
selection
Press for confirmation Press for confirmation
to select
Press or
If select on, it needs to exit
setting or calibration mode
manually Press for confirmation
14

7.6 Hi point
Set the TH (THRESHOLD) and DB (DEADBAND) of Hi (REL1). The range for TH is
0.00~20.00ppm(mg/l);while the range for DB is 0.00~3.00ppm(mg/l).
Enter “Low point alarm” set-up
Press or to
decide to activate REL1
or not.
Press or
to make REL1
terminal or lamp test
Choose “on”,
lamp “HI” will be
lighted up on for a
while
Press or to set an
appropriate hysteresis value
Press for confirmation
Press for confirmation
Press for confirmation
Press for confirmation
Press for
confirmation
Press for confirmation
Press or to set the
required alarm point
15

7.7 Lo point
Set the TH (THRESHOLD) and DB (DEADBAND) of Lo (REL2). The range for TH is
0.00~20.00ppm(mg/l);while the range for DB is 0.00~3.00ppm(mg/l).
Enter “Auto clean time” set-up
Press or to
decide to activate REL2
or not.
Press or
to make REL2
terminal or lamp test
Choose “on”,
lamp “Lo” will be
lighted up on for a
while
Press or to set an
appropriate hysteresis value
Press for confirmation
Press for confirmation
Press for
confirmation
Press for confirmation
Press or to set the
required alarm point
Press for confirmation
Press for confirmation
16

Press or
to make RELAY tes
t
Press or to set the current output
and relay hysteresis value ( unit: second)
7.8 Wash time
Set the automatic starting time and turn-off time of the washing function. If any value
is set to be 0, the instrument will automatically stop this function.
Enter “Analog output 1 (Chlorine/O3)” set-up
Press or tosettheturnoff
time of the washing function (unit: hour)
Choose “on”, lamp “WASH” will
be lighted up for a while
Press or tosetthestarting
of the washing function (unit: second)
Press for confirmation
Press for confirmation
Press for confirmation
Press for confirmation
Press for confirmation
Press or to
activate the auto clean
function or not
17
Table of contents
Other Suntex Transmitter manuals

Suntex
Suntex EC-4110 User manual

Suntex
Suntex IT-8100 User manual

Suntex
Suntex CT-6110-POL User manual

Suntex
Suntex CT-6300 User manual

Suntex
Suntex PC-3310 Installation manual

Suntex
Suntex PC-3310-RS User manual

Suntex
Suntex PC-3110-RS User manual

Suntex
Suntex TC-7310-RS-M User manual

Suntex
Suntex EC-4110-I User manual

Suntex
Suntex PC-300A Series User manual
Popular Transmitter manuals by other brands

Mercoid
Mercoid 3100 Installation and operating instructions

Crestron
Crestron DigitalMedia HD-TX-4KZ-421-CHGR manual

Comet
Comet T3117 instruction manual

Siemens
Siemens SITRANS TH400 operating instructions

Elpas
Elpas 5-LW241020-0 user guide

rojaflex
rojaflex WSTR-5 Original assembly and operating instructions