Scale: FreezingPointofPropyleneGlycol/FreezingPointofEthyleneGlycol/SpecificGravityofBatteryAcid(ChargeLevel)
Range: -50 to 32°F / -60 to 32°F / 1.15-1.40
Resolution: 2.0°F / 2.0°F / 0.01
Accuracy: ±2.0°F / ±2.0°F / ±0.01
ATC Range: 10-30°C (50-86°F)
Dimensions: 40 x 40 x 150mm (1.6 x 1.6 x 6.5”)
Weight: 235g (8.3 oz.)
Supplied With: VinylCarryingCase(1),CalibrationScrewdriver(1)
Printed in China VGMNL010320-43064
Operation Manual
Model CTX-2
Antifreeze / Coolant °F
Specifications
Cat. No. 43064
Automatic Temperature Compensation
Figure 4
Figure 3
1
2
55
55
5
44
44
4
33
33
3
General Use
Open the daylight plate and apply one or two drops of the sample solution to
the surface of the prism. Hold the prism at an angle close to parallel with the
floor so the sample will not run off of the prism.
Gently close the daylight plate over the prism. The sample solution should
spread as a thin, even layer in between the daylight plate and the prism. By
lookingthrough thedaylight plate,ensure thatthe samplesolution coversthe
ENTIRE surface of the prism. If there are bubbles and gaps or if the sample
is only on one portion of the prism, the sample solution must be reapplied
(Figure 1). Inaccuratereadings will result if the prism is not covered correctly.
Looking through the eyepiece, hold the refractometer and direct the daylight
plate upwards towards light. If the scale is not in focus, adjust it by gently
turning the eyepiece (rubber hood) either clockwise or counterclockwise. Be
careful not to overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper field
ofview will be seen asblue and the lower fieldwill be seen as white(Figure 4).
Thereading istaken atthe pointwhere theboundary lineofthe blueand white
fields crosses the scale (Figure 5). The value is Battery Charge Level (LEFT
SCALE)/the FreezingPointofPropylene Glycol(CENTERSCALE) /theFreezing
Point of Ethylene Glycol (RIGHT SCALE) of the sample.
When each measurement is complete, the sample must be cleaned from
the prism using tissue paper and water.
Figure 5
Warranty information and
registration form can be found at:
www.veegee.com/service_support
Refractometer Components
Calibration Ring
Mechanismusedtozeroorcalibratetherefractometer.
Rubber Hood
Houses and protects the focusable lens. Prevents
lightfrom entering through the eyepiece duringuse.
Rubber Grip
Insulatestherefractometeragainsthand
heatforaccurateresults.
Non-Roll Stand
Protectsagainstdamagetotheinstrumentwhen
setdown between measurements.
Daylight Plate & Prism
The prism and sample are covered by the
daylightplate duringreadings.
This refractometer is an optical instrument -- it can become damaged if dropped or handled in a rough manner.
Theprism ismade ofoptical glassand issusceptible toscratches-- donot applyany roughor abrasivematerial andtake
care when cleaning the prism.
After each use, clean the prism surface and daylight plate with a soft cloth or tissue soaked in water and wipe off with
a dry cloth or tissue.
Do not hold the refractometer under a stream of water from a faucet. Do not splash it with or dip it in water.
If the surface of the prism becomes coated with an oily solution or similar, it will repel test samples and affect readings.
If this occurs, the prism should be cleaned with a weakened detergent or similar solvent.
2 3
1
2
55
55
5
44
44
4
33
33
3
Calibration
Precautions
CalibrationofATC refractometersshouldonly beconductedwhen theprevious
calibration setting has shifted and is noticeably affecting measurements.
DO NOT PERFORM CALIBRATIONS IN THE FIELD! Calibration must
take place in a controlled environment of 20°C (68°F) using distilled water
of the same temperature. It’s recommended to allow the refractometer and
the distilled water to reach temperature equilibrium with the controlled
environment before calibration takes place.
Openthe daylight plateand apply oneor two dropsof distilled wateron tothe
surfaceof theprism. Hold theprism atan angle closeto parallelwith thefloor
so the distilled water will not run off of the prism.
Gently close the daylight plate over the prism. The distilled water should
spread as a thin, even layer in between the daylight plate and the prism. By
looking through the daylight plate, ensure that the distilled water covers the
ENTIRE surface of the prism. If there are bubbles and gaps or if the distilled
wateris onlyon oneportion ofthe prism,the distilledwater mustbe reapplied
(Figure1). Inaccuratecalibrationswill resultifthe prismisnotcovered correctly.
Looking through the eyepiece, hold the refractometer and direct the daylight
plate upwards towards light. If the scale is not in focus, adjust it by gently
turning the eyepiece (rubber hood) either clockwise or counterclockwise. Be
careful not to overturn the focusing mechanism.
When the refractometer scale is viewed through the eyepiece, the upper field
ofview will be seen asblue and the lower fieldwill be seen as white(Figure 2).
Confirmthatthe boundaryline crossesthescale at32°F/1.00S.G. (Figure3).
Ifthe boundarylinefallsaboveorbelow32°F/1.00S.G.,gentlyloosenthesetscrew
on the calibration ring with the supplied screwdriver. While looking through
the eyepiece, gently turn the calibration ring clockwise or counterclockwise
untilthe boundaryline isatzero. Once thisis achievedgently tightendown the
setscrew with thesupplied screwdriver. (NOTE:Do not over-tighten. Ifthe set
screw is over-tightened, the boundary line may shift slightly).
When calibration is complete, gently wipe the prism using tissue paper.
Thank you for purchasing this VEE GEE Refractometer. With the user in mind, VEE GEE Refractometers are built from modern
designs and, with proper care, this instrument should provide many years of reliable performance. It’s recommended this
manual is read entirely before using the refratometer for the first time.
Introduction
Figure 1
Figure 2
Figure 3
Figure 3
66
66
6
77
77
7
VEE GEE Model CTX-2 Refractometer is equipped with an Automatic Temperature Compensation (ATC) mechanism.
Refractometers that do not feature such a mechanism must be recalibrated (zeroed) when shifts in ambient temperature
occur, or make use of a temperature correction table. Model CTX-2 automatically compensates for ambient temperature
changes (10-30°C), so recalibration is not required.
With continued use, the calibration setting on ATC refractometers can shift over time . It’s good practice to periodically check
the refractometer by conducting a measurement of distilled water. If the reading is more than one subdivision from zero,
the refractometer should be manually calibrated. Please refer to the Calibration section below.
DO NOT PERFORM CALIBRATIONS IN THE FIELD! Calibrationmusttake place
in a controlled environment of 20°C (68°F) using distilled water of the same
temperature. It’s recommended to allow the refractometer and the distilled water to
reach temperature equilibrium with the controlled environment before calibration
takes place.
Automatic Temperature Compensation

Other VEE GEE Measuring Instrument manuals

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE SX-2 User manual

VEE GEE

VEE GEE SX-2 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE C10 User manual

VEE GEE

VEE GEE C10 User manual

VEE GEE BX-4 User manual

VEE GEE

VEE GEE BX-4 User manual

VEE GEE NDX-1 User manual

VEE GEE

VEE GEE NDX-1 User manual

VEE GEE BX-2 User manual

VEE GEE

VEE GEE BX-2 User manual

VEE GEE CTX-1 User manual

VEE GEE

VEE GEE CTX-1 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BX-10 User manual

VEE GEE

VEE GEE BX-10 User manual

VEE GEE SGX-1 User manual

VEE GEE

VEE GEE SGX-1 User manual

VEE GEE 43035 User manual

VEE GEE

VEE GEE 43035 User manual

VEE GEE BX-3 User manual

VEE GEE

VEE GEE BX-3 User manual

VEE GEE 43045 User manual

VEE GEE

VEE GEE 43045 User manual

VEE GEE C25L User manual

VEE GEE

VEE GEE C25L User manual

VEE GEE BX-1 User manual

VEE GEE

VEE GEE BX-1 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BTX-1 User manual

VEE GEE

VEE GEE BTX-1 User manual

VEE GEE BTX-20 User manual

VEE GEE

VEE GEE BTX-20 User manual

VEE GEE BX-20 User manual

VEE GEE

VEE GEE BX-20 User manual

VEE GEE STX-3 User manual

VEE GEE

VEE GEE STX-3 User manual

Popular Measuring Instrument manuals by other brands

Sibelmed DATOSPIR MICRO A user manual

Sibelmed

Sibelmed DATOSPIR MICRO A user manual

Innovate LM-2 user manual

Innovate

Innovate LM-2 user manual

F&F LE-03d CT400 quick start guide

F&F

F&F LE-03d CT400 quick start guide

Precision Digital Corporation ProVu PD6200 instruction manual

Precision Digital Corporation

Precision Digital Corporation ProVu PD6200 instruction manual

Anritsu MS2722C Maintenance manual

Anritsu

Anritsu MS2722C Maintenance manual

techem IST Series installation instructions

techem

techem IST Series installation instructions

Extech Instruments 380940 user guide

Extech Instruments

Extech Instruments 380940 user guide

horiba APMA-370 Operation manual

horiba

horiba APMA-370 Operation manual

Hanna Instruments HI 83746 instruction manual

Hanna Instruments

Hanna Instruments HI 83746 instruction manual

ATAGO PAL-Ramen Meister instruction manual

ATAGO

ATAGO PAL-Ramen Meister instruction manual

Kiwitron KiwiEye X001010 Installation, use and maintenance manual

Kiwitron

Kiwitron KiwiEye X001010 Installation, use and maintenance manual

Growatt Smart Meter-US installation guide

Growatt

Growatt Smart Meter-US installation guide

LIMIT DDA operating manual

LIMIT

LIMIT DDA operating manual

Dupla pH-Control Pro User instructions

Dupla

Dupla pH-Control Pro User instructions

PCB Piezotronics IMI SENSORS 66103PPZ1 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 66103PPZ1 Installation and operating manual

Ditel DINOS DT Series Quick installation guide

Ditel

Ditel DINOS DT Series Quick installation guide

Somogyi Elektronic Home SMA 2101 instruction manual

Somogyi Elektronic

Somogyi Elektronic Home SMA 2101 instruction manual

LaserLiner SensoLite 410 quick start guide

LaserLiner

LaserLiner SensoLite 410 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.