WIKA F2301 User manual

Operating instructions
Betriebsanleitung
EN
DE
Tension/compression force transducer
Models F2301, F23C1, F23S1, F2303, F2304
Zug-/Druckkraftaufnehmer
Typen F2301, F23C1, F23S1, F2303, F2304
Example: Models F2301, F2304

2
ADPR1X914096.01 06/2020 EN/DE
WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
EN
DE
Operating instructions models F2301, F23C1, F23S1, F2303, F2304
Page
3 - 48
Betriebsanleitung Typen F2301, F23C1, F23S1, F2303 ,F2304
Seite
49 - 95
© 06/2020 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®and tecsis®are registered trademarks in various countries.
WIKA®and tecsis®sind geschützte Marken in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!

3WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
Contents
Contents
1. General information 4
2. Design and function 5
2.1 Overview models F2301, F23C1, F23S1 . . . . . . . . . . . . .5
2.2 Overview model F2303 . . . . . . . . . . . . . . . . . . .6
2.3 Overview model F2304 . . . . . . . . . . . . . . . . . . .6
2.4 Description . . . . . . . . . . . . . . . . . . . . . . .7
2.5 Scope of delivery . . . . . . . . . . . . . . . . . . . . .7
3. Safety 7
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . .7
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . .8
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . .9
3.4 Responsibility of the operator . . . . . . . . . . . . . . . . .9
3.5 Personnel qualification . . . . . . . . . . . . . . . . . . 10
3.6 Personal protective equipment . . . . . . . . . . . . . . . 10
3.7 Labelling, safety marks . . . . . . . . . . . . . . . . . . 11
3.8 Ex marking . . . . . . . . . . . . . . . . . . . . . . 12
3.9 Signal jump electronics . . . . . . . . . . . . . . . . . . 13
4. Transport, packaging and storage 14
4.1 Transport. . . . . . . . . . . . . . . . . . . . . . . 14
4.2 Packaging and storage . . . . . . . . . . . . . . . . . . 14
5. Commissioning, operation 15
5.1 Precautions before mounting . . . . . . . . . . . . . . . . 15
5.2 Mounting instructions . . . . . . . . . . . . . . . . . . 15
5.3 Mounting of models F2301, F23C1 and F23S1 . . . . . . . . . . 16
5.4 Mounting of models F2301, F23C1 and F23S1 with swivel heads . . . 17
5.5 Mounting of model F2303 . . . . . . . . . . . . . . . . . 18
5.6 Mounting of model F2304 . . . . . . . . . . . . . . . . . 19
5.7 Electrical connection. . . . . . . . . . . . . . . . . . . 20
5.8 Electrical connection hazardous area . . . . . . . . . . . . . 23
6. Faults 23
7. Maintenance and cleaning 24
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . 24
7.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . 25
7.3 Recalibration . . . . . . . . . . . . . . . . . . . . . 25
8. Dismounting, return and disposal 25
8.1 Dismounting . . . . . . . . . . . . . . . . . . . . . 25
8.2 Return. . . . . . . . . . . . . . . . . . . . . . . . 26
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . 26
9. Specifications 27
9.1 Approvals . . . . . . . . . . . . . . . . . . . . . . 34
9.2 Dimensions model F2301 . . . . . . . . . . . . . . . . . 35
9.3 Dimensions models F2301 (signal jump) and F23C1 . . . . . . . . 37
9.4 Dimensions model F23S1 . . . . . . . . . . . . . . . . . 39
9.5 Dimensions model F2303 . . . . . . . . . . . . . . . . . 40
9.6 Dimensions model F2304 . . . . . . . . . . . . . . . . . 42
10. Accessories 44
10.1 Cable . . . . . . . . . . . . . . . . . . . . . . . 44
10.2 Swivel heads, dimensions in mm . . . . . . . . . . . . . . 44
Appendix: EU declaration of conformity 46

4WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
1. General information
1. General information
The tension/compression force transducers described in the operating instruc-
tions have been designed and manufactured using state-of-the-art technology. All
components are subject to stringent quality and environmental criteria during produc-
tion. Our management systems are certified to ISO 9001.
These operating instructions contain important information on handling the
instrument. Working safely requires that all safety instructions and work instructions
are observed.
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument's range of use.
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions on to the next operator or owner of the instrument.
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
The general terms and conditions contained in the sales documentation shall apply.
Subject to technical modifications.
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with
international standards.
Further information:
- Internet address: www.wika.de
- Relevant data sheet: FO 51.17 (F2301, F23C1, F23S1)
FO 51.47 (F2303)
FO 51.47 (F2304)
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
Abbreviations, definitions
2-wire The two connection lines are used for the voltage supply.
The measuring signal also provides the supply current.
3-wire Two connection lines are used for the voltage supply.
One connection line is used for the measuring signal.
UB+ Positive power supply terminal
UB- Negative power supply terminal
S+ Positive output terminal
S- Negative output terminal
UR+ Positive power supply terminal for relay (signal jump)
UR- Negative power supply terminal for relay (signal jump)
Shield Case
x-pin Pin assignment

5WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
2. Design and function
2. Design and function
2.1 Overview models F2301, F23C1, F23S1
Force introduction (spherical surface)
Counter nut
Measuring tube
Electrical connection
Swivel heads (optional)

6WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
2. Design and function
2.2 Overview model F2304
Force introduction
Counter nut
Measuring tube
Electrical connection
2.3 Overview model F2303
Force introduction
Measuring tube
Electrical connection

7WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
2.4 Description
The tension/compression force transducer is designed for measuring static and
dynamic tension and/or compression forces.The tension/compression force transdu-
cer consists of a measuring tube and a welded thin-film sensor. The measuring body
is made of stainless steel and is elastically deformed by a tension and/or compression
force introduced in the force direction. The resulting mechanical tensions are measured
by the installed thin-film sensor and output by an electrical output signal.
2.5 Scope of delivery
Tension/compression force transducer
Operating instructions
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
DANGER!
... indicates a potentially dangerous situation in the hazardous area that
can result in serious injury or death, if not avoided.
Information
... points out useful tips, recommendations and information for efficient
and trouble-free operation.
2. Design and function / 3. Safety

8WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
3. Safety
3.2 Intended use
The tension/compression force transducers are designed for measuring static and
dynamic tension and/or compression forces.
Defective instruments must be returned to the manufacturer.
Model Version
F2301
Option 1 Version with male threads
Option 2 Version with male threads and signal jump
F23C1
Option 1 Version with male threads
in acc. with ATEX directive 2014/34/EU
ATEX approval per EN 60079-0:2012 and
EN 60079-11:2012 (Ex ib)
IECEx approval per IEC 60079-0:2011 (Ed.6) and
IEC 60079-11:2011 (Ed. 6) (Ex ib)
Option 2 Version with male threads
UL approval per UL 61010-1 and CSA C22.2 NO. 61010-1
F23S1 Version with male threads
in acc. with requirements for functional safety
in acc. with the Machinery Directive 2006/42/EC, applies only in
combination with a safety control system, for example overload
protection ELMS1
F2303 Version with swivel head and female thread
F2304 Version with male thread and female thread
These instruments have been designed and tested in accordance with the relevant
safety regulations for electronic measuring instruments. Any usage outside of this is
deemed to be improper. The perfect functioning and operational safety of the transdu-
cers can only be guaranteed when complying with the instructions given in the operating
instructions. During its use, the legal and safety regulations (e.g.VDE 0100) required for
the particular application must additionally be observed. This also applies accordingly
when using accessories.Tension/compression force transducers are RoHS-compliant in
accordance with Directive 2011/65/EU Art. 2 (2) and (4) d), e) and g). Faultless and safe
operation of this transducer requires proper transport, professional storage, installation
and mounting as well as careful operation and corrective maintenance.
The instrument has been designed and built solely for the intended use described
here, and may only be used accordingly.The technical specifications contained in these
operating instructions must be observed. Improper handling or operation of the instru-
ment outside of its technical specifications requires the instrument to be taken out of

9WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
3. Safety
service immediately and inspected by an authorised service engineer.
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic fields, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary
to the intended use.
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and
injuries.
Refrain from unauthorised modifications to the instrument.
Any use beyond or different to the intended use is considered as improper use.
3.4 Responsibility of the operator
The instrument is used in the industrial sector. The operator is therefore responsible for
legal obligations regarding safety at work.
The safety instructions within these operating instructions, as well as the
safety, accident
prevention and environmental protection regulations for the application area must be
maintained.
The operator is obliged to maintain the product label in a legible condition.
To ensure safe working on the instrument, the operating company must ensure
that suitable first-aid equipment is available and aid is provided whenever required.
that the skilled electrical personnel are regularly instructed in all topics regarding
work safety, first aid and environmental protection and know the operating instruc-
tions and in particular, the safety instructions contained therein.
that the instrument is suitable for the particular application in accordance with its
intended use.
that personal protective equipment is available.

10 WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
3. Safety
3.5 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to equip-
ment.
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their techni-
cal training, know-how and experience as well as their knowledge of country-specific
regulations, current standards and directives, are capable of carrying out work on
electrical systems and independently recognising and avoiding potential hazards.The
skilled electrical personnel have been specifically trained for the work environment they
are working in and know the relevant standards and regulations. The skilled electrical
personnel must comply with current legal accident prevention regulations.
Special knowledge for working with instruments for hazardous areas:
The skilled electrical personnel must have knowledge of ignition protection types,
regulations and provisions for equipment in hazardous areas.
Special operating conditions require further appropriate knowledge, e.g. of aggressive
media.
3.6 Personal protective equipment
The requirements for the required protective equipment result from the ambient
conditions at the place of use, other products or the connection to other products.
The requisite personal protective equipment must be provided by the operating company.
The operator is in no way relieved of his obligations under labour law for the safety and
the protection of workers' health.
The design of the personal protective equipment must take into account all operating
parameters of the place of use.

11WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
12
11
3. Safety
3.7 Labelling, safety marks
Product label
Model
Serial number, TAG number
Measuring range
Pin assignment
Ingress protection per DIN EN 60259
Address
Force direction (tension or compression force)
Power supply, excitation voltage for mV/V sensor
Output signal
Product code
Date of manufacture calendar week/year
Approval-related product information
11
12
ATEX product label
Standard product label

12 WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
3.8 Ex marking
DANGER!
Danger to life due to loss of explosion protection
Non-observance of these instructions and their contents may result in the
loss of explosion protection.
Observe the safety instructions in this chapter and further explosion
instructions in these operating instructions.
Observe the information given in the applicable type examination certi-
ficate and the relevant country-specific regulations for installation and
use in hazardous areas (e.g. IEC 60079-14, NEC, CEC).
It applies to ATEX instruments with classification: EN 60079-0:2012 /
EN 60079-11:2012 (Ex ib).
Ex II 2G Ex ib IIC T4 Gb -25 °C < T
amb < +85 °C
Ex II 2G Ex ib IIC T3 Gb -25 °C < T
amb < +100 °C
Ex I M2 Ex ib I Mb -25 °C < T
amb < +85 °C
Ex II 2G Ex ib IIC T4 Gb -40 °C < T
amb < +85 °C
I M2 Ex ib I Mb (only available with cable connection)
Check whether the classification is suitable for the application. Observe the relevant
national regulations.
Certification body 0158
Type examination certificate IECEx BVS08.0028X / BVS 08 ATEX E 075 X
The normal installation instructions of the operating instructions shall apply.
Special conditions for safe use
The tension/compression force transducers are suitable for use in the following ambient
temperature ranges:
Model Temperature class Ambient temperature range
F*3C*.****.12************* T4
T3
-25 °C to +85 °C
-25 °C to +100 °C
F*3C*.****.14************* Group I -25 °C to +85 °C
F*3C*.****.15************* T4 -25 °C to +85 °C
F*3C*.****.16************* T4
T3
-40 °C to +85 °C
-40 °C to +100 °C
F*3C*.****.17************* T4
T3
-45 °C to +85 °C
-45 °C to +100 °C
F*3C*.****.21************* T4 and group I -40 °C to +85 °C
3. Safety

13WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
3.9 Signal jump electronics
Compliance with functional safety
An external safety control system independent of the transducer must monitor the safe
functioning of the transducer. The functional test (signal step change) with a signal jump
of 4 mA / 2 V is executed at an interval of 24 hours. The safety control system activates a
relay, thus changing the output signal of the transducer in a defined manner.
If the expected change in the output signal occurs, it can be assumed that the entire
signal path functions correctly. If this does not occur, then it can be concluded that there
is a fault in the signal path.
Moreover, the measuring signal should be checked by the safety control system for the
Min (A) and Max (B) signal value to ensure that any cable break or short-circuit that has
occurred is detected. The default setting of the transducer with current output 4 ... 20 mA
for overload detection is, for example:
With a fixed, set signal jump of, for example, 4 mA, then, in any operating state, the test
cycle can be triggered by activating the test relay. The upper measuring range limit of
20 mA will never be reached and thus the testing of the signal jump is enabled.
Short-circuit (B)
Cable break (A)
[mA]4
3.8 16 20 21
Signals of the signal jump electronics
Overload point
(100 % rated load)
3. Safety
The connection lead of the tension/compression force transducer models
F*3C*.****.15************* and F*3C*.****.21************* must be permanently installed.
No modification must be made to instruments operated in connection with hazardous
areas.

14 WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the tension/compression force transducer for any damage that may have been
caused by transport. Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
When unloading packed goods upon delivery as well as during
internal transport, proceed carefully and observe the symbols on the
packaging.
With internal transport, observe the instructions in chapter 4.2
“Packaging and storage”.
As precision measuring instruments, transducers require careful handling during
transport and mounting. Load impacts during transport (e.g. hitting a hard surface) can
lead to permanent damage, resulting in measuring errors in the subsequent measuring
operation.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
The measuring tube is made completely of stainless steel. The version of the accessory
cables has an ingress protection of IP67.The ingress protection IP67 is only guaranteed
in the plugged-in state. During storage the protection cap must always be on the electri-
cal connection to avoid entry of moisture and dirt.
Permissible conditions at the place of storage:
Storage temperature: -40 ... +85 °C
Humidity: 35 ... 85 % relative humidity (non-condensing)
Avoid exposure to the following factors:
Mechanical vibration, mechanical shock (putting it down hard)
Dust, dirt, and other objects may not be deposited in such a way that they form a
force shunt with the measuring tube, since this will falsify the measuring signal.

15WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
5. Commissioning, operation
5. Commissioning, operation
5.1 Precautions before mounting
Tension/compression force transducers are sensitive measuring instruments and
must be handled with appropriate care.
Before installing the tension/compression force transducer, check whether the threa-
ded bores of the installation position have been executed dimensionally accurately,
thus guaranteeing problem-free mechanical installation of the tension/compression
force transducer.
Before installation, make sure that any drilling chips have been removed from the
threaded bores.
5.2 Mounting instructions
CAUTION!
Damage to the instrument through improper installation
With the installation of the force transducer, the installation position and
the load direction must be considered.
Torsional and shear forces must be avoided. Transverse loads and
lateral forces also include the corresponding components of the
measurands that are introduced slanted.
Torsional moments, eccentric loads and transverse loads or lateral
forces cause measuring errors and may permanently damage the tensi-
on/compression force transducer.
The supplied counter nuts must never rest on the deformation element.
A gap of at least 0.5 mm must be provided.
The installation of the tension/compression force transducer must not
be made with any violent force.
During installation of the tension/compression force transducer, the
output signal (force value) must always be monitored to avoid mechani-
cal overload.
When subjected to loads in other devices, other than those planned
in the application, a change in the zero signal may result in permanent
damage.
Do not use the tension/compression force transducer as a climbing aid.
If high shear forces or lateral forces are to be expected, the use of
swivel heads is recommended (see chapter 10 “Accessories”).

16 WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
5.3 Mounting of models F2301, F23C1 and F23S1
Counter nut Measuring object
Thread Measuring object, counterpart
Force transducer Spherical surface
5. Commissioning, operation
If necessary, screw counter nuts on each
thread.
Screw the force transducer into the
measuring object. If necessary, secure
with locknut. Observe the max. torque.
Screw the counterpart of the measuring
object onto the free thread. If necessary,
secure with locknut. Observe the max.
torque. Load the force transducer.
Position the counterpart of the measuring
object against the spherical surface/face
of the force transducer. Load the force
transducer.
1
3A
2
3B

17WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
5.4 Mounting of models F2301, F23C1 and F23S1 with swivel heads
Counter nut Swivel heads
Thread Clamping system
Force transducer Retaining bolt
5. Commissioning, operation
Screw counter nuts on the thread. Screw swivel heads onto the thread.
Tighten counter nuts. Observe the max.
torque.
1
2
Place the swivel heads in the clamping
device and fix with retaining bolts.
The
retaining bolts must not restrict the free
location of the force transducer. Load the
force transducer.
34

18 WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
5. Commissioning, operation
5.5 Mounting of model F2303
Thread Clamping system
Force transducer Retaining bolt
Swivel head
Place the swivel head in the clamping
device and fix with retaining bolt.
The retai-
ning bolt must not restrict the free location
of the force transducer.
12
Screw force transducer onto thread of
the measuring object. Observe the max.
torque.
Load the force transducer.

19WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
5.6 Mounting of model F2304
5. Commissioning, operation
If necessary, screw a counter nut onto the
thread.
Screw the force transducer to the
measuring object.
Screw the counterpart of the measuring
object onto the free thread. As necessary,
secure with locknut. Observe the max.
torque. Load the force transducer.
Position the counterpart of the measuring
object against the spherical surface/face
of the force transducer. Load the force
transducer.
1
3A
2
3B
Counter nut Measuring object
Thread Measuring object, counterpart
Force transducer Spherical surface

20 WIKA operating instructions, models
F2301, F23C1, F23S1, F2303, F2304
ADPR1X914096.01 06/2020 EN/DE
EN
Output 0...10 V, 4 ... 20 mA, 3-wire
Circular connector M12 x 1, 4-pin
Output 4 ... 20 mA, 2-wire
Circular connector M12 x 1, 4-pin
5. Commissioning, operation
Circular connector M12 x 1, 4-pin
4 ... 20 mA
2-wire
4 ... 20 mA
3-wire
0 …10 V
3-wire
UB+ 111
0V/UB- 333
S+ 144
S- 333
Shield Case Case Case
5.7 Electrical connection
To prevent interferences from coupling into the system:
Use only shielded and low-capacitance measuring cables (for cables, see chapter 10
“Accessories”).
Ground the shield of the measuring cable.
Connect the cable shield with the case of the tension/compression force transdu-
cer. In the cables of the accessories, the cable shield is connected by means of the
knurled nut, thus connecting it to the case of the tension/compression force transdu-
cer (for cables, see chapter 10 “Accessories”).
Do not install measuring cables in parallel to 3-phase-current cables and control
cables.
Avoid stray fields of transformers, motors and contactors.
Transducers, amplifiers and processing or display units must not be grounded several
times.
Connect all instruments to the same protective conductor.
The pin assignment of the connector or of the cable can be found on the product label.
5.7.1 Pin assignment of analogue output, models F2301, F2303, F2304
When using extensions, only shielded and low-capacitance cables should be used.The
permitted maximum and minimum lengths of cable are defined in ISO 11898-2. Care
should be taken about the shielding.