WIKA F2822 User manual

Operating instructions
Betriebsanleitung
EN
DE
Tension/compression force transducer Pancake
Model F2822
Zug-/Druckkraftaufnehmer Flachprofil
Typ F2822
Model F2822

2
ADPR1X914113.01 09/2020 EN/DE
WIKA operating instructions, model
F2822
EN
DE
Operating instructions model F2822 Page
3 - 20
Betriebsanleitung Typ F2822 Seite
21 - 39
© 09/2020 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA®and tecsis®are registered trademarks in various countries.
WIKA®and tecsis®sind geschützte Marken in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!

3WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
Contents
Contents
1. General information 4
2. Design and function 5
2.1 Overview model F2822 . . . . . . . . . . . . . . . . . . .5
2.2 Description . . . . . . . . . . . . . . . . . . . . . . .5
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . .5
3. Safety 6
3.1 Explanation of symbols . . . . . . . . . . . . . . . . . . .6
3.2 Intended use . . . . . . . . . . . . . . . . . . . . . .6
3.3 Improper use . . . . . . . . . . . . . . . . . . . . . .7
3.4 Responsibility of the operator . . . . . . . . . . . . . . . . .7
3.5 Personnel qualification . . . . . . . . . . . . . . . . . . .8
3.6 Personal protective equipment . . . . . . . . . . . . . . . .8
3.7 Labelling, safety marks . . . . . . . . . . . . . . . . . . .9
4. Transport, packaging and storage 10
4.1 Transport. . . . . . . . . . . . . . . . . . . . . . . 10
4.2 Packaging and storage . . . . . . . . . . . . . . . . . . 10
5. Commissioning, operation 11
5.1 Precautions before commissioning . . . . . . . . . . . . . . 11
5.2 Mounting instructions . . . . . . . . . . . . . . . . . . 11
5.3 Installation of the tension/compression force transducer. . . . . . . 12
5.4 Electrical connection. . . . . . . . . . . . . . . . . . . 14
5.5 Connecting an amplifier. . . . . . . . . . . . . . . . . . 14
5.6 Pin assignment model F2822 . . . . . . . . . . . . . . . . 14
6. Faults 15
7. Maintenance and cleaning 16
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . 16
7.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . 16
7.3 Recalibration . . . . . . . . . . . . . . . . . . . . . 16
8. Dismounting, return and disposal 17
8.1 Dismounting . . . . . . . . . . . . . . . . . . . . . 17
8.2 Return. . . . . . . . . . . . . . . . . . . . . . . . 17
8.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . 17
9. Specifications 18
9.1 Approvals . . . . . . . . . . . . . . . . . . . . . . 19
10. Accessories 19
Annex: EU declaration of conformity 20

4WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
1. General information
1. General information
■
The tension/compression force transducers described in the operating instruc-
tions have been designed and manufactured using state-of-the-art technology.
All components are subject to stringent quality and environmental criteria during
production. Our management systems are certified to ISO 9001.
■
These operating instructions contain important information on handling the
instrument. Working safely requires that all safety instructions and work instructions
are observed.
■
Observe the relevant local accident prevention regulations and general safety
regulations for the instrument's range of use.
■
The operating instructions are part of the product and must be kept in the immediate
vicinity of the instrument and readily accessible to skilled personnel at any time. Pass
the operating instructions on to the next operator or owner of the instrument.
■
Skilled personnel must have carefully read and understood the operating instructions
prior to beginning any work.
■
The general terms and conditions contained in the sales documentation shall apply.
■
Subject to technical modifications.
■
Factory calibrations / DKD/DAkkS calibrations are carried out in accordance with
international standards.
Further information:
- Internet address: www.wika.de
- Relevant data sheet: FO 51.50 (F2822)
- Application consultant: Tel.: +49 9372 132-0
Fax: +49 9372 132-406
info@wika.de
Abbreviations, definitions
2-wire The two connection lines are used for the voltage supply.
The measuring signal also provides the supply current.
3-wire Two connection lines are used for the voltage supply.
One connection line is used for the measuring signal.
UB+ Positive power supply terminal
UB- Negative power supply terminal
S+ Positive output terminal
S- Negative output terminal
Shield Case

5WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
F
2.2 Description
The tension/compression force transducer is designed for measuring static and dyna-
mic tension/compression forces.The tension/compression force transducer consists
of a measuring spring on which strain gauges (SGs) are applied.The measuring body
is made of steel or corrosion-resistant stainless steel and is elastically deformed by a
tension/compression force introduced in the force direction.The resulting mechanical
tensions are measured by the strain gauges and output as an electrical output signal.
The strain gauges are arranged so that two of them are tensioned and the other two
are compressed when a force is applied to the transducer. The pick-up circuit includes
correction and compensation resistors to eliminate unwanted eects on the zero signal
and characteristic value.
2.3 Scope of delivery
■
Tension/compression force transducer
■
Operating instructions
2. Design and function
2. Design and function
2.1 Overview model F2822
Force introduction
Measuring spring
Electrical connection (cable)

6WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
These instruments have been designed and tested in accordance with the relevant
safety regulations for electronic measuring instruments. Any usage outside of this is
deemed to be improper.The perfect functioning and operational safety of the transdu-
cers can only be guaranteed when complying with the instructions given in the operating
instructions. During its use, the legal and safety regulations (e.g.VDE 0100) required for
the particular application must additionally be observed. This also applies accordingly
when using accessories.Tension/compression force transducers are RoHS-compliant
in accordance with directive 2011/65/EU. Faultless and safe operation of this transducer
requires proper transport, professional storage, installation and mounting as well as
careful operation and corrective maintenance.The instrument has been designed and
built solely for the intended use described here, and may only be used accordingly. The
technical specifications contained in these operating instructions must be observed.
Improper handling or operation of the instrument outside of its technical specifications
requires the instrument to be taken out of service immediately and inspected by an
authorised service engineer. Handle electronic precision measuring instruments with the
required care (protect from humidity, impacts, strong magnetic fields, static electricity
and extreme temperatures, do not insert any objects into the instrument or its openings).
Plugs, wires and sockets must be protected from contamination.The manufacturer shall
not be liable for claims of any type based on operation contrary to the intended use.
3. Safety
3.1 Explanation of symbols
WARNING!
... indicates a potentially dangerous situation that can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation that can result in light injuries
or damage to property or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient
and trouble-free operation.
3. Safety
3.2 Intended use
Model Version
F2822 Version for measuring ranges
0 ... 5 kN to 0 ... 500 kN

7WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
3. Safety
3.3 Improper use
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and
injuries.
▶
Refrain from unauthorised modifications to the instrument.
Any use beyond or different to the intended use is considered as improper use.
3.4 Responsibility of the operator
The instrument is used in the industrial sector. The operator is therefore responsible for
legal obligations regarding safety at work.
The safety instructions within these operating instructions, as well as the
safety, accident
prevention and environmental protection regulations for the application area must be
maintained.
The operator is obliged to maintain the product label in a legible condition.
To ensure safe working on the instrument, the operating company must ensure
■
that suitable first-aid equipment is available and aid is provided whenever required.
■
that the skilled electrical personnel are regularly instructed in all topics regarding
work safety, first aid and environmental protection and know the operating instruc-
tions and in particular, the safety instructions contained therein.
■
that the instrument is suitable for the particular application in accordance with its
intended use.
■
that personal protective equipment is available.

8WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
3. Safety
3.5 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient
Improper handling can result in considerable injury and damage to
equipment.
▶
The activities described in these operating instructions may only be
carried out by skilled personnel who have the qualifications described
below.
Skilled electrical personnel
Skilled electrical personnel are understood to be personnel who, based on their
technical training, know-how and experience as well as their knowledge of country-
specific regulations, current standards and directives, are capable of carrying out work
on electrical systems and independently recognising and avoiding potential hazards.
The skilled electrical personnel have been specifically trained for the work environment
they are working in and know the relevant standards and regulations.The skilled electri-
cal personnel must comply with current legal accident prevention regulations.
3.6 Personal protective equipment
The requirements for the required protective equipment result from the ambient
conditions at the place of use, other products or the connection to other products.
The requisite personal protective equipment must be provided by the operating
company.The operator is in no way relieved of his obligations under labour law for the
safety and the protection of workers' health.
The design of the personal protective equipment must take into account all operating
parameters of the place of use.

9WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
3. Safety
3.7 Labelling, safety marks
Product label
11
Type
Manufacturer logo
Serial number
Ingress protection per DIN EN 60529
Pin assignment
Manufacturer address
Force direction
Supply voltage
Output signal
Measuring range
Product code
11

10 WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
4. Transport, packaging and storage
4. Transport, packaging and storage
4.1 Transport
Check the tension/compression force transducer for any damage that may have been
caused by transport. Obvious damage must be reported immediately.
CAUTION!
Damage through improper transport
With improper transport, a high level of damage to property can occur.
▶
When unloading packed goods upon delivery as well as during
internal transport, proceed carefully and observe the symbols on the
packaging.
▶
With internal transport, observe the instructions in chapter 4.2
“Packaging and storage”.
As precision measuring instruments, transducers require careful handling during trans-
port and mounting. Load impacts during transport (e.g. hitting a hard surface) can lead
to permanent damage, resulting in measuring errors in the subsequent measuring
operation.
4.2 Packaging and storage
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change
in installation site, sending for repair).
Permissible conditions at the place of storage:
■
Storage temperature: See chapter 9 “Specifications”
■
Humidity: 35 ... 85 % relative humidity (non-condensing)
Avoid exposure to the following factors:
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Dust, dirt, and other objects may not be deposited in such a way that they form a
force shunt with the measuring spring, since this will falsify the measuring signal.

11WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
5. Commissioning, operation
5. Commissioning, operation
5.1 Precautions before commissioning
■
Tension/compression force transducers are sensitive measuring instruments and
must be handled with appropriate care.
■
Before installing the tension/compression force transducer, check whether the threa-
ded bores of the mounting position have been executed dimensionally accurately,
thus guaranteeing problem-free mechanical installation of the tension/compression
force transducer.
■
Before installation, make sure that any drilling chips have been removed from the
threaded bores.
5.2 Mounting instructions
CAUTION!
Damage to the instrument through improper installation
■
With the installation of the force transducer, the installation position and
the load direction must be considered.
■
Torsional and transverse forces must be avoided.Transverse loads
and lateral forces also include the corresponding components of the
measurands that are introduced slanted.
■
Torsional moments, eccentric loads and transverse loads or lateral
forces cause measuring errors and may permanently damage the
tension/compression force transducer.
■
The installation of the tension/compression force transducer must not
be made with any violent force.
■
During installation of the tension/compression force transducer, the
output signal (force value) must always be monitored to avoid mechani-
cal overload.
■
When subjected to loads in other devices, other than those planned
in the application, a change in the zero signal may result in permanent
damage.
■
Do not use the tension/compression force transducer as a climbing aid.

12 WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
5.3 Installation of the tension/compression force transducer
Force transducer Measuring object
Support surface Screws
5. Commissioning, operation
Place the force transducer correctly on the support surface.
Secure the force transducer to the support surface with screws. Observe the max.
torque.
1
2

13WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
5. Commissioning, operation
Screw the measuring object together with the force transducer. Observe the max.
torque. Load the force transducer.
3

14 WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
5. Commissioning, operation
5.4 Electrical connection
To prevent interferences from coupling into the system:
■
Use only shielded and low-capacitance measuring cables (for cables, see chapter 10
“Accessories”).
■
Ground the shield of the measuring cable.
■
Connect the cable shield with the case of the tension/compression force transdu-
cer. In the cables of the accessories, the cable shield is connected by means of
the knurled nut, thus connecting it to the case of the tension/compression force
transducer (for cables, see chapter 10 “Accessories”).
■
Do not install measuring cables in parallel to 3-phase-current cables and control
cables.
■
Avoid stray fields of transformers, motors and contactors.
■
Transducers, amplifiers and processing or display units must not be grounded
several times. Connect all instruments to the same protective conductor.
The pin assignment of the connector or of the cable can be found on the product label.
When using extensions, only shielded and low-capacitance cables should be used.The
permitted maximum and minimum lengths of cable are defined in ISO 11898-2. Care
should be taken about the shielding.
5.5 Connecting an amplifier
■
The use of an amplifier is optionally possible.
■
The amplifier is not included in delivery.
■
A version with integrated amplifier is possible as an option.
■
The tension/compression force transducers described in these operating instructions
output a signal in mV/V based on strain gauges.
■
Any DC amplifiers and carrier amplifiers designed for strain gauge measuring
systems can be used.
5.6 Pin assignment model F2822
Electrical connection - Cable assignment
Excitation voltage (+) Red
Excitation voltage (-) Black
Signal (+) Green
Signal (-) White
Shield Shield

15WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
6. Faults
CAUTION!
Physical injuries, damage to equipment
If faults cannot be eliminated by means of the measures listed, the
tension/compression force transducer must be taken out of operation
immediately.
▶
Contact the manufacturer.
▶
If a return is needed, please follow the instructions given in chapter 9.2
“Return”.
For contact details see chapter 1 “General information” or the back page
of the operating instructions.
Faults Causes Measures
No output signal No or wrong power supply,
current pulse
Rectify the power supply
Cable break Check the continuity
No or wrong output signal Wrong pin assignment Check pin assignment
Deviating zero point signal Overload, last offset, wrong
connection
Consult the manufacturer
Constant output signal when
changing force
Mechanical overload, wrong
pin assignment
Consult the manufacturer
Signal span varies EMC interference sources in
the environment; for example,
frequency converter
Shield instrument; cable
shield; remove source of
interference
Signal span drops/too small Mechanical overload Consult the manufacturer
6. Faults

16 WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
7. Maintenance and cleaning
7. Maintenance and cleaning
For contact details see chapter 1 “General information” or the back page
of the operating instructions.
7.1 Maintenance
This instrument is maintenance-free.
Repairs must only be carried out by the manufacturer.
Only use original parts (see chapter 10 “Accessories”).
7.2 Cleaning
1. Prior to cleaning, disconnect the tension/compression force transducer from the
voltage supply and dismount it.
2. Clean the tension/compression force transducer with a cloth.
Electrical connections must not come into contact with moisture!
CAUTION!
Damage to the instrument
Improper cleaning may lead to damage to the instrument!
▶
Do not use any aggressive cleaning agents.
▶
Do not use any hard or pointed objects for cleaning.
7.3 Recalibration
DKD/DAkkS certificate - official certificates:
We recommend that the tension/compression force transducer is recalibrated by the
manufacturer at regular time intervals of approx. 24 months.The basic settings will be
corrected if necessary.

17WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
8. Dismounting, return and disposal
8. Dismounting, return and disposal
8.1 Dismounting
Release the load from the tension/compression force transducer and disconnect from
power. Remove the tension/compression force transducer from the mounting.
8.2 Return
Strictly observe the following when shipping the instrument:
All instruments delivered to WIKA must be free from any kind of hazardous substances
(acids, bases, solutions, etc.) and must therefore be cleaned before being returned.
When returning the instrument, use the original packaging or a suitable transport packa-
ging.
To avoid damage:
1. Wrap the instrument in an antistatic plastic film.
2. Place the instrument, along with shock-absorbent material, in the packaging.
Place shock-absorbent material evenly on all sides of the transport packaging.
3. If possible, place a bag containing a desiccant inside the packaging.
4. Label the shipment as carriage of a highly sensitive measuring instrument.
Information on returns can be found under the heading “Service” on our
local website.
8.3 Disposal
Incorrect disposal can put the environment at risk.
Dispose of instrument components and packaging materials in an environmentally
compatible way and in accordance with the country-specific waste disposal regulations.
Do not dispose of with household waste. Ensure a proper disposal in
accordance with national regulations.

18 WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
9. Specifications
9. Specifications
Model F2822
Rated load Fnom kN 5, 10, 20, 50, 70, 100, 200, 250, 300, 400, 500
Relative linearity error dlin ±0.05 % Fnom
Relative reversibility error v ±0.05 % Fnom
Relative creep, 30 min. ±0.05 % Fnom
Relative span in unchanged mounting
position brg
±0.03 % Fnom
Relative deviation of zero signal dS, 0 ±1 % Fnom
Temperature eect on the zero signal TK0≤ ±0.03 %/10 °C
Temperature eect on the characteristic
value TKC
≤ ±0.03 %/10 °C
Limit force FL150 % Fnom
Breaking force FB200 % Fnom
Material of the measuring body Steel
Rated temperature range BT, nom -10 … +60 °C
Service temperature range BT, G -20 … +80 °C
Input resistance Re750 ±10 Ω
Output resistance Ra700 ±5 Ω
Insulation resistance Ris ≥ 5,000 MΩ/DC 100 V
Output signal (rated characteristic value)
Cnom
2.0 ±1 % mV/V
Electrical connection Cable ∅ 5 x 5.000 mm
Excitation voltage BU, nom
Standard DC 10 V (max. 15 V)
Option DC 12 … 28 V integrated or cable amplier
0(4) … 20 mA
DC 0 … 10 V
DC 0 … 5 V
Protection (acc. to IEC/EN 60529) IP66
Weight in kg
5, 10, 20, 50, 70 1.8
100, 200, 250 4.5
300, 400, 500 7

19WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
9. Specifications / 10. Accessories
Rated load
kN
Dimensions in mm
A B B1 C D E F H M
5, 10, 20, 50, 70 105 37 34 89 6.5 32 32 7.2 M16 x 1.5
100, 200, 250 125 56 46 101.8 10.5 39 39 11 M30 x 1.5
300, 400, 500 145 65 54 116.8 12.5 60 60 14 M40 x 1.5
9.1 Approvals
Logo Description Country
EU declaration of conformity
■
EMC directive
■
RoHS directive
European Union
EAC
■
EMC directive
Eurasian Economic
Community
10.Accessories
WIKA accessories can be found online at www.wika.com.
ØA
ØF
40
17.5
M
ØC
ØE(8x)
ØD(8x) 5000
H
B1
B

20 WIKA operating instructions, model F2822
ADPR1X914113.01 09/2020 EN/DE
EN
Annex: EU declaration of conformity
Table of contents
Languages:
Other WIKA Transducer manuals
Popular Transducer manuals by other brands

Mianyang Weibo Electronic
Mianyang Weibo Electronic WBP114S41 user manual

Setra Systems
Setra Systems 224 installation guide

Induction Dynamics
Induction Dynamics Solid Drive SD1g Installation & use guide

Pribusin
Pribusin IUC-7X-WTX instruction manual

Humminbird
Humminbird 530511-3_A Installation

Setra Systems
Setra Systems Multi-Sense 231RS manual

Balluff
Balluff BTL5-A11 series Technical Description, User's Guide

Balluff
Balluff BTL7 Series Condensed guide

SSET
SSET CE-F01-MS-0.5 Series instructions

HBM
HBM SLB-700A Mounting instructions

NK TECHNOLOGIES
NK TECHNOLOGIES VTD Series instructions

Daiichi Electronics
Daiichi Electronics PTT2-92A-12 instruction manual