WIKA intelliTHERM TGT70 User manual

ES
FR
DE
EN
Operating instructions
Betriebsanleitung
Mode d'emploi
Manual de instrucciones
Expansion thermometer with electrical output signal, model TGT70
Expansion thermometer with electrical output signal
Stainless steel version, model TGT70
Tensionsthermometer mit elektrischem Ausgangssignal
CrNi-Stahl-Ausführung, Typ TGT70
Thermomètres à dilatation avec signal de sortie électrique
Version tout inox, type TGT70
Termómetro a tensión con señal de salida eléctrica
Ejecución de acero inoxidable, modelo TGT70

2WIKA operating instructions model TGT70
ES
FR
DE
EN
14047194.03 03/2022 EN/DE/FR/ES
Operating instructions model TGT70
Page
3 - 16
Betriebsanleitung Typ TGT70
Seite
17 - 30
Mode d'emploi type TGT70
Page
31 - 44
Manual de instrucciones modelo TGT70
Página
45 - 57
© 2012 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Lire le mode d'emploi avant de commencer toute opération !
A conserver pour une utilisation ultérieure !
¡Leer el manual de instrucciones antes de comenzar cualquier trabajo!
¡Guardar el manual para una eventual consulta!

WIKA operating instructions model TGT70 3
EN
14047194.03 03/2022 EN/DE/FR/ES
1. General information 4
2. Safety 5
3. Specifications 8
4. Design and function 9
5. Transport, packaging and storage 10
6. Commissioning, operation 11
7. Maintenance and cleaning 13
8. Faults 14
9. Dismounting, return and disposal 15
Contents
Contents

4WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES
1. General information
■
The expansion thermometer described in the operating instructions has been
designed and manufactured using state-of-the-art technology. All components
are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001 and ISO 14001.
■
These operating instructions contain important information on handling the
instrument.Working safely requires that all safety instructions and work instructions
are observed.
■
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument‘s range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The manufacturer‘s liability is void in the case of any damage caused by using
the product contrary to its intended use, non-compliance with these operating
instructions, assignment of insufficiently qualified skilled personnel or unauthorised
modifications to the instrument.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modifications.
■
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant data sheet: TV 18.01
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
1. General information

WIKA operating instructions model TGT70 5
EN
14047194.03 03/2022 EN/DE/FR/ES
DANGER!
... identifies hazards caused by electric power. Should the safety
instructions not be observed, there is a risk of serious or fatal injury.
WARNING!
... indicates a potentially dangerous situation that can result in burns,
caused by hot surfaces or liquids, if not avoided.
Information
... points out useful tips, recommendations and information for efficient
and trouble-free operation.
2. Safety
WARNING!
Before installation, commissioning and operation, ensure that the
appropriate thermometer has been selected in terms of measuring range,
design and specific measuring conditions.
Non-observance can result in serious injury and/or damage to the
equipment.
Further important safety instructions can be found in the individual
chapters of these operating instructions.
2.1 Intended use
General-purpose expansion thermometer with electrical output signal for gaseous, liquid
and highly-viscous media.
At any point where the process temperature must be displayed locally and there is
a requirement to simultaneously transmit the signal to a central controller or remote
control room, the intelliTHERM®model TGT70 can be used.
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
1. General information / 2. Safety

6WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES
The technical specifications contained in these operating instructions must be
observed. Improper handling or operation of the instrument outside of its technical
specifications requires the instrument to be taken out of service immediately and
inspected by an authorised WIKA service engineer.
If the instrument is transported from a cold into a warm environment, the formation of
condensation may result in instrument malfunction. Before putting it back into operation,
wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
2.2 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient!
Improper handling can result in considerable injury and damage to
equipment.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
▶
Keep unqualified personnel away from hazardous areas.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and
knowledge of country-specific regulations, current standards and directives, are capab-
le of carrying out the work described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
2.3 Special hazards
WARNING!
For hazardous media such as oxygen, acetylene, flammable or toxic
gases or liquids, and refrigeration plants, compressors, etc., in addition
to all standard regulations, the appropriate existing codes or regulations
must also be followed.
2. Safety

WIKA operating instructions model TGT70 7
EN
14047194.03 03/2022 EN/DE/FR/ES
WARNING!
Residual media in the dismounted instrument can result in a risk to
persons, the environment and equipment.
Take sufficient precautionary measures.
2.4 Labelling, safety marking
Product label
Before mounting and commissioning the instrument, ensure you read the
operating instructions!
Model
2. Safety

8WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES
3. Specifications
Specifications intelliTHERM®model TGT70
Measuring principle Bourdon tube system
Nominal size in mm 63, 100
Filling medium Xylol, silicone oil or syltherm
Connection location Lower mount (LM) (radial)
Version B: back mount (axial)
Instrument version
H Instrument with capillary and rear surface mounting flange
M Instrument with capillary and surface mounting bracket
B Instrument with capillary, triangular bezel and mounting clamp
V Instrument with capillary and panel mounting flange
R Direct connection without capillary
Indication accuracy Class 2, EN 13190
Capillary Length in accordance with customer specifications (max. 10 m)
Ø 2 mm, stainless steel 1.4571, bending radius no less than 6 mm
Capillary entry lower mount
Capillary mounting Take care that the mounting is free from vibration
Case and bayonet ring Stainless steel
Connection plain, stainless steel 1.4571
Stem Ø 8 mm, stainless steel 1.4571
Active sensor length Depends on Ød and scale range
Dial Plastic sticker, white with logo
Aluminium, white, black lettering
Pointer Aluminium, black
Window Laminated safety glass
Ingress protection IP65 per EN/IEC 60529
Electronics
Output signal
Voltage output With Us= DC 5 V, ratiometric: 0.5 … 4.5 V
With Us= DC 12 … 32 V, non ratiometric: 0.5 … 4.5 V
Current output 4 … 20 mA, 2-wire
Power supply (Us)DC 5 V / DC 12 … 32 V
Electromagnetic compatibility (EMC) EN 61326 emission (group 1, class B) and
interference immunity (industrial application)
3. Specifications

WIKA operating instructions model TGT70 9
EN
14047194.03 03/2022 EN/DE/FR/ES
3.1 Output signal and permissible load
Voltage output (3-wire): RA > 5 kΩ
Current output (2-wire) 4 ... 20 mA:
RA ≤ (USIG - 10 V) / 0.02 A with RA in Ω and USIG in DC V
3.2 Electrical connection
Lateral terminal box, connector M12 or cable outlet
Pin assignment
Output signal UB+ UB- Signal
2-wire (current output) 12-
3-wire (voltage output) 123
Wire colour (for cable outlet)
UB+ UB- Signal
red black orange
Operate the instrument with shielded cable, and ground the shield on at least one end
of the lead, if the lines are longer than 30 m or laid outside of buildings.
This is an instrument for operation with extra-low voltages, which are separated from the
mains or for voltages higher than AC 50 V or DC 120 V. A connection to SELV circuits
or alternatively to circuits with a protective measure according to IEC 60364-4-41 is
recommended.
For further specifications see WIKA data sheet TV 18.01 and the order documentation.
4. Design and function
4.1 Description
The closed measuring system with liquid filling consists of a temperature sensor, a
capillary and a Bourdon tube.The volume change as a consequence of the heat applied
to the temperature sensor, and thus the pressure change in the closed system, is
transmitted through the capillary to the Bourdon tube and then indicated on the dial by
an angular deflection of the pointer shaft.
Through the combination of a mechanical measuring system and electronic signal
processing, the process temperature can be read securely, even if the power supply is lost.
USIG
Permissible range
4 ... 20 mA, 2-wire
3. Specifications / 4. Design and function

10 WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES
The built-in Bourdon tube system generates a rotational pointer movement that is
proportional to the temperature. An electronic angle encoder (non-contact, and
therefore completely free from wear and friction) determines the position of the
instrument pointer. From this, the electrical output signal is produced, proportional to the
temperature.
4.2 Scope of delivery
Cross-check scope of delivery with delivery note.
5. Transport, packaging and storage
5.1 Transport
Check the instrument for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
Storage temperature: -20 ... +60 °C
Avoid exposure to the following factors:
■
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Potentially explosive environments, flammable atmospheres
Store the instrument in its original packaging in a location that fulfils the conditions
listed above. If the original packaging is not available, pack and store the instrument as
described below:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag containing a
desiccant inside the packaging.
WARNING!
Before storing the instrument (following operation), remove any residual
media.This is of particular importance if the medium is hazardous to health,
e.g. caustic, toxic, carcinogenic, radioactive, etc
.
4. Design and function / 5. Transport, packaging and storage

WIKA operating instructions model TGT70 11
EN
14047194.03 03/2022 EN/DE/FR/ES
6. Commissioning, operation
WARNING!
▶
Only work on the gauge with the voltage disconnected.
▶
Avoid strong vibrations and mechanical impacts during operation of the
instrument.
▶
Avoid soiling and strong fluctuations of the ambient temperature.
▶
Cutting through the measuring line will destroy the instrument and thus
lead to a wrong electrical output signal.
▶
Cutting through the capillary will destroy the instrument and lead to the
leakage of the thermometric liquid (Xylol, Syltherm or silicone oil).
6.1 Installation conditions
Before mounting the sensor, check whether the sensor material used (specified in the
delivery note) is chemically resistant/neutral to the medium being measured.This also
applies to thermowells.
The working pressure at the thermowell stem is permitted up to a maximum of 25 bar
(static).
Make sure that the suitable accessories required for fastening the case of the instrument
into the panel are available. The accessories are either fitted or located in a bag
fastened to the thermometer.
■
Expose, if possible, the entire length of the stem to the temperature to be measured.
However, at least the length of the active part that corresponds to the length of the
expansion vessel (active length).
■
The temperature probe in pipelines or other measuring points must be angled as far
towards the flow as possible.
■
When using thermowells, please observe that the stem must not touch the bottom of
the thermowell.
■
Protect capillaries against bending.The smallest bending radius of the capillary
should not be less than 6 mm. Any bending or discontinuities of the capillary may
result in the failure of the instrument.
■
If the sensor is is installed in a location which is exposed to shock and vibration, it
is essential that the capillary is coiled in several loops between the last attachment
point and the sensor, and free from tension. Any excess length must also be coiled,
since cutting would render the instrument unserviceable.
■
Avoid welding or soldering and permanent fastening of the capillary as this could
seriously damage the capillary and impair the operation of the instrument.
■
The dial casing must be mounted free from vibration.
■
If necessary, it is possible to isolate the instrument from the mounting point by instal-
ling a flexible connection line between the measuring point and the thermometer and
mounting the instrument on a suitable bracket.
If this is not possible, do not exceed the following limits:
Frequency range < 150 Hz
Acceleration < 0.7 g (7 m/s2)
6. Commissioning, operation

12 WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES
6.2 Ambient conditions
Unless the ingress protection of the thermometer‘s housing is specifically identified
in the order confirmation, the instrument must be protected from humid air and other
aggressive atmospheres.
To ensure the best measuring accuracy possible, the ambient temperature on the
indicator case should be between 0 ... 40 °C. Higher or lower ambient temperatures can
cause indication errors.
6.3 Indicator check
Indicator checks should only be carried out in comparison with a more accurate instru-
ment or, if possible, with a calibrated instrument.The temperature during the check must
remain constant. Fluctuating temperatures can lead to reading errors caused by the
different response times of the sensors.
Before checking the indicators of thermometers without thermowells, a waiting time of at
least 5 minutes with the stem's full length properly inserted is required to allow tempera-
ture equalisation.
Temperature checks using thermometers with sensors assembled with thermowells
can, in many cases, only result in a reference temperature at the instrument under
test, due to permanent heat dissipation caused by the thermowell. In temperature
applications fitted with static checking thermometers (e.g. pipeline systems), permanent
offsets in reference temperatures can also be caused by the measuring path.
Ambient temperatures around the indicator case which differ substantially from room
temperature can lead to steady indication errors with constant ambient temperatures
and to varying indication errors with fluctuating ambient temperatures.
Permanent indication errors caused by the ageing of the measuring system can be
ignored, as they only account for a fraction of the indication accuracy.
6.4 Zero adjustment
A zero adjustment may only be carried out by the manufacturer or in adequately equip-
ped workshops by qualified persons.
■
The electrical connection must only be made by qualified skilled personnel.
■
Connect the switch via cable terminal box.
■
Conductor cross section max. 1.5 mm2
■
The terminal assignment is stated on the product label of the thermometer.
6. Commissioning, operation

WIKA operating instructions model TGT70 13
EN
14047194.03 03/2022 EN/DE/FR/ES
7. Maintenance and cleaning
7.1 Maintenance
This expansion thermometer is maintenance-free.
The indicator should be checked once or twice every year. To do this the instrument
must be disconnected from the process and checked using a temperature calibrator.
Repairs must exclusively be carried out by the manufacturer or by appropriately quali-
fied personnel.
7.2 Cleaning
CAUTION!
▶
Prior to cleaning, switch off and disconnect the instrument from the
mains.
▶
Clean the instrument with a moist cloth (in soapy water).
▶
Electrical connections must not come into contact with moisture.
▶
Wash or clean the dismounted instrument before returning it, in order to
protect persons and the environment from exposure to residual media.
▶
Residual media in the dismounted instrument can result in a risk to
persons, the environment and equipment.Take sufficient precautionary
measures.
For information on returning the instrument see chapter 9.2 "Returns".
7. Maintenance and cleaning

14 WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES
8. Faults
Faults Causes Measures
No output signal No power supply or cable
break
Check voltage supply and cables,
replace any defective components
found
Electronics failure due to too
high power supply or through
external voltage
Return the instrument to the
manufacturer for repair
Too high, constant
output signal upon
change in temperature
Electronics failure due to too
high power supply or through
external voltage
Return the instrument to the
manufacturer for repair
Signal span too small Power supply too low Rectify the power supply
Load too high Stay within the max. permissible load
Zero point maladjusted Return the instrument to the
manufacturer for repair
Zero point too low/
too high
Zero point maladjusted Return the instrument to the
manufacturer for repair
Output signal instable Power supply too low Rectify the power supply
CAUTION!
If faults cannot be eliminated by means of the measures listed above, shut
down the instrument immediately, and ensure that pressure and/or signal
are no longer present, and secure the instrument from being put back into
operation inadvertently.
In this case, contact the manufacturer.
If a return is needed, please follow the instructions given in chapter
9.2 "Return".
8. Faults

WIKA operating instructions model TGT70 15
EN
14047194.03 03/2022 EN/DE/FR/ES
9. Dismounting, return and disposal
WARNING!
Residual media in the dismounted instrument can result in a risk to
persons, the environment and equipment.
Take sufficient precautionary measures.
9.1 Dismounting
WARNING!
Risk of burns!
Let the instrument cool down sufficiently before dismounting!
During dismounting there is a risk of dangerously hot pressure media
escaping.
Only disconnect the expansion thermometer once the system has been depressurised!
9.2 Return
WARNING!
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous
substances (acids, bases, solutions, etc.).
When returning the instrument, use the original packaging or a suitable transport package.
To avoid damage:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag containing a desiccant inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring instrument.
Information on returns can be found under the heading "Service" on our
local website.
9.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specific waste disposal regulations.
9. Dismounting, return and disposal

16 WIKA operating instructions model TGT70
EN
14047194.03 03/2022 EN/DE/FR/ES

14047194.03 03/2022 EN/DE/FR/ES
1. Allgemeines 18
2. Sicherheit 19
3. Technische Daten 22
4. Aufbau und Funktion 23
5. Transport, Verpackung und Lagerung 24
6. Inbetriebnahme, Betrieb 25
7. Wartung und Reinigung 27
8. Störungen 28
9. Demontage, Rücksendung und Entsorgung 29
WIKA-Betriebsanleitung Typ TGT70 17
DE
Inhalt
Inhalt

14047194.03 03/2022 EN/DE/FR/ES
18 WIKA-Betriebsanleitung Typ TGT70
DE
1. Allgemeines
■
Das in der Betriebsanleitung beschriebene Tensionsthermometer wird nach dem
aktuellen Stand der Technik konstruiert und gefertigt. Alle Komponenten unterliegen
während der Fertigung strengen Qualitäts- und Umweltkriterien. Unsere Manage-
mentsysteme sind nach ISO 9001 und ISO 14001 zertifiziert.
■
Diese Betriebsanleitung gibt wichtige Hinweise zum Umgang mit dem Gerät.Voraus-
setzung für sicheres Arbeiten ist die Einhaltung aller angegebenen Sicherheitshin-
weise und Handlungsanweisungen.
■
Die für den Einsatzbereich des Gerätes geltenden örtlichen Unfallverhütungsvor-
schriften und allgemeinen Sicherheitsbestimmungen einhalten.
■
Die Betriebsanleitung ist Produktbestandteil und muss in unmittelbarer Nähe des
Gerätes für das Fachpersonal jederzeit zugänglich aufbewahrt werden.
■
Das Fachpersonal muss die Betriebsanleitung vor Beginn aller Arbeiten sorgfältig
durchgelesen und verstanden haben.
■
Die Haftung des Herstellers erlischt bei Schäden durch bestimmungswidrige Verwen-
dung, Nichtbeachten dieser Betriebsanleitung, Einsatz ungenügend qualifizierten
Fachpersonals sowie eigenmächtiger Veränderung am Gerät.
■
Es gelten die allgemeinen Geschäftsbedingungen in den Verkaufsunterlagen.
■
Technische Änderungen vorbehalten.
■
Weitere Informationen:
- Internet-Adresse: www.wika.de / www.wika.com
- zugehöriges Datenblatt: TV 18.01
- Anwendungsberater: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
Symbolerklärung
WARNUNG!
… weist auf eine möglicherweise gefährliche Situation hin, die zum Tod
oder zu schweren Verletzungen führen kann, wenn sie nicht gemieden
wird.
VORSICHT!
… weist auf eine möglicherweise gefährliche Situation hin, die zu gering-
fügigen oder leichten Verletzungen bzw. Sach- und Umweltschäden
führen kann, wenn sie nicht gemieden wird.
1. Allgemeines

14047194.03 03/2022 EN/DE/FR/ES
WIKA-Betriebsanleitung Typ TGT70 19
DE
GEFAHR!
… kennzeichnet Gefährdungen durch elektrischen Strom. Bei Nichtbeach-
tung der Sicherheitshinweise besteht die Gefahr schwerer oder tödlicher
Verletzungen.
WARNUNG!
… weist auf eine möglicherweise gefährliche Situation hin, die durch
heiße Oberflächen oder Flüssigkeiten zu Verbrennungen führen kann,
wenn sie nicht gemieden wird.
Information
… hebt nützliche Tipps und Empfehlungen sowie Informationen für einen
effizienten und störungsfreien Betrieb hervor.
2. Sicherheit
WARNUNG!
Vor Montage, Inbetriebnahme und Betrieb sicherstellen, dass das richtige
Thermometer hinsichtlich Messbereich, Ausführung und spezifischen
Messbedingungen ausgewählt wurde.
Bei Nichtbeachten können schwere Körperverletzungen und/oder
Sachschäden auftreten.
Weitere wichtige Sicherheitshinweise befinden sich in den einzelnen
Kapiteln dieser Betriebsanleitung.
2.1 Bestimmungsgemäße Verwendung
Universell einsetzbares Tensionsthermometer mit elektrischem Ausgangssignal für
gasförmige, flüssige und hochviskose Messstoffe.
Überall dort, wo die Prozesstemperatur vor Ort angezeigt werden muss und gleichzei-
tig eine Signalübertragung an die zentrale Steuerung oder Fernwarte gewünscht wird,
findet das intelliTHERM®Typ TGT70 seinen Einsatz.
Das Gerät ist ausschließlich für den hier beschriebenen bestimmungsgemäßen
Verwendungszweck konzipiert und konstruiert und darf nur dementsprechend verwen-
det werden.
1. Allgemeines / 2. Sicherheit

14047194.03 03/2022 EN/DE/FR/ES
20 WIKA-Betriebsanleitung Typ TGT70
DE
Die technischen Spezifikationen in dieser Betriebsanleitung sind einzuhalten. Eine
unsachgemäße Handhabung oder ein Betreiben des Gerätes außerhalb der techni-
schen Spezifikationen macht die sofortige Stilllegung und Überprüfung durch einen
autorisierten WIKA-Servicemitarbeiter erforderlich.
Wird das Gerät von einer kalten in eine warme Umgebung transportiert, so kann durch
Kondensatbildung eine Störung der Gerätefunktion eintreten.Vor einer erneuten
Inbetriebnahme die Angleichung der Gerätetemperatur an die Raumtemperatur abwar-
ten.
Ansprüche jeglicher Art aufgrund von nicht bestimmungsgemäßer Verwendung sind
ausgeschlossen.
2.2 Personalqualifikation
WARNUNG!
Verletzungsgefahr bei unzureichender Qualifikation!
Unsachgemäßer Umgang kann zu erheblichen Personen- und
Sachschäden führen.
▶
Die in dieser Betriebsanleitung beschriebenen Tätigkeiten nur durch
Fachpersonal nachfolgend beschriebener Qualifikation durchführen
lassen.
▶
Unqualifiziertes Personal von den Gefahrenbereichen fernhalten.
Fachpersonal
Das Fachpersonal ist aufgrund seiner fachlichen Ausbildung, seiner Kenntnisse der
Mess- und Regelungstechnik und seiner Erfahrungen sowie Kenntnis der landesspezi-
fischen Vorschriften, geltenden Normen und Richtlinien in der Lage, die beschriebenen
Arbeiten auszuführen und mögliche Gefahren selbstständig zu erkennen.
Spezielle Einsatzbedingungen verlangen weiteres entsprechendes Wissen, z. B. über
aggressive Medien.
2.3 Besondere Gefahren
WARNUNG!
Bei gefährlichen Messstoffen wie z. B. Sauerstoff, Acetylen, brennbaren
oder giftigen Stoffen, sowie bei Kälteanlagen, Kompressoren etc.
müssen über die gesamten allgemeinen Regeln hinaus die einschlägigen
Vorschriften beachtet werden.
2. Sicherheit
Table of contents
Languages:
Other WIKA Measuring Instrument manuals

WIKA
WIKA DIH50 User manual

WIKA
WIKA CTD9300 Series User manual

WIKA
WIKA GDM-100-T User manual

WIKA
WIKA 732.14.100 per ATEX User manual

WIKA
WIKA CPH6200 User manual

WIKA
WIKA TC52-M User manual

WIKA
WIKA TR12 Series Operating instructions

WIKA
WIKA A-IAI-1 User manual

WIKA
WIKA 230.15 User manual

WIKA
WIKA SERIES CPG 2500 User manual
Popular Measuring Instrument manuals by other brands

VWR
VWR OX 4100 H operating manual