Victor VC9807A+ User manual

DIGITAL
DIGITAL
DIGITAL
DIGITAL
MULTIMETER
MULTIMETER
MULTIMETER
MULTIMETER
OPERATION
OPERATION
OPERATION
OPERATION
MANUAL
MANUAL
MANUAL
MANUAL
1.
1.
1.
1.
GENERAL
GENERAL
GENERAL
GENERAL
The meter is a stable 4 1/2 multimeter with LCD display, driven by battery. It's widely
used on measuring DCV,
ACV,
DC current, AC current, resistance, conductance,
capacitance, diode, transistor, continuity test and frequency, and has the function of
overload pro tection, auto power off and data hold, etc.
2.
2.
2.
2.
SAFETY
SAFETY
SAFETY
SAFETY
Meet with IEC1010 standard
Read the safety note before operation.
2-1 .Do not input a limited value over-ranged.
2-2
.
The voltage below 36V is safety.
To
avoid electric shock, check whether the test lead s
are connected correctly, whether the insulation is good when measuring over 36 V DCV or
25
V
ACV.
2-3 .The test leads should be removed from the testing point when changing function and
range.
2-4 .Choose correct function and range to avoid wrong operation.
2-5 .Do not take measurement if the battery is not fit on or rear cover ked.
2-6. Do not input voltage when measuring resistance.
2-7. Remove the test leads and turn the power off before replacing battery or fuse.
2- 8 .Specification of safety signal
“ ” existing dangerous volt “ ” GND “ ” dual insulation
“ ” operator must refer to the manual “ ” low battery indication
3.
3.
3.
3.
SPECIFICATION
SPECIFICATION
SPECIFICATION
SPECIFICATION
3-1
3-1
3-1
3-1
.GENERAL
GENERAL
GENERAL
GENERAL
3-1-1 .Displaying: LCD display
3-1-2 .Max. displaying :19999 (4 1/2 )auto-polarity display
3-1-3 .Measuring method :dual-slope A/D transfer
3-1-4 .Sa mpling rate :approx. 2- 3 times/sec
3-1-5. Max. common mode voltage: 500V DC/AC RMS
3-1-6 .Over-range display :the MSD displays “1”or “- 1 ”
3-1-7 .Low battery display
:
“ ” displays
3-1-8 .Operation :(0 ~40) ℃ , relative humidity <80% R.H
3-1-9. Storage: (-10 ~50) ℃
,
relative humidity <80% R.H
3-1-10. power: 9V battery (NEDA1604/6F22 or equivalent)
3-1-11. Dimension: 19 2 mm × 95 mm × 48 mm
3-1-12. Weight: approx. 455 g (including battery)
3-2
3-2
3-2
3-2
.TECHNICAL
TECHNICAL
TECHNICAL
TECHNICAL
SPECIFICATION
SPECIFICATION
SPECIFICATION
SPECIFICATION
3-2-1 .Accuracy : ± ( a% ×reading +digit ) TEM :( 23 ±5 ) ℃ , <75% R.H ,one year
3-2-2 .TECHNICAL
DATA
3-2-2- 1 .DCV
Input impedance :1 0M Ωfor all ranges
Over - range protection :250 V DCV or AC peak factor at 2 00mV range,
1000 V DCV or AC peak factor at other range
Range
Accuracy
Resolution
200 mV
±(0 .1 %+ 5 )
0.01mV
2V
0.0001 V
20V
0.001 V
200V
0.01 V
1000V
±(0. 2 %+5)
0.1 V
3-2-2-2.ACV
Input resistance: 2 M Ω for all ranges
Overload protection: Range 200mV: DC 250V or AC peak value.
Other ranges: DC 1000V or AC peak value
Frequency response :below 200V :(40~400)H z ,750V range: (40~200)H z
Displaying :1): sine wave RMS (mean value response )
2): Some can
’
t turn to “ 0 ” which is normal phenomenon; it can
’
t affect the
test accuracy
3-2-2-3.DCA
Max. measuring volt drop: 200 mV
Max. input current: 2 0A (less than 10 seconds )
Overload protection: 0.2A / 250V self resume fuse ; 20A range without fuse.
3-2-2-4.ACA
Max. measuring volt drop :200 m V
Max. input current :2 0A (less than 1 0 seconds )
Overload protection :0.2A / 250V self resume fuse ,2 0A range without fuse.
Frequency response :40H z ~200Hz
Displaying :sine wave RMS (mean value response )
3-2-2- 5 . RESISTANCE (Ω)
Open voltage :less than 3 V
Overload protection :250V DCV or ACV peak factor
Note
Note
Note
Note
:
In 200 Ωrange
,
should make the test leads short, and measure the resistance of th e
wire, then, subtracts from the actual measuring value.
Range
Accuracy
Resolution
2V
± (0. 8 %+ 2 5)
1m V
20V
10m V
200V
100m V
750 V
± ( 1.0 %+ 2 5)
1 V
Range
Accuracy
Resolution
2 00u A
±(0.5%+4)
0. 0 1 u A
2mA
0.0001mA
20mA
0.001m A
200mA
±(0.8%+6)
0.01m A
20A
±(2.0%+15)
0.001 A
Range
Accuracy
Resolution
200mA
±(1. 5 %+25)
0.1mA
20A
±(2.5%+35)
1 0 mA
Range
Accuracy
Resolution
200 Ω
±(0. 4 %+1 0 )
0. 0 1 Ω
2k Ω
±( 0.4 %+ 5)
0. 1 Ω
20k Ω
1Ω
200k Ω
10 Ω
2M Ω
1 00 Ω
200M Ω
±【5 % (reading-10.00) + 30 】
10 k Ω
3-2-2-6.CAPACITANCE
Testing frequency: 400Hz
Overload protection :36V DCV or
AC
V peak factor ;630mA/250V fuse protection
3-2-2-7.FREQUENCY
Input sensitivity :120 m V RMS
Overload protection :250V DC or AC peak factor (less than 10sec.)
Note: 200kHz maybe can
’
t return to “ 0 ” ,this phenomenon is normal, it can
’
t affect
the test accuracy.
3-2-2-8.DIODE AND CONTINUITY TEST
Overload protection :250V DC or AC peak factor
Warning: do not input any voltage value at this range for safety!
3-2-2-9. hFE
3-2-2-10. Conductance (resistance range: 10 , 000M Ω-10M Ω )
Open voltage is approx:3V
4.operation
4.operation
4.operation
4.operation
4-1
4-1
4-1
4-1
.Fr
Fr
Fr
Fr
o
o
o
o
nt
nt
nt
nt
panel
panel
panel
panel
1.LCD :display the measured value
2 -1 Power switch :turn on or off the power
2-2. Backlight key: turn on or off the power
2-3. hFE testing terminal :for measuring
transistor hFE
2-4. HOLD key :press the key down, the
measured value will be hold and “H”
signal appears; press it down again, “H”
signal disappears and out of the HOLD mode.
2-5. Continuity buzzer key
3.Knob switch: for changing function and range
4. Testing terminal for 20A current
5. Testing terminal for less than 200 mA current and conductance
Range
Accuracy
Resolution
20nF
±( 4.0 %+ 5 0)
10 pF
2uF
100 0 pF
200uF
10nF
Range
Accuracy
Resolution
20 k H z
±( 1.5 %+ 2 5 )
10H z
200kHz
100Hz
RANGE
DISPLAYING VALUE
TEST CONDITION
D iode forward volt drop
Forward DC current is
approx. 1 mA ,
backward voltage is
approx. 3V
Buzzer rings, the impedance
between the two testing points is
less than (70 ±20) Ω
O pen voltage is
approx. 3V
Range
Scope
Testing condition
hFE NPN or PNP
0~1000
Base DC current is approx.
10u A ,Vce is approx. 3V
Range
Accuracy
Resolution
(0 . 1-100)nS
±( 1 .0%+ 30)
0. 1nS

6. Common ground
7. Terminal for
V oltage
,
resistance, and frequency testing
4-2
4-2
4-2
4-2
.DCV
DCV
DCV
DCV
measurement
measurement
measurement
measurement
4-2-1 .Apply the black test leads to “COM ”terminal and the red one to “V/ Ω/Hz ”
terminal.
4-2-2 .Switch the knob to a proper DCV range, then, connect the test leads
cross the measured circuit, the voltage and polarity of the point which connected
with the red test lead will be displayed.
NOTE
NOTE
NOTE
NOTE
:
:
:
:
1.If the voltage under measured is unclear beforehand, should set the range knob to the
highest range, then, switch to a proper range according to the displaying value.
.
If LC D
displays
“1
”
,
it means overrange, the range knob must be switched to a higher rang e.
2.D o not input a voltage over 1000V
,
or the meter might be damaged.
3.Be careful when measuring high voltage circuit.
4-3
4-3
4-3
4-3
.ACV
ACV
ACV
ACV
measurement
measurement
measurement
measurement
4-3-1 .Apply the black test leads to “COM ”terminal and the red one to “V/ Ω/Hz ”
terminal.
4-3-2
.
Set the range knob to a prop er ACV range, then, connect the test leads
cross to the measured circuit.
NOTE:
NOTE:
NOTE:
NOTE:
1.If the range under measured is unclear beforehand, should set the range knob to the
highest range, then, switch to a proper range according to displaying value. .If LCD
displays “1”
,
it means overrange, should set the range knob to a higher range.
2. The remained digits do not affect on measuring accuracy.
3. D o not input a voltage over 1000V rms
,
or,
the meter might be damaged.
4.Be careful when measuring high voltage circuit.
4-4
4-4
4-4
4-4
.DC
DC
DC
DC
current
current
current
current
measurement
measurement
measurement
measurement
4-4-1
.
Apply the black test leads to “COM
”
terminal and the red one to “m A
”
ter minal
(Max.200 mA
) ,
or the red test lead to “20A ”terminal (max. 20A ).
4-4-2 .Set the range knob to a proper DCA current range, then, connect the meter to the
measured circuit, the measured current value and the current polarity which the
red test lead connect will be displayed on LCD.
NOTE:
NOTE:
NOTE:
NOTE:
1.If the range under measured is unknown beforehand, should set the range knob to the
highest range, then, switch to a proper range according to the displaying value ; If LCD
displays “1
”
,it means overrange, the range knob must be set to a higher range.
2.T he max. input current is 200 mA or 20A (subject to the position where the red test
lead insert
) ,
large current may blow the fuse. Be careful especially at 20A range,
because there is no fuse protection at this range, large current may heat the circuit,
even damage the meter.
4-5
4-5
4-5
4-5
.AC
AC
AC
AC
current
current
current
current
measurement
measurement
measurement
measurement
4-5-1 .Apply the black test lead to “COM ”terminal ,and the red one to “m A ”ter m
(max. 200 mA
) ,
or the red teat lead to “20A ”terminal (max. 20A
) ;
4-5-2 .Set the range knob to a proper AC
A
current range, then, connect the meter to the
circuit under measured.
NOTE:
NOTE:
NOTE:
NOTE:
1
.
If the current range under measure is unknown beforehand, should set the range knob
t
o
the highest range, then, switch to the proper range according to the displaying value. If
LCD displays “1
”
,it means overrange, must set the range knob to a higher range.
2
.
T he max. input current is 200 mA or
20A
(subject to the position where the red test lea d
insert
) ,
large current might blow the fuse
,
be careful especially at 20A range, becaus e
there is no fuse protection at this range, large current may heat the circuit, even damage
the meter.
3. The remained digits do not affect on measuring accuracy.
4-6
4-6
4-6
4-6
.Resistance
Resistance
Resistance
Resistance
measurement
measurement
measurement
measurement
4-6-1 .Apply the black test lead to “COM ”terminal ,and the red one to “V/ Ω/Hz ”
terminal.
4-6-2 .Set the range knob to a proper resistance range, connect the test leads across the
resistance under measured.
NOTE:
NOTE:
NOTE:
NOTE:
1
.
If the resistance value is over the selected range value, “1
”
displays, thus, should se t to
a higher range. When measuring value is over 1M Ω,the reading needs a few seconds
to be stable. It's normal for high resistance measuring.
2.When input terminal is open-circuit, overload displays.
3.Before measuring in-line resistance, be sure that power is off and all capacitance are
released.
4.D o not input voltage at this range.
4-7
4-7
4-7
4-7
.Capacitance
Capacitance
Capacitance
Capacitance
measurement
measurement
measurement
measurement
Set the knob to proper capacitance range, and insert the capacitor under tested.
NOTE:
1.If the capacitance is unknown beforehand, should set the range knob to the highest
range
,
then, switch to a proper range according to the displaying value.
If
“1
”
displa ys,
it means overrange, should set to a higher range.
2.Before taking measurement, LCD displaying might not be zero, the remaining will
become smaller and smaller, it can be ignored. .
3. If creeps seriously or the capacitor is breakdown when measuring large capacitance,
some digits are displayed and unstable.
4. Before measuring, should release the capacitor completely to avoid damage.
5. unit: 1 uF =10 00 nF 1nF=10 00 pF
4-8.
4-8.
4-8.
4-8.
Transistor
Transistor
Transistor
Transistor
hFE
hFE
hFE
hFE
4-8-1. Set the range to “hFE ”range.
4-8-2. Select NPN or PNP
,
insert separately emitter
,
base and collector to proper
terminal.
4-9
4-9
4-9
4-9
.Diode
Diode
Diode
Diode
and
and
and
and
continuity
continuity
continuity
continuity
test
test
test
test
4-9-1 .Apply the black test lead to “COM ”terminal ,and the red one to “V/ Ω/Hz ”
terminal (the polarity of the red test lead is “+
”
).
4-9-2 .Set the range knob to “
”
,connect the test leads to the diode under measured ,
and the reading is close to the value of diode forward voltage drop.
4-9-3 .Connect the test leads to two points of circuit under measured, if the built-in buzze r
sounds, the resistance between two points is less than approx. (70 ±20) Ω.
NOTE:
NOTE:
NOTE:
NOTE:
Do not input voltage at “ ” range to avoid damage the meter.
4-10
4-10
4-10
4-10
.Frequency
Frequency
Frequency
Frequency
measurement
measurement
measurement
measurement
4-10-1 .Apply the test leads or shield cable to “COM ”and “V/ Ω/Hz ”terminal.
4-10-2 .Switch the range knob to frequency range, connect the test leads across to the
signal source or the measured load.
NOTE:
NOTE:
NOTE:
NOTE:
1.When input is over 1 0V rms ,reading is possible but maybe overrange.
2.It's better to use shielding cable on noisy condition when measuring small signal.
3.Be careful when measuring high volt circuit.
4.Do not input a voltage over 250V DC or AC peak factor to avoid to damage the meter.
4-11
4-11
4-11
4-11
.Conductance
Conductance
Conductance
Conductance
measurement
measurement
measurement
measurement
4-11-1 .Apply the black test leads to “COM ”terminal, and the red one to “mA ”
terminal.
4-11-2
.
Set the range knob
to
“nS
”
range, connect the test leads to the insulating resistan ce
under measured.
4-12
4-12
4-12
4-12
.DATA
DATA
DATA
DATA
HOLD
HOLD
HOLD
HOLD
PRESS “HOLD ”key, the current data will be hold on LCD.
4-13.
4-13.
4-13.
4-13.
Auto
Auto
Auto
Auto
Power
Power
Power
Power
Off
Off
Off
Off
After working (20 ±10) minutes, the meter will be auto power off and to be sleeping
mode. Press POWER key twice to restart the power.
4-14.
4-14.
4-14.
4-14.
BACKLIGHT
BACKLIGHT
BACKLIGHT
BACKLIGHT
Press B/L key to turn on the backlight, and press it again to turn off.
NOTE: When turning on backlight, the working current will be enlarged, it leads to
shorten battery usage and enlarge accuracy of some functions.
5
5
5
5
.Maintenance
Maintenance
Maintenance
Maintenance
Do not try to modify the electric circuit.
5-1 . Keep the meter away from water, dust and shock.
5-2 . Do not store and operate the meter under the condition of high temperature, high
humidity, combustible, explosive and strong magnetic place.
5-3 . W ipe the case with a damp cloth and detergent, do not use abrasives and alcohol.
5-4 . I f do not operate for a long time, should take out the battery to avoid leakage
5-5 . Change fuse
When replacing fuse, to use the same type and specification
6.
6.
6.
6.
If
If
If
If
the
the
the
the
meter
meter
meter
meter
does
does
does
does
not
not
not
not
work
work
work
work
properly,
properly,
properly,
properly,
check
check
check
check
the
the
the
the
meter
meter
meter
meter
as
as
as
as
following:
following:
following:
following:
MB-9807-62
CONDITIONS
WAY
TO SOLVE
NO DISPLAYING
●T he power is not turned on
●HOLD key
●Replace battery
symbol displays
●Replace battery
NO CURRENT
INPUT
●Replace fuse
BIG ERROR
●Replace battery
Table of contents
Other Victor Multimeter manuals

Victor
Victor 86B Parts list manual

Victor
Victor 98A User manual

Victor
Victor VC9801A User manual

Victor
Victor VC890D User manual

Victor
Victor 6000counts User manual

Victor
Victor 81A User manual

Victor
Victor VC97 User manual

Victor
Victor VR-81D User manual

Victor
Victor VC88E User manual

Victor
Victor 88C User manual

Victor
Victor VC97 User manual

Victor
Victor 8165 User manual

Victor
Victor VC921 User manual

Victor
Victor VC3280 User manual

Victor
Victor VC70D User manual

Victor
Victor VC9808 User manual

Victor
Victor VC9808+ User manual

Victor
Victor VC9806+ User manual

Victor
Victor 606B-II User manual

Victor
Victor 81D User manual