2.2.1 Analog Inputs
2.2.2 Clock Input
2.2.3 Control Inputs
2.2.3.1 Output Enable
2.2.3.2 MSB Control
2.2.3.3 Power Down
2.2.3.4 External Reference
2.2.4 Power
Circuit Description
The EVM can be configured to provide the ADC with either transformer-coupled or differential amplifierinputs from a single-ended source. The inputs are provided via SMA connectors J15 and J16 fortransformer-coupled input, and J1, J2, J7, and J9 for differential amplifier input. To set up for one of theseoptions, the EVM must be configured as follows:1. For a 1:1 transformer-coupled input to the ADC, a single-ended source is connected to J15 for channelA and J16 for channel B. R5, R6, R8, and R9 must not be installed. This is the default configuration forthe EVM.2. For single-ended inputs into the differential amplifiers, sources are connected to J2 for channel A andJ9 for channel B. R5, R6, R8, and R9 must be installed, and R74, R75, R78, and R79 removed.3. For a differential input into the amplifiers, the positive source of channel A is connected to J2 and thenegative source to J1. The positive source of channel B is connected to J9 and the negative source toJ7. R5, R6, R8, and R9 must be installed, and R74, R75, R78, and R79 removed. R19 and R22 mustbe replaced with 383- Ωresistors, and R12 and R14 must be replaced with 54.9- Ωresistors for propertermination.
To provide a clock to the ADS5231/32/37, apply a 3.3-V, 1.65-V offset, square-wave input to J3. This inputis terminated with a 50- Ωresistor to ground. Caution: the ADS5231/32/37 does not work with asinusoidal clock source.
The EVM has five discrete inputs to control the operation of the device.
With jumper W4 installed between pins 2 and 3, channel A of the ADC is enabled. This channel isdisabled with the jumper installed between pins 1 and 2. With jumper W2 installed between pins 2 and 3,channel B is enabled. Channel B is disabled with the jumper installed between pins 1 and 2.
With jumper W9 installed between pins 2 and 3, the outputs of the ADC are straight binary. The outputsare 2s complement with jumper W9 installed between pins 1 and 2.
With jumper W3 installed between pins 1 and 2, the ADC is in power-down mode.
With jumper W1 installed between pins 1 and 2, the ADC is in internal reference mode. The device is inexternal reference mode with jumper W1 installed between pins 2 and 3. When using the externalreference mode, install R3 and R4. Use the potentiometers R17 and R18 to adjust the corresponding topreference and bottom reference. For convenience, two test points, TP6 and TP7, allow monitoring thevoltage while using the external reference mode.
Power is supplied to the EVM via banana jack sockets. Separate connections are provided for a 3.3-Vanalog supply (J11 and J10), 3.3-V driver supply, external buffer supply (J17 and J18), and ±5-V amplifiersupply (J14, J13, and J12).
SBAU111B – May 2006 – Revised November 2007 ADS5231/32/37 EVM 7Submit Documentation Feedback