UE Systems 100-UP 10000SD User manual

U
L
T
R
A
P
R
O
B
E
1
0
0
0
0
I
N
S
T
R
U
C
T
I
O
N
M
A
N
U
A
L
V 3.2.01

2
Table of Contents
1. Introduction
Ultrasound Technology Instruction Scholarship Certificate 5
2. Ultraprobe Kit
Basic Components 5
A. Plug-In Modules 6
Scanning Module 6
Contact Module
3. Accessories 6
A. Standard Accessories 6
Headset 6
WTG-1 Warble Tone Generator 6
Rubber Focusing Probe 7
Stethoscope Extension Kit 7
BCH-10L Battery Charger 7
BCH-WTG Battery Charger 7
RAS/RAM Magnetic Transducer 7
B. Optional Accessories 7
LRM-9/10 7
CFM-9/10 7
UWC-9/10 7
DHC-2 Headphones 7
TFSM 7
TFCM 7
UFMTG-1991 7
WTG-2SP Warble Pipe Threaded Tone Generator 7
BCH-10L 7
BCH-WTG2 8
HTS-10 8
LLA 8
RAS/RAM-MT 8
3. Overview
A. Pistol Grip Housing 9
Display Panel 9
Trigger On/Off Switch 9
I/O Port 9
Line Input 10
Battery Compartment/Battery 10
Wrist Strap 10
Sensitivity Control Dial 10
Storage Entry Button 10
Head Set Jack 10
Recharge Jack 10
Line Input Jack 10
4. User Instructions 11
Trisonic- Scanning Module 11
Method of Airborne Detection 11
Headset 11
Rubber Focusing Probe 11
Stethoscope Module 11
Stethoscope Extension Kit 11

3
Charging The UP10000 11
Warble Tone Generator 12
Charging The Warble Tone Generator 12
Helpful Hints 12
Auto Shutdown Battery Feature 12
Resetting The On Board Computer 12
6. Operational Overview 13
Main Display Navigation 13
Main Display Information 13
Bargraph Display 14
Sensitivity Control Dial 14
Sensitivity / Volume Adjust 14
Frequency Adjust 14
Function Field Description 15
Yellow Enter Button 15
Storage Display Navigation 15
Storage Display Information 16
ABCD Navigation 16
Entering A Test Condition In Storage Display 17
Entering Text Using Text Editor 17
Alarm Enable/Disable 18
Generic Info 18
Record Wav 18
CF Info 19
To Store A Reading 19
Storing ABCD Information 19
Freezing a Reading on the Display Panel 19
7. Setup Mode
Overview Of Setup Mode 20
How To Use The Setup Mode 20
Menu 01 Data Transfer 20
Menu 02 Display Screen 21
Menu 03 Application Select 21
Menu 04 Module Type Select 22
Menu 05 Instrument Setup (Manual or Auto) 22
Menu 06 Alarm Enable/Disable 22
Menu 07 Set Record Time 22
Menu 08 Record WAV on Alarm 22
Menu 09 Select Function List 23
Menu 10 Text Editor Select 23
Menu 11 MORE 23
Menu 12 Default Settings 23
Menu 13 User Sensitivity Default 24
Menu 14 User Frequency Default 24
Menu 15 Frequency Adjust (Yes or No) 24
Menu 16 Restore Lists 25
Menu 17 Set Shut Off Mode 25
Menu 18 Set Turn Off Time 25
Menu 19 Set WAV Sample Rate 25
Menu 20 Display Response 25
Menu 21 Comm Flow Control 26
Menu 22 Line Input 26
Menu 23 Sensitivity Mode 26
Menu 24 Headphone Volume 27
Menu 25 Enter Inspector ID 27

4
Menu 26 Set Time and Date 27
Menu 27 Date Format Select 27
Menu 28 Cal Due Date 28
Menu 29 Standard or Metric Units 28
Menu 30 Display Mode Select 28
Menu 31 dB Scale Type Select 28
Menu 32 dB Offset Value 29
Menu 33Auto Store to CF 29
Menu 34 Upgrade UE10K PGM 29
Menu 35 BACK 30
8. How To Use The Application Select Modes 30
A. Generic 30
B. Valves 30
C. Bearings 31
D. Electrical 31
E. Steam 32
F. Leaks 33
9. General Applications 34
Leak Detection, How To Locate Leaks, Shielding Techniques, Low Level 34 - 36
Leaks
UltratoneTM 37
Electric Arc, Corona, Tracking Detection 38
Monitoring Bearing Wear 39
Detecting Bearing Failure 40
Comparative and Historical Trending 40
Lack Of Lubrication 40
Over Lubrication 41
Slow Speed Bearings 41
FFT Interface 41
General Mechanical Trouble Shooting 41
Locating Faulty Steam Traps 42
General Steam/Condensate/Flash Steam Confirmation 42
Inverted Bucket, Float & Thermostatic, Thermodynamic and Thermostatic 43
Inspecting Valves 44
ABCD Method 45
Underground Leaks 45
Leakage Behind Walls 46
Partial Blockage 46
Flow Direction 46
Ultrasound Technology 47
Ultraprobe 10,000 Specifications 48
Instructions For Setting The Combination Lock On Carrying Case 49

5
WELCOME TO THE WONDERFUL WORLD OF AIRBORNE ULTRASOUND INSPECTION
Congratulations, you are about to experience the ultimate in airborne/structure borne ultrasonic inspection. Your Ultraprobe 10000 is
equipped with features that will help you inspect in the most challenging environment.
1. INTRODUCTION
Your Ultraprobe 10000 is a versatile instrument with many features that will make your inspections easy, fast and accurate. As with
any new instrument, it is important to review this manual before you begin inspections.
While simple to use as a basic inspection tool, there are many powerful features that when understood, will open up a world of
opportunities for inspection and data analysis.
ULTRASOUND TECHNOLOGY INSTRUCTION SCHOLARSHIP CERTIFICATE:
Your Ultraprobe 10000 has many applications ranging from leak detection and electrical inspection to mechanical inspection.
It may be used to trend, analyze or just find a problem. How it is used is up to you. As you gain knowledge and learn how many
modes of inspection you can cover, you might want to extend your knowledge by enrolling in one of the many training courses offered
by UE Training Systems, Inc. A Scholarship certificate is available to you.
2. ULTRAPROBE 10000 KIT
Pistol-Grip
Housing
WTG-1 Warble
Tone Generator
Rubber Focusing Probe
Long Range
Module
Stethoscope Extension Kit
Stethoscope
Module
Trisonic Scanning
Module
Battery
RAS/RAM
Magnetic
Transducer &
Cable
Headphones
Battery Recharger

6
A. PLUG-IN MODULES
TRISONICTM SCANNING MODULE:
This module is utilized to receive air-borne ultrasound such as the ultrasounds emitted by pressure/vacuum leaks and electrical
discharges. There are four prongs at the rear of the module. For placement, align the prongs with the four corresponding jacks in the
front end of the pistol housing and plug in. The TrisonicTM Scanning Module has a phased array of three piezoelectric transducers to
pick up the airborne ultrasound. This phased array focuses the ultrasound on one "hot spot" for directionality and effectively
intensifies the signal so that minute ultrasonic emissions can be detected.
STETHOSCOPE (CONTACT) MODULE:
This is the module with the metal rod. This rod is utilized as a "waveguide" in that it is sensitive to ultrasound that is generated
internally such as within a pipe, bearing housing or steam trap. Once stimulated by ultrasound, it transfers the signal to a piezoelectric
transducer located directly in the module housing. The module is shielded to provide protection from stray RF waves that have a
tendency to effect electronic receiving and measurement. The Stethoscope (Contact) Module can be effectively utilized in practically
any environment ranging from airports to broadcasting towers. It is equipped with low noise amplification to allow for a clear,
intelligible signal to be received and interpreted. For placement align the four prongs on the back with the corresponding receptacles in
the front of the pistol and plug in.
3. ACCESSORIES
A. STANDARD ACCESSORIES
HEADSET:
This heavy-duty headset is designed for use with or without a hardhat and can block out intense sounds often found in industrial
environments so that the user may easily hear the sounds received by the ULTRAPROBE. In fact, these headphones provide over 23
dB of noise attenuation.
WTG-1 WARBLE TONE GENERATOR:
The WTG-1 Tone Generator is an ultrasonic transmitter designed to flood an area with ultrasound. It is used for a special type of leak
test. 'When placed inside an empty container or on one side of a test item, it will flood that area with an intense ultrasound that will not
penetrate any solid but will flow through any existing fault or void. By scanning with the TrisonicTM Scanning Module, empty
containers such as pipes, tanks, windows, doors, bulkheads or hatches can be instantly checked for leakage. This Tone Generator is a
WARBLE TONE GENERATOR. This internationally patented transmitter sweeps through a number of ultrasonic frequencies in a
fraction of a second to produce a strong, recognizable "Warble" signal. The warble tone prevents a standing wave condition which can
produce false readings and provides for a consistency of testing in practically any material.
Stethoscope (Contact) Module
Trisonic
TM
Scanning Module

7
RUBBER FOCUSING PROBE:
The Rubber Focusing Probe is a cone-shaped rubber shield. It is used to block out stray ultrasound and to assist in narrowing the field
of reception of the “Trisonic" Scanning Module.
STETHOSCOPE EXTENSION KIT:
This consists of three metal rods that will enable a user to reach up to 31 additional inches (78.7 cm) with the Stethoscope Probe.
BCH-10L BATTERY CHARGER:
This is the standard battery charger for the UP10000 with a line input of 120VAC @ 60Hz. The charging time is about 4 hours. (For
countries with 220 V/ 50 Hz, adaptors are available for the BCH 10L.)
BCH-WTG BATTERY CHARGER:
This is the standard battery charger for all Warble Tone Generators that are used with the UP10000. The line input is 120 VAC @
60Hz and the charging time is about 4 hours. (For countries with 220V/50Hz, the BCH-102 is considered a “Standard Accessory”.)
LRM-9/10:
A cone shaped scanning module that increases the detection distance above standard scanning modules. The LRM-9 is ideal for high
voltage inspection and for locating leaks at great distances.
RAS/RAM-MT REMOTE MAGNETIC TRANSDUCER
The RAS/RAM- MT is a magnetically mountable contact probe with cable. The probe is applied to a test surface and the RAM
(Remote Access Module) is plugged into the front end of the Ultraprobe.
B. OPTIONAL ACCESSORIES
CFM-9/10:
A scanning module used for close proximity leak detection in pressure and vacuum systems.
UWC-9/10:
The UWC-9/10, Ultrasonic Waveform Concentrator, doubles the detection distance. The UWC-9/10 is great for corona, tracking and
arc detection at safe distances. Includes carrying case
DHC-2:
Headphone headset is for Standard Applications that do not require the use of a hard hat
TFSM: Telescoping Flexible Scanning Module: A flexible scanning probe that is bent to accommodate odd scanning angles. The
telescoping action helps scan hard to reach areas.
TFCM: Telescoping Flexible Stethoscope (Contact) Module: A flexible contact probe for structure borne inspection that is bent to
accommodate odd test angles. The telescoping action helps get to hard to reach areas.
UFMTG-1991:
The UFMTG 1991 is a multi directional warble tone generator. It has a high power output with a circular transmission pattern of
360.
WTG-2SP WARBLE PIPE THREADED TONE GENERATOR:
A Warble Tone Generator that is used in test conditions where it is not possible to physically place the standard WTG-1 Warble Tone
Generator, such as in pipes or in certain heat exchangers or tanks. Features: 1” NPT male threaded nipple with adapters for ¾” and ½”
female nipple with a 10 turn amplitude adjustment dial. Metric adapters available.

8
UE-POD:
Battery Recharge Pod docking station for charging Ultraprobe Batteries (Lithium Ion only). This optional pod will charge a battery
that has been removed from the Ultraprobe 10,000 and is also used to charge extra batteries.
BCH-WTG2:
This is an optional battery charger for use with all Warble Tone Generators. The line input is 220 VAC @ 50Hz and the charging
time is about 8 hours.
HTS-10:
Holster set for the UP10000. Includes a utility belt and two holsters, one for the UP10000 and one extra module and rubber focusing
probe; the other for accessories.
LLA:
Liquid Leak Amplifier case. It holds 12 8 oz. Bottles of Liquid Leak Amplifier (For Ultrasonic Bubble Testing).
\

9
4. OVERVIEW
A. PISTOL-GRIP HOUSING
DISPLAY PANEL:
In the Operation Mode the Display Panel will show inspection parameters such as intensity levels (as dB and as a bar graph),
Frequency, Battery Level, Sensitivity Level, Sensitivity Value, “Display Mode” and Function Select Field. Intensity levels are shown
simultaneously as a numeric dB value and on a sixteen segmented bar graph (with each segment representing 3 dB). The Frequency is
adjustable from 20 kHz to 100 kHz. These represent the range of frequency selection capable with the Ultraprobe. The most common
frequency used for general leak detection or electrical inspection is 40 kHz. These frequencies may be "tuned in" when performing
inspections with any of the Ultraprobe’s plug-in modules. The Display Mode indicates the operation mode of the instrument. This is
indicated on the display panel as RT for Real Time,PH for Peak Hold,SS for Snap Shot, or RTO for Real Time “Offset”,PHO for
Peak Hold “Offset” and SSO for Snap Shot “Offset”. (For additional information about Offset see dB Scale Select in the Set Up
Mode Section.
TRIGGER ON/OFF TRIGGER SWITCH:
The Ultraprobe is always "off' until the trigger switch is pressed. To operate, press the trigger. To turn the instrument off, release the
trigger. The time it takes for the instrument to turn off after release of the trigger can be set in the Set Turn Off Time mode described
later. NOTE: To save the battery charge, if the instrument is set in the “Trigger Shut Off Mode (set up 17), it will automatically turn
off after approximately 5 minutes.
I/O PORT:
This is the RS232 port for upload/download information transfer, including heterodyned sound output. Align the pins from the cable
and plug in. Note: before downloading data be sure the cable is connected to both the I/O port and to the computer or other hardware
(vibration analyzer, tape recorder, etc.). This connection may also be used for sound recording, vibration analyzer and notebook
computer download. In this manner, the sounds, as they are recorded, can be simultaneously listened to through the headphones.
Sensitivity Control Dial Yellow Enter Button
Display Panel
I/O CONNECTOR DIAGRAM
Pin 1 – N/A
Pin 2 - RX
Pin 3 - TX
Pin 4 - DTR
Pin 5 - GND
Pin 6 – N/A
Pin 7 - RTS
Pin 8 - CTS
Pin 9 – N/A

10
LINE INPUT:
This Input is designed to receive signals from other instruments such as Temperature readings from infrared thermometers. The
Instrument will display the readings on the Display Panel.
QUICK CHANGE BATTERY:
To insert the battery, align the battery with the handle (arrow pointing towards the trigger) and push in until the clips snap in place.
To remove the battery, push firmly on the battery clips with the fingers of one hand while holding your other hand under the handle
to catch the released battery.
BATTERY:
This Ultraprobe uses a lithium ion battery with no memory problems. A full charge will take about 4 hours, however you may charge
the unit at any time for short intervals or for a longer period. If it is kept on charge over 4 hours, there will be no harm to the battery.
NOTE: When the effective battery charge is used up the instrument shuts down and a message to recharge the battery will be
displayed in the display panel.
WRIST STRAP:
To protect the instrument, against being dropped accidentally, use the wrist strap.
SENSITIVITY CONTROL DIAL:
This is one of the most important controls in the unit. In the operation mode it can be used to adjust the sensitivity. When clicked or
spun, it can be used to scroll around the screen or to select operation categories or change frequency. In the Set Up Mode it moves the
cursor and by clicking sets notations.
YELLOW ENTER BUTTON:
This yellow button is used to enter and exit the various operations such as set up mode, text editor, store data or record WAV files.
HEAD SET JACK:
This is where you plug in the headset. Be sure to plug it in firmly until it clicks. This can also be used as an output for a vibration
analyzer or notebook computer when used with a proper miniphone to BNC connector.
RECHARGE JACK :
This Jack receives the plug from the recharger. The recharger is designed to plug into a standard electrical receptacle.
LINE INPUT JACK:
This jack is provided to connect other instrument outputs to the Ultraprobe 10000.
Recharge Jack
Line In
I/O Port
Battery
Battery Clip

11
5. USERS INSTRUCTIONS
TRISONICTM SCANNING MODULE:
1. This module plugs into the front end of the instrument.
2. Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH)
and plug in.
3. For general use position the frequency selection to 40 kHz.
4. Start to scan the test area.
METHOD OF AIRBORNE DETECTION:
The method of air borne detection is to go from the "gross to the fine". Start at maximum sensitivity (S=70), constantly reducing the
sensitivity and following the bar graph amplitude display to the loudest point. If there is too much ultrasound in the area, reduce the
sensitivity, place the RUBBER FOCUSING PROBE (described below) over the scanning module and proceed. Every time the sound
level rises to a point where it is difficult to follow, reduce the sensitivity again and again until you are able to follow the test sound to
its' loudest point.
HEADSET:
The DHC-HH headphones are designed to be worn with hard hats. To use, simply plug the headset cord into the headset Jack on the
pistol housing, and place the headphones over your ears.
RUBBER FOCUSING PROBE :
To use, simply slip it over the front of the scanning module or the contact module.
NOTE: To prevent damage to the module plugs, always remove the module BEFORE attaching and/or
removing the Rubber Focusing Probe.
STETHOSCOPE MODULE:
1. Align the pins located at the rear of the module with the four jacks in the front end of the Metered Pistol Housing (MPH)
and plug in.
2. Touch test area.
As with the SCANNING MODULE, go from the "gross" to the "fine". Start at maximum sensitivity (S=70) on the Sensitivity
Selection dial and proceed to reduce the sensitivity until a satisfactory sound level is achieved. At times it may be necessary to utilize
the STETHOSCOPE MODULE with the sensitivity level at or near maximum. Occasionally when in this situation stray ultrasound
may interfere with clear reception and be confusing. If this occurs, place the RUBBER FOCUSING PROBE over the Stethoscope
probe to insulate against the stray ultrasound.
STETHOSCOPE EXTENSION KIT:
1. Remove the Stethoscope Module from the Metered Pistol Housing.
2. Unscrew the metal rod in the Stethoscope Module.
3. Look at the thread of the rod you just unscrewed and locate a rod in the kit that has the same size thread - this is the "base piece".
4. Screw the Base Piece into the Stethoscope Module.
5. If all 31" (78.7 cm) are to be utilized, locate the middle piece. (This is the rod with a female fitting at one end) and screw this
piece into the base piece.
6. Screw third "end piece" into middle piece.
7. If a shorter length is desired, omit step 5 and screw "end piece" into "base piece".
TO CHARGE THE UP10000:
1. Plug recharger cable into recharger jack on the UP10000 and then plug the recharger into a wall receptacle.
2. Make sure that the LED on the charger is blinking when recharging.
3. The LED remains solid when the battery is charged. The instrument may stay connected to the charger without
damaging the battery. Charge time is approximately 4 hours.
4. WARNING: Use the supplied UE Systems recharger (BCH-10L) only. Use of unauthorized rechargers will void the warranty
and may degrade or damage the battery.

12
WARBLE TONE GENERATOR (UE-WTG-1):
1. Turn Tone Generator on by selecting either "LOW" for a low amplitude signal (usually recommended for small containers) or
"HIGH" for high amplitude. In the High position, the Warble Tone Generator will cover up to 4,000 cubic feet (113.3 cu. meters)
of unobstructed space.
2. When the Tone Generator is on, a red light (located below the recharge jack in the front) flickers.
3. Place the Warble Tone Generator within the test item/container and seal or close it. Then scan the suspect areas with the
TrisonicTM Scanning Module in the Ultraprobe and listen for where the "warble" ultrasound penetrates.
As an example, if the item to be tested is the seal around a window, place the Warble Tone Generator on one side of the window,
close it and proceed to scan on the opposite side.
4. To test the condition of the Warble Tone Generator battery, set to the LOW INTENSITY position and listen to the sound
through the Ultraprobe at 40 kHz. A continuous warbling sound should be heard. If a "beeping" is heard instead, then a full
recharge of the Warble Tone Generator is indicated.
TO CHARGE THE WARBLE TONE GENERATOR:
1. Plug recharger cable into recharger jack on the Warble Tone Generator and then plug the recharger into a wall receptacle.
2. Make sure that the LED on the charger is lit when recharging.
3. The LED turns OFF when the battery is charged.
HELPFUL HINTS:
Before you begin your inspection activities, it is suggested that you review the applications section to become familiar with the basic
inspection methods.
Here are some helpful hints that should prove useful in some difficult situations.
If you can’t read your display panel while testing:
1. Trigger-release-
Trigger method: Take the reading while holding the trigger in. Immediately release the trigger and the reading
will be set. Once the trigger is pulled again, the instrument will return to normal operation . NOTE: This operation can only be
performed when the instrument SHUTOFF mode is set at TIMED.
2. Snap Shot Method: If you know you are going to be in a situation where you won’t be able to read the display panel, go into
Setup Mode and spin to Display Mode (Menu 09). Select Snap Shot and return to Operation Mode. This will hold your reading
even while you continue to hold the trigger in. For a quick grab, take a reading, press the trigger in to hold it. For an update or
new reading, simply release and press the trigger switch.
AUTO-SHUTDOWN BATTERY FEATURE
The Ultraprobe 10000 is equipped with an auto-shutdown feature, which enables the instrument to provide accurate measurement
throughout the effective battery charge. If the Ultraprobe 10000 goes into the auto-shutdown mode, A message in the Display Panel
will read “RECHARGE BATTERY”, no sound will be heard in the headphones and measurements will not be displayed on the
display panel. To restore the Ultraprobe 10000 to its’ normal operating mode, recharge the instrument using the BCH-10L Battery
Charger (BCH-10L2 for 220 V operation). If the Compact Flash Card is installed, on power-down a prompt will appear to store
records to the compact flash card. The instrument can be preset to automatically store all records onto the card automatically at
shutdown.
RESETTING THE ON BOARD COMPUTER:
For security purposes, there is no reset switch on the instrument. Should it be necessary to reset the instrument:, disconnect the battery
for one (1) minute and then reconnect the battery or enter SETUP Mode and spin to DEFAULT SETTINGS(Menu 12) and choose the
YES prompt to re-set to factory default settings. WARNING: Selecting Default Settings erases all records stored in the
instrument.

13
6. OPERATIONAL OVERVIEW
The Ultraprobe 10000 is an ultrasonic inspection information, storage, sound event recording and retrieval system in a pistol housing.
There are two modes that are important to understand:
OPERATION MODE:
The operation mode will be described in detail under the operation mode section. In this mode you will perform all inspection actions
such as scanning, probing, “Click and Spin” activities, sound recording and data storage. NOTE: “Click” operations require pressing
a dial. “Spin” operations require turning a dial.
SET UP MODE:
The setup mode will be described in detail under the Set Up Mode section,. There are 24 menu options that will be described in that
section. To use the set up mode, you must “click” to the bottom of the display column, “spin” to “setup mode and press the yellow
button labeled “enter”. For more details, see the SETUP Mode section.
A. OPERATION MODE
1. MAIN DISPLAY NAVIGATION
Moving around the Display Panel is easy. Basically three clicks (presses) of the Sensitivity Dial will move the cursor on the
Display Panel to three key spots: the Decibel Indicator, the Frequency Indicator and the Function Selection Indicator. When
an indicator is selected, or “active”, it will blink. When the cursor is moved to the Decibel Indicator and the Decibel Indicator
is blinking, the Sensitivity Dial may be spun (rotated) clockwise or counter clockwise to adjust the sensitivity/volume of the
instrument. When the cursor is moved to the Frequency Indicator and the Frequency Indicator is blinking, the Sensitivity
Dial may be spun to change the frequency. When blinking, the Function Selection, located on the bottom of the screen can be
changed by spinning the Sensitivity Dial up or down to select a specific feature which may be entered by pressing the yellow
Enter Button.
Main Display Information:
When the trigger is pressed to turn the instrument on, the Display Panel will display intensity levels simultaneously on a bar graph and
as a numerical decibel value. The current selected frequency will also be shown. Remaining Battery Charge is shown in the upper
right corner. Immediately under the dB reading is the Sensitivity value which may be used as a reference for comparing readings or
for specialized trending purposes. The letters RT,SS or PK note the current display mode. RT indicates that the instrument is
running in “Real Time”,SS indicates “Snap Shot” and PK indicates “Peak Hold”. Should the instrument be set in the Offset Mode,
then the letters RO,SO and PO will be displayed. Along the bottom right you will see the record number (identified as 001-399 Rec).
On the bottom line of the display panel, you will find the Function Select Indicator: “Press Enter for:” (you can use this to select
any of the following function fields.) The default function fields are:
Click for dB
Click for Function Options
Click for Frequency
002 Rec
Storage Location
(Record Number)

14
a. Setup Menu : The various screen and operation functions can be set in this mode
b. Text Editor (ON/OFF): When on text notations can be set using up to 16 alphanumeric characters.
c. Alarm (DISABLED ENABLED): Set (enable) alarm levels in dB or turn the alarm off (disable).
d. GENERIC Info: This function provides inspection information. It can also be set to display specific application information
(for example: Leak Info, Bearing Info, Steam Info, Electric Info, Valve Info)
e. Storage Display: The display panel will split in half. The top half will show current test data while the bottom will display
stored information.
f. Record WAV: The Ultraprobe 10000 is configured to record heterodyned ultrasounds with this function setting.
g. CF INFO (Compact Flash): Information of stored recorded WAV files and record time can be viewed here.
h. STORE Record: Use this function to store inspection data. This is a quick way to data log if inspection data is to be saved
without viewing previously stored data (as in Storage Display).
i. Exit: Press the Yellow ENTER Button and you will EXIT back to Operation Mode.
BAR GRAPH DISPLAY:
The bar graph has 16 segments. Each segment equals 3 dB.
40 dB
SENSITIVITY CONTROL DIAL:
This is the main control dial. It will function as a cursor control as well as a sensitivity and frequency controller.
TO ADJUST THE SENSITIVITY/VOLUME:
1. Look at the meter. To adjust the sensitivity, the dB (decibel) indicator must blink. The kHz (frequency indicator) must be
steady (not blinking).
2. 2If the frequency indicator is blinking, click in the Sensitivity Control Dial until the frequency indicator is steady and the
dB (decibel) indicator blinks.. This indicates that you are now able to adjust the sensitivity.
3. Once in the Sensitivity mode, turn the Sensitivity Control Dial clockwise to increase the sensitivity and counter clockwise
to decrease the sensitivity.
4. The Sensitivity Control Dial increases/decreases the sensitivity of the instrument simultaneously with the sound level in the
headphones. When in range, it moves the bar graph up and down and will adjust the headset volume.
NOTE: the instrument needs to be in range for accurate testing.
5. If the sensitivity is too low, a blinking arrow pointing to the right will appear and there will be no numeric dB value visible
in the display panel. If this occurs, increase the sensitivity until the arrow disappears (in low level sound environments the
arrow will blink continuously and It will not be possible to achieve a dB indication until a higher intensity sound level is
sensed).
6. If the sensitivity is too high, a blinking arrow pointing to the left will appear and there will be no numeric dB value visible
on the display panel. Reduce the sensitivity until the arrow disappears and the numeric dB value is shown.
NOTE: The blinking arrow indicates the direction in which the Sensitivity Control Dial is to be turned.
7. The Sensitivity Control Dial controls the bar graph display.
8. Each click of the Sensitivity Control Dial changes the sensitivity / volume by 1 dB
TO ADJUST THE FREQUENCY:
1. Look at the meter. The kHz indicator must blink to be able to tune the frequency.
2. If it is not blinking, “Click” in the Sensitivity Control Dial until the kHz indicator blinks.
3. When the kHz indicator blinks, change the frequency by rotating the Sensitivity Control Dial up (clockwise) or down
(counter clockwise).
Real Time = RT
Snap Shot = SS
Peak Hold = PK
60 dB 40 kHz
S=69
SS
Press ENTER
for Setup Menu
60 dB 40 kHz
S=69
PK
Press ENTER
for Setup Menu
60 dB 40 kHz
S=69
RT
Press ENTER
for Setup Menu

15
FUNCTION FIELD DESCRIPTION
a. SETUP MENU : The various screen and operation functions can be set in this mode
b. TEXT EDITOR (ON/OFF): When on text notations can be set using up to 16 alphanumeric characters.
c. ALARM (DISABLED ENABLED): Set (enable) alarm levels in dB or turn the alarm off (disable).
d. GENERIC INFO: This function provides inspection information. It can also be set to display specific application
information (for example: Leak Info, Bearing Info, Steam Info, Electric Info, Valve Info)
e. STORAGE DISPLAY: The display panel will split in half. The top half will show current test data while the bottom will
display stored information.
f. RECORD WAV: The Ultraprobe 10000 is configured to record heterodyned ultrasounds with this function setting.
g. CF INFO (Compact Flash): Information of stored recorded WAV files and record time can be viewed here.
h. STORE RECORD: Use this function to store inspection data. This is a quick way to data log if inspection data is to be
saved without viewing previously stored data (as in Storage Display).
i. EXIT: Press the yellow ENTER button and you will EXIT back to Operation Mode.
YELLOW ENTER BUTTON:
The Yellow Enter Button allows you to enter and exit specific menu operations. A message on the bottom of the Display Panel
will note when to press this button.
2. STORAGE DISPLAY NAVIGATION
Moving around the Display Panel is easy. Basically four clicks (presses) of the Sensitivity Control Dial will move the cursor on
the Display Panel to four key spots: the Decibel Indicator, the Frequency Indicator,Storage Location Indicator and the
Function Selection Indicator. When an indicator is selected, or “active”, it will blink. When the cursor is moved to the Decibel
Indicator and the Decibel Indicator is blinking, it may be spun (rotated) clockwise or counter clockwise to adjust the
sensitivity/volume of the instrument. When the cursor is moved to the Frequency Indicator and the Frequency Indicator is
blinking, the Sensitivity Control Dial may be spun to change the frequency. The Storage Location (Record Number) can be
changed up or down when it is blinking by spinning the Sensitivity Control Dial. When blinking, the Function Selection,
located on the bottom of the screen can be changed by spinning the Sensitivity Control Dial up or down to select a specific
feature which may be entered by pressing the Yellow Enter Button. Note: If a storage location has been used during a test
run, it will be highlighted in black with white letters as shown:
On the bottom line of the Display Panel, you will find the Function Select Indicator: “Press Enter for:” (you can use this to
enter any of the following function fields) The default function fields are:
a. EXIT MAIN: Exits to main display.
b. EDIT TEXT: When on text notations can be set using up to 16 alphanumeric characters.
c. MORE: View additional test information including Date, Time and Module used.
d. PARAM INFO: View or change inspection details. It can also be set in the Applications Select menu for specific application
information such as Leak Info, Bearing Info, Steam Info, Electric Info, Valve Info)
Storage Location
(Record Number) Click for Frequency
Click for dB
Click for Function selection
007

16
e. RECORD WAV: The Ultraprobe 10000 is configured to record heterodyned ultrasounds with this function setting.
f. CF INFO (Compact Flash): Information of stored recorded WAV files and record time can be viewed here.
g. STORE REC: Use this function to store inspection data. This is a quick way to data log if inspection data is to be saved
without viewing previously stored data (as in Storage Display).
Storage Display Information:
When in the Storage Display mode, the basic inspection parameter information can be viewed and stored if desired. To view
Parameter Information when in the Storage Display mode:
1. “Click” the Sensitivity Control Dial until the “Press ENTER For” blinks.
2. Spin to Param INFO and Press (Click) the Yellow ENTER Button to enter
3. The information display will read: “Test Results:” and “Temperature:”
PARAMETER INFO (in Storage Display only)
The Parameter Info screen displays the test data relevant to a selected application (ex: bearings, valves, leak, steam). This
screen may be used to enter additional data such as test results or temperature. An extension of the Parameter Information
screen is the MORE Function Selection. This will display additional inspection information: Date, Time, Module used and
Offset Value (if an offset value has been set). See the APPLICATIONS section for details. Note: this is the same information as
will be seen in the basic application information screens. i.e.: GENERIC INFO or BEARING INFO.
3. ABCD NAVIGATION
Note: The ABCD Display will only work if the “Valve” application has been enabled in the Setup Mode. This is described in
SETUP, “Menu 03 Application Select”. To select the ABCD Screen, after the Valve Application has been selected, Select
Menu 02,Display Screens in the Setup Mode and spin to ABCD.
Moving around the Display Panel is easy. Basically three clicks (presses) of the Sensitivity Control Dial will move the cursor
on the Display Panel to three key spots: the Decibel Indicator, the Frequency Indicator, and the Function Selection Indicator.
When an indicator is selected, or “active”, it will blink. When the cursor is moved to the Decibel Indicator and the Decibel
Indicator is blinking, it may be spun (rotated) clockwise or counter clockwise to adjust the sensitivity/volume of the instrument.
When the cursor is moved to the Frequency Indicator and the Frequency Indicator is blinking, the Sensitivity Control Dial
may be spun to change the frequency. When blinking, the Function Selection, located on the bottom of the screen can be
changed by spinning the Sensitivity Control Dial up or down to select a specific feature which may be entered by pressing the
Yellow Enter Button. Each A,B,C and D dB entry will be stored separately by clicking the Yellow ENTER button to store.
On the bottom line of the display panel, you will find the Function Select Indicator: “Press Enter for:” (you can use this to
enter any of the following default fields) The default function fields are:
a. STORE (A) K. CF INFO (COMPACT FLASH)
b. STORE (B) L. STORE RECORD AND EXIT
c. STORE (C)
d. STORE (D)
e. SETUP MENU
f. TEXT EDITOR (ON/OFF)
g. ALARM (DISABLED ENABLED)
h. VALVE INFO
i. STORAGE DISPLAY
j. RECORD WAV
Click for dB
Click for Frequency
Click for Function selection

17
HOW TO USE THE OPERATIONAL FEATURES
ENTERING A TEST CONDITION IN STORAGE DISPLAY
1. Spin to Test, the condition default is OK and will blink slowly. If it is necessary to change the test condition:
2. Click with the Sensitivity Control Dial and the condition (OK) will blink rapidly.
3. Spin to select a condition indication. The choices are; OK,CHK (for check) and a series of TR1-TR9, which can be used as
codes for other conditions. These codes can be changed in the Ultratrend software and uploaded to the Ultraprobe.
4. To Set, Click the Sensitivity Control Dial (the condition will blink slowly)
5. Spin to Temperature should a Temperature entry be desired
6. Click on the Temperature setting (it will blink rapidly) Spin to the left to select a “100” reading (ex 200, 300) and to the
Right to select a “tens” reading (ex: 50)
7. Click to set
8. Press Enter to store the information or
9. Spin to EXIT and click to exit.
Additional information can be viewed by clicking the Sensitivity Control Dial until the ENTER For Blinks. Spin to MORE and
press the Yellow Enter Button to enter. There you will see the following information represented as an example below:
ENTERING TEXT USING TEXT EDITOR:
TO ENTER TEXT IN THE TEXT FIELD:
1. There is space to enter 16 characters of alphanumeric text.
2. To enter text, the Text Edit feature must be set to ON (refer to SET UP MODE 08). If text is ON, click the Sensitivity
Control Dial to the bottom of the Display Panel until Press ENTER For is blinking. Spin to Edit Text and press Enter.
3. The text icon will blink. If the field has no entry, the screen will be blank.
4. The Sensitivity Control Dial can be used to scroll through the alphabet, A-Z , a space character and then through numbers 0-
9. move back through numbers (9-0) and then back down the alphabet (Z-A).
5. Spin the Sensitivity Dial clockwise to move up through numbers and then through the alphabet starting with Aor spin
counter clockwise to move back through numbers (9-0) and then back down the alphabet (Z-A).
6. To enter a character, spin the Sensitivity Control Dial to the desired character and then click to move the cursor to the next
of 16 positions.
7. Continue until through or until all 16 fields are filled.
8. If an error in recording a letter or number occurs, click the Sensitivity Control Dial and the cursor will move to the right.
Continue clicking the Sensitivity Dial and the cursor will “wrap” around to the right until the desired location is reached. As
explained above, spin the Sensitivity Dial until the corrected entry is displayed and “click” the Sensitivity Control Dial to
enter the text character.
9. When the text is correct, click the Yellow Enter Button to save and store the text. The instrument will return to the
Operation Mode.Gear 12345
Press ENTER to Exit
Press ENTER GENERIC
To store 001
Test Results: OK
Temperature: 150 F
001 43dB
S=33 30KhZ
----STORED DATA-----
62 dB 30 kHz RT
Sens 69 Alarm 84
ENTER for Param INFO
DATE: 01/01/03
TIME 08:30:43
Module Type: SCM
Offset: 00 dB
Press ENTER to Store
Press Sens. to Exit

18
ALARM ENABLE/DISABLE
While in the Operation mode, this function selection will display Alarm Disabled if no alarm level has been set. If an alarm
level has been set, it will show the set level in dB. To enable the alarm, see ALARM ENABLE/DISABLE in the SETUP
MODE below.
WARNING: If the alarm value is changed, the new dB alarm value will be stored in the record and will overwrite the previous value.
The alarm dB level stored in the current record can be viewed in the main display. It can be changed. To do so:
1. Click down to the bottom of the Display Panel until PRESS ENTER FOR:… is blinking
2. Spin to Alarm (the dB value will be shown)
3. Click to Enter
4. Spin to the desired dB alarm level
5. Click the Yellow ENTER Button to exit
GENERIC INFO:
1. To view Generic Info, click the Sensitivity Control Dial to the bottom of the Display Panel until “Press ENTER For” is
blinking. Spin to Generic Info and press Enter.
2. To enter a test condition:
3. Spin to Test the condition default is OK and will blink slowly. If it is necessary to change:
4. Click with the SensitivityControl Dial and the condition (OK) will blink rapidly.
5. Spin to select a condition indication. The choices are; OK,CHK (for check) and a series of TR1-TR9 which can be use as
codes for other conditions.
6. To Set, Click the Sensitivity Control Dial (the condition will blink slowly)
7. Spin to Temperature should a Temperature entry be desired
8. Click on the Temperature setting (it will blink rapidly) Spin to the left to select a “100” reading (ex 200, 300) and to the
Right to select a “tens” reading (ex: 50)
9. Click to set
10. Press Enter to store the information or
11. Spin to EXIT and click to exit.
Record WAV
To Record a WAV (sound file) click the Sensitivity Control Dial to the bottom of the Display Panel until “Press ENTER For”
is blinking. Spin to Record WAV and press Enter.
Be sure the Compact Flash Card is inserted into the slot on the Ultraprobe and keep the Trigger Switch pulled on (or set the
Timed Off mode as described in Setup Menu 16 and 17 to shut off at a pre-determined time, ex: 30 seconds.) or set to Trigger
Off mode (see setup mode 17 Shut Off Mode).
1. When ready to record, click the Sensitivity Dial and Hold it in for the duration of time you wish to record.
2. When through, release the Sensitivity Control Dial.
3. You can preset the time of recording in Setup Mode 7 for: Manual (as above), 5, 10, 15, 20,25,30 seconds.
4. Press Enter to playback/review the sound you have recorded.
5. Press Sensitivity Control Dial to Continue on to the next screen
6. Press Enter to Save the recording or if you do not wish to save the recording, press Sensitivity Control Dial to Exit.
7. When through saving a recording, press the Sensitivity Control Dial to Exit.
Press ENTER GENERIC
To store 001
Test Results: OK
Temperature: 350 F
60 dB 40 kHz
S=69
RT
Press ENTER
for Generic Info
60 dB 40 kHz
S=69
RT
Press ENTER
for Alarm
Set Alarm Value
Current Alarm Value
( 120 dB )
Press ENTER
for Setup Menu

19
CF INFO
This feature allows you to view file names stored on the Compact Flash Card. When a sound is recorded and a WAV file is
saved, two files are written to the Compact Flash Card: a text file which contains all stored record fields for the specific record
number associated with the WAV file and the WAV file. When CF info is selected, 5 selections will be viewed on the screen.
You may click the Sensitivity Dial to select/enter any of these.
1. Free Space & Info (This will display the number of Bytes Free and the Remaining Recording time on the card).
2. Display Files (The files stored on the CF card)
3. Delete Files
To delete a file:
a. Click to enter, the File Number will blink.
b. Press Sens to DELETE or Spin to the next file (the file you wish to delete)
c. If you don’t want to delete a file, press ENTER to exit.
4. Delete ALL Files
5. EXIT to MAIN
WARNING: Never remove or insert the Compact Flash Card with the instrument turned on.
TO STORE A READING:
1. Continue to “Click” the Sensitivity Control Dial to move the indicator to the bottom of the meter until PRESS ENTER
FOR: is blinking. Then Spin the dial until STORE Record is displayed. press the Yellow Enter Button. This puts the
instrument in the Data Storage mode. In the data storage mode the Display Panel changes.
2. The screen will display the following: Save To Record (the current record number is displayed, (for example 003) (This
allows you to determine whether to store the information in this particular record location or to change the location.) Turn
SENS.: Change (If you want to change the store location this prompts you to turn (Spin) the Sensitivity Control Dial to
change the location). Press the Sensitivity Control Dial to EXIT (If you do not wish to store the information, Click the
Sensitivity Control Dial to return to operation mode) Click (Press) Enter to Store (If you want to store the information,
Click Enter, a screen will pop up to confirm the information has been stored.
3. To view previously stored information, Click the Sensitivity Control Dial to the bottom of the Display Panel until “PRESS
ENTER FOR” is blinking and spin to Storage Display. Press Enter to view the stored data. To view information in other
storage locations, click the Sensitivity Control Dial until the storage number blinks. The storage number in the Storage
Display mode is located in the upper left corner. Spin the Sensitivity Control Dial to the desired storage location.
4. If you wish to view all stored information during inspections, you may set the screen to Storage Display.
Click the Sensitivity Control Dial until “PRESS ENTER FOR” is blinking. 2. If Storage Display is not shown, spin until
Storage Display is shown and blinking. 3. Click the Yellow Enter button. The Display Screen will be split. All
inspection data will be displayed on the top half while the stored information will be displayed on the bottom half of the
Display Screen.
STORING ABCD INFORMATION
To Store information when the ABCD Function has been set:
1. When Click ENTER for Store (A) is blinking, press ENTER
2. Then Click ENTER for Store (B) will then blink, press ENTER
3. Then Click ENTER for Store (C) will blink, press ENTER
4. Then Click ENTER for Store (D) will blink, press ENTER
5. The function will then shift to STORE DATA
6. Press ENTER and all four dB levels will be stored at the selected Record Number.
FREEZING A READING ON THE DISPLAY PANEL
If in a test situation where it is difficult to view the display panel, when in the Timed Off Mode the decibel reading can be frozen for
viewing. If using the Stethoscope Module, continue to touch the test point and then release the trigger. Remove the instrument and
view the panel. Be sure to view within the set timed off period (ex: within 5 seconds if set for 5 second off) If in the scanning mode,
point at target, release the trigger as described.
6. SETUP MODE
42 dB 25 kHz
002 Rec
S=60 A: 0dB C: 0dB
B: 0dB D: 0dB
Press ENTER
for Store (A)

20
1 Data Transfer 18 Set Turn Off Time
2 Display Screens 19 Set WAV Sample Rate
3 Application Select 20 Display Response
4 Module type select 21 Communication Flow Control
5 Instrument Setup (Manual or Auto) 22 Line Input Select
6 Alarm Enable/Disable 23 Sensitivity Mode
7 Set Record Time 24 Headphone Volume
8 Record WAV on Alarm 25 Enter Inspector ID
9 Select Function List 26 Set time and Date
10 Text Editor Select 27 Date Format Select
11 MORE (moves to second level) 28 Cal Due Date
SECOND LEVEL SET UP MODE 29 Standard or Metric Units
12 Default Settings 30 Display Mode Select
13 User Sensitivity Default 31 dB Scale Type Select
14 User Frequency Default 32 dB Offset Value
15 Frequency Adjust (Yes or No) 33 Auto Store to CF
16 Restore Lists 34 Upgrade UE10K PGM
17 Set Shut Off Mode 35 BACK (moves to first level)
OVERVIEW OF SETUP MODE
There are 35 functions that can be set to either change the display screens or the format of data entry and collection.
HOW TO USE THE SETUP MODE
There are two ways to enter the Set Up Mode:
a. While in Operation Mode
1. Click down to move the cursor to the bottom of the Display Screen until PRESS ENTER FOR is blinking
2. Spin to Setup Mode
3. Click the Yellow ENTER button
b. While the instrument is off
1. Click both the Yellow ENTER button and the Sensitivity Control Dial at the same time. Only after these two
controls have been pressed, then squeeze the trigger. NOTE: Hold the Trigger in during any of the Set Up
Mode operations.
2. When in the first Menu mode: Data Transfer, you may move to any of the other Menu modes by spinning the
Sensitivity Dial up or down (clockwise or counter clockwise).
3. When the desired Menu mode is reached, Click the Sensitivity Control Dial in.
4. You may Spin and Click to enter and exit any Menu mode in the Set Up mode as long as the trigger on/off
switch is pressed.
Menu 01 Data Transfer
NOTE: Before downloading data, be sure the Ultraprobe is connected to the computer.
To download data from the Ultraprobe to your computer:
1. Follow the steps to Enter the Setup Mode as described earlier.
2. The first selection to be displayed on the screen will be Menu 01,Data Transfer.
3. Click the Sensitivity Control Dial to enter. You Can output data from either the Serial Port or from the CF card (Note: you
will only transfer txt data, not .wav files to Ultratrend DMS)
4. Spin to SERIAL OUTPUT. SERIAL Output will blink.
5. Connect the RS 232 cable to the computer and to the Ultraprobe
6. Open Ultratrend DMS
7. Select Communication
8. Configure the Port
9. Test the Connection
10.Select Download Group and all the data will be transferred to the PC. (NOTE: For software management, refer to Ultratrend
DMSTm Instructions.)
Table of contents
Other UE Systems Analytical Instrument manuals
Popular Analytical Instrument manuals by other brands

DHC
DHC BT101 TEST PROCEDURES / OPERATING INSTRUCTIONS

LaserLiner
LaserLiner LaserRange-Master T2 manual

Senix
Senix ULTRA-30 Series quick start guide

Insight Vision
Insight Vision IV2 owner's manual

Smith-Root
Smith-Root eDNA Citizen Scientist Assembly and use instructions

Bruker
Bruker M1 MISTRAL user manual