WIKA F1811 User manual

Operating instructions
Betriebsanleitung
EN
DE
Model F1821 Model F1848
Model F1811 Model F1814 Model F1818
Compression force transducer
Models F1811, F1814, F1818, F1821, F1848
Druckkraftaufnehmer
Typen F1811, F1814, F1818, F1821, F1848

2
79106246.02 11/2022 EN/DE
WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
DE Betriebsanleitung, Typen F1811, F1814,
F1818, F1821, F1848 Seite 27 - 51
© 11/2022 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®is a registered trademark in various countries.
WIKA®ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Operating instructions, models F1811,
F1814, F1818, F1821, F1848 Page 4 - 26

3WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
Contents
Contents
1. General information 4
2. Design and function 5
2.1 Overview. . . . . . . . . . . . . . . . . . . . . . . .5
2.2 Description . . . . . . . . . . . . . . . . . . . . . . .5
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . .5
3. Safety 6
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . .6
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . .6
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . .7
3.4 Responsibility of the operator . . . . . . . . . . . . . . . . .7
3.5 Personal protective equipment . . . . . . . . . . . . . . . .8
3.6 Personnel qualification . . . . . . . . . . . . . . . . . . .8
3.7 Labelling, safety marks . . . . . . . . . . . . . . . . . . .9
4. Transport, packaging and storage 10
4.1 Transport. . . . . . . . . . . . . . . . . . . . . . . 10
4.2 Packaging and storage . . . . . . . . . . . . . . . . . . 10
5. Commissioning, operation 11
5.1 Precautions before commissioning . . . . . . . . . . . . . . 11
5.2 Mounting instructions . . . . . . . . . . . . . . . . . . 11
5.3 Installation of the compression force transducer . . . . . . . . . 12
5.4 Electrical connection. . . . . . . . . . . . . . . . . . . 12
6. Faults 14
7. Maintenance and cleaning 15
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . 15
7.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . 15
7.3 Recalibration . . . . . . . . . . . . . . . . . . . . . 15
8. Dismounting, return and disposal 15
8.1 Dismounting . . . . . . . . . . . . . . . . . . . . . 16
8.2 Return. . . . . . . . . . . . . . . . . . . . . . . . 16
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . 16
9. Specifications 17
9.1 Approvals . . . . . . . . . . . . . . . . . . . . . . 23
10. Accessories 23
10.1 Cable amplifier, model B1940 . . . . . . . . . . . . . . . 23
Annex: EU declaration of conformity 24

4WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
1. General information
1. General information
■
The compression force transducers described in the operating instructions have
been designed and manufactured using state-of-the-art technology. All components
are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001.
■
These operating instructions contain important information on handling the
instruments.Working safely requires that all safety instructions and work instructions
are observed.
■
Observe the relevant local accident prevention regulations and general safety
regulations for the instruments' range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions on to the next user or owner of the instrument.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modifications.
■
Factory calibrations / DAkkS calibrations are carried out in accordance with
international standards.
Further information:
- Internet address: www.wika.com
- Relevant data sheet: FO 51.56 (F1811)
FO 51.57 (F1814)
FO 51.58 (F1818)
FO 51.59 (F1821)
FO 51.76 (F1848)
- Application consultant: Tel.: +49 9372 132-0
info@wika.com
Abbreviations, definitions
2-wire The two connection leads are used for the voltage supply.
The measuring signal also provides the supply current.
3-wire Two connection leads are used for the voltage supply.
One connection lead is used for the measuring signal.
UB (+) Positive power supply terminal
UB (-) Negative power supply terminal
S (+) Positive signal output
S (-) Negative signal output
Shield Case
SG Strain gauge

5WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
F
2. Design and function
2. Design and function
The compression force transducer is designed for measuring static and dynamic
compression forces.The compression force transducer consists of a measuring spring
on which strain gauges are applied.The measuring spring is made of corrosion-resistant
stainless steel and is elastically deformed by a tension/compression force introduced
in the force direction.The resulting mechanical tensions are measured by the strain
gauges and output as an electrical output signal.
The strain gauges are arranged so that two of them are tensioned and the other two are
compressed when a force is applied to the transducer.
The transducer circuit includes correction and compensation resistors to eliminate
unwanted effects on the zero signal and characteristic value.
Force introduction (F) on crowned surface
Measuring spring
Electrical connection
2.1 Overview
2.2 Description
2.3 Scope of delivery
■
Compression force transducer
■
Operating instructions

6WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
Model
Version
Design
F1811 Standard version, measuring ranges 0 ... 2.5 kN to 0 ... 300 kN
[0 ... 562 lbf to 0 ... 67,443 lbf]
F1814 Miniature design, measuring ranges 0 ... 30 N to 0 ... 1,000 N
[0 ... 6.74 lbf to 0 ... 224.8 lbf]
F1818 Miniature design, measuring ranges 0 ... 50 N to 0 ... 20 kN
[0 ... 11.24 lbf to 0 ... 4.5 lbf]
F1821 Standard version, measuring ranges 0 ... 200 N to 0 ... 100 kN
[0 ... 45 lbf to 0 ... 22.48 lbf]
F1848 Standard version, measuring ranges from 1 t [from 2,205 lbs]
Force transducer for measuring compression forces, for example in presses.The
product is designed for use both outdoors and inside buildings.
Compression force transducers can measure forces in one direction. The measured
force is output as an electrical signal. These devices are designed for operation in
industrial environment. In other environments, e.g. residential or commercial, they may
interfere with other equipment. In this case, the operator may be required to implement
appropriate measures.
Only use the compression force transducers in applications that are within the technical
performance limits (e.g. max. ambient temperature, material compatibility, etc.). For
performance limits, see chapter 9 “Technical data”.
The compression force transducers are not approved for use in explosive areas or for
safety applications.
3. Safety
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient
and trouble-free operation.
3.2 Intended use

7WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
3. Safety
The compression force transducers are designed exclusively for the intended use which
is described here and may only be used accordingly.
Handle electronic precision measuring instruments with the required care (protect from
moisture, impacts, strong magnetic elds, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
This instrument is intended for connection to an external PELV circuit that meets the
requirements of UL/IEC/EN 61010-1 Section 9.4 (limited-energy circuit).The manufactu-
rer shall not be liable for claims of any type based on operation contrary to the intended
use.The compression force transducers are designed for indoor and outdoor use, at
altitudes of to 2,500 m [8,202 feet] above sea level.The pollution degree in accordance
with UL certication is Degree 4: “Electrical equipment for outdoor use”.to the intended
use.
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and
injuries.
▶
Refrain from unauthorised modifications to the instrument.
Any use beyond or different to the intended use is considered as improper use.
3.4 Responsibility of the operator
The instrument is used in the industrial sector. The operator is therefore responsible for
legal obligations regarding safety at work.
The safety instructions within these operating instructions, as well as the safety, accident
prevention and environmental protection regulations for the application area of the
instrument must be maintained.
The operator is obliged to maintain the product label in a legible condition.
To ensure safe working on the instrument, the operating company must ensure
■
that suitable first-aid equipment is available and aid is provided whenever required
■
that the skilled electrical personnel are regularly instructed in all topics regarding
occupational safety, first aid and environmental protection and know the operating
instructions and, in particular, the safety instructions contained therein
■
that the instrument is suitable for the particular application in accordance with its
intended use
■
that personal protective equipment is available.

8WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
3.6 Personal protective equipment
The requirements for the required protective equipment result from the ambient
conditions at the place of use, other products or the connection to other products.
The requisite personal protective equipment must be provided by the operating
company.The operator is in no way relieved of his obligations under labour law for the
safety and the protection of workers’ health.
The design of the personal protective equipment must take into account all operating
parameters of the place of use.
3.5 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to
property.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their
technical training, know-how and experience as well as their knowledge of country-
specific regulations, current standards and directives, are capable of carrying out work
on electrical systems and independently recognising and avoiding potential hazards.
The skilled electrical personnel have been specifically trained for the work environment
they are working in and know the relevant standards and regulations. The skilled
electrical personnel must comply with current legal accident prevention regulations.
3. Safety

9WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
3. Safety
3.7 Labelling, safety marks
Product label
11
Model
Manufacturer's logo
Serial number
Ingress protection per IEC/EN 60529
Pin assignment
Manufacturer's address
Force direction
Excitation voltage
Output signal
Rated force
Article number
Before mounting and commissioning the instrument, ensure you read
the operating instructions!
11

10 WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the compression force transducer for any damage that may have been caused by
transport. Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during internal
transport, proceed carefully and observe the symbols on the packaging.
▶
With internal transport, observe the instructions in chapter 4.2
“Packaging and storage”.
As precision measuring instruments, transducers require careful handling during
transport and mounting. Load impacts during transport (e.g. hitting a hard surface) can
lead to permanent damage, resulting in measured errors in the subsequent measuring
operation.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
Permissible conditions at the place of storage:
■
Storage temperature: -20 ... +80 °C [-4 … +176 °F]
■
Humidity: 35 ... 85 % relative humidity (non-condensing)
Avoid exposure to the following factors:
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Dust, dirt, and other objects may not be deposited in such a way that they form a
force shunt with the measuring spring, since this will falsify the measuring signal.

11WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
5. Commissioning, operation
5. Commissioning, operation
5.1 Precautions before commissioning
■
Compression force transducers are sensitive measuring instruments and must be
handled with appropriate care.
■
With the commissioning of compression force transducers, make sure they are on a
level, clean and grease-free contact surface.The flatness shall be 0.02 mm [0.0008 in].
■
The contact surface must have a hardness of at least HRC 43, be ground or lapped,
have a surface roughness of Rz 4 and must not bend under load.
5.2 Mounting instructions
CAUTION!
Damage to the instrument through improper commissioning
Torsional torques, eccentric loading and transverse loading or lateral
forces cause measured errors and may permanently damage the
compression force transducer. Incorrect loading can lead to permanent
damage resulting in a zero point offset in the unloaded state.
▶
The force introduction must always be applied centrally to the crowned
surface of the compression force transducer.
▶
Torsional and transverse forces must be avoided. Transverse loading
and lateral forces also include the corresponding components of the
measurands that are introduced slanted.
▶
During installation of the compression force transducer, the output
signal (force value) must always be monitored to avoid mechanical
overloading.
▶
The compression force transducer should only be loaded with the
intended type of load-bearing and in the intended force direction.
▶
Overloading must be excluded at all times.
▶
The rated temperature range must be observed.Temperature-related
measured errors must be excluded.

12 WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
F
5. Commissioning, operation
5.3 Installation of the compression force transducer
■
Place the compression force tranducer
with its underside on the contact
surface of the measuring object.
■
Optionally depending on model, the
compression force transducer can
also be fastened to the contact surface
or screws of the test object with
suitable adhesive.
■
Load the compression force transducer.
Compression force transducer
Contact surface
5.4 Electrical connection
A shielded, low-capacitance measuring cable is attached to the force transducer.
To prevent interferences from coupling into the system, observe the following
information:
■
A cable amplifier can be fitted optionally.
■
Ground the shield of the measuring cable.
■
Avoid stray fields of transformers, motors and contactors.
■
When using extensions, only shielded and low-capacitance cables should be used.
The permitted maximum and minimum lengths of cable are defined in ISO 11898-2.
Care should be taken also to ensure a high-quality connection of the shielding.
■
Do not install measuring cables in parallel to 3-phase-current cables and control
cables.
■
Transducers, amplifiers and processing or display units must not be grounded
several times.
The pin assignments of the cable and the cable amplifier (fitted optionally) can be found
on the product label.

13WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
Electrical connection
Conductor Cable colour
UB (+) Red
UB (-) Black
S (+) Green
S (-) White
Shield Shield
5. Commissioning, operation
Pin assignment
Ex
c+
Sig+
Ex
c-
Sig-
UB (+)
S (+)
UB (-)
S (-)
Shield
Electrical connection
Conductor Cable Colour
UB (+) Red
UB (-) Black
S (+) Green
Shield Yellow-green
Model F1848
red
green
black
white
Electrical connection, M12 x 1, 4-pin
Conductor
4 ... 20 mA / 20 ... 4 mA
(redundant)
Connector 1 Connector 2
UB (+) 1 1
UB (-) 3 3
S (+): Channel 1
4 -
S (+): Channel 2
- 4
Shield Case Case
Output signal 4 ... 20 mA, 3-wire
Version: analogue output, redundant, opposing
Version: cable output
S (+) green
UB (-) black
UB (+) red
Version: cable output
Models F1811, F1814, F1818, F1821

14 WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
6. Faults
CAUTION!
Physical injuries, damage to property
If faults cannot be eliminated by means of the measures listed, the
compression force transducer must be taken out of operation immediately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter
8.2 “Return”.
For contact details, please see chapter 1 “General information” or the back
page of the operating instructions.
Faults Causes Measures
No output signal No or wrong supply voltage,
current pulse
Correct the supply voltage
Cable break Check the continuity
No or wrong output signal Wrong pin assignment Check cable assignment
Deviating zero point signal Overload, load oset, wrong
connection
Consult the manufacturer
Constant output signal
when changing force
Wrong pin assignment Check cable assignment
Mechanical overloading Consult the manufacturer
Signal span varies EMC interference sources
in the environment, e.g.,
frequency converter
Shield instrument; cable
shielding; remove source of
interference
Signal span drops/too small Mechanical overloading Consult the manufacturer
Defective instruments must be returned to the manufacturer.
6. Faults

15WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
7. Maintenance and cleaning
7. Maintenance and cleaning
7.1 Maintenance
This instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 “Accessories”).
7.2 Cleaning
1. Prior to cleaning, disconnect the compression force transducer from the voltage
supply and dismount it.
2. Clean the compression force transducer with a cloth.
Electrical connections must not come into contact with moisture!
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects for cleaning.
7.3 Recalibration
DAkkS certicate - ocial certicates:
We recommend that the compression force transducer is regularly recalibrated by
the manufacturer, with time intervals of approx. 24 months.The basic settings will be
corrected if necessary.
For contact details, please see chapter 1 “General information” or the back
page of the operating instructions.

16 WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
8. Dismounting, return and disposal
8. Dismounting, return and disposal
8.1 Dismounting
Release the load from the compression force transducer, disconnect from the power
and remove from the contact surface.
8.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances
(acids, bases, solutions, etc.) and must therefore be cleaned before being returned.
When returning the instrument, use the original packaging or a suitable transport
packaging.
To avoid damage:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag containing a desiccant inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring instrument.
Information on returns can be found under the heading “Service” on our
local website.
8.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specific waste disposal regulations.
Do not dispose of with household waste. Ensure a proper disposal in
accordance with national regulations.

17WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
9. Specifications
9. Specifications
Model F1811
Rated force Fnom kN 2,5 5 10 20 30 50 100 200 300
Rated force Fnom lbf 552 1,124 2,250 4,500 6,744 11,240 22,500 45,000 67,443
Relative linearity error dlin 0.1 % Fnom
Relative creep, 30 min. at
Fnom
0.1 % Fnom
Relative reversibility
error v
0.1 % Fnom
Relative span in
unchanged mounting
situation brg
0.1 % Fnom
Relative deviation of zero
signal dS, 0
±2 % Fnom
Temperature eect on
zero signal TK0
0.05 % Fnom/10 K
Temperature eect on
characteristic value TKC
0.05 % Fnom/10 K
Force limit FL150 % Fnom
Breaking force FB200 % Fnom
Material of the measuring
body
Stainless steel
Rated temperature range
BT, nom
-10 … +60 °C [-14 … +140 °F]
Operating temperature
range BT, G
-20 … +80 °C [-4 … +176 °F]
Input resistance Re750 ±10 Ω
Output resistance Ra700 ±5 Ω
Insulation resistance Ris ≥ 5,000 MΩ/DC 100 V
Output signal (rated
characteristic value) Cnom
1.5 ±0.15 mV/V
Electrical connection Cable Ø 5 × 5,000 mm [Ø 0.197 in × 197 in]
Voltage supply DC 10 V (max. 15 V)
Ingress protection
(per IEC/EN 60529)
IP66
Weight
2.5 kN ... 100 kN 1.2 kg [2.65 lbs]
200 kN 1.6 kg [3.53 lbs]
300 kN 3.2 kg [7.05 lbs]

18 WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
9. Specifications
Dimensions in mm [in]
Model F1811
SR = spherical radius
Rated force
in kN [lbf]
Dimensions in mm [in]
ØA B C ØD SR E F G ØK H L M
2.5 [552]
5 [1,124]
10 [2,250]
20 [4,500]
30 [6,744]
50 [11,240]
100 [22,500]
82
[3.23]
44
[1.73]
32
[1.26]
22
[0.87]
50
[1.97]
14
[0.55]
13
[0.51]
68
[2.67]
M8 5,000
[197]
15.7
[0.62]
200
[45,000]
100
[3.94]
48
[1.89]
36
[1.41]
28
[1.10]
100
[3.94]
14
[0.55]
13
[0.51]
80
[3.15]
M8 5,000
[197]
15.7
[0.62]
300
[67,443]
126
[4.96]
54
[2.12]
40
[1.57]
35
[1.38]
100
[3.94]
14
[0.55]
13
[0.51]
101
[3.98]
M10 5,000
[197]
15.7
[0.62]

19WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
9. Specifications
Dimensions in mm [in]
SR = spherical radius
Model F1814
Model F1814
Rated force Fnom N50 100 200 300 500 1,000
Rated force Fnom lbf 11.24 22.5 45 67.44 112 225
Relative linearity error dlin 1 % Fnom
Relative reversibility error v 0.5 % Fnom
Relative span in unchanged
mounting situation brg
0.5 % Fnom
Limit force FL120 % Fnom
Breaking force FB150 % Fnom
Material of the measuring body
30 N Aluminium
≥ 50 N Stainless steel
Service temperature range BT, G -20 … +80 °C [-4 … +176 °F]
Input resistance Re1,030 ±80 Ω
Output resistance Ra1,030 ±80 Ω
Insulation resistance Ris ≥ 5,000 MΩ/DC 100 V
Output signal (rated characteristic
value) Cnom
1.0 ±0.1 mV/V
Electrical connection Cable Ø 2 × 3,000 mm [Ø 0.1 in x 118 in]
Voltage supply DC 5 V (max. 7 V)
Ingress protection (per IEC/EN 60529)
IP64
Weight 0.1 kg [0.22 lbs]
Rated force
in N [lbf]
Dimensions in mm [in]
ØA ØB SR C D E F G ØH L
50 [11.24]
100 [22.5]
1.5
[0.06]
10
[0.39]
5
[0.19]
6
[0.24]
5.2
[0.20]
50
[1.97]
150
[5.90]
8
[0.31]
3,000
[118]
200 [45]
300 [67.44]
2
[0.08]
8
[0.31]
500 [112]
1,000 [225]
2.5
[0.1]
10
[0.39]

20 WIKA operating instructions, models F1811, F1814, F1818, F1821, F1848
EN
79106246.02 11/2022 EN/DE
9. Specifications
Model F1818
Rated force Fnom kN 0.05 0.1 0.2 0.5 1 2 5
Rated force Fnom lbf 11.24 22.5 45 112 225 450 1,124
Relative linearity error dlin 0.5 % Fnom
Relative reversibility error v 0.5 % Fnom
Relative span in unchanged mounting
situation brg
0.1 % Fnom
Relative deviation of zero signal dS, 0
±2 % Fnom
Force limit FL150 % Fnom
Breaking force FB200 % Fnom
Material of the measuring body Stainless steel
Operating temperature range BT, G -20 … +80 °C [-4 … +176 °F]
Input resistance Re350 ±10 Ω
Output resistance Ra350 ±5 Ω
Insulation resistance Ris ≥ 5,000 MΩ/DC 100 V
Output signal (rated characteristic
value) Cnom
1.5 ±0.15 mV/V
Electrical connection Cable Ø 2 × 3,000 mm [Ø 0.1 in × 118 in]
Voltage supply DC 5 V (max. 10 V)
Ingress protection (per IEC/EN 60529) IP65
Weight 0.1 kg [0.22 lbs]
Dimensions in mm [in]
SR = spherical radius
Model F1818
Rated force in
kN [lbf]
Dimensions in mm [in]
ØA ØB ØC
SR
D
E F G H L M
0.05 [11.24] / 0.1 [22.5]
0.2 [45] / 0.5 [112]
1 [225] / 2 [450]
5 [1,124]
16.5
[0.65]
5
[0.19]
20
[0.79]
20
[0.79]
11
[0.43]
8.5
[0.33]
5
[0.19]
7.5
[0.26]
3.000
[118]
M2.5
This manual suits for next models
4
Table of contents
Languages:
Other WIKA Transducer manuals
Popular Transducer manuals by other brands

turck
turck IMX12-FI01-1SF-1I1R Series quick start guide

Mianyang Weibo Electronic
Mianyang Weibo Electronic WBP214P75 user manual

Tecsis
Tecsis P3230 instruction manual

HBM
HBM CFT+ Series operating manual

GE
GE BAKER HUGHES DPS 5000 Series instruction manual

Omega Engineering
Omega Engineering PX791 Series user guide