manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. VEE GEE
  6. •
  7. Measuring Instrument
  8. •
  9. VEE GEE BX-1 User manual

VEE GEE BX-1 User manual

Range: Brix 0.0-32.0%
Resolution: 0.2%
Accuracy: ±0.2%
Dimensions: 40 x 40 x 165mm (1.6 x 1.6 x 6.5”)
Weight: 240g (8.5 oz.)
Supplied With: Vinyl Carrying Case (1), Plastic Transfer Pipet (1)
Printed in China VGMNL010320-43001
Operation Manual
Model BX-1
Brix 0-32%
Specifications
Cat. No. 43001
Temperature Correction Table
For Refractometers Calibrated At 20°C (l=589nm)
Source: ICUMSA, 1974
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following tempera-
ture correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at
exactly 20°C, follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if
ambient temperature changes, simply apply the temperature correction values listed below. For example, a reading
of Brix 40.0% at 25°C is corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.
Warranty information and
registration form can be found at:
www.veegee.com/service_support
Refractometer Components
Calibration Ring
Mechanism used to zero or calibrate the refractometer,
without the requirement of additional tools.
Rubber Hood
Houses and protects the focusable lens. Prevents
light from entering through the eyepiece during use.
Rubber Grip
Insulates the refractometer
against hand heat for accurate
results.
Non-Roll Stand
Protects against damage to the
instrument when set down between
measurements.
Daylight Plate & Prism
The prism and sample are covered by
the daylight plate during readings.
This refractometer is an optical instrument -- it can become damaged if dropped or handled in a rough manner.
The prism is made of optical glass and is susceptible to scratches -- do not apply any rough or abrasive
material and take care when cleaning the prism.
After each use, clean the prism surface and daylight plate with a soft cloth or tissue soaked in water and
wipe off with a dry cloth or tissue.
Do not hold the refractometer under a stream of water from a faucet. Do not splash it with or dip it in water.
If the surface of the prism becomes coated with an oily solution or similar, it will repel test samples and
affect readings. If this occurs, the prism should be cleaned with a weakened detergent or similar solvent.
2 3
1
2
5
4
3
Calibration
Precautions
Calibration should be conducted at the start of each day or when any
shifts in ambient temperature occur. If recalibration is impractical, refer
to the directions and temperature correction table on the following page.
For standard calibration procedures, please follow the directions below.
Open the daylight plate and apply one or two drops of distilled water on to
the surface of the prism. Hold the prism at an angle close to parallel with
the oor so the distilled water will not run off of the prism.
Gently close the daylight plate over the prism. The distilled water should
spread as a thin, even layer in between the daylight plate and the prism.
By looking through the daylight plate, ensure that the distilled water covers
the ENTIRE surface of the prism. If there are bubbles and gaps or if the
distilled water is only on one portion of the prism, the distilled water must
be reapplied (Figure 1). Inaccurate calibrations will result if the prism is
not covered correctly.
Looking through the eyepiece, hold the refractometer and direct the daylight
plate upwards towards light. If the scale is not in focus, adjust it by gently
turning the eyepiece (rubber hood) either clockwise or counterclockwise.
Be careful not to overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper
eld of view will be seen as blue and the lower eld will be seen as white
(Figure 2). Conrm that the boundary line crosses the scale at “0” (Figure 3).
If the boundary line falls above or below zero, gently loosen the screw on the
calibrationring.Whilelookingthroughtheeyepiece,gently turnthecalibration
ring clockwise or counterclockwise until the boundary line is at zero. Once
this is achieved gently tighten down the set screw. (NOTE: Do not over-tight-
en. If the set screw is over-tightened, the boundary line may shift slightly).
When calibration is complete, gently wipe the prism using tissue paper.
Thank you for purchasing this VEE GEE Refractometer. With the user in mind, VEE GEE Refractometers are built
from modern designs and, with proper care, this instrument should provide many years of reliable performance.
It’s recommended this manual is read entirely before using the refractometer for the rst time.
Introduction
Figure 1
Figure 2
Figure 3
Figure 4
Figure 3
1
2
5
4
3
General Use
Open the daylight plate and apply one or two drops of the sample solution
to the surface of the prism. Hold the prism at an angle close to parallel
with the oor so the sample will not run off of the prism.
Gently close the daylight plate over the prism. The sample solution should
spread as a thin, even layer in between the daylight plate and the prism.
By looking through the daylight plate, ensure that the sample solution
covers the ENTIRE surface of the prism. If there are bubbles and gaps or if
the sample is only on one portion of the prism, the sample solution must
be reapplied (Figure 1). Inaccurate readings will result if the prism is not
covered correctly.
Looking through the eyepiece, hold the refractometer and direct the daylight
plate upwards towards light. If the scale is not in focus, adjust it by gently
turning the eyepiece (rubber hood) either clockwise or counterclockwise.
Be careful not to overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper
eld of view will be seen as blue and the lower eld will be seen as white
(Figure 4). The reading is taken at the point where the boundary line of the
blue and white elds crosses the scale (Figure 5). The value is the Brix%
reading of the sample.
When each measurement is complete, the sample must be cleaned from
the prism using tissue paper and water.
Calibration Ring Set Screw
Secures the calibration ring in place to
protect from any shifts which could occur
during use.
Figure 5
Figure 3
6
7

Other VEE GEE Measuring Instrument manuals

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE 43045 User manual

VEE GEE

VEE GEE 43045 User manual

VEE GEE SGX-1 User manual

VEE GEE

VEE GEE SGX-1 User manual

VEE GEE C10 User manual

VEE GEE

VEE GEE C10 User manual

VEE GEE BX-4 User manual

VEE GEE

VEE GEE BX-4 User manual

VEE GEE CTX-1 User manual

VEE GEE

VEE GEE CTX-1 User manual

VEE GEE BTX-20 User manual

VEE GEE

VEE GEE BTX-20 User manual

VEE GEE BX-2 User manual

VEE GEE

VEE GEE BX-2 User manual

VEE GEE NDX-1 User manual

VEE GEE

VEE GEE NDX-1 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE STX-3 User manual

VEE GEE

VEE GEE STX-3 User manual

VEE GEE C25L User manual

VEE GEE

VEE GEE C25L User manual

VEE GEE CTX-2 User manual

VEE GEE

VEE GEE CTX-2 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-3 User manual

VEE GEE

VEE GEE BX-3 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE 43035 User manual

VEE GEE

VEE GEE 43035 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-10 User manual

VEE GEE

VEE GEE BX-10 User manual

VEE GEE SX-2 User manual

VEE GEE

VEE GEE SX-2 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

Popular Measuring Instrument manuals by other brands

OWL KIT-FO4B-WSMDst Quick reference guide

OWL

OWL KIT-FO4B-WSMDst Quick reference guide

multicomp pro MP001186 instruction manual

multicomp pro

multicomp pro MP001186 instruction manual

KROHNE OPTISWITCH 3000 C operating instructions

KROHNE

KROHNE OPTISWITCH 3000 C operating instructions

Hanna Instruments HI 93717 instruction manual

Hanna Instruments

Hanna Instruments HI 93717 instruction manual

Stefan Mayer Instruments AM502 Series instruction manual

Stefan Mayer Instruments

Stefan Mayer Instruments AM502 Series instruction manual

Freedom R8100 SERIES Operator's manual

Freedom

Freedom R8100 SERIES Operator's manual

Brandson 302735/20170214ZH006 user manual

Brandson

Brandson 302735/20170214ZH006 user manual

East Tester ET430 Instructions for use

East Tester

East Tester ET430 Instructions for use

PCB Piezotronics 3743F1110G Installation and operating manual

PCB Piezotronics

PCB Piezotronics 3743F1110G Installation and operating manual

Keller CellaTemp PA 1 Series quick guide

Keller

Keller CellaTemp PA 1 Series quick guide

Kobold DSS Series Operating	 instruction

Kobold

Kobold DSS Series Operating instruction

Vega VEGAPULS 66 operating instructions

Vega

Vega VEGAPULS 66 operating instructions

Beta Screen PRESTO SPECTRO user manual

Beta Screen

Beta Screen PRESTO SPECTRO user manual

CARLO GAVAZZI WM22-DIN manual

CARLO GAVAZZI

CARLO GAVAZZI WM22-DIN manual

KaWe 12.20600.001 user manual

KaWe

KaWe 12.20600.001 user manual

Virone EM-1 manual

Virone

Virone EM-1 manual

Hydreka 801 Operation manual

Hydreka

Hydreka 801 Operation manual

Evoqua V10K instructions

Evoqua

Evoqua V10K instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.