Vega VEGABAR 81 User manual

Quick setup guide
Pressure transmitter with metallic
measuring cell
VEGABAR 81
Slave sensor for electronic differential pres-
sure
With SIL qualication
Document ID: 48050

2
Contents
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
Contents
1 For your safety
1.1 Authorised personnel ....................................................................................................... 3
1.2 Appropriate use................................................................................................................ 3
1.3 Warning about incorrect use............................................................................................. 3
1.4 General safety instructions ............................................................................................... 3
1.5 CE conformity................................................................................................................... 3
1.6 SIL qualication according to IEC 61508.......................................................................... 4
1.7 Permissible process pressure .......................................................................................... 4
2 Product description
2.1 Conguration.................................................................................................................... 5
3 Mounting
3.1 General instructions for use of the instrument .................................................................. 6
3.2 Ventilation and pressure compensation............................................................................ 6
3.3 Combination Master - Slave ............................................................................................. 6
4 Connecting to power supply
4.1 Connecting....................................................................................................................... 9
4.2 Single chamber housing................................................................................................. 10
4.3 Connection example ...................................................................................................... 11
5 Set up with the display and adjustment module
5.1 Insert display and adjustment module............................................................................ 12
5.2 Parameter adjustment .................................................................................................... 13
6 Supplement
6.1 Technical data ................................................................................................................ 18
Information:
This quick setup guide enables quick setup and commissioning of
your instrument.
You can nd supplementary information in the corresponding, more
detailed Operating Instructions Manual as well as the Safety Manual
that comes with instruments with SIL qualication. These manuals are
available on the supplied DVD or in the download area of "www.vega.
com".
Operating instructions VEGABAR 81 - Slave sensor for elec-
tronic dierential pressure with SIL qualication: Document-ID
48045
Safety Manual VEGABAR series 80 - Two-wire 4 … 20 mA/HART
with SIL qualication: Document-ID 48369
Editing status of the quick setup guide: 2016-02-04

3
1 For your safety
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
1 For your safety
1.1 Authorised personnel
All operations described in this operating instructions manual must
be carried out only by trained specialist personnel authorised by the
plant operator.
During work on and with the device the required personal protective
equipment must always be worn.
1.2 Appropriate use
VEGABAR 81 is a slave sensor for electronic dierential pressure
measurement.
You can nd detailed information about the area of application in
chapter "Product description".
Operational reliability is ensured only if the instrument is properly
used according to the specications in the operating instructions
manual as well as possible supplementary instructions.
1.3 Warning about incorrect use
Inappropriate or incorrect use of the instrument can give rise to
application-specic hazards, e.g. vessel overll or damage to system
components through incorrect mounting or adjustment. Also the pro-
tective characteristics of the instrument can be inuenced.
1.4 General safety instructions
This is a high-tech instrument requiring the strict observance of stand-
ard regulations and guidelines. The user must take note of the safety
instructions in this operating instructions manual, the country-specic
installation standards as well as all prevailing safety regulations and
accident prevention rules.
The instrument must only be operated in a technically awless and
reliable condition. The operator is responsible for trouble-free opera-
tion of the instrument.
During the entire duration of use, the user is obliged to determine the
compliance of the necessary occupational safety measures with the
current valid rules and regulations and also take note of new regula-
tions.
1.5 CE conformity
The device fullls the legal requirements of the applicable EC guide-
lines. By axing the CE marking, we conrm successful testing of the
product.
You can nd the CE Certicate of Conformity in the download section
of our homepage.

4
1 For your safety
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
1.6 SIL qualication according to IEC 61508
The Safety Integrity Level (SIL) of an electronic system is used to as-
sess the reliability of integrated safety functions.
For detailed specication of the safety requirements, multiple SIL
levels are specied according to safety standard IEC 61508. You can
nd detailed information in chapter "Functional safety (SIL)"
The instrument meets the specications of IEC 61508: 2010 (Edi-
tion 2). It is qualied for single-channel operation up to SIL2. The
instrument can be used homogeneously redundant up to SIL3 in
multi-channel architecture with HFT 1.
1.7 Permissible process pressure
The permissible process pressure is specied on the type label with
"Process pressure", see chapter "Conguration". For safety reasons,
this range may not be exceeded. This applies even if a measuring cell
with a measuring range (order-related) higher than the permissible
pressure range of the process tting is installed.

5
2 Product description
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
2 Product description
2.1 Conguration
The type label contains the most important data for identication and
use of the instrument:
2
1
13
14
15
12
10
11
5
3
6
4
7
8
9
Fig. 1: Layout of the type label (example)
1 Instrument type
2 Product code
3 Field for approvals
4 Power supply and signal output, electronics
5 Protection rating
6 Measuring range
7 Permissible process pressure
8 Material, wetted parts
9 Order number
10 Serial number of the instrument
11 Data-Matrix-Code for smartphone app
12 Symbol of the device protection class
13 ID numbers, instrument documentation
14 Reminder to observe the instrument documentation
15 SILidentication
The type label contains the serial number of the instrument. With it
you can nd the following instrument data on our homepage:
• Product code (HTML)
• Delivery date (HTML)
• Order-specic instrument features (HTML)
• Operating instructions, quick setup guide and Safety Manual at the
time of shipment (PDF)
• Test certicate (PDF) - optional
Go to www.vega.com, "VEGA Tools" and "Instrument search". Enter
the serial number.
Alternatively, you can access the data via your smartphone:
• Download the smartphone app "VEGA Tools" from the "Apple App
Store" or the "Google Play Store"
• Scan the Data Matrix code on the type label of the instrument or
• Enter the serial number manually in the app
Type label
Serial number - Instru-
ment search

6
3 Mounting
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
3 Mounting
3.1 General instructions for use of the instrument
Protect your instrument against moisture ingress through the following
measures:
• Use the recommended cable (see chapter "Connecting to power
supply")
• Tighten the cable gland
• When mounting horizontally, turn the housing so that the cable
gland points downward
• Loop the connection cable downward in front of the cable gland
This applies mainly to outdoor installations, in areas where humidity is
expected (e.g. through cleaning processes) and on cooled or heated
vessels.
3.2 Ventilation and pressure compensation
The lter element is mounted into the electronics housing. It has the
following functions:
• Ventilation of the electronics housing
• Atmospheric pressure compensation (with relative pressure meas-
uring ranges)
→ Turn the housing so that the lter element points downward after
the instrument is installed. This provides better protection against
buildup.
1 2 3
4 4 4
Fig.2:Positionofthelterelement-non-Ex,Ex-iaversion
1 Housing plastic, stainless steel precision casting
2 Housing aluminium
3 Housing stainless steel, electropolished
4 Filter element
With the following instruments a blind plug is installed instead of the
lter element:
• Instruments in protection IP 66/IP 68 (1 bar) - ventilation via capil-
laries in non-detachable cable
• Instruments with absolute pressure
3.3 Combination Master - Slave
In principle, any sensor combination within the VEGABAR 80 series is
allowed. The following requirements must be fullled:
Protection against mois-
ture
Instruments in non-Ex
and Ex-ia version

7
3 Mounting
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
• Conguration, Master sensor suitable for electronic dierential
pressure
• Pressure type is identical for both sensors, i.e. relative pressure/
relative pressure or absolute pressure/absolute pressure
• Master sensor measures the higher pressure
• Measurement setup as shown in the following chapters
The measuring range of each sensor is selected such that it ts the
measuring loop. For this, the max. recommended turn down must be
noted. See chapter "Technical data". It is absolutely necessary the the
measuring ranges of Master and Slave correspond.
Measurement result = Measured value of Master (total pres-
sure) - measured value of Slave (static pressure)
Depending on the application, individual combinations can result, see
following examples:
Data
Application: Level measurement
Medium: Water
Vessel height: 12 m, hydrostatic pressure = 12 m x 1000 kg/m3x
9.81 m/s2 = 117.7 kPa = 1.18 bar
Superimposed pressure: 1 bar
Total pressure: 1.18 bar + 1 bar = 2.18 bar
Instrument selection
Nominal measuring range Master: 2.5 bar
Nominal measuring range Slave: 1 bar
Turn Down: 2.5 bar/1.18 bar = 2.1 : 1
Data
Application: Level measurement
Medium: Water
Vessel height: 500 mm, hydrostatic pressure = 0.50 m x 1000 kg/m3x
9.81 m/s2 = 4.9 kPa = 0.049 bar
Superimposed pressure: 350 mbar = 0.35 bar
Total pressure: 0.049 bar + 0.35 bar = 0.399 bar
Instrument selection
Nominal measuring range Master: 0.4 bar
Nominal measuring range Slave: 0.4 bar
Turn Down: 0.4 bar /0.049 bar = 8.2 : 1
Data
Application: Dierential pressure measurement
Medium: Gas
Static pressure: 0.8 bar
Dierential pressure on orice: 50 mbar bar = 0.050 bar
Total pressure: 0.8 bar bar + 0.05 bar = 0.85 bar
Example - large vessel
Example - small vessel
Example - orice in
pipeline

8
3 Mounting
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
Instrument selection
Nominal measuring range Master: 1 bar
Nominal measuring range Slave: 1 bar
Turn Down: 1 bar/0.050 bar = 20 : 1
The measuring result (level, pressure dierence) as well as measured
value Slave (static or superimposed pressure) are outputted by the
sensor. Depending on the instrument version, output as 4 … 20 mA
signal or digitally via HART, Probus PA or Foundation Fieldbus.
Output measured values

9
4 Connecting to power supply
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
4 Connecting to power supply
4.1 Connecting
The connection to the Master sensor is carried out through spring-
loaded terminals in the respective housing. For this, use the supplied,
confectioned cable. Solid cores as well as exible cores with cable
end sleeves are inserted directly into the terminal openings.
In case of exible cores without end sleeves, press the terminal from
above with a small screwdriver, the terminal opening is then free.
When the screwdriver is released, the terminal closes again.
Information:
The terminal block is pluggable and can be removed from the
electronics. To do this, lift the terminal block with a small screwdriver
and pull it out. When reinserting the terminal block, you should hear it
snap in.
You can nd further information on the max. wire cross-section under
"Technical data/Electromechanical data"
Proceed as follows:
1. Unscrew the housing lid
2. Loosen compression nut of the cable gland and remove blind
plug
3. Remove approx. 10 cm (4 in) of the cable mantle, strip approx.
1 cm (0.4 in) insulation from the individual wires or use supplied
connection cable
4. Insert the cable into the sensor through the cable entry
Fig. 3: Connection steps 5 and 6
5. Insert the wire ends into the terminals according to the wiring plan
6. Check the hold of the wires in the terminals by lightly pulling on
them
Connection technology
Connection procedure

10
4 Connecting to power supply
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
7. Connect the screen to the internal ground terminal, connect the
external ground terminal to potential equalisation
8. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
9. Unscrew the blind plug on the Master, screw in the supplied cable
gland
10. Connection cable to Master, see steps 3 to 8
11. Screw the housing lid back on
The electrical connection is nished.
4.2 Single chamber housing
The following illustration applies to the non-Ex, Ex-ia and Ex-d-ia
version.
5678 4
2
1
connect to Master
Fig. 4: Wiring plan VEGABAR 81 Slave sensor
1 To the Master sensor
2 Ground terminal for connection of the cable screen1)
Electronics and terminal
compartment
1) Connect screen here. Connect ground terminal on the outside of the housing
to ground as prescribed. The two terminals are galvanically connected.

11
4 Connecting to power supply
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
4.3 Connection example
5678
4
2
3
1
4
connect to Master
5
12
+
( ) (-)
678
4...20mA
Fig.5:Connectionexample,electronicdierentialpressure
1 Master sensor
2 Slave sensor
3 Connection cable
4 Supply and signal circuit, Master sensor
Connection between master and sensor is carried out according to
the table:
Master sensor Slave sensor
Terminal 5 Terminal 5
Terminal 6 Terminal 6
Terminal 7 Terminal 7
Terminal 8 Terminal 8
Connection example,
electronic dierential
pressure

12
5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
5 Set up with the display and adjustment
module
5.1 Insert display and adjustment module
The display and adjustment module can be inserted into the sensor
and removed again at any time. You can choose any one of four dier-
ent positions - each displaced by 90°. It is not necessary to interrupt
the power supply.
Proceed as follows:
1. Unscrew the housing lid
2. Place the display and adjustment module on the electronics in the
desired position and turn it to the right until it snaps in.
3. Screw housing lid with inspection window tightly back on
Disassembly is carried out in reverse order.
The display and adjustment module is powered by the sensor, an ad-
ditional connection is not necessary.
Fig. 6: Installing the display and adjustment module in the electronics compart-
ment of the single chamber housing

13
5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
1 2
Fig. 7: Installing the display and adjustment module in the double chamber
housing
1 In the electronics compartment
2 In the terminal compartment
Note:
If you intend to retrot the instrument with a display and adjustment
module for continuous measured value indication, a higher lid with an
inspection glass is required.
5.2 Parameter adjustment
A parameter change with SIL qualied instruments must always be
carried out as follows:
• Unlock adjustment
• Change parameters
• Lock adjustment and verify modied parameters
This ensures that all modied parameters have been deliberately
changed.
Unlock adjustment
The instrument is shipped in locked condition.
To prevent unintentional or unauthorized adjustment, the instrument
is protected (locked) against all parameter changes while in normal
operating condition.
For each parameter change you have to enter the PIN of the instru-
ment. In delivery status, the PIN is "0000".
Operating sequence with
SIL

14
5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
Change parameters
You can nd a description below the respective parameter.
Lock adjustment and verify modied parameters
You can nd a description below the parameter "Setup - Lock adjust-
ment".
1. Go to the menu "Setup" via the display and adjustment module.
2. In this menu item you activate/deactivate the slave sensor for
electronic dierential pressure and select the application, e.g.
level
3. Select in the menu item "Units" the adjustment unit of the instru-
ment, e.g. "bar".
4. Depending on the application, carry out the adjustment e.g. in the
menu items "Min. adjustment" and "Max. adjustment".
VEGABAR 81 always measures pressure independently of the pro-
cess variable selected in the menu item "Application". To output the
selected process variable correctly, an allocation of the output signal
to 0 % and 100 % must be carried out (adjustment).
With the application "Level", the hydrostatic pressure, e.g. with full
and empty vessel, is entered for adjustment. A superimposed pres-
sure is detected by the slave sensor and automatically compensated.
See the following example:
Change setup parameters
Parameterization example

15
5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
2
1
100%
0%
5 m
(196.9
")
4
3
Fig. 8: Parameter adjustment example "Min./max. adjustment, level measure-
ment"
1 Min. level = 0 % corresponds to 0.0 mbar
2 Max. level = 100 % corresponds to 490.5 mbar
3 VEGABAR 81
4 VEGABAR 81 - Slave sensor
If these values are not known, an adjustment with lling levels of e.g.
10 % and 90 % is also possible. By means of these settings, the real
lling height is then calculated.
The real product level during the adjustment is not important, because
the min./max. adjustment is always carried out without changing the
product level. These settings can be made ahead of time without the
instrument having to be installed.
1. Set the requested damping of the output signal in the menu item
"Damping".
2. Select the output characteristics in the menu item "Current out-
put".
To detect parameterization errors reliably, safety-relevant parameters
must be veried before saving them into the instrument.
1. Enter PIN
In delivery status, the PIN is "0000".
2. Character string comparison
You then have to carry out the character string comparison. This is
used to check the character presentation.
3. Serial number acknowledgement
Afterwards you conrm that the serial number of your instrument was
carried over correctly. This is used to check device communication.
4. Verify parameters
Conrm the modied values one after the other.
Further steps
Lock setup - adjustment

16
5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
If the described process of parameter adjustment was run through
completely and correctly, the instrument will be locked and hence
ready for operation.
Menu - Setup
Menu item Parameter Default setting
Measurement loop
name
Sensor
Application (SIL) Application Level
Slave for electronic dierential pres-
sure
Deactivated
Units Unit of measurement mbar (with nominal measuring range ≤400 mbar)
bar (with nominal measuring ranges ≥1 bar)
Static pressure bar
Position correction
(SIL)
0.00 bar
Adjustment (SIL) Distance (with density and interface) 1.00 m
Zero/Min. adjustment 0.00 bar
0.00 %
Span/Max. adjustment Nominal measuring range in bar
100.00 %
Damping (SIL) Integration time 0.0 s
Linearization (SIL) Linear
Current output (SIL) Current output - Mode Output characteristics
4 … 20 mA
Reaction when malfunctions occur
≤ 3.6 mA
Current output - Min./Max. 3.8 mA
20.5 mA
Lock adjustment
(SIL)
Last setting
Menu - Display
Menu item Default setting
Menu language Order-specic
Displayed value 1 Current output in %
Displayed value 2 Ceramic measuring cell: Measuring cell temperature in °C
Metallic measuring cell: Electronics temperature in °C
Display format 1 and 2 Number of positions after the decimal point, automatically
Backlight Switched on
Menu and parameter
overview

17
5 Set up with the display and adjustment module
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
Menu - Diagnosis
Menu item Parameter Default setting
Sensor status -
Peak value Pressure Actual measured value
Temperature Actual temperature values from measuring cell,
electronics
Simulation -
Menu - Additional adjustments
Menu item Parameter Default setting
PIN 0000
Date/Time Actual date/Actual time
Copy instrument set-
tings
-
Special parameters No reset
Scaling Scaling size Volume in l
Scaling format 0 % corresponds to 0 l
100 % corresponds to 0 l
Current output Current output - Meas. variable Lin. percent - Level
Current output - Adjustment 0 … 100 % correspond to 4 … 20 mA
HART mode Address 0
DP ow element Unit m3/s
Adjustment 0.00 % correspond to 0.00 m3/s
100.00 %, 1 m3/s
Menu - Info
Menu item Parameter
Device name VEGABAR 8.
Instrument version Hardware and software version
Factory calibration date Date
Sensor characteristics Order-specic characteristics

18
6 Supplement
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212
6 Supplement
6.1 Technical data
Electromechanical data - version IP 66/IP 67
Options of the cable entry
ƲCable entry M20 x 1.5, ½ NPT
ƲCable gland M20 x 1,5; ½ NPT (cable ø see below table)
Ʋ Blind plug M20 x 1.5; ½ NPT
Ʋ Closing cap ½ NPT
Material cable
gland
Material seal
insert
Cable diameter
4.5 … 8.5 mm 5 … 9 mm 6 … 12 mm 7 … 12 mm 10 … 14 mm
PA black NBR – ● ● –●
PA blue NBR – ● ● –●
Brass, nickel-
plated
NBR ● ● ● – –
Stainless steel NBR – ● ● –●
Wire cross-section (spring-loaded terminals)
Ʋ Massive wire, stranded wire 0.2 … 2.5 mm² (AWG 24 … 14)
Ʋ Stranded wire with end sleeve 0.2 … 1.5 mm² (AWG 24 … 16)
Interface to the Master sensor
Data transmission Digital (I²C-Bus)
Connection cable to the master sensor
Ʋ Conguration four wires, one suspension wire, one breather capillary,
screen braiding, metal foil, mantle
ƲWire cross-section 0.5 mm² (AWG 20)
ƲWire resistance < 0.036 Ω/m (0.011 Ω/ft)
ƲStandard length 5 m (16.40 ft)
ƲMax. length 25 m (82.02 ft)
Ʋ Min. bending radius at 25 °C/77 °F 25 mm (0.985 in)
Ʋ Diameter approx. 8 mm (0.315 in)
ƲColour Blue
Voltage supply for the complete system through Master
Operating voltage
ƲUB min 12 V DC
ƲUB min - lighted display and adjustment
module
12 V DC
ƲUB max Depending on signal output and version

19
Notes
VEGABAR 81 • Slave sensor for electronic dierential pressure
48050-EN-160212

Printing date:
VEGA Grieshaber KG
Am Hohenstein 113
77761 Schiltach
Germany
48050-EN-160212
All statements concerning scope of delivery, application, practical use and operat-
ing conditions of the sensors and processing systems correspond to the information
available at the time of printing.
Subject to change without prior notice
© VEGA Grieshaber KG, Schiltach/Germany 2016
Phone +49 7836 50-0
Fax +49 7836 50-201
E-mail: [email protected]
www.vega.com
Other manuals for VEGABAR 81
16
Table of contents
Other Vega Accessories manuals

Vega
Vega VEGAWAVE 61 Instruction Manual

Vega
Vega VEGAFLEX 82 User manual

Vega
Vega VEGAFLEX 82 User manual

Vega
Vega VEGAPULS 63 User manual

Vega
Vega VEGAPULS 69 User manual

Vega
Vega VEGAPULS 63 User manual

Vega
Vega POINTRAC 31 User manual

Vega
Vega VEGACAP 63 User manual

Vega
Vega ATEX Installation instructions

Vega
Vega DP-1 User manual

Vega
Vega VEGAPULS 66 User manual

Vega
Vega VEGAFLEX 61 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega VEGAPULS 65 User manual

Vega
Vega VEGABAR 82 User manual

Vega
Vega VEGAFLEX 81 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega VEGAPULS 62 User manual

Vega
Vega VEGASON 61 User manual

Vega
Vega FIBERTRAC 32 User manual