Vega SI-11 User manual

www.vega-absolute.ru
PULSE COUNTER
VEGA SI-11
USER MANUAL
DOCUMENT REVISION
FIRMWARE VERSION
26 2.4

Vega SI-11 / User Manual
2 Revision 26 of 25 August 2021
CONTENTS
INTRODUCTION......................................................................................................................................................................... 3
1 DEVICE PURPOSE AND OPERATION PRINCIPAL.............................................................................................................. 4
Device purpose ....................................................................................................................................................................... 4
Operation algorithm .............................................................................................................................................................. 4
Functional ................................................................................................................................................................................ 5
Marking..................................................................................................................................................................................... 6
2 SPECIFICATION ....................................................................................................................................................................... 7
Device Specification .............................................................................................................................................................. 7
Default Device Settings ......................................................................................................................................................... 8
3 OPERATION.............................................................................................................................................................................. 9
Device Appearance ................................................................................................................................................................ 9
Contacts Description ........................................................................................................................................................... 10
LED Indications ..................................................................................................................................................................... 11
Mounting Recomendations................................................................................................................................................ 12
4 COMMUNICATION PROTOCOL ........................................................................................................................................ 14
Vega SI-11 pulse counter transmits the following types of packets .......................................................................... 14
1. Packet with current readings ..................................................................................................................................14
2. «Alarm» packet ..........................................................................................................................................................15
3. Packet with time correction request......................................................................................................................15
4. Settings packet...........................................................................................................................................................16
Vega SI-11 pulse counter receives packets of the following types ............................................................................ 17
1. Real-time clock adjustment.....................................................................................................................................17
2. Packet with request of settings...............................................................................................................................17
3. Packet with settings ..................................................................................................................................................17
5 STORAGE AND TRANSPORTATION REQUIREMENTS ................................................................................................... 19
6 CONTENT OF THE PACKAGE ............................................................................................................................................. 20
7 WARRANTY............................................................................................................................................................................. 21

Vega SI-11 / User Manual
3
Revision 26 of 25 August 2021
INTRODUCTION
This manual is designated for Vega SI-11 pulse counter (hereinafter – the counter)
manufactured by Vega-Absolute OOO and provides information on powering and activation
procedure, control commands and functions of the counter.
This manual is targeted at specialists familiar with installation work fundamentals of
electronic and electrical equipment.
Vega-Absolute OOO reserves the right to make changes to the manual related to the
improvement of equipment and software, as well as to eliminate typos and inaccuracies,
without prior notice.

Vega SI-11 / User Manual
4 Revision 26 of 25 August 2021
1 DEVICE PURPOSE AND OPERATION PRINCIPAL
DEVICE PURPOSE
Vega SI-11 pulse counter is designed for counting of pulses incoming to 4 independent
inputs, further accumulating and transmitting of this information to the LoRaWAN®network.
In addition, Vega SI-11 can be used as a security device - anyone of four inputs can be
configured as security input.
The pulse counter can be used for any utilities’ meters and industrial equipment with
pulse output of herkon type or open-drain type.
Equipment with NAMUR pulse output is not supported
The counter is powered by a SAFT LS 17500 battery with capacity 3600 mAh.
The device is powered by a non-rechargeable lithium-thionyl chloride
(LiSOCl2) battery
Attempts to charge the battery may result in fire
OPERATION ALGORITHM
Vega SI-11 operates in modes listed below:
“Storage” – is a mode for storing and transporting. In this mode the device does not
communicate regularly with the network.
“Active” – is a main mode of device operation.
Before start using you need to switch the device out of “Storage” mode.
Vega SI-11 supports two activation methods in the LoRaWAN®network - ABP and OTAA.
Select one of the methods via the “Vega LoRaWAN Configurator” application (see “User Manual”
on the program).
ABP method. After pressing the start button, the device immediately starts working in the
"Active" mode.
OTAA method. After pressing the start button, the device makes three attempts to
connect to the network within the set frequency plan. After the activation in the LoRaWAN®
network is confirmed, the device sends a signal (LED flashing for 5 seconds) and switches to
the "Active" mode. If all attempts fail, the counter will continue to accumulate data and will
attempt to connect to the network every 6 hours.

Vega SI-11 / User Manual
5
Revision 26 of 25 August 2021
Hold the start button pressed (min. 5 seconds) to switch the device from the "Active"
mode back to the "Storage" mode.
When switching to the "Storage" mode, all readings from the pulse inputs
accumulated in the device memory are reset
The device forms the data packet with current state with a configurable period from 5
minutes to 24 hours. The packets stored in the device memory and transmitting during the next
communication session with the LoRaWAN®network.
Examples
If the data collection period is set to 24 hours the packet is formed at 00.00 on the
internal clock of the device
If the data collection period is 12 hours then at 00.00 and at 12.00, and so on.
Data collection period when measurements are out of specified temperature range set
independently and may be equal from 5 minutes to 24 hours.
The adjustable data transfer period can be from 5 minutes to 24 hours. When beginning
of communication session, the device starts sending packets with readings from the earliest
packet. The time of data transmitting cannot be specified, it’s defined in random way for every
device in chosen period of transmission from the moment of connection to the network.
Example
Transmission period is 30 minutes, and device was started at 16:40 by the internal device
clock. In random way the device calculate data transmitting time and set it at 16:41 in the half-
hour period from 16:40 to 17:10. Thus, packets from this device will transmit at 16:41, at 17:11,
at 17:41, at 18:11 and so on every 30 minutes by the internal device clock.
The internal clock is set automatically when you connect to the device through USB,
also can be adjust via LoRaWAN®network.
FUNCTIONAL
Vega SI-11 pulse counter is A class device (LoRaWAN®classification) and has the
following features:
ADR support (Adaptive Data Rate)
Sending of confirmed packets
Two operating modes: "Active" and "Storage"
Inputs can be switched to “security” mode for connection to external leakage
and safety sensors, etc.

Vega SI-11 / User Manual
6 Revision 26 of 25 August 2021
Queue for sending packets if delivery fails
Time referencing of readings by internal clock
Communication in case of security inputs actuation
Temperature measurement
Charge measuring of the built-in battery (%)
MARKING
Device marked with sticker that contain the next information:
Device name;
DevEUI;
Release Date.
Sticker located in three places – on device case, in factory certificate and on the packing
box.
Besides, there is an additional sticker located on the packing box which contains:
Information about firmware version;
QR-code containing DevEUI and keys for device registration in network via
OTAA method.

Vega SI-11 / User Manual
7
Revision 26 of 25 August 2021
2 SPECIFICATION
DEVICE SPECIFICATION
MAIN
Double-contact digital inputs up to 4
Maximum input frequency 200 Hz
Security inputs up to 4
USB-port mini-USB, type B
Operating temperatures -40…+85 °С
Built-in temperature sensor yes
LORAWAN®
LoRaWAN®class A
Quantity of LoRa channels
16
Frequency plans supported by default
RU868, EU868, KZ865, custom (EU868
based)
Frequency plans available as order option
IN865, AS923, AU915, KR920, US915
Activation type ABP or OTAA
Communication period 5, 15, 30 minutes, 1, 6, 12 or 24 hours
Data collection period 5, 15, 30 minutes, 1, 6, 12 or 24 hours
Memory amount for storing packets 200 packets
Type of the LoRaWAN antenna internal
Sensitivity -138 dBm
Radio coverage in restrained urban conditions up to 5 km
Radio coverage within line of sight up to 15 km
Transmitter power by default 25 mW (configurable)
Maximum transmitter power 100 mW
POWER
Battery
3600 mAh SAFT LS 17500
Calculated number of packets sent by the device
with default settings
80 000
CASE
Housing dimensions, no more than 90 х 49 х 46 mm
Ingress protection rating IP65
Mounting Clamp fastening to the support, DIN-rail,
wall-mounting
PACKAGE
Dimensions 95 х 50 х 46 mm
Weight 0.093 kg

Vega SI-11 / User Manual
8 Revision 26 of 25 August 2021
DEFAULT DEVICE SETTINGS
PARAMETER
VALUE
Frequency plan RU868
Activation type OTAA
Adaptive Data Rate ON
Confirmed Uplinks OFF
Rx 1 Delay
1 second
Join Accept Delay 5 seconds
Uplink number of transmissions 1
Data rate DR0
Power 14 dBm
Communication period 24 hours
Data collection period 24 hours
Time zone UTC +00:00
Inputs operate in mode pulse
For changing the device settings, you need to connect to it with “Vega LoRaWAN
Configurator” application. You can download app on the iotvega.com site in SOFT section as
well as User Manual for configurator. Go to the app page.

Vega SI-11 / User Manual
9
Revision 26 of 25 August 2021
3 OPERATION
DEVICE APPEARANCE
Vega SI-11 is represented in small plastic case which has four screws and mounting for
DIN-rail.
1 – screw ø 2 mm х 8 mm, cross
2 – DIN-rail with mounting holes ø 3 mm
3 – silicone gasket without through holes, ensuring the protection rating of the device
case IP65.
All of elements for manage and indication as well as connecting contacts are placed on
the board inside the case.
4 – launch button
5 – LED indicator
6 – contacts
7 – USB-port
1 2
3
4
5
6
7

Vega SI-11 / User Manual
10 Revision 26 of 25 August 2021
CONTACTS DESCRIPTION
When connecting a radio modem to external devices, consider the internal circuitry of
its pulse inputs shown below.
De-bouncing logics with 5 msec time constant is integrated in the pulse counter. The
pulse counting is carried out for frequencies up to 200 Hz.
reed switch;
mechanical pushbutton;
open-collector output.
Equipment with NAMUR pulse output is not supported
Polarity effects only "open collector" circuits.
To reset all readings from the pulse inputs accumulated in the device memory you need
to switch the device to the "Storage" mode by holding the start button pressed (min. 5 seconds).
Pulse inputs can be configured for work in the "Security" mode via the “Vega LoRaWAN
Configurator” application. In this mode, the device does not count pulses at the "Security" input,
but only monitors its status. Should the "Security" input trigger, the device is activated and sends
an alarm message to the network.
The maximum possible frequency of sending alarm packets is one per 1 second.

Vega SI-11 / User Manual
11
Revision 26 of 25 August 2021
LED INDICATIONS
There is one red LED on the board. The indication is only used when the device is
activating in the LoRaWAN®network and when the operating modes are changing.
LED SIGNAL
MEANING
Short flashings Linking to the network in progress
One long flashing
during 5 sec
The device has been successfully
connected to the network and is in
active mode
Three flashings each by
1 sec
Linking to the network has been failed
or the device switched to the
“Storage” mode
In case of connection attempt fail, the device will continue to accumulate
data and will attempt to connect to the network every 6 hours
LED location on the
board

Vega SI-11 / User Manual
12 Revision 26 of 25 August 2021
MOUNTING RECOMENDATIONS
To provide the stable radio between the gateway and the end device it is recommended
avoiding the device installation in the places which are barriers for the radio signal getting
through like a reinforced floors and walls, a basement, an underground facilities and wells, a
metal case etc.
The necessary stage for the network deploying including a big quantity of end devices is
a radio planning work with nature experiments.
Before starting mounting work, you must make sure that the latest
firmware version is installed on the equipment
For mounting you will need:
cross-shaped screwdriver ;
awl;
wire stripper;
laptop.
Step by step mounting be like:
1. Setting the devices and connecting them to the network are usually carried out in the
office (see Network Deployment Manual).
2. Determination of suitable places for mounting at the object with a network tester.
3. De-energizing the connected equipment, metering devices, etc.
4. Making holes in the silicone gasket for wires - strictly according to the number of
wires. It must be remembered that the wire must be of circular cross-section and no
more than 3 mm in diameter.
When removing the sealant, as well as when installing wires of a
different diameter or cross-
section, device performance may
deteriorate until failure due to moisture entering the case
5. Connecting all necessary wires to the SI-11 contacts.
6. Device launching – switching to the “Active” mode and registration in the network.

Vega SI-11 / User Manual
13
Revision 26 of 25 August 2021
7. By the laptop you can make sure that the device successfully sends the data.
8. Before assembling the device, it is necessary to reset the pulses accumulated during
testing and connection by switching the device to the "Storage" mode by pressing
the button for 5 seconds.
9. Start the device by pressing launch button.
10. Assembling the device.
11. DIN-rail mounting or another available way to mount the device on the object.
Location of the
launch button on
the board

Vega SI-11 / User Manual
14 Revision 26 of 25 August 2021
4 COMMUNICATION PROTOCOL
This part describes the SI-11 communication protocol with LoRaWAN®network.
In fields consisting of several bytes, the little endian byte order is used
VEGA SI-11 PULSE COUNTER TRANSMITS THE FOLLOWING TYPES OF PACKETS
1. Packet with current readings
Sent regularly on LoRaWAN port 2.
Size in bytes
Field description
Data type
1 byte
Packet type, this packet == 1
uint8
1 byte
Battery charge, %
uint8
1 byte
Values of basic settings (bit field)
uint8
4 bytes Reading time for values in this packet (unixtime UTC) uint32
1 byte Temperature,
°
С int8
4 bytes Input 1 reading (depending on the type – number of pulses or
status: 0 – open, 1 - closed)
uint32
4 bytes
Input 2 reading (depending on the type – number of pulses or
status: 0 – open, 1 - closed)
uint32
4 bytes
Input 3 reading (depending on the type – number of pulses or
status: 0 – open, 1 - closed)
uint32
4 bytes Input 4 reading (depending on the type – number of pulses or
status: 0 – open, 1 - closed)
uint32
The counter has internal clock and calendar; time and date are factory set. When sending
a packet with the current readings, the device uses the data taken at the nearest time, which is
multiple to the interval according to data collection period:
1 hour period: the readings of the beginning of the current hour are sent;
6 hours period: 00:00, 06:00, 12:00, 18:00 readings are sent;
12 hours period: 00:00, 12:00 readings are sent;
24 hours period: the readings of 00:00 of the current day are sent.

Vega SI-11 / User Manual
15
Revision 26 of 25 August 2021
"Values of basic settings" bit field decoding
Bit
Meaning
0 bit
Activation type: 0 - OTAA, 1 – ABP
1, 2,3 bits
Communication period:
|1 == 0|2==0|3==0| - 5 minutes
|1 == 1|2==0|3==0| - 15 minutes
|1 == 0|2==1|3==0|- 30 minutes
|1 == 1|2==1 |3==0|- 1 hour
|1 == 0|2==0|3==1|- 6 hours
|1 == 1|2==0|3==1|- 12 hours
|1 == 0|2==1|3==1|- 24 hours
4 bits
First input type: 0 – pulse, 1 - security
5 bits
Second input type: 0 – pulse, 1 - security
6 bits Third input type: 0 – pulse, 1 - security
7 bits Fourth input type: 0 – pulse, 1 - security
Time zone considered during collection data from an external meter.
2. «Alarm» packet
Sent in case of security input closure on LoRaWAN port 2.
Size in bytes
Field description
Data type
1 byte Packet type, this packet == 2 uint8
1 byte Battery charge, % uint8
1 byte
Values of basic settings (bit field)
uint8
1 byte
Input No, where the alarm is detected
uint8
4 bytes
Packet time (unixtime UTC)
uint32
4 bytes Input 1 reading (depending on the type – number of pulses or
status: 0 – open, 1 - closed)
uint32
4 bytes Input 2 reading (depending on the type – number of pulses
or status: 0 – open, 1 - closed)
uint32
4 bytes Input 3 reading (depending on the type – number of pulses
or status: 0 – open, 1 - closed)
uint32
4 bytes
Input 4 reading (depending on the type – number of pulses
or status: 0 – open, 1 - closed)
uint32
When sending alarm packet, the device uses the data taken at moment of alarm triggers.
3. Packet with time correction request
Sent every seven days on LoRaWAN port 4.
Size in bytes
Field description
Data type
1 byte
Packet type, this packet == 255
uint8
4 bytes Time of the modem at a moment of the packet transmission
(unixtime UTC)
uint32

Vega SI-11 / User Manual
16 Revision 26 of 25 August 2021
After receiving this type of package, the application can send to modem the packet with
time correction.
4. Settings packet
Transmitting on LoRaWAN port 3 when settings request command received, or device
connected to the network.
Size in bytes
Field description
Data type
1 byte Packet type, this packet == 0 uint8
2 bytes ID of parameter uint16
1 byte Data length (len) uint8
len bytes Parameter value -----
2 bytes ID of parameter uint16
1 byte Data length (len) uint8
len bytes Parameter value -----
… … …
2 bytes ID of parameter uint16
1 byte Data length (len) uint8
len bytes Parameter value -----

Vega SI-11 / User Manual
17
Revision 26 of 25 August 2021
VEGA SI-11 PULSE COUNTER RECEIVES PACKETS OF THE FOLLOWING TYPES
1. Real-time clock adjustment
Sent by application on LoRaWAN port 4.
Size in bytes
Field description
Data type
1 byte Packet type, this packet = 255 uint8
8 bytes The value in seconds for which you need to adjust the time.
Can be positive or negative
int64
2. Packet with request of settings
Sent by application on LoRaWAN port 3.
Size in bytes
Field description
Data type
1 byte
Packet type, this packet == 1
uint8
Answering this packet, the device sends the packet with settings.
3. Packet with settings
Is identical to such packet from device.
Size in bytes
Field description
Data type
1 byte Packet type, this packet == 0 uint8
2 bytes ID of parameter uint16
1 byte Data length (len) uint8
len bytes Parameter value -----
2 bytes ID of parameter uint16
1 byte Data length (len) uint8
len bytes Parameter value -----
… … …
2 bytes ID of parameter uint16
1 byte Data length (len) uint8
len bytes Parameter value -----
The package with settings sent to the device may not contain all the settings supported
by the device, but only the part that needs to be changed.
Table of ID of SI-11 parameters and these possible values
ID of
parameter
Description
Data
length
Possible values
4 Confirmed uplinks 1 byte 1 – confirmed
2 – unconfirmed

Vega SI-11 / User Manual
18 Revision 26 of 25 August 2021
8 Uplinks number of transmissions 1 byte from 1 to 15
12 Input 1 mode 1 byte 1 – pulse
2 – guard
13
Input 2 mode
1 byte
1 – pulse
2 – guard
14 Input 3 mode 1 byte 1 – pulse
2 – guard
15 Input 4 mode 1 byte 1 – pulse
2 – guard
16 Communication period 1 byte 1 – 1 hour
2 – 6 hours
3 – 12 hours
4 – 24 hours
5 – 5 minutes
6 – 15 minutes
7 – 30 minutes
49
Data collection period
1 byte
1 – 1 hour
2 – 6 hours
3 – 12 hours
4 – 24 hours
5 – 5 minutes
6 – 15 minutes
7 – 30 minutes
55 Time zone, in minutes 2 bytes from -720 to 840

Vega SI-11 / User Manual
19
Revision 26 of 25 August 2021
5 STORAGE AND TRANSPORTATION REQUIREMENTS
Vega SI-11 pulse counter shall be stored in the original packaging in heated room at
temperatures +5 °С to +40 °С and relative humidity less than 85%.
The counter shall be transported in covered freight compartments of all types at any
distance at temperatures -40 °C to +85 °C.
Table of contents
Other Vega Measuring Instrument manuals

Vega
Vega ALT-5 User manual

Vega
Vega VEGACAL 67 User manual

Vega
Vega VEGAPULS 68 User manual

Vega
Vega VEGACAP 67 User manual

Vega
Vega VEGAPULS 6X User manual

Vega
Vega VEGAFLEX 80 User guide

Vega
Vega VEGAPULS 66 User manual

Vega
Vega VEGABAR 81 User manual

Vega
Vega VEGAPULS 61 User manual

Vega
Vega VEGAVIB 63 User manual

Vega
Vega LMPVL5TN User manual

Vega
Vega VEGAFLEX 81 User manual

Vega
Vega VEGASON 63 User manual

Vega
Vega VEGAPULS 63 User manual

Vega
Vega VEGAMIP R61 User manual

Vega
Vega VEGAWAVE 63 User manual

Vega
Vega VEGAPULS 68 User manual

Vega
Vega VEGAFLEX 86 User manual

Vega
Vega VEGAPULS 64 User manual

Vega
Vega VEGASWING 51 User manual