
14
3OPERATING PROCEDURES
3.1 PRINCIPLE OF MEASUREMENT
The new patented 9091C Dissolved Hydrogen Analyzer electrochemical hydrogen sensor
functions in a potentiostatic mode with an open three-electrode system. The measuring
electrode is made of solid titanium, coated with platinum. The counter electrode is made
of stainless steel and the reference electrode of silver/silverchloride. The use of an open
measuring electrode means, that there is a direct contact between the sample and the
measuring electrode. A membrane is not required.
For the measurement of dissolved molecular hydrogen (H2), the fluid medium flows into
the coaxial measurement cell between the platinum measuring electrode and counter
electrode. If the measuring electrode has reached the correct potential, hydrogen is
oxidized in the interphase boundary layer. The electrochemical reaction may be
represented qualitatively as:
H2+ 2H2O 2H3O+ + 2e-
The characteristic potential for this reaction- the potential at which hydrogen oxidation
occur in preference to other competing reactions- is measured by holding the potential of
the measuring electrode steady with a potentiostat and comparing the value with the
value of the reference electrode. Under these conditions the rate of the electrochemical
reaction is limited by the rate of diffusion of hydrogen molecules to the electrode. As
long as we remain within this ‘diffusion-limiting-current’ area, the electrochemical cell
delivers an electric current, whose size depends linearly on the concentration of dissolved
hydrogen. This current, which is measured by the instrument’s electronic processing unit,
is known as the diffusion-limiting current, Idiff:
cH2= k ·diff
This diffusion-limiting current depends mainly on the mass transport of the
electrochemically active components, and hence on the hydrodynamics in the measuring
cell. It therefore changes significantly with the temperature and flow rate through the
measuring cell. In practice the flow rate and temperature vary over wide ranges, but the
instrument corrects for these changes using empirical correlations.
A second important influence on the measurement is the condition of the electrodes. In
particular, the adsorptive and oxidative state of the electrode surface layers. This is
independent of the flow conditions. It depends instead on the voltage between the
electrode and the electrolyte, and on the material of which the electrode is made.
Appearing flow-and temperature effects of the measuring signal are recorded with the aid
of the installed flow meter and the temperature sensor and can be compensated, if it is
necessary.