Ref.No.:032011
For Model with thermocouple socket
The non-contact infrared thermometer is the ultimate tool for temperature measurements. It can measure from 100 to 1800oC
(212 ~3272oF) in the distance of 100:1 DS (Distance to Spot) ratio. Furthermore, the memory mode can record and display the
specific measurement with emissivity. There are many mathematical modes for the Infrared function.
* Please remember to keep away from children and don’t use it for safety related applications.
(Default Screen)
FUNCTION
Press Emissivity key ( 3) for setting the emissivity
Press Emissivity key ( 3), then press Up key ( 4) or Down key ( 2) to set the emissivity, then press Mode key
the probe measurement, please hold down the Up key ( 4) or Down key ( 2).
After measure high temp, the probe may remain HOT for a while.
** The thermometer will automatically shut off if left idle for more than 60sec.ADD VALUE
Press Up key ( 4) forLOCK mode ON/OFF. The lock mode is particularly useful for continuous
monitoring of temperatures for up to 60 minutes.
In MAX, MIN, DIF,
AVG mode: Press Down key ( 2) foroCor oF transferred.
and press Up key ( 4) for backlight functionON/OFF
In all modes: First
hold on the Meas.
key (
8) and press Down key ( 2) forlaser functionON/OFF.
(With flash light be turned on in the same time as laser )
CAUTION
1. WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE TWO LASER BEAMS–PERMANENTEYE DAMAGE MAY RESULT.
2. USE EXTREME CAUTION WHEN OPERATING THE LASER.
3. NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4. KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
It should be stored at room temperature. The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times,
care should be taken when cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry
before using the thermometer. Do not submerge any part of the thermometer.
LCD ERROR MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and LAL.
「」「」
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’
message remains please contact the Service Department for further assistance.
「」「 」 ‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’: measurements are not
possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AA, 1.5V batteries. Please note: It is
important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
noitcnufnacSderarfnItcatnoc-noNmetI Thermocouple Probe Scan function
(K type; a bead wire included.)
Measurement Range 100 to +1800 oC(212 to +3272oF) K Type: -64 to +1400°C (-83.2 to +2552°F)
Bead : -50 to +250°C(-58 to +482°F)
Operating Range 0 to +50°C (32 to +122°F)
Accuracy (Tamb=23±3°C) ±2% of reading +/-1% of reading or 1°C (1.8°F) whichever is
greater (Test under Tamb=23±6°C)
Emissivity Range 0.95 default – adjustable 0.1 to 1 step .01
Resolution 0.1°C/0.1°F at -83.2 to 999.9(°C/°F), otherwise 1°C/1°F
Response Time (90%) 1 sec.
Distance:Spot 100:1
Memory Detailed 24 Memories with Temperature & Emissivity
).thgiLkcaBdnaresaLtuohtiw,enilaklA(esusuounitnocsruoh041nim,081.pyTefiLyrettaB
Dimensions 233.8*207.5*60.6mm(9.20×8.17×2.38 inch)
Weight 550 grams(19.40 oz) including batteries (AA*2pcs)
Note:Under an electromagnetic field of 3V/m from 700 to 1,000 MHz the maximum error is +/-30°C (54 oF), from 1.4〜1.5 GHz the maximum error is +/- 6 oC (10.8 oF ).
EMC/RFI: Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approximately 3 volts per
meter, but the performance of the instrument will not be permanently affected.
www.thermoworks.com
1762 W. 20 S. #100 | Lindon, UT 84042
801-756-7705
IR-PRO-100 Industrial Hi-Temp Infrared Thermometer - Operating Instructions
Press measurement trigger and release to store the reading that you want to save. (Note – while in Memory (M00)
function readings are not taken). Once the reading is shown on the display, press the MODE button until the M00
(Memory) is displayed. To store the reading press the measurement trigger and the current reading along with the
emissivity setting will be saved to the M01 location and all other readings will be bumped up one memory location
to allow the current reading to be saved. (Note – the current reading can only be stored to one location).
To view stored readings press the MODE button until M00 (Memory) is displayed. Press the Up and Down button
to view readings. To clear all memory data, press and hold the UP and Down arrows simultaneously for 2
seconds while in the Memory function.
Simply aim the thermometer at the
measure target with Lens (11) and press
Meas. key (8) to display the surface
temperature. The Distance:Spot is 100:1.
The two laser points are the reference for
the target spot size. Please make sure
the target area is within the field of view.