manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ulrich medical
  6. •
  7. Medical Equipment
  8. •
  9. ulrich medical golden gate CS 3715 Configuration guide

ulrich medical golden gate CS 3715 Configuration guide

Other ulrich medical Medical Equipment manuals

ulrich medical Obelisc CS 2931 Series Configuration guide

ulrich medical

ulrich medical Obelisc CS 2931 Series Configuration guide

ulrich medical Pezo PEEK cage Series Configuration guide

ulrich medical

ulrich medical Pezo PEEK cage Series Configuration guide

ulrich medical uCentrum User manual

ulrich medical

ulrich medical uCentrum User manual

ulrich medical neon3 CS 3921 Series Configuration guide

ulrich medical

ulrich medical neon3 CS 3921 Series Configuration guide

ulrich medical tezo titanium cage Series User manual

ulrich medical

ulrich medical tezo titanium cage Series User manual

ulrich medical ADDplus anterior distraction device plus User manual

ulrich medical

ulrich medical ADDplus anterior distraction device plus User manual

ulrich medical tezo CS 3460 Series Configuration guide

ulrich medical

ulrich medical tezo CS 3460 Series Configuration guide

ulrich medical Obelisc User manual

ulrich medical

ulrich medical Obelisc User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

CS 3715-01
Einsetzinstrument
Expansion instrument
CS 3715-02
Halter für Distraktionsinstrument
Holder
CS 3715-03
Distrahierstab
Expansion rod
CS 3715-04
Drehknopf
Turning knob
CS 3715-05
Fixierstab
Locking rod
1
2
3
4
5
golden gate™anterior plate system
Montage- und Demontageanleitung mit speziellen Reinigungshinweisen
Assembly and disassembly instructions with special cleaning instructions
Einsetzinstrument
Expansion instrument
Seite 1 I Page 1
CS 3715
WS 3785-01 R1/2016-01
Montage des
Distraktionsinstrumentes
Der Distrahierstab (CS 3715-
03) wird in den Halter (CS
3715-02) gesteckt.
Assembling the
expansion instrument
Slide the expansion rod (CS
3715-03) into the holder (CS
3715-02).
Montageanleitung l Assembly instructions
ulrich GmbH & Co. KG
Buchbrunnenweg 12 l 89081 Ulm l Germany
Telefon/Phone: +49 (0)731 9654-0
Fax national/international: +49 (0)731 9654-2705/2805
[email protected] l www.ulrichmedical.com
Bitte stellen Sie sicher, dass Sie stets die aktuellste systembezogene „OP-Technik“ und UH 1100 „Aufbereitungshandbuch Implantate
und Instrumente“ verwenden und beachten. Diese können Sie jederzeit kostenfrei herunterladen unter: www.ifu.ulrichmedical.com
Please ensure that you are using and observing the most current “surgical technique” of this system and UH 1100 „Processing
manual implants and instruments“ at all times. These can be downloaded any time free of charge at: www.ifu.ulrichmedical.com
golden gate™anterior plate system
Montage- und Demontageanleitung mit speziellen Reinigungshinweisen
Assembly and disassembly instructions with special cleaning instructions
Einsetzinstrument
Expansion instrument
Seite 2 I Page 2
CS 3715
WS 3785-01 R1/2016-01
Montage des
Distraktionsinstrumentes
Anschließend wird der
Drehknopf (CS 3715-04)
hinten auf den Distrahierstab
aufgesteckt. Dabei sollte
der Distrahierstab von vorne
gehalten werden.
Der Fixierstab (CS 3715-05)
wird in den Distrahierstab
eingesetzt.
Das Distraktionselement
(CS 3715-01) wird auf
die Führungsschiene des
Halters aufgesteckt. Die
Zähne des Distrahierstabes
greifen in die Kerben des
Distraktionselements. Durch
Drehen des Fixierstabs wird
das Distraktionselement mit
dem Halter fixiert. Durch
Drehen des Drehknopfes
wird das Distraktionselement
distrahiert bzw. komprimiert.
Vor Gebrauch muss der
Distraktionsmechanismus
auf einwandfreie und leichte
Funktion überprüft werden.
Assembling the
expansion instrument
Put the turning knob (CS
3715-04) on the back of the
expansion rod while hold-
ing the expansion rod in the
front.
Then insert the locking rod
(CS 3715-05) into the expan-
sion rod.
Slide the expansion element
(CS 3715-01) on to the
guiding bar of the holder.
The teeth of the expansion
rod grab the notches of the
expansion element. Turn the
locking rod to fix the expan-
sion element to the holder.
Turning the turning knob
causes the distraction and
accordingly compression of
the expansion element. Check
the expansion mechanism
for easy and proper function
before using.
Montageanleitung l Assembly instructions
ulrich GmbH & Co. KG
Buchbrunnenweg 12 l 89081 Ulm l Germany
Telefon/Phone: +49 (0)731 9654-0
Fax national/international: +49 (0)731 9654-2705/2805
[email protected] l www.ulrichmedical.com
Bitte stellen Sie sicher, dass Sie stets die aktuellste systembezogene „OP-Technik“ und UH 1100 „Aufbereitungshandbuch Implantate
und Instrumente“ verwenden und beachten. Diese können Sie jederzeit kostenfrei herunterladen unter: www.ifu.ulrichmedical.com
Please ensure that you are using and observing the most current “surgical technique” of this system and UH 1100 „Processing
manual implants and instruments“ at all times. These can be downloaded any time free of charge at: www.ifu.ulrichmedical.com
Montage- und Demontageanleitung mit speziellen Reinigungshinweisen
Assembly and disassembly instructions with special cleaning instructions
Seite 3 I Page 3
Zur Reinigung und Lagerung
auf dem Sieb muss das
Distraktionsinstrument
demontiert werden.
Demontage des
Distraktionsinstruments
Durch Drehen des Fixierstabs
(CS 3715-05) wird das Distrak-
tionselement (CS 3715-01)
gelöst und vom Halter (CS
3715-01) entfernt. Der Fixier-
stab wird nach hinten aus
dem Halter gezogen.
Der Drehknopf (CS 3715-04)
wird zusammen mit dem
Distrahierstab (CS 3715-03)
nach hinten aus dem Halter
gezogen. Anschließend wird
der Drehknopf vom Distra-
hierstab abgezogen.
Demontage der Aufätze
für Distraktionselement
Falls auf dem Distraktionsele-
ment zur Realisierung einer
größeren Distraktionshöhe
ein Aufsatz (CS 3715-07 bzw.
CS 3715-08) montiert wurde,
muss dieser vor der Reinigung
mit dem Schraubendreher-
Schaft (CS 3712-01) entfernt
werden.
The inserter must be disas-
sem-bled for cleaning and
storage on the tray.
Disassembling the expan-
sion instrument
Turn the locking rod (CS
3715-05) to release the
expansion element (CS 3715-
01) and remove it from the
holder (CS 3715-01). Pull the
locking rod out of the holder.
The turning knob (CS 3715-
04) and the expansion rod
(CS 3715-03) will be removed
back-words of the holder.
Afterwards the turning knob
will be disassembled from the
expansion rod.
Disassembling the
extension block
If an extension block (CS
3715-07 or CS 3715-08) was
assembled to realize a greater
expansion height it has to be
removed by using the Shaft
for screwdriver (CS 3712-01)
before cleaning.
Demontageanleitung l Disassembly instructions
ulrich GmbH & Co. KG
Buchbrunnenweg 12 l 89081 Ulm l Germany
Telefon/Phone: +49 (0)731 9654-0
Fax national/international: +49 (0)731 9654-2705/2805
[email protected] l www.ulrichmedical.com
Bitte stellen Sie sicher, dass Sie stets die aktuellste systembezogene „OP-Technik“ und UH 1100 „Aufbereitungshandbuch Implantate
und Instrumente“ verwenden und beachten. Diese können Sie jederzeit kostenfrei herunterladen unter: www.ifu.ulrichmedical.com
Please ensure that you are using and observing the most current “surgical technique” of this system and UH 1100 „Processing
manual implants and instruments“ at all times. These can be downloaded any time free of charge at: www.ifu.ulrichmedical.com
golden gate™anterior plate system
Montage- und Demontageanleitung mit speziellen Reinigungshinweisen
Assembly and disassembly instructions with special cleaning instructions
Einsetzinstrument
Expansion instrument
CS 3715
WS 3785-01 R1/2016-01
Montage- und Demontageanleitung mit speziellen Reinigungshinweisen
Assembly and disassembly instructions with special cleaning instructions
Seite 4 I Page 4
Manuelle Reinigung
Alle Lumen und Hohlräume
müssen sorgfältig und voll-
ständig gespült werden.
Visuelle Kontrolle
Zum Abschluss muss geprüft
werden, ob sichtbare Ver-
schmutzungen vorhanden
sind. In diesem Fall müssen
die vorhergehenden Schritte
wiederholt werden.
Pflege des
Distraktionsinstrumentes
Zur Pflege sollten bewegliche
Teile wie Gelenke, Scharniere
und Gewindeteile mit geeigne-
tem Pflegeöl behandelt werden.
Manual cleaning
All lumens and cavities must
be rinsed out carefully and
thoroughly.
Visual inspection
Finally, the instrument must
be checked for any visible
contamination. The steps
described above must be
repeated if contamination is
found.
Caring for the expansion
instrument
For proper maintenance, treat
moving parts (such as joints,
hinges and threaded elements)
with a suitable oil.
Spezielle Reinigungshinweise l Special cleaning instructions
ulrich GmbH & Co. KG
Buchbrunnenweg 12 l 89081 Ulm l Germany
Telefon/Phone: +49 (0)731 9654-0
Fax national/international: +49 (0)731 9654-2705/2805
[email protected] l www.ulrichmedical.com
Bitte stellen Sie sicher, dass Sie stets die aktuellste systembezogene „OP-Technik“ und UH 1100 „Aufbereitungshandbuch Implantate
und Instrumente“ verwenden und beachten. Diese können Sie jederzeit kostenfrei herunterladen unter: www.ifu.ulrichmedical.com
Please ensure that you are using and observing the most current “surgical technique” of this system and UH 1100 „Processing
manual implants and instruments“ at all times. These can be downloaded any time free of charge at: www.ifu.ulrichmedical.com
golden gate™anterior plate system
Montage- und Demontageanleitung mit speziellen Reinigungshinweisen
Assembly and disassembly instructions with special cleaning instructions
Einsetzinstrument
Expansion instrument
CS 3715