manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. VEE GEE
  6. •
  7. Measuring Instrument
  8. •
  9. VEE GEE BX-2 User manual

VEE GEE BX-2 User manual

Range:
Resolution:
Accuracy:
Dimensions:
Weight:
Supplied With:
Optional Oil Standards:
Brix 28.0-62.0%
0.2%
±0.2%
40 x 40 x 150mm (1.6 x 1.6 x 5.9”)
238g (8.4 oz.)
VinylCarryingCase(1),PlasticTransferPipet(1)
Brix 60.0%, ¼ oz. Bottle (cat. no. 45000-OS025)
Printed in China VGMNL010320-43003
Operation Manual
Model BX-2
Brix 28-62%
Specifications
Cat. No. 43003
Temperature Correction Table
For Refractometers Calibrated At 20°C (l=589nm)
Source: ICUMSA, 1974
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following temperature
correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at exactly 20°C,
follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if ambient temperature
changes, simply apply the temperature correction values listed below. For example, a reading of Brix 40.0% at 25°C is
corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.
Warranty information and
registration form can be found at:
www.veegee.com/service_support
Refractometer Components
Calibration Ring
Mechanismusedtozeroor calibrate therefractometer,
withouttherequirementof additionaltools.
Rubber Hood
Houses and protects the focusable lens. Prevents
lightfromenteringthroughtheeyepieceduringuse.
Rubber Grip
Insulatestherefractometeragainsthand
heatforaccurateresults.
Non-Roll Stand
Protectsagainstdamagetotheinstrumentwhen
setdownbetweenmeasurements.
Daylight Plate & Prism
The prism and sample are covered by the
daylightplate during readings.
This refractometer is an optical instrument -- it can become damaged if dropped or handled in a rough manner.
Theprism ismadeof opticalglass and issusceptible toscratches-- donot applyanyrough orabrasivematerial andtake
care when cleaning the prism.
After each use, clean the prism surface and daylight plate with a soft cloth or tissue soaked in water and wipe off with
a dry cloth or tissue.
Do not hold the refractometer under a stream of water from a faucet. Do not splash it with or dip it in water.
If the surface of the prism becomes coated with an oily solution or similar, it will repel test samples and affect readings.
If this occurs, the prism should be cleaned with a weakened detergent or similar solvent.
2 3
1
2
5
4
3
Calibration
Precautions
Calibration should be conducted at the start of each day or when any shifts in
ambient temperature occur. If recalibration is impractical, refer to the directions
and temperature correction table on the following page. For standard calibration
procedures, please follow the directions below.
Open the daylight plate and apply one or two drops of a liquid standard on to
the surface of the prism [a Brix 50.0% oil standard is used for this example.
A prepared sucrose solution (weight of sucrose vs. distilled water) of a
known concentration (minimum Brix 28.0%) may also be used for
calibration purposes]. Hold the prism at an angle close to parallel with the floor
so the liquid standard will not run off of the prism.
Gently close the daylight plate over the prism. The liquid standard should spread
as a thin, even layer in between the daylight plate and the prism. By looking
through the daylight plate, ensure that the liquid standard covers the ENTIRE
surface of the prism. If there are bubbles and gaps or if the liquid standard is only
on one portion of the prism, the liquid standard must be reapplied (Figure 1).
Inaccurate calibrations will result if the prism is not covered correctly.
Lookingthrough theeyepiece, hold therefractometer anddirect the daylightplate
upwards towards light. If the scale is not in focus, adjust it by gently turning the
eyepiece (rubber hood) either clockwise or counterclockwise. Be careful not to
overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper field of
viewwillbe seen asblueandthe lowerfieldwillbe seen aswhite(Figure2). Confirm
thatthe boundarylinecrosses thescaleat “50”(orthe valueofthe liquidstandard
being used) (Figure 3).
If the boundary line falls above or below “50”, gently loosen the set screw on the
calibration ring. While looking through the eyepiece, gently turn the calibration
ring clockwise or counterclockwise until the boundary line is at “50.” Once this
is achieved gently tighten down the set screw. (NOTE: Do not over-tighten. If the
set screw is over-tightened, the boundary line may shift slightly).
When calibration is complete, gently wipe the prism using tissue paper.
Thank you for purchasing this VEE GEE Refractometer. With the user in mind, VEE GEE Refractometers are built from modern
designs and, with proper care, this instrument should provide many years of reliable performance. It’s recommended this
manual is read entirely before using the refratometer for the first time.
Introduction
1
2
5
4
3
General Use
Open the daylight plate and apply one or two drops of the sample solution to the
surface of the prism. Hold the prism at an angle close to parallel with the floor so
the sample will not run off of the prism.
Gently close the daylight plate over the prism. The sample solution should spread
as a thin, even layer in between the daylight plate and the prism. By looking
through the daylight plate, ensure that the sample solution covers the ENTIRE
surface of the prism. If there are bubbles and gaps or if the sample is only on one
portion ofthe prism, thesample solutionmust bereapplied (Figure 1). Inaccurate
readings will result if the prism is not covered correctly.
Lookingthrough theeyepiece, hold therefractometer anddirect the daylightplate
upwards towards light. If the scale is not in focus, adjust it by gently turning the
eyepiece (rubber hood) either clockwise or counterclockwise. Be careful not to
overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper field of
view will be seen as blue and the lower field will be seen as white (Figure 4). The
reading is taken at the point where the boundary line of the blue and white fields
crosses the scale (Figure 5). The value is the Brix% reading of the sample.
When each measurement is complete, the sample must be cleaned from the
prism using tissue paper and water.
Calibration Ring Set Screw
Securesthecalibrationringin place toprotect
fromanyshiftswhichcouldoccur during use.
6
7
Figure 1
Figure 2
Figure 3
Figure 4
Figure 3
Figure 5
Figure 3

Other VEE GEE Measuring Instrument manuals

VEE GEE C10 User manual

VEE GEE

VEE GEE C10 User manual

VEE GEE 43045 User manual

VEE GEE

VEE GEE 43045 User manual

VEE GEE SGX-1 User manual

VEE GEE

VEE GEE SGX-1 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE CTX-2 User manual

VEE GEE

VEE GEE CTX-2 User manual

VEE GEE BX-3 User manual

VEE GEE

VEE GEE BX-3 User manual

VEE GEE CTX-1 User manual

VEE GEE

VEE GEE CTX-1 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE 43035 User manual

VEE GEE

VEE GEE 43035 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-1 User manual

VEE GEE

VEE GEE BX-1 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE C25L User manual

VEE GEE

VEE GEE C25L User manual

VEE GEE NDX-1 User manual

VEE GEE

VEE GEE NDX-1 User manual

VEE GEE STX-3 User manual

VEE GEE

VEE GEE STX-3 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE BX-10 User manual

VEE GEE

VEE GEE BX-10 User manual

VEE GEE SX-2 User manual

VEE GEE

VEE GEE SX-2 User manual

VEE GEE BTX-20 User manual

VEE GEE

VEE GEE BTX-20 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE BX-4 User manual

VEE GEE

VEE GEE BX-4 User manual

Popular Measuring Instrument manuals by other brands

Wet Labs ECO FL user guide

Wet Labs

Wet Labs ECO FL user guide

Research Instruments HandiLaz Mini quick start guide

Research Instruments

Research Instruments HandiLaz Mini quick start guide

YATO YT-7307 quick start guide

YATO

YATO YT-7307 quick start guide

S+S Regeltechnik PREMASGARD 814 Series Operating and mounting instructions

S+S Regeltechnik

S+S Regeltechnik PREMASGARD 814 Series Operating and mounting instructions

Trotec T500 operating manual

Trotec

Trotec T500 operating manual

vinmetrica SC-100A user manual

vinmetrica

vinmetrica SC-100A user manual

IMV Easi-Scan: Go instruction manual

IMV

IMV Easi-Scan: Go instruction manual

Jet 310605 operating instructions

Jet

Jet 310605 operating instructions

Factair F3000 operating instructions

Factair

Factair F3000 operating instructions

AST TI-1800 user manual

AST

AST TI-1800 user manual

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 user guide

Agena BIOSCIENCE

Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 user guide

SSE Sharky 773 user guide

SSE

SSE Sharky 773 user guide

Muller Ziegler EZD manual

Muller Ziegler

Muller Ziegler EZD manual

Quarton ML-100 instructions

Quarton

Quarton ML-100 instructions

ATAGO 4440 instruction manual

ATAGO

ATAGO 4440 instruction manual

Brodie BiRotor Plus B27X Installation & operation manual

Brodie

Brodie BiRotor Plus B27X Installation & operation manual

Hanna Instruments HI 9811-5 instruction manual

Hanna Instruments

Hanna Instruments HI 9811-5 instruction manual

KPS KPS-PA30 instruction manual

KPS

KPS KPS-PA30 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.