TESTO Saveris 2 User manual

WiFi data logger system · testo Saveris 2
Instruction manual

2

1 Contents
3
1 Contents
1Contents .......................................................................................3
2System..........................................................................................5
2.1 System overview.........................................................................5
3First steps ....................................................................................6
4Licences .......................................................................................6
4.1 testo Saveris 2 App.....................................................................6
5WiFi data logger...........................................................................8
5.1 Technical data ............................................................................8
5.1.1 Measurement-specific data ..................................................8
5.1.2 General data.........................................................................9
5.1.3 WiFi-specific data ...............................................................10
5.1.4 Technical data for a secure wireless LAN..........................10
5.1.5 Authorizations.....................................................................11
5.1.5.1 European Union, EFTA countries................................12
5.1.5.2 USA/Canada................................................................17
5.1.5.3 Japan ...........................................................................17
5.1.5.4 Australia .......................................................................17
5.1.5.5 China............................................................................17
5.1.5.6 Korea ...........................................................................18
5.1.5.7 Russia ..........................................................................18
5.1.5.8 Chile.............................................................................18
5.1.5.9 Singapore ....................................................................18
5.1.5.10 Ukraine.......................................................................18
5.1.5.11 United Arab Emirates.................................................18
5.2 Safety and the environment......................................................19
5.2.1 Safety .................................................................................19
5.2.2 Protecting the environment ................................................19
5.3 Wall bracket ..............................................................................20
5.4 Display and control elements....................................................21
5.4.1 Overview ............................................................................21
5.4.2 Display icons ......................................................................22

1 Contents
4
6User interface.............................................................................23
7Dashboard..................................................................................25
7.1 Measuring points ......................................................................25
7.2 Active alarms ............................................................................25
8Analysis & reports..................................................................... 25
8.1 Data analysis ............................................................................25
8.2 Automatic reports......................................................................26
9Alarms ........................................................................................ 27
9.1 Alarm list ................................................................................... 27
9.2 Alarm settings ........................................................................... 28
9.3 System warnings ......................................................................29
10 Configuration ............................................................................. 30
10.1 Users .....................................................................................30
10.2 User roles ..............................................................................31
10.3 Account ID ............................................................................. 32
10.4 Measuring point groups.........................................................32
10.5 Areas .....................................................................................33
10.6 WiFi data loggers .................................................................. 34
10.7 Firmware updates.................................................................. 35
11 User management ..................................................................... 35
11.1 User settings.......................................................................... 35
11.2 Account information............................................................... 35
11.3 Change password ................................................................. 35
11.4 Logoff..................................................................................... 35
12 Command bar ............................................................................ 36
12.1 Open Quick Start Guide ........................................................36
12.2 Open Online Help..................................................................36
12.3 Open system messages........................................................36
13 System and status information................................................ 36
13.1 Display system status............................................................ 36

2 System
5
2 System
2.1 System overview
The testo Saveris 2 WiFi data logger system is the modern
solution for monitoring temperature and humidity values in
storerooms and work rooms.
The system consists of WiFi data loggers and a Cloud data
storage system. The testo Saveris 2 WiFi data loggers record
temperature and humidity reliably at adjustable intervals and
transmit the readings directly to the Testo Cloud via WLAN. The
readings stored in the Cloud can be analyzed at any time,
anywhere, using an internet-enabled smartphone, tablet or PC.
Limit value violations are immediately reported via e-mail, or
optionally via SMS.
Similarly, WiFi data loggers can be programmed via the Testo
Cloud and analyses set up.
Record readings
with WiFi data
loggers
WLAN router (at
customer site)
Save the
readings in the
Testo Cloud
Call up readings
on any terminal
device: PC,
smartphone,
tablet

4 Licences
6
3 First steps
The Quick Start Guide is provided to assist you when you take
your first steps with the testo Saveris 2.
It can help you with logging in WiFi data loggers.
• In the command bar, click on the icon to open the
Quick Start Guide.
4 Licences
When purchasing your Saveris 2 system you will receive a free
basic licence.
Some measuring functions which are displayed in your Saveris 2
system are inactive. By expanding your licence you can activate
and use these functions.
In the header you can see which licence you are currently using.
Purchase licence
By clicking on the licence display in the header you will be
redirected to the licence shop. Here you can choose and
purchase your licence expansion.
Or
Click on the symbol next to the inactive measuring function
. This also opens the licence shop to expand your
licence.
4.1 testo Saveris 2 App
The testo App for the testo Saveris 2 monitoring system.
The free testo App offers you practical functions for
commissioning, network analysis and receiving Saveris 2 push
messages.
Specifications of the testo Saveris 2 App
Loggers are commissioned via the configuration wizard.

4 Licences
7
Note
In order to be able to establish a connection via WLAN, you need
a tablet or smartphone with the Testo App already installed on it.
You can get the App for Android devices in the Play Store.
Compatibility:
Requires Android 4.3 or later
Please note that, due to IOS system limitations, the network
analysis and logger commissioning can only be carried out on
Android devices
Network analysis
Use the network analysis to analyze your WLAN network in
advance in order to check the requirements for operating the
testo Saveris 2 loggers correctly.
You can also optionally create and send status reports.
Alerts
With the testo App, you can receive all Saveris 2 push alerts and
see an overview.
Purchase the testo Saveris 2 App
This link will take you to the Google Play Store, where you can
download the testo App onto your mobile terminal device free of
charge.

5 WiFi data logger
8
5 WiFi data logger
5.1 Technical data
5.1.1 Measurement-specific data
WiFi
data
logger
s
T1 T2 T3 H1 H2
Sensor
type
1x NTC
internal
2x NTC
or door
contact
external
2x
thermocoupl
e type K/J/T
external
1x
NTC/humi
dity
internal
1x
NTC/hu
midity
external
Measu
ring
range
-30 to +5
0 °C
-50 to +1
50 °C
K: -195 to +1
350 °C
J: -100 to +7
50 °C
T: -200 to +4
00 °C
-30 to +50
°C
0 to 100
%RH 1
Probe-
specific
Accura
cy
±1 digit
± 0.5 °C ± 0.5 °C ± (0.5 °C +
0.5% of
reading)
± 0.5 °C
±2 %RH
Probe-
specific
Resolu
tion
0.1 °C 0.1 °C 0.1 °C
(-200 to +999
.9 °C)
1 °C
(residual
range)
0.1 °C
0.1%
0.1 °C
0.1%
1Not for condensing atmosphere. For continuous application in high humidity
(> 80 %RH at ≤30 °C for > 12 h, > 60 %RH at > 30 °C for > 12 h), please
contact us (www.testo-international.com).

5 WiFi data logger
9
5.1.2 General data
WiFi data
loggers
T1 T2 T3 H1 H2
Operating
temperature
-30 to +50 °C
Storage
temperature
(including
batteries)
-40 to +70 °C
Protection class IP 65 IP 65 IP 54 IP 30 IP 54
Measurement
rate
1 min to 24 h (15 minutes default)
Communication
rate
1 min to 24 h (15 minutes default)
Memory 10,000 readings/channel
Battery life 24 months 2
at +25 °C, 15 min measuring cycle, 15 min day
communication cycle (8 h/day) and 90 min night
communication cycle (16 h/day)
at -30 °C and 15 min measuring cycle, 15 min day
communication cycle (8 h/day) and 90 min night
communication cycle (16 h/day) with Energizer
batteries 0515 0572
Voltage supply 4 x AA AlMn batteries
Mains unit optional
for temperatures below -10 °C please use Energizer
batteries 0515 0572
Dimensions 95 x 75 x 30.5 mm
2typical value, depending on the WLAN infrastructure

5 WiFi data logger
10
WiFi data
loggers
T1 T2 T3 H1 H2
Weight
(including
batteries)
240 g
EN 12830-
compliant
Yes Yes No No No
Door contact No Optional No No No
Wall bracket Included in delivery
5.1.3 WiFi-specific data
WiFi data
loggers
T1 T2 T3 H1 H2
Radio
frequency
2.4 GHz (IEEE 802.11 b/g/n)
Encryption
method
WEP, WPA (TKIP), WPA2 (TKIP, AES, CCMP)
5.1.4 Technical data for a secure wireless LAN
Ports
The testo Saveris 2 WiFi data loggers use the MQTT protocol,
which communicates via port TCP 1883.
These UDP port approvals are also required:
• Port 53 (DNS name resolution)
• Port 123 (NTP time synchronization)
All ports must be able to communicate only externally to the
Cloud. No bi-directional port approvals are necessary.
Ping
The default gateway, which must be communicated to the probe
via DHCP or manually, must answer the PING request of the WiFi
data logger.

5 WiFi data logger
11
Note
During the first configuration, it is possible to select whether
DHCP or Static IP is used (select expert mode for the
corresponding information)
testo Saveris 2 application
The testo Saveris 2 application is accessible via a normal, up-to-
date browser (www). The standard TCP ports http (80) and https
(443) are used.
5.1.5 Authorizations
Please note the following country-specific information for the
product authorization.

5 WiFi data logger
12
5.1.5.1 European Union, EFTA countries
T1 (0572 2001):

5 WiFi data logger
13
T2 (0572 2002):

5 WiFi data logger
14
T3 (0572 2003):

5 WiFi data logger
15
H1 (0572 2004):

5 WiFi data logger
16
H2 (0572 2005):

5 WiFi data logger
17
5.1.5.2 USA/Canada
This instrument complies with Part 15C of the FCC Rules and
Industry Canada RSS-210 (revision 8). Commissioning is subject
to the following two conditions: (1) This instrument must not
cause any harmful interference and (2) this instrument must be
able to cope with interference, even if this has undesirable effects
on operation.
5.1.5.3 Japan
T1 (0572 2001):
T2 (0572 2002):
T3 (0572 2003):
H1 (0572 2004):
H2 (0572 2005):
5.1.5.4 Australia
E 1561
5.1.5.5 China
T1 (0572 2001): CMIIT ID: 2014DP5734
T2 (0572 2002): CMIIT ID: 2014DP5735
T3 (0572 2003): CMIIT ID: 2014DP5736
H1 (0572 2004): CMIIT ID: 2014DP5732
H2 (0572 2005): CMIIT ID: 2014DP5733

5 WiFi data logger
18
5.1.5.6 Korea
Instrumen
t model
Saveris
2 T1
Saveris
2 T2
Saveris
2 T3
Saveris
2 H1
Saveris
2 H2
Identificati
on number
MSIP-
CRM-
Toi-
Saveris2
T1
MSIP-
CRM-
Toi-
Saveris2
T2
MSIP-
CRM-
Toi-
Saveris2
T3
MSIP-
CRM-Toi-
Saveris2
H1
MSIP-
CRM-Toi-
Saveris2
H2
Applicant Testo Instruments (Shenzen) Co., Ltd.
Manufactu
rer
Testo Instruments (Shenzen) Co., Ltd.
Country of
manufactu
re
China
Note
5.1.5.7 Russia
Authorized.
5.1.5.8 Chile
Authorized.
5.1.5.9 Singapore
Authorized.
5.1.5.10 Ukraine
Authorized.
5.1.5.11 United Arab Emirates
Authorized.

5 WiFi data logger
19
5.2 Safety and the environment
5.2.1 Safety
• Only operate the product properly, for its intended purpose
and within the parameters specified in the technical data.
Do not apply any force.
• Only carry out maintenance and repair work on this
instrument described in the documentation. Follow the
prescribed steps exactly. Use only original spare parts from
Testo.
5.2.2 Protecting the environment
• Dispose of faulty rechargeable batteries/spent batteries in
accordance with the valid legal specifications.
• At the end of its useful life, send the product to the
separate collection for electric and electronic devices
(observe local regulations).

5 WiFi data logger
20
5.3 Wall bracket
The wall bracket ensures that the WiFi data logger is kept
securely in place.
Mounting materials are not included in the delivery. Select
mounting materials that are appropriate for the required mounting
location.
1 Wall bracket with notches for mounting materials (screws
or cable ties)
2 Safety bolt
3 Lock
4 Storage compartment for USB port sealing plug
5 Storage compartment for probe socket sealing plug
(left/right)
6 USB cable holder, folding: push down on the spot marked
with a circle using a screwdriver and slide cable holder
downwards.
Other manuals for Saveris 2
2
Table of contents
Other TESTO Measuring Instrument manuals