4. WARNING: Use the supplied UE Systems recharger (BCH-2) only. Use of unauthorized rechargers will void the warranty and may
degrade or damage the battery.
WHEN TO RECHARGE: When the red low level indicator light goes on, IMMEDIATELY recharge the Ultraprobe. If the instrument
is not used for a week or more, recharge it. If the Ultraprobe is not used for a few days, it can be used without recharging, however, for
best results, it is advisable to recharge it as a "booster" for about an hour before using.
TO CHARGE THE Ultraprobe 2000 WTG-1
1. Insert Tone Generator plug (yellow) into Warble Tone Generator Recharge jack and then plug the recharger into a wall receptacle.
2. Make sure that the LED on the charger is blinking when recharging.
3. The LED remains solid when the battery is charged. The instrument may stay connected to the charger without damaging the
battery. Charge time is approximately 4 hours.
4. WARNING: Use the supplied UE Systems recharger (BCH-2) only. Use of unauthorized rechargers will void the warranty and may
degrade or damage the battery.
2. TRISONIC` SCANNING MODULE
This module is utilized to receive air-borne ultrasound such as the ultrasounds emitted by pressure leaks and electrical discharges.
There are three prongs at the rear of the module. For placement, align the prongs with the three corresponding jacks in the front end of
the metered pistol housing and plug in.
The TrisonicTM Scanning Module has a phased array of three piezoelectric transducers to pick up the air-borne ultrasound. This phased
array focuses the ultrasound on one "hot spot" for directionality and effectively intensifies the signal so that minute ultrasonic
emissions may be detected.
To use the Trisonic Scanning Module:
1. Plug in to front end.
2. Select the LOG position on the meter selection dial.
3. For general use position the frequency selection dial to the "fixed-band" mode.
4. Start with the sensitivity selection dial at maximum (10).
5. Start to scan the test area. a. The method of air borne detection is to go from the "gross to the fine". If there is too much ultrasound
in the area, reduce the sensitivity, place the RUBBER FOCUSING PROBE (described below) over the scanning module and
proceed to follow the test sound to its' loudest point constantly reducing the sensitivity and following the meter. RUBBER
FOCUSING PROBE: The Rubber Focusing Probe is a cone-shaped rubber shield. It is used to block out stray ultrasound and to
assist in narrowing the field of reception of the Trisonic- Scanning Module. To use, simply slip it over the front of the scanning
module or the contact module. NOTE: To prevent damage to the module plugs, always remove the module BEFORE attaching and
removing the Rubber focusing Probe.
Trisonic Scanning Module
3. CONTACT (STETHOSCOPE) MODULE
This is the module with the metal rod. This rod is utilized as a "waveguide" in that it is sensitive to ultrasound that is generated
internally such as within a pipe, bearing housing, steam trap or wall. Once stimulated by ultrasound, it transfers the signal to a
piezoelectric transducer located directly in the module housing. This module is shielded to provide protection from stray RF waves
that have a tendency to effect electronic receiving. It is equipped with low noise amplification to allow for a clear, intelligible signal to
be received and interpreted.