WIKA TC52-M User manual

Operating instructions
Betriebsanleitung
Mode d'emploi
Manual de instrucciones
EN
DE
FR
ES
Magnetic surface thermocouple for high temperatures
Model TC52-M
Magnetisches Oberflächen-Thermoelement für hohe
Temperaturen, Typ TC52-M
Thermocouple de surface magnétique pour hautes
températures, type TC52-M
Termopar magnético de superficie para altas temperaturas
Modelo TC52-M
Magnetic surface thermocouple, model TC52-M

2
14502900.01 12/2021 EN/DE/FR/ES
WIKA operating instructions model TC52-M
EN
DE
FR
ES
© 12/2021 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d'emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!
Operating instructions model TC52-M
Page
3 - 24
Betriebsanleitung Typ TC52-M
Seite
25 - 46
Mode d'emploi type TC52-M
Page
47 - 68
Manual de instrucciones modelo TC52-M
Página
69 - 90

3WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
Contents
Contents
1. General information 4
2. Design and function 4
3. Safety 5
4. Transport, packaging and storage 8
5. Commissioning, operation 9
6. Faults 19
7. Maintenance, cleaning and recalibration 19
8. Dismounting, return and disposal 21
9. Specifications 22

4WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
1. General information
■
The high-temperature magnetic surface thermocouple described in the operating
instructions has been manufactured using state-of-the-art technology. All components
are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the instrument.
Working safely requires that all safety instructions and work instructions are observed.
■
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions onto the next operator or owner of the instrument.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modifications.
■
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: TE 66.52
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
2. Design and function
2.1 Description
The magnetic thermocouple assembly is designed as a non-welded option for vessel wall
temperature measurement.The circular magnet design incorporates an insulation barrier
from radiant heat along with a sensor tip designed to maximise surface contact to ensure
accurate temperature measurement.
The neck tube can be manufactured to length based on insulation thickness.The spring-
loaded neck tube ensures sensor contact with the vessel wall and allows for maintenance
without the use of tools above the insulation/cladding of the vessel.
This document describes instruments in standard version. For applications in hazardous
areas special instrument versions are required.
1. General information / 2. Design and function

5WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
For further information for operation in hazardous areas, see the additional information for
the corresponding ignition protection type (separate document).
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
2.2 Scope of delivery
Cross-check scope of delivery with delivery note.
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious injury
or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries or
damage to property or the environment, if not avoided.
WARNING!
... indicates a potentially dangerous situation that can result in burns, caused
by hot surfaces or liquids, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and
trouble-free operation.
3.2 Intended use
This thermocouple is used for temperature measurement in industrial applications.
The instrument is not permitted to be used in hazardous areas!
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The technical specifications contained in these operating instructions and the related data
sheets must be observed. Improper handling or operation of the instrument outside of its
technical specifications requires the instrument to be taken out of service immediately and
inspected by an authorised WIKA service engineer.
2. Design and function / 3. Safety

6WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
The manufacturer shall not be liable for claims of any type based on operation contrary to
the intended use.
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
▶
Refrain from unauthorised modifications to the instrument.
▶
Do not use the instrument within hazardous areas.
▶
Do not use the instrument with abrasive or viscous media.
Any use beyond or different to the intended use is considered as improper use.
Do not use this instrument in safety or emergency stop devices.
3.4 Responsibility of the operator
The instrument is used in the industrial sector. The operator is therefore responsible for
legal obligations regarding safety at work.
The safety instructions within these operating instructions, as well as the safety, accident
prevention and environmental protection regulations for the application area must be
maintained.
The operator is obliged to maintain the product label in a legible condition.
To ensure safe working on the instrument, the operating company must ensure
■
that suitable first-aid equipment is available and aid is provided whenever required.
■
that the operating personnel are regularly instructed in all topics regarding work safety,
first aid and environmental protection and know the operating instructions and in
particular, the safety instructions contained therein.
■
that the instrument is suitable for the particular application in accordance with its
intended use.
■
that personal protective equipment is available.
3.5 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to property.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
3. Safety

7WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their technical
training, know-how and experience as well as their knowledge of country-specific
regulations, current standards and directives, are capable of carrying out work on
electrical systems and independently recognising and avoiding potential hazards.The
skilled electrical personnel have been specifically trained for the work environment they
are working in and know the relevant standards and regulations.The skilled electrical
personnel must comply with current legal accident prevention regulations.
Operating personnel
The personnel trained by the operator are understood to be personnel who, based on their
education, knowledge and experience, are capable of carrying out the work described and
independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
3.6 Personal protective equipment
The personal protective equipment is designed to protect the skilled personnel from
hazards that could impair their safety or health during work. When carrying out the various
tasks on and with the instrument, the skilled personnel must wear personal protective
equipment.
Follow the instructions displayed in the work area regarding personal protective
equipment!
The requisite personal protective equipment must be provided by the operating company.
Wear safety goggles!
Protect eyes from flying particles and liquid splashes.
Wear protective gloves!
Protect hands from friction, abrasion, cuts or deep injuries and also from
contact with hot surfaces and aggressive media.
Wear ear defenders!
Protect ears from noise.
Ear defenders are required in case other appropriate measures cannot
prevent the operating personnel from being put at risk.
3. Safety

8WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
3.7 Labelling, safety marks
Product labels (examples)
Model
Serial number
Information on version (measuring element, measuring range...)
Sensor in accordance with standard (thermocouple): grounded
Transmitter model (only for design with transmitter)
Year of manufacture
Sensor symbol
■
ungrounded = ungrounded welded
■
grounded = welded to the sheath (grounded)
■
quasi grounded = The thermometer is, due to its low isolation clearances between
sensor and sheath, to be considered as grounded.
Approval-related data (non-Ex version = no specification)
Before mounting and commissioning the instrument, ensure you
read the operating instructions!
TR10-B
1 x Pt100 / B / 3 (F) -50 ... +250 °C
IEC 60751
Made in Germany
2014
1102AB12
T32.1S.0NI 4 ... 20 mA -50 ... +250 °C
D-63911 Klingenberg
HART ®
TR10-A-IICZ
1102AB12
1 x Pt100 / B / 3
-50 ... +250 °C
D = 6 mm
525 mm
(F)
D-63911 Klingenberg
IEC 60751
II 3G Ex nA IIC T1 ... T6 Gc X
WARNING! DO NOT OPEN WHILE ENERGISED!
II 3G Ex nA IIC T1 ... T6 Gc X
II 3D Ex tc IIIC T440°C ... T80°C Dc X
Tamb T6/T5/T4-T1: -20 ... +55/+70/+80 °C
Tamb T80/95/130-440°C: -20 ... +55/+70/+80 °C
L = 1 µH/m, C = 200 pF/m
Made in Germany 2014
WARNING! POTENTIAL ELECTROSTATIC CHARGING HAZARD!
II 3G Ex nA IIC T1 ... T6 Gc X
II 3D Ex tc IIIC T440°C ... T80°C Dc X
Tamb T6/T5/T4-T1: -20 ... +55/+70/+80 °C
Tamb T80/95/130-440°C: -20 ... +55/+70/+80 °C
L = 1 µH/m, C = 200 pF/m
Ex n / Ex d
Ex i
EAC (landesspezifisches Zusatzschild)
ΒΗИМАНИЕ!
ПОТЕНЦИАЛЬНЫЙ РИСК НАКОПЛЕНИЯ
ЭЛЕКТРОСТАТИЧЕСКОГО ЗАРЯДА!
СДЕЛАНО В ГЕРМАНИИ
TR10-B
1 x Pt100 / B / 3 (F) -50 ... +250 °C
IEC 60751
Made in Germany
2015
11012345
T32 4 ... 20 mA -50 ... +250 °C
TR10-A
11012345
1 x Pt100 / B / 3
-50 ... +250 °C
D = 6 mm
525 mm
(F)
D-63911 Klingenberg
HART ®
TC10-B
1 x Type K / 1 / . 0 ... 1260 °C
IEC 60584
Made in Germany
2015
11012345
T32 4 ... 20 mA -50 ... +250 °C
D-63911 Klingenberg
HART ®
Made in Germany 2015
D-63911 Klingenberg
IEC 60751
TC10-A
11012345
1 x Type K / 1 / .
0 ... 1260 °C
D = 2 mm
Made in Germany 2015
D-63911 Klingenberg
IEC 60584
525 mm
3. Safety

9WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
4. Transport, packaging and storage
4.1 Transport
Check the instrument for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during internal
transport, proceed carefully and observe the symbols on the packaging.
▶
With internal transport, observe the instructions in chapter 4.2 “Packaging
and storage”.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in instrument malfunction. Before putting it back into operation,
wait for the instrument temperature and the room temperature to equalise.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
Permissible conditions at the place of storage:
■
Storage temperature: 0 ... 70 °C [32 ... 158 °F]
■
Humidity: 35 ... 85 % relative humidity (no condensation)
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Hazardous environments, flammable atmospheres
Store the instrument in its original packaging in a location that fulfils the conditions listed
above. If the original packaging is not available, pack and store the instrument as described
below:
1. Place the instrument together with shock absorbing material in the packaging.
2. If stored for a prolonged period of time (more than 30 days), place a bag containing a
desiccant inside the packaging.
4. Transport, packaging and storage

10 WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5. Commissioning, operation
Only use original parts.
WARNING!
Physical injuries and damage to property and the environment caused
by hazardous media
Upon contact with hazardous media (e.g. oxygen, acetylene, flammable
or toxic substances), harmful media (e.g. corrosive, toxic, carcinogenic,
radioactive), and also with refrigeration plants and compressors, there is a
danger of physical injuries and damage to property and the environment.
Should a failure occur, aggressive media with extremely high temperature
and under high pressure or vacuum may be present at the instrument.
▶
For these media, in addition to all standard regulations, the appropriate
existing codes or regulations must also be followed.
▶
Wear the requisite protective equipment (see chapter 3.6 “Personal
protective equipment”).
5.1 Mechanical mounting
Ensure there is enough MI cable or flex armour to reach the termination location for each
magnet thermocouple sensor.
■
Avoid passing thermocouple wire or flex armour through hot zones.
■
Ensure all surfaces attachment areas are cleaned by appropriate methods.
■
When grinding always use clean abrasive grinding materials that have not been used on
other materials.
■
Wire brushes or wheels are not appropriate tools for cleaning the surface.
■
Test fit the magnet thermocouple sensor at the desired location to ensure adequate
length of MI cable or flex armour was provided.
■
Ensure all bendings are correct and sufficient allowance is provided for movement.
5. Commissioning, operation

11WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5. Commissioning, operation
5.1.1 Remove outer cladding (if applicable)
Mark the area for installation with a punch or another suitable marking.
At the marked location start by removing the outer cladding using a 100 mm [4 in] hole saw.
Once the cladding has been penetrated STOP drilling. Remove cladding and discard.

12 WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5. Commissioning, operation
5.1.2 Removal of insulation
WARNING!
Risk of penetrating the vessel surface
The centring pilot drill must be removed to avoid drilling through the vessel
wall.Vessel wall damage can compromise the integrity and wall thickness of
the vessel.
▶
Ensure proper speed and feed of drilling with removal of insulation to
measure depth of cut as required.
▶
Remove pilot drill as described below.

13WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
Continue drilling through the insulation. Drill by steps of approximately 25 mm [1 in] and
remove debris as required. Check that you have not reached the vessel surface after each
step.
5. Commissioning, operation

14 WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
Once the vessel surface has been reached with the hole saw, stop drilling and clean out
completely the cavity.
5. Commissioning, operation

15WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5.1.3 Vessel surface cleaning
Using a hand drill with auxillary side handle attachment begin cleaning the surface using a
50 mm [2 in] coarse grit sanding pad or coarse grit flap disc.
Surface must have scale and rust removed to provide proper adhesion of magnet. A bright
finish of the surface metal is preferred.
WARNING!
Risk of over grinding the vessel wall thickness
Vessel wall thickness is critical to proper operations. Over griding can cause
a weakened or thin section of the vessel wall.
▶
Use proper grinding methods and tools.
▶
Only remove scale and rust.
▶
Do not penetrate the vessel surface!
5. Commissioning, operation

16 WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5.1.4 Magnet installation
Remove the keeper from the magnet and discard. Check surface of magnet for any debris.
Place the magnet onto the vessel surface at the middle of the cavity.
5.1.5 Insulation packing
The top of the magnet must be packed with insulation under the cladding. Using a high-
temperature fiber blanket, e.g. Kaowool®, lightly pack the top of the magnet around the
stem.
5. Commissioning, operation

17WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5. Commissioning, operation
5.1.6 Cladding installation
A section of cladding must be installed over the magnet to provide environment protection.
The section of cladding is approximately 200 x 200 mm [8 x 8 in]. The middle of the
cladding section must have a hole to fit over the neck of the magnet assembly.
Place the section of cladding over the neck of the magnet assembly. Secure the cladding in
place using self tapping screws.

18 WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5.1.7 Sensor installation
Obtain the sensor and check that the spring and knurl fitting are secured on the sensor.
Insert sensor into the neck of the magnet assembly. Spring should align with the top of the
neck.
Insert the knurl fitting on top of the spring and tighten into the neck. The MI cable or flex
armour can be routed to the termination location to complete electrical connection.
5. Commissioning, operation

19WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
5.2 Electrical mounting
■
Only connect cables and lines in an unpowered state!
■
Fine-stranded leads must be finished with end splices.
■
Cables used must comply with the respective requirements in terms of stability and
temperature, strain and mobility. Over-working the cable may cause fraying of the wire.
■
Connection cables must be run so that any mechanical damage is prevented. This
includes areas where wear may occur resulting in fraying of the wire.
■
When bending connection cables it cannot be bent within 25 mm [1 in] of the transition
region.
■
Connection cables are not to be in contact with hot surfaces (i. e. hot manifolds).Where
this cannot be achieved proper selection of electrical wire, transition or high-temperature
sleeve shall be used.
Colour coding of thermocouple cable
Thermocouple Standard Positive Negative
Type J IEC 60584 Black White
ASTM E230 White Red
DIN 43714 Red Blue
Type K IEC 60584 Green White
ASTM E230 Yellow Red
Type E IEC 60584 Violet White
ASTM E230 Violet Red
Type T IEC 60584 Brown White
ASTM E230 Blue Red
Type N IEC 60584 Pink White
ASTM E230 Orange Red
5. Commissioning, operation

20 WIKA operating instructions model TC52-M
14502900.01 12/2021 EN/DE/FR/ES
EN
6. Faults
CAUTION!
Physical injuries and damage to property and the environment
If faults cannot be eliminated by means of the measures listed, the instrument
must be taken out of operation immediately.
▶
Ensure that pressure or signal is no longer present and protect against
accidental commissioning.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter 8.2
“Return”.
For contact details, please see chapter 1 “General information” or the back
page of the operating instructions.
Faults Causes Measures
No signal/cable break Mechanical load too high or
overtemperature
Replace probe with one of a
suitable design
Erroneous measured
values
Sensor drift caused by
overtemperature
Replace probe with one of a
suitable design
Erroneous measured
values (too low)
Entry of moisture into cable or
measuring insert
Replace probe with one of a
suitable design
Erroneous measured
values and response
times too long
Wrong mounting geometry, e.g.
contact pressure too small or
heat dissipation too high
Loosen compression fitting at the
neck tube, press the probe harder
on the surface, hold the probe
in this position and tighten the
compression fitting
Deposits on the sensor Remove deposits
Erroneous measured
values
(of thermocouples)
Parasitic voltages (thermal
voltages, galvanic voltage) or
wrong equalisation line
Use suitable compensating cable
Display of measured
value jumps
Cable break in connection cable
or loose contact caused by
mechanical overload
Replace probe or measuring insert
with a suitable design, for example
equipped with a strain relief or a
thicker conductor cross-section
Signal interference Stray currents caused by electric
fields or earth loops
Use shielded connection cables,
and increase the distance to
motors and power cables
Earth loops Eliminate potential differences by
using galvanically isolated repeater
power supplies or transmitters
6. Faults
Table of contents
Languages:
Other WIKA Measuring Instrument manuals

WIKA
WIKA GDM-RC-100-T User manual

WIKA
WIKA TR Series User manual

WIKA
WIKA 891.34 User manual

WIKA
WIKA FLC-2300 User manual

WIKA
WIKA FLC-608 User manual

WIKA
WIKA intelliTHERM TGT70 User manual

WIKA
WIKA mensor CPG2400 User manual

WIKA
WIKA TC80 User manual

WIKA
WIKA DG-10 User manual

WIKA
WIKA F73.100 User manual

WIKA
WIKA 212.20.100 User manual

WIKA
WIKA 732.14.100 per ATEX User manual

WIKA
WIKA 821 User manual

WIKA
WIKA NG 100 User manual

WIKA
WIKA 732.14 User manual

WIKA
WIKA DELTA-trans DPGT40 User manual

WIKA
WIKA CTD9300 Series User manual

WIKA
WIKA TC82 Operating instructions

WIKA
WIKA IPT-2 Series User manual

WIKA
WIKA A-10 User manual