1If the device has powered down, press the On/Off button, and after
several seconds the display will look something like…
You now need to enter the subject’s: height, age, gender, and weight,
in that order.
2To enter height,
press the Up or Down buttons to set the height to between 0 and
250cm (or between 0 and 99 inches if in USA mode),
press the Enter button to continue
3To enter age,
press the Up or Down buttons to set the age to between 0 and 99,
press the Enter button to continue
Note that if a value of 0 is entered for either height or age the rest of the
demographic entry will be skipped and the device goes straight to test mode. If
this occurs no predicted parameters will appear on the printouts.
Note that to return to the previous screen, press the Up and Down buttons together.
4To enter gender,
press the Up or Down buttons to set the gender to 0 for male, or
1 for female,
press the Enter button to continue.
5To enter weight,
press the Up or Down buttons to set the weight to between 0 kg
and 250 kg (or between 0 and 550 lbs if in USA mode),
press the Enter button to continue, and after several seconds
the display will look like ...
You are now ready to perform the test.
Note that it is only possible to perform a test when the Blow icon is displayed so always
watch for it, it will be accompanied by two beeps.
6To perform test,
fit a disposable SafeTway mouthpiece (single use only) onto the micro, and give the
following instructions to the subject so that testing is performed properly:
Note that it is advisable to also perform an animated demonstration of the test technique, using a mouthpiece, prior
to the subject performing the test.
a) Sit upright, fit the noseclip and relax (the use of a nose-clip is recommended),
b) Hold the unit, keeping it away from your mouth. Do not jerk the unit (this can de-stabilize the unit).
c) Inhale as deeply as possible, then insert the mouthpiece carefully into your mouth, not like a trumpet, but
clamping it gently between your teeth.
d) Seal your lips round the mouthpiece and keep your tongue down.
e) Blow out as fast and as long as possible. Keep blowing for at least 6 seconds. (From the very beginning of the
blow the operator should encourage the subject to keep going in a lively fashion. Keep eye contact with the
subject).
f) After 6-s instruct the subject to rest – say ‘Well done!’
cont ...
Page 3
... performing the FVC test
1
2
3
4
5