manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. VEE GEE
  6. •
  7. Measuring Instrument
  8. •
  9. VEE GEE BX-20 User manual

VEE GEE BX-20 User manual

Range: Brix 0.0-20.0%
Resolution: 0.1%
Accuracy: ±0.1%
Dimensions: 40 x 40 x 185mm (1.6 x 1.6 x 7.3”)
Weight: 285g (10.0 oz.)
Supplied With: VinylCarryingCase(1),PlasticTransferPipet(1)
Printed in China VGMNL010320-43011
Operation Manual
Model BX-20
Brix 0-20%
Specifications
Cat. No. 43011
Temperature Correction Table
For Refractometers Calibrated At 20°C (l=589nm)
Source: ICUMSA, 1974
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following temperature
correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at exactly 20°C,
follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if ambient temperature
changes, simply apply the temperature correction values listed below. For example, a reading of Brix 40.0% at 25°C is
corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.
Warranty information and
registration form can be found at:
www.veegee.com/service_support
Refractometer Components
Calibration Ring
Mechanismusedtozeroorcalibratetherefractometer,
withouttherequirementofadditional tools.
Rubber Hood
Houses and protects the focusable lens. Prevents
lightfrom entering through the eyepieceduringuse.
Rubber Grip
Insulatestherefractometeragainsthand
heatfor accurate results.
Non-Roll Stand
Protectsagainstdamagetotheinstrumentwhen
setdown betweenmeasurements.
Daylight Plate & Prism
The prism and sample are covered by the
daylightplate during readings.
This refractometer is an optical instrument -- it can become damaged if dropped or handled in a rough manner.
Theprism ismadeof opticalglass and issusceptible toscratches-- donot applyanyrough orabrasivematerial andtake
care when cleaning the prism.
After each use, clean the prism surface and daylight plate with a soft cloth or tissue soaked in water and wipe off with
a dry cloth or tissue.
Do not hold the refractometer under a stream of water from a faucet. Do not splash it with or dip it in water.
If the surface of the prism becomes coated with an oily solution or similar, it will repel test samples and affect readings.
If this occurs, the prism should be cleaned with a weakened detergent or similar solvent.
2 3
1
2
55
55
5
44
44
4
33
33
3
Calibration
Precautions
Calibration should be conducted at the start of each day or when any shifts
in ambient temperature occur. If recalibration is impractical, refer to the
directions and temperature correction table on the following page.
For standard calibration procedures, please follow the directions below.
Openthe daylightplate andapply oneor two drops of distilledwater onto the
surfaceof theprism. Hold theprismat anangle closeto parallel withthe floor
so the distilled water will not run off of the prism.
Gently close the daylight plate over the prism. The distilled water should
spread as a thin, even layer in between the daylight plate and the prism. By
looking through the daylight plate, ensure that the distilled water covers the
ENTIRE surface of the prism. If there are bubbles and gaps or if the distilled
wateris onlyonone portionof theprism, thedistilledwater mustbereapplied
(Figure1). Inaccuratecalibrationswill resultiftheprism is notcoveredcorrectly.
Looking through the eyepiece, hold the refractometer and direct the daylight
plate upwards towards light. If the scale is not in focus, adjust it by gently
turning the eyepiece (rubber hood) either clockwise or counterclockwise. Be
careful not to overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper field
ofview will beseen as blueand the lowerfield willbe seen aswhite (Figure 2).
Confirm that the boundary line crosses the scale at “0” (Figure 3).
Ifthe boundaryline fallsaboveorbelow zero,gentlyloosen the setscrewon the
calibrationring. Whilelooking throughtheeyepiece, gentlyturn thecalibration
ringclockwise orcounterclockwiseuntil theboundaryline isat zero. Oncethis
is achieved gently tighten down the set screw. (NOTE: Do not over-tighten. If
the set screw is over-tightened, the boundary line may shift slightly).
When calibration is complete, gently wipe the prism using tissue paper.
Thank you for purchasing this VEE GEE Refractometer. With the user in mind, VEE GEE Refractometers are built from modern
designs and, with proper care, this instrument should provide many years of reliable performance. It’s recommended this
manual is read entirely before using the refratometer for the first time.
Introduction
Figure 1
Figure 2
Figure 3
Figure 4
Figure 3
1
2
55
55
5
44
44
4
33
33
3
General Use
Open the daylight plate and apply one or two drops of the sample solution to
the surface of the prism. Hold the prism at an angle close to parallel with the
floor so the sample will not run off of the prism.
Gently close the daylight plate over the prism. The sample solution should
spread as a thin, even layer in between the daylight plate and the prism. By
lookingthrough thedaylightplate, ensurethat thesamplesolution coversthe
ENTIRE surface of the prism. If there are bubbles and gaps or if the sample
is only on one portion of the prism, the sample solution must be reapplied
(Figure 1). Inaccurate readings will result if the prism is not covered correctly.
Looking through the eyepiece, hold the refractometer and direct the daylight
plate upwards towards light. If the scale is not in focus, adjust it by gently
turning the eyepiece (rubber hood) either clockwise or counterclockwise. Be
careful not to overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper field
ofview will beseen as blueand the lowerfield willbe seen aswhite (Figure 4).
Thereading istakenat thepoint where theboundary lineofthe blueand white
fields crosses the scale (Figure 5). The value is the Brix% reading of the
sample.
When each measurement is complete, the sample must be cleaned from
the prism using tissue paper and water.
Calibration Ring Set Screw
Securesthecalibrationringin place toprotect
fromanyshiftswhichcouldoccur during use.
Figure 5
Figure 3
66
66
6
77
77
7

Other VEE GEE Measuring Instrument manuals

VEE GEE CTX-1 User manual

VEE GEE

VEE GEE CTX-1 User manual

VEE GEE SX-2 User manual

VEE GEE

VEE GEE SX-2 User manual

VEE GEE BX-4 User manual

VEE GEE

VEE GEE BX-4 User manual

VEE GEE BTX-20 User manual

VEE GEE

VEE GEE BTX-20 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE 43035 User manual

VEE GEE

VEE GEE 43035 User manual

VEE GEE SGX-1 User manual

VEE GEE

VEE GEE SGX-1 User manual

VEE GEE C10 User manual

VEE GEE

VEE GEE C10 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-1 User manual

VEE GEE

VEE GEE BX-1 User manual

VEE GEE STX-3 User manual

VEE GEE

VEE GEE STX-3 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE BX-3 User manual

VEE GEE

VEE GEE BX-3 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE NDX-1 User manual

VEE GEE

VEE GEE NDX-1 User manual

VEE GEE CTX-2 User manual

VEE GEE

VEE GEE CTX-2 User manual

VEE GEE C25L User manual

VEE GEE

VEE GEE C25L User manual

VEE GEE 43045 User manual

VEE GEE

VEE GEE 43045 User manual

VEE GEE BX-2 User manual

VEE GEE

VEE GEE BX-2 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-10 User manual

VEE GEE

VEE GEE BX-10 User manual

Popular Measuring Instrument manuals by other brands

Power Fist 8714941 user manual

Power Fist

Power Fist 8714941 user manual

Axio MET AX-355 instruction manual

Axio MET

Axio MET AX-355 instruction manual

Promax RANGER Neo 2 user manual

Promax

Promax RANGER Neo 2 user manual

Pro-tec PLM-8000 Operation manual

Pro-tec

Pro-tec PLM-8000 Operation manual

AquaLab Series 3 Operator's manual

AquaLab

AquaLab Series 3 Operator's manual

Applied Instruments Super Buddy Signal Level Meter Application note

Applied Instruments

Applied Instruments Super Buddy Signal Level Meter Application note

Aqualytic OxiDirect BOD-System instruction manual

Aqualytic

Aqualytic OxiDirect BOD-System instruction manual

Electronics International Inc SR-8A Operating and installation instructions

Electronics International Inc

Electronics International Inc SR-8A Operating and installation instructions

Delta LTL3000 user manual

Delta

Delta LTL3000 user manual

Air TEc SUTO S600 Instruction and operation manual

Air TEc

Air TEc SUTO S600 Instruction and operation manual

Unit UT311 operating manual

Unit

Unit UT311 operating manual

vanEssen TD-Diver DI8 Series product manual

vanEssen

vanEssen TD-Diver DI8 Series product manual

Promax OTDR Series user manual

Promax

Promax OTDR Series user manual

AutoMeter BCT-460 Controls and functions

AutoMeter

AutoMeter BCT-460 Controls and functions

Edgetech COM.AIR Operator's manual

Edgetech

Edgetech COM.AIR Operator's manual

Abus TVVR33600 quick guide

Abus

Abus TVVR33600 quick guide

Check-line DTS-200 operating manual

Check-line

Check-line DTS-200 operating manual

bf1 systems 1SYSTEM TPMS user guide

bf1 systems

bf1 systems 1SYSTEM TPMS user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.